K Number
K094030

Validate with FDA (Live)

Date Cleared
2010-12-23

(358 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FORE-SIGHT® Oximeter. Model MC-2000 Series is indicated for the continuous noninvasive monitoring of regional hemoglobin oxygen saturation of blood in the brain and skeletal muscle. It is intended for use in any individual at risk for reducedflow or no-flow ischemic states.

When used with Large sensors, the FORE-SIGHT Oximeter is indicated for use on the brain of adults and children over 40 kg. When used with Medium sensors, the FORE-SIGHT Oximeter is indicated for use on the brain of small adults and children between 4 and 80 kg. When used with Small sensors, the FORE-SIGHT Oximeter is indicated for use on the brain of infants and neonates ≤ 8 kg.

When used with Medium sensors, the FORE-SIGHT Oximeter is indicated for use on skeletal muscle of infants, children and adolescents between 5 and 50 kg.

The prospective clinical value of data from the FORE-SIGHT Oximeter has not been demonstrated in disease states and these data should not be used as a sole basis for diagnosis or therapy.

Device Description

The FORE-SIGHT Oximeter, MC-2000 Series, measures brain and skeletal muscle tissue oxygen saturation.

The FORE-SIGHT Oximeter consists of an optical transducer containing a laser light source and photodiode detectors, and a graphic display with user interface. The noninvasive, reflection mode, optical transducer is placed on the forehead (brain) or skeletal muscle, via a disposable sensor to determine tissue oxygenation. The FORE-SIGHT Oximeter is safe to use because it is designed to operate as a class 1 laser product, the safest FDA laser classification. Additional safety features include a laser interlock system, designed to prevent laser operation in case the optical transducer is not securely attached to the subject.

AI/ML Overview

Here's an analysis of the provided text to extract information about the acceptance criteria and the study that proves the device meets those criteria:

1. Table of Acceptance Criteria and Reported Device Performance:

Criterion CategorySpecific Criterion from DocumentReported Device Performance (FORE-SIGHT Oximeter)
Safety StandardsUL 60601-1 Safety testingTested to this standard
CAN/CSA C22.2 No. 601.1-M90Tested to this standard
IEC 60601-1 Safety of Medical Electrical EquipmentTested to this standard
EN 60601-1 Safety of Medical Electrical EquipmentTested to this standard
IEC 60601-1-1 Safety of Medical Electrical SystemsTested to this standard
IEC 60601-1-2: 2001 EMC Emissions and ImmunityTested to this standard
IEC 60601-1-4 Programmable Electrical Medical SystemsTested to this standard
IEC 60601-1-8 Alarm SystemsTested to this standard
IEC 60825-1 Safety of Laser ProductsTested to this standard (Class 1 laser product)
Performance (Skeletal Muscle)Precision (1 Standard deviation) against CO-oximetry± 4.9 %
Comparability/Superiority to predicate devices' precisionComparable or better than predicate devices when compared to CO-oximetry measurements

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Description: "Pediatric Subject Skeletal Muscle Validation: CO-oximetry data from invasive arterial and venous blood samples from sites representative of skeletal muscle were respectively compared to FORE-SIGHT Oximeter values. FORE-SIGHT sensors were placed over skeletal muscle of adolescents, children, and infants undergoing examination in a catheterization laboratory."
  • Sample Size: The document does not explicitly state the exact number of subjects (sample size) used in this clinical validation. It mentions "adolescents, children, and infants."
  • Data Provenance: The study was conducted in a "catheterization laboratory," implying an institutional setting. The country of origin is not specified, but the submission is from a US company (CAS Medical Systems, Inc., Branford, CT, USA) to the FDA, suggesting it was likely conducted in the US or a country with comparable medical standards. It is a prospective study because it involved placing sensors on subjects and collecting data during an ongoing medical examination.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • The ground truth for the skeletal muscle validation was established using CO-oximetry data from invasive arterial and venous blood samples. This is an objective, direct measurement, not an expert-based ground truth. Therefore, the concept of "number of experts" or "qualifications" for establishing this specific ground truth does not apply in the same way it would for image interpretation tasks. The expertise would lie in the medical professionals performing the catheterization and CO-oximetry.

4. Adjudication Method for the Test Set:

  • Since the ground truth was invasive CO-oximetry, there was no adjudication method (like 2+1 or 3+1 consensus) required for establishing the truth itself. The comparison was statistical: FORE-SIGHT Oximeter values versus CO-oximetry values.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not described in the provided text. The study focused on validating the device's accuracy against a gold standard (CO-oximetry), not on evaluating human reader performance with or without the device.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done:

  • Yes, the clinical testing described is essentially a standalone performance study of the FORE-SIGHT Oximeter. It compares the device's StO2 values directly to the CO-oximetry ground truth without involving human interpretation of the device's output as part of the primary outcome measure.

7. The Type of Ground Truth Used:

  • The type of ground truth used for the skeletal muscle validation was invasive CO-oximetry measurements from arterial and venous blood samples. This is a highly objective and direct measure of oxygen saturation.

