K Number
K112784

Validate with FDA (Live)

Date Cleared
2011-11-22

(57 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RENASYS AB Abdominal Dressing Kit with Soft Port is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome.

The RENASYS AB Abdominal Dressing Kit with Soft Port is intended for use in acute hospital settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.

Device Description

The RENASYS™ AB Abdominal Dressing Kit with Soft Port consists of two large hydrophobic reticulated polyurethane foam dressings that incorporate several cuts to facilitate custom sizing if needed. Also included in the kit are a polyurethane organ protection layer, six transparent film drapes and a Soft Port suction port assembly with tubing that attaches to the exudate canister. The kit is designed specifically for abdominal wounds and is supplied sterile, single use.

The modification to this kit is the inclusion of a Soft Port suction port, to replace the existing suction port.

The RENASYS AB Abdominal Dressing Kit is used in conjunction with Smith & Nephew RENASYS EZ and RENSASYS EZ PLUS negative pressure wound therapy pumps and canister kits, which have been cleared under 510(k) numbers K082426 and K091470.

AI/ML Overview

The provided text describes a 510(k) summary for the RENASYS™ AB Abdominal Dressing Kit with Soft Port. It focuses on demonstrating substantial equivalence to predicate devices through non-clinical bench testing. Therefore, the requested information regarding acceptance criteria and studies largely pertains to these bench tests, not clinical performance or AI algorithm validation.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Implied from “Meets required specifications”)Reported Device Performance
Comparable performance with clottable blood to existing suction portComparable performance patterns demonstrated (when tested with clottable blood).
Performs to specification under excessive weightPerforms to specification (under excessive weight, blockage by compression/folding, or particulate challenge).
Performs to specification when blocked by compression, foldingPerforms to specification (under excessive weight, blockage by compression/folding, or particulate challenge).
Performs to specification when subjected to particulate challengePerforms to specification (under excessive weight, blockage by compression/folding, or particulate challenge).
Effectively removes exudate at predetermined flow rate for a minimum of 48 hoursEffectively removes exudate from the abdomen at the predetermined flow rate for a minimum of 48 hours.
"Quick-click" connector establishes a secure connection to exudate canister tubingEstablishes a secure connection to the exudate canister tubing.

2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: The document does not specify exact sample sizes (e.g., number of tests per criterion). It states "Laboratory testing was completed" and "Testing to verify".
  • Data Provenance: The tests are described as "Laboratory testing" or "Bench" tests, implying they were conducted in a controlled lab environment. No information on country of origin or whether it was retrospective/prospective in a clinical sense is provided as these are non-clinical studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This question is not applicable to the provided document. The ground truth for the non-clinical bench tests would be the measured physical performance against predefined engineering specifications, not expert interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This question is not applicable to the provided document. Adjudication methods like 2+1 or 3+1 are used for expert consensus on clinical data. For bench tests, performance is typically measured directly against specifications.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable to the provided document. The device is a medical dressing kit, not an AI-powered diagnostic tool. Therefore, an MRMC study or AI assistance is not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable to the provided document. The device is a medical dressing kit, not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical bench tests, the "ground truth" is the predefined engineering specifications and measurable performance characteristics of the device components (e.g., flow rate, connection security, performance under stress, biocompatibility).

8. The sample size for the training set

This question is not applicable to the provided document. The device is a medical dressing kit, not an AI algorithm that requires a training set.

9. How the ground truth for the training set was established

This question is not applicable to the provided document. The device is a medical dressing kit, not an AI algorithm that requires a training set.

