K Number
K112735

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-11-02

(409 days)

Product Code
Regulation Number
870.5100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Tamarin Blue® RX PTCA Dilatation Catheters are indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion.

Device Description

The Tamarin Blue® PTCA RX Dilatation Catheter is a standard Rapid Exchange (RX) PTCA catheter with a single lumen proximally, a dual lumen distally, a semi-compliant inflatable balloon and a soft, tapered distal tip to aid in crossing tight stenoses. One lumen of the catheter's dual lumen is used for inflation and deflation of the balloon, and the other lumen allows guide wire access through the distal part of the catheter. The maximum recommended guide wire diameter is 0.014". A luer lock fitting (Hub) allows connection with an inflation device. Two radiopaque markers on the guide wire lumen tubing provide visual reference points for balloon positioning across the stenosis. The balloon material expands to a known diameter at specific pressure as defined in the compliance table supplied with the catheter.

AI/ML Overview

Here is an analysis of the provided 510(k) summary regarding the Tamarin Blue® PTCA RX Dilatation Catheter, focusing on acceptance criteria and the supporting study information:

Acceptance Criteria and Device Performance

The provided document describes a medical device, the Tamarin Blue® PTCA RX Dilatation Catheter, and its non-clinical testing. It asserts that the device met all the predetermined acceptance criteria of design verification and validation. However, the document does not explicitly list the acceptance criteria values or the specific reported device performance values for each test. It only states that the device successfully met these unlisted criteria.

Therefore, a table with specific acceptance criteria and reported performance cannot be created from the provided text. The document broadly states the types of tests performed.

Test CategoryAcceptance Criteria (Not explicitly stated)Reported Device Performance (Not explicitly stated)
BiocompatibilityWould involve meeting standards like ISO 10993 (e.g., cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, material-mediated pyrogenicity, implantation). Specific limits for extractables or biological responses would be predefined.The device materials were confirmed biocompatible based on testing results, indicating they met the requirements of ISO 10993 Part 1, 2, 4, 5, 10, 11, 12, ASTM F756-00, and 21 CFR 58 (GLP regulations).
Balloon ComplianceSpecific diameter at specific pressure (e.g., nominal pressure, rated burst pressure). Defined compliance curve within a specified tolerance from the predicate or design specification.The balloon material expands to a known diameter at specific pressure as defined in the compliance table supplied with the catheter (table not provided in summary, but implies data exists and was accepted).
Balloon Burst PressureMinimum burst pressure values for different balloon diameters and lengths, along with a specified safety factor.Device passed testing for balloon burst pressure. This implies the observed burst pressures met or exceeded the predefined minimums.
Balloon Fatigue EnduranceAbility to withstand multiple inflation/deflation cycles (e.g., number of cycles at specific pressure without failure, typically 3 or more cycles).Device passed testing for balloon fatigue (repeated balloon inflation) endurance, implying it met the required number of cycles without failure.
Balloon Inflation/Deflation PerformanceTime to inflate and deflate within specified limits; complete deflation allowing re-insertion/repositioning.Device passed testing for balloon inflation/deflation performance.
Bond StrengthsMinimum tensile or shear strength values for critical bonds (e.g., balloon-to-catheter shaft, hub-to-catheter shaft) to prevent dislodgement during use.Device passed testing for bond strengths.
Catheter Dimensions/ProfileVerification of outer diameter, balloon length, shaft diameter, guidewire lumen diameter, distal tip taper against design specifications.Device passed testing for catheter dimensions and balloon profile.
Catheter Body Min. Burst StrengthMinimum pressure at which the catheter shaft maintains integrity.Device passed testing for catheter body minimum burst strength.
Device PreparationCompatibility with specified guidewires (e.g., 0.014" diameter) and introducer sheaths; smooth passage without damage.Device passed testing for device preparation (guidewire and introducer compatibility).
Flexibility and KinkAbility to navigate tortuous anatomy without kinking or permanent deformation. Measured by kink resistance at a certain bending radius or force.Device passed testing for flexibility and kink, indicating it meets the required performance for navigation.
TorqueAbility to transmit rotational forces from the proximal end to the distal end efficiently (e.g., torque response, 1:1 torqueability).Device passed testing for torque.
RadiopacityVisibility of markers and potentially the balloon under fluoroscopy, typically compared to a reference material or having a minimum visibility score.Two radiopaque markers provide visual reference points, and the device passed testing for radiopacity, confirming their visibility.
Coating IntegrityAdherence and durability of any coating, typically assessed after simulated use or handling. No delamination or flaking.Device passed testing for coating integrity.
Particulate EvaluationNumber and size of particulates released from the device, ensuring it's below acceptable limits (e.g., ISO 10993-12, or specific limits for medical devices).Device passed testing for particulate evaluation, indicating any released particulates are within acceptable limits.

