K Number
K112810

Validate with FDA (Live)

Date Cleared
2011-10-26

(29 days)

Product Code
Regulation Number
866.5775
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BioPlex® 2200 Anti-CCP kit is a multiplex flow immunoassay intended for the semiquantitative detection of IgG antibodies to Cyclic Citrullinated Peptide (CCP) in human serum or plasma (EDTA and sodium heparin). Detection of CCP antibodies is used as an aid in the diagnosis of rheumatoid arthritis and should be used in conjunction with other clinical findings and laboratory results.

The BioPlex® 2200 Anti-CCP kit is intended for use with the Bio-Rad BioPlex® 2200 System.

Device Description

The BioPlex® 2200 Anti-CCP kit uses multiplex flow immunoassay, a methodology that permits simultaneous detection and identification of antibodies in a single tube. "CCP IgG" is an acronym for the detection of IgG antibodies to Cyclic Citrullinated Peptide.

One (1) population of fluorescent beads is coated with antigens associated with cyclic citrullinated peptide (CCP). The BioPlex® 2200 System combines an aliquot of patient sample, sample diluent, and bead reagent into a reaction vessel and incubates the mixture. After a wash cycle to remove unbound antibody, anti-human IgG conjugated to phycoerythrin is added and the mixture is incubated at 37°C. Excess conjugate is removed in another wash cycle and the washed beads are re-suspended in wash buffer. The bead mixture then passes through the detector. The identity of the assay and control beads is determined by the fluorescence embedded in the surface of the bead and the amount of immobilized antibody is determined by the fluorescence of the anti-IgG reporter conjugate. Raw data are calculated in relative fluorescence intensity (RFI).

Three additional dyed beads, Internal Standard Bead (ISB), Serum Verification Bead (SVB) and Reagent Blank Bead (RBB) are present in each reaction mixture to verify detector response, the addition of serum or plasma to the reaction and the absence of significant non-specific binding. Refer to the BioPlex® 2200 System Operation Manual for more information.

The instrument is calibrated using a set of six (6) distinct calibrator vials, supplied separately by Bio-Rad Laboratories.

AI/ML Overview

Here's an analysis of the provided text regarding the BioPlex® 2200 Anti-CCP Kit, focusing on the acceptance criteria and the study conducted to prove it meets them.

Based on the provided document, this 510(k) is a Special 510(k) for a device modification, not an initial clearance for a new device. The modification is very specific: changing the frequency of Reagent Pack Quality Control (QC) testing from "once per pack and per day" to "once per day or per new reagent pack lot." Therefore, the "study" described is primarily focused on demonstrating that this change in QC frequency does not negatively impact the device's performance or safety.


Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance (with modified QC frequency)
No compromise in device performance or safety due to modified QC frequency (Implicit, derived from the purpose of a Special 510(k) for a QC change and the risk analysis). The essential implicit acceptance criterion for this Special 510(k) is that the change in QC frequency maintains the safety and effectiveness of the device as previously cleared (K093954), meaning its "performance" in terms of clinical utility (semi-quantitative detection of IgG antibodies to CCP as an aid in RA diagnosis) remains unaffected. This would typically be demonstrated by showing that the modified QC procedure effectively controls the process and prevents the release of incorrect results.The document states: "Based on the conclusion of the risk management report, the modified QC procedure fulfills the requirements of the specifications of the design control process. Therefore, the performance of the modified QC test frequency is substantially equivalent to the current cleared kit." While specific performance metrics (e.g., sensitivity, specificity, precision, accuracy) are not explicitly provided or re-evaluated in this modification document, the conclusion of "substantial equivalence" regarding the performance of the modified QC test frequency implies that the critical performance characteristics of the assay itself (as established by the original K093954 clearance) are maintained. The risk analysis (FMEA) would have justified that the reduced QC frequency does not introduce unacceptable errors.

Study Details Pertaining to this Special 510(k)

The document describes a Risk Analysis rather than a traditional clinical or analytical performance study with patient samples. The "study" here is a design control activity and risk management exercise to justify the change in QC frequency.

  1. Sample size used for the test set and the data provenance: Not applicable in the context of this specific Special 510(k) submission. No clinical or analytical test set with biological samples is described for this modification. The "test set" in this context refers to the parameters and outcomes evaluated in the FMEA.

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The "ground truth" for the FMEA would be the expected performance characteristics of the device and the potential failure modes in a manufacturing/quality control context, as understood by Bio-Rad's internal experts (e.g., quality engineers, design engineers, regulatory affairs, manufacturing personnel). These qualifications are not specified in the public summary.

