Intended Use

The Terumo Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.

Device Description

The Terumo Aspiration Catheter is a dual lumen rapid exchange catheter. The guidewire lumen is used to facilitate passage of a guide wire which must not exceed 0.014" (0.36 mm) in diameter. The larger extraction lumen allows the removal of thrombus (thrombi) by use of the included aspiration syringe through the extension line. The catheter has a proximal stiff region and a distal flexible region that is coated with hydrophilic polymer which generates lubricity when wet. On the distal tip a radiopaque marker band is incorporated. The proximal end of the catheter is equipped with a standard luer adapter to facilitate the attachment of the included extension line, stopcock and syringes. The provided stylet can be inserted in the catheter to assist in the delivery of the catheter to the vascular lesion. The included flushing tool is used to flush the guide wire lumen in preparation for use. A filter basket is included for assistance in filtering the blood removed during the procedure for laboratory analysis or any thrombosis.

AI/ML Overview

The Terumo Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature. The acceptance criteria and supporting studies are outlined below.

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance Criteria (Bench Testing)Reported Device Performance
ISO 10555-1 Conformance Tests
Surface inspectionConformance to ISO 10555-1 standardsPassed
Corrosion resistanceConformance to ISO 10555-1 standardsPassed
Force at break (shaft, hub)Minimal force at break of 4N (more stringent than 3N requirement of ISO 10555-1)Passed
Freedom from leakageConformance to ISO 10555-1 standardsPassed
Shelf Life Performance Tests (t=0 vs. t=shelf life)
Catheter force at breakNo significant changes in performanceMet all specifications; no significant changes observed
Catheter lubricityNo significant changes in performanceMet all specifications; no significant changes observed
Dimensional verificationNo significant changes in performanceMet all specifications; no significant changes observed
Stylet coil bond tensile strengthNo significant changes in performanceMet all specifications; no significant changes observed
Stylet to connector tensile strengthNo significant changes in performanceMet all specifications; no significant changes observed
Extension line tensile strengthNo significant changes in performanceMet all specifications; no significant changes observed
In-House Comparative Performance Tests
Kink resistanceEquivalent or superior to predicate devices (PRONTO V3, Medtronic Export XT)Equivalent or superior to predicate devices
Aspiration rateEquivalent or superior to predicate devicesEquivalent or superior to predicate devices
Thrombus aspiration capabilityEquivalent or superior to predicate devicesEquivalent or superior to predicate devices
PushabilityEquivalent or superior to predicate devicesEquivalent or superior to predicate devices
TrackabilityEquivalent or superior to predicate devicesEquivalent or superior to predicate devices
Accessory Conformance Tests
StyletConformance to ISO 10555-1 standardsPassed
Stylet connectorConformance to ISO 594-1 & 2 standardsPassed
Aspiration syringeConformance to ISO 7886-1 standardsPassed
Aspiration syringe luerConformance to ISO 594-1 & 2 standardsPassed
Extension LineConformance to ISO 10555-1 standardsPassed
Extension Line luers and stopcockConformance to ISO 594-1 & 2 standardsPassed
Flushing toolConformance to ISO 594-1 & 2 standardsPassed
Biocompatibility Tests
SensitizationMeets the requirements of ISO 10993-10Meets the requirements
HemolysisNon-hemolytic (ASTM F756)Non-hemolytic
Dog thrombo-resistanceNon-thrombogenic (ISO 10993-4)Non-thrombogenic
Complement ActivationMeets the requirements of ISO 10993-4Meets the requirements
Ames AssayMeets the requirements of ISO 10993-3Meets requirements
Lymphoma Forward MutationNon-mutagenic (ISO 10993-3)Non-mutagenic
Bone Marrow MicronucleusNon-clastogenic (ISO 10993-3)Non-clastogenic
Rabbit PyrogenNon-pyrogenic (ISO 10993-11)Non-pyrogenic
CytotoxicityNon-cytotoxic (ISO 10993-5)Non-cytotoxic
Intracutaneous reactivityMeets the requirements of ISO 10993-10Meets requirements
Acute Systemic toxicityNon-toxic (ISO 10993-11)Non-toxic
Physicochemical profile (aged, sterile)Meets the requirements of USP 661Meets the requirements
Cytotoxicity (aged, sterile)Non-cytotoxic (ISO 10993-5)Non-cytotoxic
Hemolysis (aged, sterile)Non-hemolytic (ASTM F756)Non-hemolytic
Sterilization and Residuals
Sterility Assurance Level (SAL)10⁻⁶ (validated per ANSI/AAMI/ISO 11135)Achieved 10⁻⁶ SAL
Ethylene Oxide (EO) residuals≤ 4 mg per device (per ANSI/AAMI/ISO 10993-7)Will meet requirements (prior to use)
Ethylene Chlorohydrin (ECH) residuals≤ 9 mg per device (per ANSI/AAMI/ISO 10993-7)Will meet requirements (prior to use)
Shelf Life36-months (based on performance testing and package integrity validation)36-months

