K Number
K112928

Validate with FDA (Live)

Date Cleared
2012-04-17

(197 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. It is for over-the-counter use.

Device Description

White Nitrile Medical Examination gloves powder free, non sterile, as described in this 510(k) Notification is substantially equivalent to the current class I patient examination gloves with product Code LZA (21CFR 880.6250). It meets all the specifications in ASTM D6319-10, Standard Specification for Nitrile Examination Gloves. They are made from nitrile latex compound, white colour, powder free and non sterile.

AI/ML Overview

The provided text describes a 510(k) summary for Nitrile Examination Gloves and does not contain information about a study with acceptance criteria for a device involving AI or expert review. Instead, it details the characteristics and performance of medical gloves against established ASTM standards and biocompatibility tests.

Therefore, I cannot provide the requested information about acceptance criteria for an AI device, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance, as these topics are not present in the document.

However, I can extract the acceptance criteria and reported device performance for the Nitrile Examination Gloves based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (Standard)Reported Device Performance (RS WHITE Nitrile Medical Examination Gloves Powder Free)
DimensionASTM D 6319-10Meets
Physical PropertiesASTM D 6319-10Meets
Freedom from pinholesASTM D 6319-10Meets
Powder-FreeASTM D 6319-10Meets
Biocompatibility
Primary SkinConsumer Product Safety Commission,Passes
Irritation TestTitle 16, Chapter II, Part 1500
Dermal SensitizationISO 10993-10:2010(E)Passes

Explanation for Missing Information:

The request implicitly assumes the device is an AI-driven medical device requiring clinical studies with human expert input. The provided text, however, relates to a Class I medical device (Nitrile Examination Gloves) where "acceptance criteria" are based on compliance with established industry standards (ASTM) and biocompatibility tests, rather than clinical efficacy studies in the context of diagnostic or therapeutic AI.

Therefore, the following points cannot be addressed from the given document:

  • Sample size used for the test set and the data provenance: Not applicable, as this is a product-testing conformity document, not a clinical study on a dataset.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for glove performance is defined by the technical specifications in the ASTM standard and the results of laboratory tests.
  • Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  • If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The ground truth is defined by the objective physical and chemical properties and performance markers specified in ASTM D 6319-10 and biocompatibility testing protocols.
  • The sample size for the training set: Not applicable. Manufacturing processes are typically controlled by quality assurance and quality control testing, not machine learning training sets.
  • How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Appendix 2

510(k) Summary Nitrile Examination Gloves Powder Free, White colour

1.0 Submitter:

K11292.8

Company Name: Riverstone Resources Sdn. Bhd.

Company Address: Lot 55& 56, No. 13&15 Jalan Jasmin 2 Kawasan Perindustrian Bukit Beruntung, 48300 Bukit Beruntung Selangor, Malaysia.

Contact Person: Ms Chong Chu Mee

Telephone No: +603-60283033

Fax No: + 603-60283022

2.0 Preparation Date: 5th Jan 2012

3.0 Name of the Device

Trade Name/ Proprietary Name: RS WHITE Nitrile Medical Examination Gloves (Powder Free)

Device Name: Nitrile Patient Examination gloves

Device Classification Name: Patient Examination gloves (21 CFR 880.6250)

Device Class: Class I

Product Code : Nitrile - LZA

4.0 Identification of The Legally Marketed Device:

Class I patient Examination gloves, Powder Free, LZA which meets all the

requirement of ASTM D 6319-10 and FDA 21 CFR 880.6250.

It is equivalent to K 100603, RS Safe Blue Nitrile Medical Examination Gloves

Powder Free (Non-Sterile) except for colour.

{1}------------------------------------------------

5.0 Description of Device

White Nitrile Medical Examination gloves powder free, non sterile, as described in this 510(k) Notification is substantially equivalent to the current class I patient examination gloves with product Code LZA (21CFR 880.6250). It meets all the specifications in ASTM D6319-10, Standard Specification for Nitrile Examination Gloves. They are made from nitrile latex compound, white colour, powder free and non sterile.