{0}------------------------------------------------

NOV 2 2 2011

K112784

10f3

510(k) Summary

General Information
Submitters Name/Address:Smith & Nephew, Inc.970 Lake Carillon DriveSuite 110St. Petersburg, FL 33716
Establishment Registration Number:3006760724
Contact Person:Laura D. ReynoldsDirector, Regulatory Affairs
Phone Number:(727) 329-7702
Date Prepared:November 1, 2011
Trade Name:RENASYS™™ AB Abdominal Dressing Kit with SoftPort
Generic/Common Name:NPWT Abdominal Wound Dressing Kit
Classification Name:Mesh, Surgical, Polymeric, 21 CFR 878.3300
Product Classification/Code:Class II, FTL
Predicate Device Information
510(k) #DeviceManufacturerClearance Date
K100787RENASYS™ F/ABAbdominal Dressing KitSmith & Nephew, Inc.9/17/2010
K110647RENASYS Foam andGauze NPWT WoundDressing Kits with SoftPortSmith & Nephew, Inc.6/22/2011

Device Description

The RENASYS™ AB Abdominal Dressing Kit with Soft Port consists of two large hydrophobic reticulated polyurethane foam dressings that incorporate several cuts to facilitate custom sizing if needed. Also included in the kit are a polyurethane organ protection layer, six transparent film drapes and a Soft Port suction port assembly with tubing that attaches to the exudate canister. The kit is designed specifically for abdominal wounds and is supplied sterile, single use.

The modification to this kit is the inclusion of a Soft Port suction port, to replace the existing suction port.

The RENASYS AB Abdominal Dressing Kit is used in conjunction with Smith & Nephew RENASYS EZ and RENSASYS EZ PLUS negative pressure wound therapy pumps and canister kits, which have been cleared under 510(k) numbers K082426 and K091470.

{1}------------------------------------------------

2 of 2

Indications for Use

The RENASYS AB Abdominal Dressing Kit with Soft Port is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome.

The RENASYS AB Abdominal Dressing Kit with Soft Port is intended for use in acute hospital settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.

Non-Clinical Tests (Bench)

Design verification testing has been conducted to verify the Soft Port suction device meets the required specifications and functions equivalent to the existing suction port component in the kit. Testing verified that the Soft Port meets the design specifications and demonstrated substantial equivalence to the predicate device.

Summary of testing conducted:

  • Laboratory testing was completed to confirm the Soft Port and the existing suction port demonstrate comparable performance patterns when tested with clottable blood.
  • Testing to verify that the Soft Port performs to specification when it is placed under excessive ・ weight, becomes blocked by compression, folding, or is subjected to particulate challenge.
  • -Testing to verify the Soft Port assembly will effectively remove exudate from the abdomen at the predetermined flow rate for a minimum of 48 hours.
  • Testing to verify that the new "quick-click" connector establishes a secure connection to the exudate canister tubing.

The following biocompatibility testing for all kit components has been successfully completed per applicable parts of ISO 10993:

Kit ComponentTests Completed
FoamCytotoxicity
Irritation
Sensitization
Organ Protection LayerCytotoxicity
Irritation
Sensitization
Implantation
Sub-acute Toxicity
Genotoxicity
Transparent Film DrapeCytotoxicity
Irritation
Sensitization
Soft Port Suction AssemblyGenotoxicity
Cytotoxicity
Irritation
Sensitization
Sub-acute Toxicity

{2}------------------------------------------------

4112784

3.83

Conclusion

In establishing substantial equivalence to the currently marketed devices, Smith & Nephew, Inc evaluated the indications for use, materials, technology and product specifications of the device. Performance testing has been successfully completed to demonstrate that the RENASYS AB Abdominal Dressing Kit with Soft Port is substantially equivalent to the marketed device and is appropriate for the intended use.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle in flight. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

NOV 2 2 2011

Smith & Nephew, Inc. % Ms. Laura D. Reynolds Director, Regulatory Affairs 970 Lake Carillon Drive, Suite 110 St. Petersburg, Florida 33716

Re: K112784

Trade/Device Name: RENASYS™ AB Abdominal Kit with Soft Port Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL Dated: November 01, 2011 Received: November 02, 2011

Dear Ms. Reynolds:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

Page 2 -- Ms. Laura D. Reynolds

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K1|2784

Device Name: RENASYS™ AB Abdominal Kit with Soft Port

Indications for Use:

The RENASYS AB Abdominal Kit with Soft Port is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome.

The RENASYS AB Abdominal Kit with Soft Port is intended for use in acute hospital settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Kurech MXIM

al. Orthopedic,

510(k) Number K112784

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.