2. Sample Size Used for the Test Set and Data Provenance

The provided document describes bench testing and biocompatibility testing, not a clinical study with a "test set" in the context of patient data.

  • Bench Testing: The sample sizes for each specific bench test (e.g., force at break, aspiration rate) are not explicitly stated. However, the tests were conducted using both real-time aged and accelerated aged samples to assess performance over the specified shelf life.
  • Biocompatibility Testing: The number of samples for each biocompatibility test (e.g., Sensitization, Hemolysis, Pyrogen) is not explicitly stated. The tests were performed on non-aged (sterile) and aged (sterile) samples.
  • Data Provenance: The studies are described as "in-house bench testing" and "biocompatibility testing." This implies the data were generated in a laboratory setting, likely at the manufacturer's (Terumo Medical Corporation) or a contracted testing facility's site, and are not derived from human patients or a specific country of origin in the clinical sense. These are pre-market studies to demonstrate safety and performance characteristics.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. The reported studies are bench and biocompatibility tests, which do not involve establishing ground truth by human experts in the context of diagnostic interpretation or clinical outcomes. The "ground truth" for these tests refers to established scientific and engineering standards (e.g., ISO, ASTM).

4. Adjudication Method for the Test Set

Not applicable. As these are bench and biocompatibility tests, there is no human adjudication process involved in the data collection or interpretation in the way there would be for a clinical trial or image analysis study. The results are compared against predefined pass/fail criteria from international and industry standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The provided document details pre-market bench and biocompatibility studies for a physical medical device (aspiration catheter). It does not describe an AI/software as a medical device or diagnostic imaging study that would typically involve an MRMC comparative effectiveness study to assess human reader performance with and without AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. The Terumo Aspiration Catheter is a physical medical device, not an algorithm or software. Therefore, there is no "standalone" algorithm performance study described.

7. Type of Ground Truth Used

The "ground truth" for the various tests were defined by:

  • International Standards: ISO 10555-1, ISO 594-1 & 2, ISO 7886-1, ISO 10993 (various parts), ISO 11135, ISO 11607-1, ISO 11607-2.
  • National Standards: ASTM F756, ANSI/AAMI/ISO 11135, ANSI/AAMI/ISO 10993-7, USP 661.
  • Manufacturer Specifications: In-house bench testing compared the device's performance to predicate devices and likely internal design specifications that reflect expected performance characteristics (e.g., "equivalent or superior to predicate devices" for kink resistance, aspiration rate, etc.).

8. Sample Size for the Training Set

Not applicable. The studies described are for physical device testing and do not involve machine learning algorithms that require a "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, for the same reason as above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 9, 2021

Terumo Medical Corporation % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Saint Paul. Minnesota 55114

Re: K112382

Trade/Device Name: Terumo Aspiration Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: OEZ

Dear Mark Job:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 14, 2011. Specifically, FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075. Gregory. Oconnell(@FDA.HHS.gov.

Sincerelv.

