K Number
K103110

Validate with FDA (Live)

Date Cleared
2011-01-10

(94 days)

Product Code
Regulation Number
870.5100
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TREK™ RX, MINI TREK™ RX, TREK™ OTW, and MINI TREK™ OTW Coronary Dilatation Catheters are indicated for:
a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion;
b) balloon dilatation of a coronary artery occlusion for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction;
c) balloon dilatation of a stent after implantation (balloon models 2.0 mm - 5.0 mm only).

Device Description

The TREK RX and MINI TREK RX Coronary Dilatation Catheters are a rapid exchange co-axial design with a balloon at the distal tip. The balloon segment expands to a known diameter and length at a specific inflation pressure and has radiopaque marker(s) under the balloon to aid in positioning the balloon in a stenosis. The coaxial shaft consists of a tubular inner and outer member. The inner member permits the use of a guide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. The outer lumen provides for inflation and deflation of the balloon with contrast fluid. The proximal shaft consists of a hypotube with a hub on the proximal end, a tapered distal section ending distal to the guide wire notch junction, along with brachial and femoral markers.

The TREK OTW and MINI TREK OTW Coronary Dilatation Catheters are an over-the-wire co-axial design with a balloon at the distal tip. The balloon segment expands to a known diameter and length at a specific inflation pressure and has radiopaque marker(s) under the balloon to aid in positioning the balloon in a stenosis. The coaxial shaft consists of a tubular inner and outer member. The inner lumen permits the use of a guide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. The outer lumen provides for inflation and deflation of the balloon with contrast fluid. Along the proximal portion of the shaft are brachial and femoral markers to aid in gauging the catheter's position relative to the guiding catheter tip when introducing the catheter through the guiding catheter. An adaption arm is located at the proximal end to provide access to the inflation lumen and guide wire lumen and allows connection with an inflation device.

AI/ML Overview

The provided document describes a medical device, the TREK™ RX, MINI TREK™ RX, TREK™ OTW, and MINI TREK™ OTW Coronary Dilatation Catheters, and its substantial equivalence to predicate devices. However, the document does not contain information about an AI/algorithm-driven device or a study involving human readers and AI assistance. Therefore, it is not possible to fill out many of the requested fields related to AI performance, ground truth, and expert evaluation.

Based on the information provided regarding the Coronary Dilatation Catheters:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Catheter PreparationMet requirementsMet acceptance criteria
Catheter Profile (crossing and refolded)Met requirementsMet acceptance criteria
Balloon Inflation and DeflationMet requirementsMet acceptance criteria
Balloon Fatigue and Balloon Fatigue Within a StentMet requirementsMet acceptance criteria
Balloon Rupture and Balloon Rupture Within a StentMet requirementsMet acceptance criteria
Balloon ComplianceMet requirementsMet acceptance criteria
Catheter Bond Tensile StrengthMet requirementsMet acceptance criteria
Catheter Coating ParticulateMet requirementsMet acceptance criteria
Kink and FlexibilityMet requirementsMet acceptance criteria
TorqueMet requirementsMet acceptance criteria
RadiopacityMet requirementsMet acceptance criteria
Biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogen, complement activation)Met requirementsMet acceptance criteria

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not explicitly stated. The document mentions "The TREK RX, MINI TREK RX, TREK OTW and MINI TREK OTW Coronary Dilatation Catheters were subjected to the following in vitro bench tests..." suggesting that a representative number of units for each catheter type underwent testing.
  • Data Provenance: In vitro bench tests. The country of origin for the data is not specified, but the submission is to the FDA in the USA. The tests are "retrospective" in the sense that they are performed on manufactured devices as part of a product release or qualification process.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This device is a physical medical device, not an AI/algorithm-driven one that requires expert-established ground truth from medical images or data. The "ground truth" for these tests is based on engineering specifications and regulatory guidance.

4. Adjudication method for the test set:

  • Not applicable. Adjudication methods like 2+1 or 3+1 are used for expert consensus in diagnostic studies (often for AI devices). For physical device testing, performance is objectively measured against predetermined specifications.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI/algorithm-driven device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an AI/algorithm-driven device.

7. The type of ground truth used:

  • For the physical tests, the "ground truth" is defined by the acceptance criteria derived from engineering specifications, industry standards, and regulatory guidance documents (e.g., Guidance for Industry and FDA Staff - Class II Special Controls for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010).

8. The sample size for the training set:

  • Not applicable. This is not an AI/algorithm-driven device.

9. How the ground truth for the training set was established:

  • Not applicable. This is not an AI/algorithm-driven device.