Study Information

The provided document describes a non-clinical study involving various in vitro tests to demonstrate the safety and effectiveness of the device. It is not a clinical study involving human subjects or AI algorithms.

  1. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for any of the individual non-clinical tests. The document only lists the types of tests performed.
    • Data Provenance: The tests are non-clinical (laboratory-based) and conducted according to international standards (ISO, ASTM) and US federal regulations (21 CFR 58 GLP). The country of origin of the data is not explicitly stated, but the sponsor is Natec Medical Ltd. from Mauritius.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This document describes non-clinical engineering and materials testing, not a study requiring expert consensus for diagnostic ground truth.

  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. Non-clinical physical and chemical tests have predefined objective pass/fail criteria, not subjective interpretation requiring adjudication.

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. No MRMC or AI-related study was performed or mentioned. This device is a physical medical device (catheter), not an AI diagnostic tool.

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI algorithm.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For non-clinical tests, the "ground truth" is typically defined by pre-established engineering specifications, material standards, and regulatory requirements. For example, a bond strength test would have a minimum force requirement (ground truth) that the device must meet. Biocompatibility has specific toxicological thresholds.

  7. The sample size for the training set: Not applicable. There is no training set mentioned as this is not an AI/machine learning study.

  8. How the ground truth for the training set was established: Not applicable. There is no training set.

{0}------------------------------------------------

510(k) SUMMARY 807.92(c)

K112735

NOV - 2 2012

SPONSOR

Fax:

Natec Medical Ltd. Company Name: Maeva Centre Building - Ebene Business Park Company Address: Reduit, Mauritius +230 466 30 54 Telephone: +230 466 67 70

Xavier De Buchere Contact Person: Quality Manager info@natec-medical.com

Summary Preparation Date: October 08, 2012

DEVICE NAME

Trade Name:Tamarin Blue® PTCA RX Dilatation Catheter
Common/Usual Name:PTCA RX Catheter
Classification Name:Catheters, transluminal coronary angioplasty, percutaneous
Regulation Number:870.5100
Product Code:LOX
Device Class:Class II

PREDICATE DEVICE

Legally Marketed Equivalent Devices:

  • · Maverick 2 PTCA Catheter Boston Scientific Corporation P860019
  • · Sprinter Legend Rapid Exchange Balloon Dilatation Catheter Medtronic Ireland -P790017
  • = Trek ™ RX Coronary Dilatation Catheter Abbott Vascular Inc. K103110

DEVICE DESCRIPTION

The Tamarin Blue® PTCA RX Dilatation Catheter is a standard Rapid Exchange (RX) PTCA catheter with a single lumen proximally, a dual lumen distally, a semi-compliant inflatable balloon and a soft, tapered distal tip to aid in crossing tight stenoses. One lumen of the catheter's dual lumen is used for inflation and deflation of the balloon, and the other lumen allows guide wire access through the distal part of the catheter. The maximum recommended guide wire diameter is 0.014". A luer lock fitting (Hub) allows

{1}------------------------------------------------

connection with an inflation device. Two radiopaque markers on the guide wire lumen tubing provide visual reference points for balloon positioning across the stenosis. The balloon material expands to a known diameter at specific pressure as defined in the compliance table supplied with the catheter. The device is available in the balloon diameters and lengths shown in the table below:

Diameter (mm)
1.502.002.252.502.753.003.253.504.00
Length (mm)
8
12
15
20
25
30

It will be supplied sterile and is intended for one time use.