8. The Sample Size for the Training Set:

  • The document does not provide information on the sample size used for the training set of the FORE-SIGHT Oximeter's algorithm. The provided information relates to the validation (testing) performed to demonstrate substantial equivalence for the 510(k) submission.

9. How the Ground Truth for the Training Set Was Established:

  • The document does not provide information on how the ground truth for any potential training set was established. While it states "Slight modifications of algorithm" compared to a predicate, the details of the original algorithm development or any subsequent retraining are not included here.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "CASMED" in a bold, sans-serif font. The letters are all capitalized and black. There is a registered trademark symbol to the right of the letter "D".

DEC 2 3 2010

5101k):

K194030

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

Submitter:CAS Medical Systems, Inc.
Address:44 East Industrial Rd. Branford CT. 06405 USA
Contact:Ron Jeffrey - Director, Regulatory Affairs, Phone - (203) 488-6056Fax - (203) 488-9438Email - rjeffrey@casmed.com
Prepared:December 23, 2010
Trade Name:FORE-SIGHT® Absolute Tissue Oximeter
Common Name:Tissue Oximeter
Classification Name:Oximeter, Tissue Saturation
Classification #:21 CFR (870.2700)
Product Code:MUD
510kK094030

{1}------------------------------------------------

DESCRIPTION

The FORE-SIGHT Oximeter, MC-2000 Series, measures brain and skeletal muscle tissue oxygen saturation.

The FORE-SIGHT Oximeter consists of an optical transducer containing a laser light source and photodiode detectors, and a graphic display with user interface. The noninvasive, reflection mode, optical transducer is placed on the forehead (brain) or skeletal muscle, via a disposable sensor to determine tissue oxygenation. The FORE-SIGHT Oximeter is safe to use because it is designed to operate as a class 1 laser product, the safest FDA laser classification. Additional safety features include a laser interlock system, designed to prevent laser operation in case the optical transducer is not securely attached to the subject.

FORE-SIGHT Oximeter Indications for Use

The FORE-SIGHT® Oximeter, Model MC-2000 Series is indicated for the continuous noninvasive monitoring of regional hemoglobin oxygen saturation of blood in the brain and skeletal muscle. It is intended for use in any individual at risk for reduced-flow or no-flow ischemic states.

When used with Large sensors, the FORE-SIGHT Oximeter is indicated for use on the brain of adults and children over 40 kg. When used with Medium sensors, the FORE-SIGHT Oximeter is indicated for use on the brain of small adults and children between 4 and 80 kg. When used with Small sensors, the FORE-SIGHT Oximeter is indicated for use on the brain of infants and neonates ≤ 8 kg.

When used with Medium sensors, the FORE-SIGHT Oximeter is indicated for use on skeletal muscle of infants, children and adolescents between 5 and 50 kg.

The prospective clinical value of data from the FORE-SIGHT Oximeter has not been demonstrated in disease states and these data should not be used as a sole basis for diagnosis or therapy.

FORE-SIGHT Oximeter Technology Compared to Predicate Devices

The FORE-SIGHT Oximeter compares substantially to one or more of the cited predicate devices in that they use fundamentally the same optical operating principle, called diffuse reflectance spectroscopy. All cited monitors use light to probe a cross-section of tissue microvasculature (i.e., a mixed bed of arterioles, capillaries and venules). The FORE-SIGHT Oximeter and predicate devices analyze light returning from the sampled tissue, after having passed through tissues, for hemoglobin in its oxygenated and deoxygenated forms. All cited monitors calculate oxygen saturation. This value reflects the percentage of oxygenated hemoglobin in the sampled tissue.

{2}------------------------------------------------

EQUIVALENCE (Predicate Device)

The FORE-SIGHT® Oximeter, MC-2000 Series, is equivalent to the following devices:

    • CAS Medical Systems Cerebral Oximeter (K091452)
    • Somanetics Corporation INVOS® 5100C Cerebral/Somatic Oximeter (K082327)
  • ** Hutchinson Technology InSpectra™ StO2 Tissue Oxygenation Monitor, Model 650 (K100915)
    • Spectros Corporation T-Stat™ 303 Microvascular Tissue Oximeter (K081233)
Comparison 'lable
------------------------------------annon de film and contraction of the comments of consisted on the first and the forme to the first andBut the contract and the contribution to call be claim and the closes of the closes of the closes of the closes of the closes of the closes of the closes of the closes of theNO. OF ALL OLL BLACK OF CLASS OF CHARGES.---------------------------------------------------------------------------------
Predicate DeviceSimilarities to FORE-SIGHTK 094030Differences from FORE-SIGHTK 094030
CAS MedicalSystemsFORE-SIGHTOximeter(K091452)Same laser light source/energy; Same four wavelengths; Same sensor / monitor design; Same physiological purpose; Cerebral tissue oxygen saturationmeasurement; Same "cerebral" IFU; Same small, medium and largesensor sizes for cerebral; Same cerebral performance.Slight modifications of algorithm; Slight user interface changes; Added skeletal muscle monitoringwith medium sensors for specificpatient groups/weights
SomaneticsCorporationINVOS 5100COximeter(K082327)Tissue oximeter; Cerebral and somatic monitoring; Same hospital area of use; Same patient population groups forcerebral monitoring; Same parameter measured; Adhesive sensor.IFU includes cerebral for allindividuals sensor size dependant; Infants, children and adolescentpopulation for skeletal muscles; LED light source, two wavelengths,LEDs in sensor; Trend monitor <2.5 kg.
HutchinsonTechnologyInSpectra StO2TissueOxygenationMonitor(K100915)Tissue oximeter; Four wavelengths; Fiber optics used; IFU includes skeletal muscle; Adhesive sensor.LED light source; Specific patient population notidentified.
SpectrosCorporation T-StatMicrovascularTissue Oximeter(K081233)Tissue oximeter, Broad bandwidth white light; Fiber optics used; IFU states microvascular tissuespaces; Various needle or probe- likesensors.Visible white light source; Specific monitoring sites notidentified.

{3}------------------------------------------------

Non-Clinical Performance Testing to Demonstrate Substantial Equivalence

The FORE-SIGHT Oximeter. MC-2000 Series has been tested to the following standards in accordance with CAS Medical Svstems Product Performance Specifications:

  • UL 60601-1 Safety testing for use of the UL Classified mark; .
  • CAN/CSA C22.2 No. 601.1-M90; .
  • IEC 60601-1 Safety of Medical Electrical Equipment: .
  • EN 60601-1 Safety of Medical Electrical Equipment: .
  • IEC 60601-1-1 Safety of Medical Electrical Systems: .
  • . IEC 60601-1-2: 2001 Safety of Medical Electrical Equipment with regard to EMC Emissions and EMC Immunity:
  • . IEC 60601-1-4 Safety of Programmable Electrical Medical Systems;
  • IEC 60601-1-8 Safety of Alarm Systems for Medical Equipment/Systems; .
  • IEC 60825-1: Safety of Laser Products (with amendments A1 and A2) .

In addition to the above laboratory tests, CAS has conducted a full program of individual hardware, software and systems monitor and sensor verification and validation studies.

Clinical Testing to Demonstrate Substantial Equivalence

Pediatric Subject Skeletal Muscle Validation: CO-oximetry data from invasive arterial and venous blood samples from sites representative of skeletal muscle were respectively compared to FORE-SIGHT Oximeter values. FORE-SIGHT sensors were placed over skeletal muscle of adolescents, children, and infants undergoing examination in a catheterization laboratory.

Conclusions Drawn from Clinical and Non-Clinical Testing

When compared to CO-oximetry measurements, the precision (1 Standard deviation) of FORE-SIGHT Oximeter values (StO>) were ± 4.9 % for skeletal muscle in human subjects. The precision of FORE-SIGHT Oximeter values were comparable or better than the predicate devices. when compared to CO-oximetry measurements.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

CAS Medical Systems, Inc. % Mr. Ron Jeffrey Director, Regulatory Affairs 44 East Industrial Road Branford, CT 06405

DEC 2 3 2010

Re: K094030

Trade/Device Name: Fore-Sight Absolute Cerebral and Somatic Oximeter, MC-2000 Series Regulation Number: CFR 21 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: MUD Dated: November 8, 2010 Received: November 9, 2010

Dear Mr. Jeffrey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{5}------------------------------------------------

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Dubmal Falls

Malvina B. Evdelman. M. Director Division of Ophthalmic. Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

DEC 2 3 2010

510(k) Number:

K 094030

Device Name:

FORE-SIGHT® Absolute Tissue Oximeter Model MC-2000 Series.

Indications for Use:

The FORE-SIGHT® Oximeter. Model MC-2000 Series is indicated for the continuous noninvasive monitoring of regional hemoglobin oxygen saturation of blood in the brain and skeletal muscle. It is intended for use in any individual at risk for reducedflow or no-flow ischemic states.

When used with Large sensors, the FORE-SIGHT Oximeter is indicated for use on the brain of adults and children over 40 kg. When used with Medium sensors, the FORE-SIGHT Oximeter is indicated for use on the brain of small adults and children between 4 and 80 kg. When used with Small sensors, the FORE-SIGHT Oximeter is indicated for use on the brain of infants and neonates ≤ 8 kg.

When used with Medium sensors, the FORE-SIGHT Oximeter is indicated for use on skeletal muscle of infants, children and adolescents between 5 and 50 kg.

The prospective clinical value of data from the FORE-SIGHT Oximeter has not been demonstrated in disease states and these data should not be used as a sole basis for diagnosis or therapy.

Prescription Use _ v _ _ (Part 21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) (Division Sign Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

K094030 510(k) Number -

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).