Summary of the Study:

The study was not for an AI/algorithm. Instead, it was a series of in vitro bench tests and biocompatibility tests conducted on the TREK™ RX, MINI TREK™ RX, TREK™ OTW, and MINI TREK™ OTW Coronary Dilatation Catheters. The purpose of these tests was to demonstrate that the devices met all predetermined acceptance criteria and performed similarly to predicate devices. The acceptance criteria were established based on the requirements outlined in the "Guidance for Industry and FDA Staff - Class II Special Controls for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010" and other relevant engineering and safety standards. The study concluded that the new devices "met all acceptance criteria and performed similarly to the predicate devices," with "no new safety or effectiveness issues...raised." This allowed the FDA to determine that the devices were "substantially equivalent" to legally marketed predicate devices.

{0}------------------------------------------------

/C10 3110

JAN 1 0 2011

1

510(k) Summary

The 510(k) Summary is submitted in accordance with 21 CFR §807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990.

.

1. SUBMITTER'S NAMEAbbott Vascular Inc.
2. SUBMITTER'S ADDRESS26531 Ynez Road, Temecula, CA 92591
3. TELEPHONE(951) 914-3243
4. FAX(951) 914-0339
5. CONTACT PERSONSuzanne Redman
6. DATE PREPAREDNovember 9, 2010
7. DEVICE TRADE NAMETREK™ RX Coronary Dilatation CatheterMINI TREK™ RX Coronary Dilatation CatheterTREK™ OTW Coronary Dilatation CatheterMINI TREK™ OTW Coronary DilatationCatheter
8. DEVICE COMMON NAME• Coronary Dilatation Catheter• Percutaneous Transluminal CoronaryAngioplasty (PTCA) Catheter
9. DEVICE CLASSIFICATIONNAMEPTCA Catheter (LOX)
10. PREDICATE DEVICE NAMEVOYAGER® RX Coronary Dilatation CatheterVOYAGER® NC Coronary Dilatation CatheterVOYAGER® OTW Coronary Dilatation Catheter

{1}------------------------------------------------

11. DEVICE DESCRIPTION

TREK RX and MINI TREK RX Coronary Dilatation Catheter 11.1

The TREK RX and MINI TREK RX Coronary Dilatation Catheters are a rapid exchange co-axial design with a balloon at the distal tip. Table 1 provides a matrix of the balloon diameters and lengths available.

BalloonDiameter(mm)Balloon Length
6mm8mm12mm15mm20mm25mm30mm
MINI TREK RX
1.50XXXXX
2.00XXXXXXX
TREK RX
2.25XXXXXXX
2.50XXXXXXX
2.75XXXXXXX
3.00XXXXXXX
3.25XXXXXXX
3.50XXXXXXX
3.75XXXXXXX
4.00XXXXXXX
4.50XXXX
5.00XX

Table 1 TREK RX and MINI TREK RX Balloon Sizes

The balloon segment expands to a known diameter and length at a specific inflation pressure and has radiopaque marker(s) under the balloon to aid in positioning the balloon in a stenosis. The coaxial shaft consists of a tubular inner and outer member. The inner member permits the use of a guide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. The outer lumen provides for inflation and deflation of the balloon with contrast fluid. The proximal shaft consists of a hypotube with a hub on the proximal end, a tapered distal section ending distal to the guide wire notch junction, along with brachial and femoral markers.

TREK OTW and MINI TREK OTW Coronary Dilatation Catheter 11.2

The TREK OTW and MINI TREK OTW Coronary Dilatation Catheters are an over-thewire co-axial design with a balloon at the distal tip. Table 2 provides a matrix of the balloon diameters and lengths available.

{2}------------------------------------------------

Table 2TREK OTW and MINI TREK OTW Balloon Sizes
BalloonDiameter(mm)Balloon Length
6mm8mm12mm15mm20mm25mm30mm
MINI TREK OTW
1.50XXXXX
2.00XXXXXXX
TREK OTW
2.25XXXXXXX
2.50XXXXXXX
2.75XXXXXXX
3.00XXXXXXX
3.25XXXXXXX
3.50XXXXXXX
3.75XXXXXXX
4.00XXXXXXX
4.50XX
5.00XX

The balloon segment expands to a known diameter and length at a specific inflation pressure and has radiopaque marker(s) under the balloon to aid in positioning the balloon in a stenosis. The coaxial shaft consists of a tubular inner and outer member. The inner lumen permits the use of a guide wire to facilitate advancement of the catheter to and through the stenosis to be dilated. The outer lumen provides for inflation and deflation of the balloon with contrast fluid. Along the proximal portion of the shaft are brachial and femoral markers to aid in gauging the catheter's position relative to the guiding catheter tip when introducing the catheter through the guiding catheter. An adaption arm is located at the proximal end to provide access to the inflation lumen and guide wire lumen and allows connection with an inflation device.