  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. This refers to consensus among experts, which isn't described for the FMEA process in this summary. The FMEA process itself is a systematic analysis.

  4. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an in vitro diagnostic (IVD) kit, not an AI-assisted diagnostic imaging or interpretation tool.

  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is an IVD kit used with an instrument, not a standalone algorithm.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For this specific modification, the "ground truth" is implied to be the established safety and effectiveness of the predicate device (BioPlex® 2200 Anti-CCP Kit, K093954) and the expected control afforded by the QC procedure. The FMEA evaluates potential deviations from this "ground truth" (i.e., failure modes) and their impact.

  7. The sample size for the training set: Not applicable. This is not an AI/ML device, and no "training set" is mentioned for this modification.

  8. How the ground truth for the training set was established: Not applicable.


Summary of the "Study" (Design Control Activities) for this Modification

The "study" conducted for this Special 510(k) primarily comprised:

  • Failure Mode and Effect Analysis (FMEA): This systematically identified potential risks and impacts associated with the change in QC frequency.
    • Purpose: To "facilitate, capture, and quantify potential impacts of false positive or negative patient results."
    • Metric: Risk Priority Number (RPN) – a quantitative measure combining severity, occurrence, and detection of potential risks.
    • Outcome: Specific mitigations were recommended if the RPN exceeded a chosen threshold.
  • Risk Management Report: Concluded that the modified QC procedure "fulfills the requirements of the specifications of the design control process." This implies that the FMEA demonstrated that the proposed change does not introduce unacceptable risks or compromise the device's performance.

The conclusion of these activities was that "the performance of the modified QC test frequency is substantially equivalent to the current cleared kit." This statement is the direct evidence that the acceptance criterion (maintaining safety and effectiveness despite the QC change) was met.

{0}------------------------------------------------

Bio-Rad Laboratories, Inc Special 510(k): Device Modification BioPlex 2200 Anti-CCP Kit

Confidential

BioPlex® 2200 Anti-CCP Kit 510(k) Summary

510(k) Number_K112810

Date Prepared: September 19, 2011

Introduction

Bio-Rad Laboratories hereby submits this Special 510(k) in accordance with the Dio-Rud Laboratories neroof backers of 807.92. This summary of 510(k) safety and requirements of STAD i on provides detail as a basis for a determination of substantial equivalence for the BioPlex® 2200 Anti-CCP Kit.

Submitter name, address and contact

SubmitterContact Person
Bio-Rad Laboratories, IncJuang Wang
BioPlex DivisionRegulatory Affairs Representative
5500 E. Second StreetPhone: (510)741-4609
Benicia, CA 94510Fax: (510)741-4650

Device name and Classification

Product Trade NameBioPlex® 2200 Anti-CCP Kit
Common NameMulti-Analyte Detection System --Anti-CCP IgG
Classification nameAntibodies, anti-Cyclic Citrullinated Peptide (CCP)
Device ClassClass II
Classification PanelImmunology
Regulation Number21 CFR 866.5775
Product CodeNHX

Legally Marketed Predicate Device

BioPlex® 2200 Anti-CCP Kit, K093954

Intended Use/Indications For Use

The BioPlex® 2200 Anti-CCP kit is a multiplex flow immunoassay intended for the semiquantitative detection of IgG antibodies to Cyclic Citrullinated Peptide (CCP) in human serum or plasma (EDTA and sodium heparin). Detection of CCP antibodies is used as an aid in the diagnosis of rheumatoid arthritis and should be used in conjunction with other clinical findings and laboratory results.

{1}------------------------------------------------

Bio-Rad Laboratories, Inc Special 510(k): Device Modification BioPlex 2200 Anti-CCP Kit

Confidential

The BioPlex® 2200 Anti-CCP kit is intended for use with the Bio-Rad BioPlex® 2200 System.

Device Description

The BioPlex® 2200 Anti-CCP kit uses multiplex flow immunoassay, a methodology that I ne DIOT for " Les be a resembly but permits simultaneous detection and identification groat. y reseilbodies in a single tube. "CCP IgG" is an acronym for the detection of lgG antibodies to Cyclic Citrullinated Peptide.

One (1) population of fluorescent beads is coated with antigens associated with cyclic citrullinated peptide (CCP). The BioPlex® 2200 System combines an aliquot of patient sample, sample diluent, and bead reagent into a reaction vessel and incubates the mixture sample, After a wash cycle to remove unbound antibody, anti-human IgG conjugated to phycoerythrin is added and the mixture is incubated at 37°C. Excess conjugate is proved in another wash cycle and the washed beads are re-suspended in wash buffer. The bead mixture then passes through the detector. The identity of the assay and control beads is determined by the fluorescence embedded in the surface of the bead and the amount of immobilized antibody is determined by the fluorescence of the anti-lgG reporter conjugate. Raw data are calculated in relative fluorescence intensity (RFI).