Digitally signed by Gregory W. O'connell -S

Gregory W. O'connell -S Date: 2021.02.09 08:16:35 -05'00

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DEC 1 4 2011

Terumo Medical Corporation c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25" Street, NW Buffalo, MN 55313

Re: K112382/S1

Trade/Device Name: Terumo Aspiration Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE Dated: November 14, 2011 Received: November 15, 2011

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device we have reviewed your bocured the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosure) to regard to the Medical Device Amendments, or to conninered processified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, therefore, manot of the Act include requirements for annual registration, listing of general controls proficions of ractice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (bee above) inxisting major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may found in the Code of reactive concerning your device in the Federal Register. publish furtier announcements conceential equivalence determination does not mean Please be advised that I DA 3 issualled of a succe complies with other requirements of the Act that I DA has made a decemmation administered by other Federal agencies. You must or any rederal statutes and regulations aincluding, but not limited to: registration and listing

{2}------------------------------------------------

Page 2 - Mr. Mark Job

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing

practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR

regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, .

W. S. Allehen

S سر Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known):K112382
------------------------------------

Terumo Aspiration Catheter Device Name: Device

Indications For Use:

The Terumo Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.

Image /page/3/Picture/5 description: The image shows a document stamp. The stamp includes the text "FDA CDRH DMC" on the top line. The date "AUG 1 8 2011" is printed in the middle of the stamp. The word "Received" is printed on the bottom line.

Prescription Use __ X (Part 21 CFR 801 Subpart D) * AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M.A. Hibben

(Division Sign-Off) Division of Cardiovascular Devices

112582 510(k) Number

{4}------------------------------------------------

DEC 1 4 2011

510(k) Summary

A. Device Name

Proprietary NameTerumo Aspiration Catheter
Classification NameEmbolectomy Catheter (as per 870.5150)
Common NameEmbolectomy Catheter, Aspiration Catheter
Product CodeDXE

B. Intended Use

The Terumo Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.

The predicate devices are the PRONTO V3 Extraction Catheter (K083784) and the Medtronic Export XT Catheter (K061958).

The PRONTO V3 Extraction Catheter (K083784) is indicated for removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.

The Medtronic Export XT Catheter (K061958) is indicated for:

  • Removal/aspiration of embolic material (thrombus/debris) from vessels of the arterial . system, and
  • To subselectively infuse/deliver diagnostics or therapeutic agents with or without . vessel occlusion.

The Terumo Aspiration Catheter has the same indication for use as the Pronto V3 predicate device. Unlike the Export XT device, the Terumo Aspiration Catheter is not indicated for the infusion or delivery of diagnostic or therapeutic agents. This difference does not alter the safety or effectiveness of the Terumo Aspiration Catheter when used as indicated.

C. Device Description

The Terumo Aspiration Catheter is a dual lumen rapid exchange catheter. The guidewire lumen is used to facilitate passage of a guide wire which must not exceed 0.014" (0.36 mm) in diameter. The larger extraction lumen allows the removal of thrombus (thrombi) by use of the included aspiration syringe through the extension line. The catheter has a proximal stiff region and a distal flexible region that is coated with hydrophilic polymer which generates lubricity when wet. On the distal tip a

{5}------------------------------------------------

radiopaque marker band is incorporated. The proximal end of the catheter is equipped with a standard luer adapter to facilitate the attachment of the included extension line, stopcock and syringes. The provided stylet can be inserted in the catheter to assist in the delivery of the catheter to the vascular lesion. The included flushing tool is used to flush the guide wire lumen in preparation for use. A filter basket is included for assistance in filtering the blood removed during the procedure for laboratory analysis or any thrombosis.

D. Principle of Operation / Technology

The Terumo Aspiration Catheter is operated manually or by manual process. This is the same principle of operation of the predicate devices.