6.0 Intended use of the Device

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. It is for over-the-counter use.

7.0 Summary of the Technological Characteristics of the Devices compared to the Predicate Device for substantial equivalent discussion

There is no different technology characteristics compared to the predicate device except colour. Gloves are made from nitrile latex compound, White colour, powder free and non sterile. It is equivalent to K 100603, RS Safe Blue Nitrile Medical Examination Gloves Powder Free (Non-Sterile) except for colour.

{2}------------------------------------------------

CharacteristicsRS WHITE NitrileMedical ExaminationGloves (Powder Free)RS Safe Blue NitrileMedical ExaminationGloves Powder Free(Non-Sterile) K 100603
Product CodeLZALZA
Intended useA powder free patientexamination glove is adisposable deviceintended for medicalpurposes that is wornon the examiner's handor finger to preventcontamination betweenpatient and examiner.The device is for over-the-counter use.A powder free patientexamination glove is adisposable deviceintended for medicalpurposes that is wornon the examiner's handor finger to preventcontamination betweenpatient and examiner.The device is for over-the-counter use.
Material useNitrile latex compoundNitrile latex compound
ColourWhiteBlue
SterilityNon sterileNon sterile
DimensionsMeets ASTM D6319-10Meets ASTM D6319-00a
Physical propertiesMeets ASTM D6319-10Meets ASTM D6319-00a
Freedom from pinholesMeets ASTM D6319-10Meets ASTM D6319-00a
Powder-FreeMeets ASTM D6319-10Meets ASTM D6319-00a
Biocompatability test -Primary Skin IrritationTestPassesPasses
Dermal SensitizationPassesPasses
Assay
CharacteristicsStandardsDevice Performance
RS WHITE Nitrile Medical
Examination Gloves
Powder Free (Non-Sterile)
DimensionASTM D 6319-10Meets
Physical PropertiesASTM D 6319-10Meets
Freedom from pinholesASTM D 6319-10Meets
Powder-FreeASTM D 6319-10Meets
BiocompatabilityPrimary Skin Irritation TestConsumer Product SafetyCommission, Title 16,Chapter II, Part 1500Dermal SensitizationAssay- ISO 10993-10:2010(E)Passes

{3}------------------------------------------------

RS WHITE Nitrile Medical Examination Gloves Powder Free possesses the following technological characteristics compared to ASTM or Equivalent standards:

8.0 Conclusion

It can be concluded that White Nitrile Medical Examination Gloves (Powder Free) will perform according to the glove performance standards ASTM D 6319-10, biocompatibility requirement and FDA requirements and the labeling claims for the product.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Chong Chu Mee Technical Manager River Stone Resources Sdn. Bhd. Lot 55 & 56, NO. 13 & 15, Jalan Jasmin 2 Kawasan Perindustrian Bukit Beruntung, 48300 Bukit Beruntung Selangor, Malaysia

· Re: K112928

Trade/Device Name: RS WHITE Nitrile Medical Examination Gloves (Powder Free)

Regulation Number: 21 CFR 880.6250

Regulation Name: Patient Examination Glove

Regulatory Class: I

Product Code: LZA Dated: March 30, 2012 Received: April 4, 2012

Dear Ms. Chu Mee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

APR 1 7 2012

{5}------------------------------------------------

Page 2 - Ms. Chu Mee

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

G. Anthony D. Watson, B.S., M.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

· Enclosure

{6}------------------------------------------------

Appendix 1

INDICATION FOR USE

Riverstone Resources Sdn. Bhd.

510(K) Number: K112928

Device Name:

Applicant:

RS WHITE Nitrile Medical Examination Gloves

(Powder Free)

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Dokie faugulis

(Division Sign-Off) (Division-Sign-On)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K112928

Section 1A-1

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.