Three additional dyed beads, Internal Standard Bead (ISB), Serum Verification Bead (SVB) and Reagent Blank Bead (RBB) are present in each reaction mixture to verify , detector response, the addition of serum or plasma to the reaction and the absence of significant non-specific binding. Refer to the BioPlex® 2200 System Operation Manual for more information.

The instrument is calibrated using a set of six (6) distinct calibrator vials, supplied separately by Bio-Rad Laboratories.

Similarities and Differences

Similarities

FeatureModified Device
Intended Use/Indications For UseNo Change
Kit componentsNo Change
Technical SpecificationsNo Change
Fundamental Scientific TechnologyNo Change

{2}------------------------------------------------

Bio-Rad Laboratories, Inc Special 510(k): Device Modification BioPlex 2200 Anti-CCP Kit

Confidential

Differences

The only difference of the BioPlex® 2200 Anti-CCP kit is to modify QC testing from each reagent pack to once per day as stated in the Instructions For Use (IFU) of the BioPlex® 2200 Anti-CCP reagent kit.

FeatureModifiedPredicate
Frequency ofReagent Pack QCTestingQC once per day or per newreagent pack lotQC once per pack and perday

Summary of Design Control Activities

A Failure Mode and Effect Analysis (FMEA) was used to facilitate, capture, and quantify potential impacts of false positive or negative patient results. The Risk Priority Number (RPN) is a quantitative measure of the combined effects of severity, occurrence, and detection of potential risks. Specific mitigations are recommended that may include changes to the design or formulation if the RPN score exceeds a chosen threshold.

The Design Control Activities include Risk Analysis method to identify the verification and validation activities required, test used and acceptance criteria.

Based on the conclusion of the risk management report, the modified QC procedure fulfills the requirements of the specifications of the design control process. Therefore, the performance of the modified QC test frequency is substantially equivalent to the current cleared kit.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is positioned to the right of a circular seal that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter of the circle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Bio-Rad Laboratories c/o Mr. Juang Wang Regulatory Affairs Representative BioPlex 2200 Division 5500 E. Second Street Benicia, CA 94510

OCT 2 6 2011

Re: K112810

Trade/Device Name: Bio-Plex® 2200 Anti-CCP Kit Regulation Number: 21 CFR § 866.5775 Regulation Name: Rheumatoid factor immunological test system Regulatory Class: Class II Product Code: NHX Dated: September 26, 2011 Received: September 27, 2011

Dear Mr. Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice

{4}------------------------------------------------

Page 2 - Mr. Juang Wang

requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Maria M. Chan, Ph.D.

Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indication(s) for use

510(k) Number (if known): _ K 112810

Device Name: BioPlex® 2200 Anti-CCP Kit

Indications for Use:

The BioPlex® 2200 Anti-CCP kit is a multiplex flow immunoassay intended for the semi-quantitative detection of IgG antibodies to Cyclic Citrullinated Peptide (CCP) in human serum or plasma (EDTA and sodium heparin). Detection of CCP antibodies is used as an aid in the diagnosis of rheumatoid arthritis and should be used in conjunction with other clinical findings and laboratory results.

The BioPlex 2200 Anti-CCP kit is intended for use with the Bio-Rad BioPlex 2200 system. .

The BioPlex 2200 Anti-CCP Calibrator Set is intended for the calibration of the BioPlex 2200 Anti-CCP Reagent Pack.

The BioPlex 2200 Anti-CCP Control Set is intended for use as an assayed quality control to monitor the overall performance of the BioPlex 2200 instrument and BioPlex 2200 Anti-CCP Reagent Pack in the clinical laboratory. The performance of the BioPlex 2200 Anti-CCP Control Set has not been established with any other Anti-CCP assay.

Prescription Usex
---------------------

AND/OR

Over-the-Counter Use
------------------------

(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)(Please do not write below this line-Continue on another page of needed)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign Off

Division Sign-Off( Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K112810

§ 866.5775 Rheumatoid factor immunological test system.

(a)
Identification. A rheumatoid factor immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the rheumatoid factor (antibodies to immunoglobulins) in serum, other body fluids, and tissues. Measurement of rheumatoid factor may aid in the diagnosis of rheumatoid arthritis.(b)
Classification. Class II (performance standards).