E. Design / Materials

The Terumo Aspiration Catheter submitted in this 510(k), the PRONTO V3 extraction catheter cleared under K083784 and the Medtronic Export XT Aspiration Catheter cleared under K061958 have the same principle design features:

  • Dual lumen for rapid exchange over a guidewire .
  • Wire braided tubing for enforced strength ●
  • Distal marker band for high visibility under fluoroscopy .
  • External hydrophilic coating for easy access to vascular lesion ●

All three devices have a three-layered construction comprising a stainless steel wire or mesh sandwiched in between an inner and outer elastomer layer.

Pronto V3Export XTTerumo AspirationCatheter
OuterLayerPoly-ether-block-amide,Nylon 12Poly-amide-elastomerPolyamide Elastomer,Nylon 12
BraidStainless SteelStainless SteelStainless Steel
InnerLayerPoly-olefinNot disclosedPFA (per-fluoroalkoxy)resin

The Terumo Aspiration Catheter in this submission uses similar types of materials as the predicate devices. The results of biological compatibility and bench testing show that the differences in materials between the Terumo Aspiration Catheter and the predicate devices do not raise any new issues of safety and effectiveness.

{6}------------------------------------------------

F. Specifications

The Terumo Aspiration Catheter submitted in this 510(k), the PRONTO V3 extraction catheter cleared under K083784 and the Medtronic Export XT Aspiration Catheter cleared under K061958 have similar device specifications. Differences in specifications between the devices do not raise any new issues of safety and effectiveness.

CatheterLengthMaximumO.D.CompatibleGuidingCatheter sizeCompatibleGuidewiresizeExtractionlumen area
TerumoAspirationCatheter140 cm1.70 mm6 Fr.0.014"0.85 mm2
1.90 mm7 Fr.1.28 mm2
Pronto V3140 cm1.60 mm6 Fr.0.014"0.93 mm2
Export XT140 cm1.73 mm6 Fr.0.014"0.85 mm2
1.83 mm7 Fr.1.27mm2

G. Performance

1. Conformance to ISO 10555-1

The Terumo Aspiration Catheter successfully passed the following performance tests, demonstrating conformance with ISO 10555-1:

ISO 10555-1 Compliance Testing of the Terumo Aspiration Catheter (aged, sterile)
TestTest methodsResult
Surface inspectionISO 10555-1Passed
Corrosion resistanceISO 10555-1Passed
Force at break (shaft, hub)ISO 10555-1Passed
Freedom from leakageISO 10555-1Passed

The above tests were conducted using both real-time aged and accelerated aged samples.

A minimal force at break of 4N was applied to distal marker and distal tip bonding (based on ISO10555-3). This is more stringent than 3N requirement of 10555-1. No other deviations or exclusions were taken from methods defined in ISO 10555-1.

{7}------------------------------------------------

2. Potential changes over shelf life

To evaluate the potential for performance changes over the stated shelf life, the following additional bench testing was conducted using non-aged (t=0) samples:

  • Catheter force at break .
  • Catheter lubricity .
  • Dimensional verification .
  • Stylet coil bond tensile strength .
  • Stylet to connector tensile strength 0
  • Extension line tensile strength: .

All non-aged samples met all specifications and no significant changes in performance were observed between the t=0 and the t=shelf life samples.

3. Comparison to predicates

The following in-house bench testing was conducted to demonstrate equivalent performance with the predicate devices.

In-House Comparative Performance Testing
TestMethod
Kink resistanceDimensional measurement (distance and radius)of kinked, fixed-length sample of catheter shaft.
Aspiration rateMeasurement of time required to aspirate knownvolume of viscous fluid.
Thrombus aspiration capabilityGram weight measurement of aspiratedsimulated clot.
PushabilityMeasurement of resistance encounteredtraversing a guiding catheter
TrackabilityMeasurement of resistance and distance travelledthrough anatomical (PCTA) model

Performance characteristics of the Terumo Aspiration Catheter were found to be equivalent or superior to the predicate devices.

4. Accessories

Components/accessories were also tested as indicated in the following table. Each component successfully passed all required tests. No exclusions or deviations were taken for any of the applied standards. All testing was conducted using both real-time aged and accelerated aged samples.