DEVICE INTENDED USE

The Tamarin Blue® PTCA RX Dilatation Catheters are indicated for Balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion.

COMPARISON OF TECHNICAL CHARACTERISTICS 807.92(a)(6)

The Tamarin Blue® PTCA RX Dilatation Catheter uses similar technology, materials and methods of operation as the approved predicates: Maverick2 PTCA Catheter (P860019), Sprinter Legend RX (P790017/S096 & K103095) and Trek ™ RX device (K103110). The properties of the Tamarin Blue® PTCA RX Catheter to treat disease are identical to predicate PTCA balloon dilatation catheters.

Additionally, the Tamarin Blue® PTCA RX Dilatation Catheter incorporates substantially equivalent indications for use, design and dimensional and performance specifications as those found with the aforementioned predicate devices.

NONCLINICAL TESTS

SAFETY and EFFECTIVENESS

BIOCOMPATIBILITY

All materials used in the Tamarin Blue® PTCA RX Dilatation Catheter are biocompatible based on biocompatibility testing results. The device has been tested according to ISO 10993 Part 1, 2, 4, 5, 10, 11, 12, ASTM F756-00 and 21 CFR 58 (GLP regulations).

{2}------------------------------------------------

PERFORMANCE DATA

The safety and effectiveness of the Tamarin Blue® PTCA RX Dilatation Catheter has been evaluated in the following non-clinical tests:

  • · Biocompatibility
  • · Balloon compliance
  • · Balloon burst pressure
  • · Balloon fatigue (repeated balloon inflation) endurance
  • = Balloon inflation/deflation performance
  • · Bond strengths
  • · Catheter dimensions and balloon profile
  • Catheter body minimum burst strength
  • · Device preparation (guide wire and introducer compatibility)
  • Flexibility and Kink
  • · Torque
  • · Radiopacity
  • · Coating integrity
  • · Particulate evaluation

CONCLUSION

The Tamarin Blue® PTCA RX Dilatation Catheter met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, and test protocols.

The Tamarin Blue® PTCA RX Dilatation Catheter has the same intended use, similar design and technological characteristics, equivalent performance properties, identical sterilization and packaging, same mode of operation, and no new safety or effectiveness issues. Therefore, the Tamarin Blue® PTCA RX Dilatation Catheter is considered substantially equivalent to the aforementioned predicate devices.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

NOV 2 0 2012

Natec Medical Ltd c/o Judy Danielson Senior Regulatory Consultant CardioMed Device Consultants, LLC 5523 Research Park Drive, Suite 360 Baltimore, MD 21228

Re: K112735

Trade/Device Name: Tamarin Blue® PTCA RX Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Catheters, transluminal coronary angioplasty, percutaneous Regulatory Class: Class II Product Code: LOX Dated (Date on orig SE letter): October 11, 2012 Received (Date on orig SE letter): October 15, 2012

Dear Ms. Danielson:

This letter corrects our substantially equivalent letter of November 2, 2012.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Matthew G. 201 Hillebrenner """,""","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","","",

  • for Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{5}------------------------------------------------

Indications for Use Statement

510(k) Number:K112735
Device Name:Tamarin Blue® RX PTCA Dilatation Catheter
Indications for Use:The Tamarin Blue® RX PTCA Dilatation Catheters are indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

TUS. Killelu

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number KL12735

§ 870.5100 Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter.

(a)
Standard PTCA Catheter —(1)Identification. A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion.(2)
Classification. Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
Cutting/scoring PTCA Catheter —(1)Identification. A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
Classification. Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.