12. INDICATIONS FOR USE

The TREK™ RX, MINI TREK™ RX, TREK™ OTW, and MINI TREK™ OTW are indicated for:

a) balloon dilatation of the stentotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion;

b) balloon dilatation of a coronary artery occlusion of the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction;

c) balloon dilatation of a stent after implantation (balloon models 2.0 mm - 5.0 mm only).

TECHNOLOGICAL CHARACTERISTICS 13.

{3}------------------------------------------------

Comparisons of the new and predicate devices show that the technological characteristics such as product performance, design and intended use are substantially equivalent to the current marketed predicate devices.

PERFORMANCE DATA 14.

The TREK RX, MINI TREK RX, TREK OTW and MINI TREK OTW Coronary Dilatation Catheters were subjected to the following in vitro bench tests according to the requirements of Guidance for Industry and FDA Staff - Class II Special Controls for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010:

  • ◆ Catheter Preparation
  • Catheter Profile (crossing and refolded) .
  • Balloon Inflation and Deflation �
  • Balloon Fatigue and Balloon Fatigue Within a Stent .
  • Balloon Rupture and Balloon Rupture Within a Stent ●
  • Balloon Compliance .
  • ◆ Catheter Bond Tensile Strength
  • . Catheter Coating Particulate
  • . Kink and Flexibility
  • . Torque
  • . Radiopacity

Biocompatibility testing included cyctotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogen, and complement activation.

These in vitro bench and biocompatibility tests demonstrated that the TREK RX, MINI TREK RX, TREK OTW and MINI TREK OTW Coronary Dilatation Catheters met all acceptance criteria and performed similarly to the predicate devices. No new safety or effectiveness issues were raised during the testing program and therefore, these devices may be considered substantially equivalent to the predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with three lines extending upwards, possibly representing growth or progress.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Abbott Vascular, Inc c/o Ms. Suzanne Redman Principal regulatory Affairs Associates 26531 Ynez Road Temecula, CA 92591

JAN 1 0 2011

Re: K103110

Trade/Device Name: TREK™ RX Coronary Dilatation Catheter MINI TREK™ RX Coronary Dilatation Catheter TREKTM OTW Coronary Dilatation Catheter MINI TREKTM OTW Coronary Dilatation Clatheter Regulation Number: 21 CFR 870.5100 Regulation Name: PTCA Catheters Regulatory Class: Class II (two) Product Code: LOX Dated: December 29, 2010 Received: December 30, 2010

Dear Ms. Redman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactnent date of the Medical Device Amendinents, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do trequire approval of a provisions of the Federal Food, Drug,
You may, therefore, market the douise of i You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misting and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

:

and the contraction of the comments of the comments of

{5}------------------------------------------------

Page 2 – Ms. Suzanne Redman

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

una R. Vi. Vanes

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

JAN 1 0 2011

INDICATIONS FOR USE

510(k) Number (if known):K103110
Device Names:TREKTM RX Coronary Dilatation CatheterMINI TREKTM RX Coronary Dilatation CatheterTREKTM OTW Coronary Dilatation CatheterMINI TREKTM OTW Coronary Dilatation Catheter
Indicationsfor Use:The TREKTM RX, MINI TREKTM RX, TREKTM OTW, and MINITREKTM OTW Coronary Dilatation Catheters are indicated for: a)balloon dilatation of the stenotic portion of a coronary artery orbypass graft stenosis for the purpose of improving myocardialperfusion;b) balloon dilatation of a coronary artery occlusion for the purpose ofrestoring coronary flow in patients with ST-segement elevationmyocardial infarction;c) balloon dilatation of a stent after implantation (balloon models 2.0mm - 5.0 mm only)

X Prescription Use_ (Per 21 CFR 801.109)

OR

Over-The-Counter_ (Optional Format 1-1-96)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

smmaR.L.hmes4

(Division Sign-Off) (Division Sign-On)
Division of Cardiovascular Devices Page _ of ___

510(k) Number_k 10 310(k) Number

§ 870.5100 Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter.

(a)
Standard PTCA Catheter —(1)Identification. A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion.(2)
Classification. Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
Cutting/scoring PTCA Catheter —(1)Identification. A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
Classification. Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.