{8}------------------------------------------------

Conformance Testing of Accessory Components (aged, sterile)
ComponentTest MethodsResult
StyletISO 10555-1Passed
Stylet connectorISO 594-1 & 2Passed
Aspiration syringeISO 7886-1Passed
Aspiration syringe luerISO 594-1 & 2Passed
Extension LineISO 10555-1Passed
Extension Line luers and stopcockISO 594-1 & 2Passed
Flushing toolISO 594-1 & 2Passed

H. Biocompatibility and Sterilization

The Terumo Aspiration Catheter is classified as an Externally Communicating Device, Circulating Blood, Limited Contact (≤ 24h). Blood contacting materials were tested in accordance with FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing". No deviations or exclusions were taken from the methods defined in the applicable parts of ISO 10993. The Terumo Aspiration Catheter successfully passed all of the following biocompatibility tests:

Biocompatibility Testing of the Terumo Aspiration Catheter (non-aged, sterile)
TestTest methodsResult
SensitizationISO 10993-10Meets the requirements
HemolysisASTM F756Non-hemolytic
Dog thrombo-resistanceISO 10993-4Non-thrombogenic
Complement ActivationISO 10993-4Meets the requirements
Ames AssayISO 10993-3Meets requirements
Lymphoma Forward MutationISO 10993-3Non-mutagenic
Bone Marrow MicronucleusISO 10993-3Non-clastogenic
Rabbit PyrogenISO 10993-11Non-pyrogenic
CytotoxicityISO 10993-5Non-cytotoxic
Intracutaneous reactivityISO 10993-10Meets requirements
Acute Systemic toxicityISO 10993-11Non-toxic

{9}------------------------------------------------

Biocompatibility Testing on the Terumo Aspiration Catheter (aged, sterile) Result Test Test methods Meets the requirements USP 661 Physicochemical profile ISO 10993-5 Non-cytotoxic Cytotoxicity ASTM F756 Non-hemolytic Hemolysis

In addition, the following screening tests were conducted on the accelerated aged, sterile device to demonstrate that aging does not affect the device's biocompatibility:

Again, no deviations or exclusions were taken from the methods defined in the applicable standards.

Sterilization conditions have been validated according to ANSI / AAMI / ISO 11135, Sterilization of Health Care Products– Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices to provide a Sterility Assurance Level (SAL) of 10°.

Residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) residuals will meet requirements for limited exposure devices (contact up to 24 hours) prior to use, based on ANSI/AAMI/ISO 10993-7, Biological Evaluation of medical devices- Part 7: Ethylene Oxide Sterilization residuals. Residual EO will not exceed 4 mg per device and residual ECH will not exceed 9 mg per device.

The Terumo Aspiration Catheter is individually packaged in a peel package composed of polyester-polyethylene laminated film and Tyvek. Expiration dating for the Terumo Aspiration Catheter is 36-months based on performance testing in accordance with ISO 10555-1, of both real-time aged (>3years) and accelerated aged samples, and validation of package integrity in accordance with ISO11607-1 and ISO11607-2.

{10}------------------------------------------------

Terumo Clinicál Supply Co., Ltd. Premarket Notification - Terumo Aspiration Catheter Section II. 510(K) Summary

I. Substantial Equivalence

The Terumo Aspiration Catheter submitted in this 510(k) is substantially equivalent in intended use, design, technology/principles of operation, and materials to the Pronto V3 extraction catheter which was cleared under K083784 and the Medtronic Export XT Aspiration Catheter which was cleared under K061958. Differences between the devices do not raise any issues of safety or effectiveness.

J. Submitter Information

Prepared By:Mr. Daniel R. PlonskiRegulatory Affairs Specialist
Prepared For:Terumo Medical Corporation950 Elkton Blvd.Elkton, MD 21921Phone: (410) 392-7395Fax: (410) 398-6079Email: daniel.plonski@terumomedical.com
Date Prepared:December 12, 2011

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).