K Number
K112302

Validate with FDA (Live)

Date Cleared
2011-12-21

(132 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Proclear Multifocal Toric (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with nondiseased eyes that are myopic or hyperopic which, possess astigmatism to -5.75 diopters or less, and are presbyopic.

Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with nondiseased eyes that are myopic or hyperopic and are presbyopic. The lenses may be worn by persons who exhibit astigmatism of 0.75 diopters or less that does not interfere with visual acuity.

Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic which possess astigmatism corrections to -5.75 diopters.

Proclear Sphere and Asphere: (omafilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in non-aphakic persons with nondiseased eyes that are myopic or hyperopic and exhibit astigmatism of 2.00D or less that does not interfere with visual acuity.

Proclear Toric: (omafilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in non-aphakic persons with non-diseased eyes that are myopic or hyperopic The lens may be worn by persons who have astigmatism of 5.00D or less.

Proclear Multifocal: (omafilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, and astigmatism) and presbyopia in non-aphakic persons with non-diseased eyes. The lens may be worn by persons who have astigmatism of 2.00D or less that does not interfere with visual acuity.

Proclear (omafilcon A) Soft (hydrophilic) Contact Lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only).

Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients.

Device Description

The Proclear lens is composed of polymer of 2-hydroxy-ethylmethacrylate and 2-metacryloloyoxyethyl phosphorylcholine cross linked with ethylene glycol dimethacrylate. The lenses are tinted blue from edge to edge for visibility purposes. The Proclear (omafilcon A) Soft (hydrophilic) contact lenses are a hemispherical shell. The physical properties and available dimensions are unchanged from predicate 510ks.

Proclear Toric and Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are back surface toric.

Proclear Multifocal and Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are available as a multifocal lens with an aspherical front surface and spherical back surface for the correction of visual acuity in presbyopic persons who are myopic or hyperopic. The multifocal lens has two multifocal zones, as well as the edge shape being optimized to provide comfort without sacrificing tensile strength.

Proclear Multifocal Toric (omafilcon A) Soft (Hydrophilic) Contact Lens front surface is aspherical, with the anterior surface having a toric generated surface for the purpose of correcting vision in an eve that is astigmatic.

Proclear Sphere and Asphere: (omafilcon A) Soft (hydrophilic) Contact Lenses. The sphere lenses have spherical optical zone and asphere lens have an aspherical optical zone. This aspheric optical zone design (front surface) controls and limits the amount of longitudinal spherical aberration generated by the lens across the power range

AI/ML Overview

The provided document describes a 510(k) premarket notification for CooperVision's Proclear (omafilcon A) Soft Contact Lenses. This submission aims to demonstrate substantial equivalence to previously cleared predicate devices based on modifications to the hydration and inspection processes.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

Tested Characteristics SummaryAcceptance CriteriaResults
Total Extractables, Water content, Dk, Light Transmittance, refractive index tested per ISO 18369-4:2006Equivalent to predicate lens and meet tolerances in ISO 18369-2:2006Pass
Cytotoxicity Test ISO 10993-5:1999All 3 monolayers exposed to the test article show no grade greater than grade 2 (reactivity mild)Pass
ISO Ocular Irritation ISO 10993-10:2002Test extract shows no significant irritation over the reagent control during the observation period.Pass
Systemic Toxicity Study ISO 10993-11:1996During the observation period, none of the animals treated with the individual test extract exhibit a significantly greater reaction than the control animals.Pass
ISO Contact Lens 22 Day Ocular Irritation Study ISO 9394Scores from macroscopic and microscopic ocular examinations equivalent between test and control eyes.Pass

Study to Prove Device Meets Acceptance Criteria:

The document clearly states that no clinical data is required because the "technical characteristics, formulation, manufacturing, and sterilization processes of this lens are not changing and are equivalent to omafilcon A soft contact lenses currently marketed by CooperVision."

Instead, the submission relies on non-clinical tests to demonstrate that the modifications to hydration and inspection processes do not alter the lens's safety and performance characteristics.

Here's a breakdown of the studies conducted:

  • Physicochemical Studies:

    • Study Design: These studies tested properties such as Total Extractables, Water content, Dk (oxygen permeability), Light Transmittance, and Refractive Index according to ISO 18369-4:2006.
    • Acceptance Criteria: The results needed to be equivalent to the predicate lens and meet tolerances specified in ISO 18369-2:2006.
    • Results: The device "Pass"ed all these tests, indicating "no change to physicochemical properties of the lenses."
  • Toxicology Studies:

    • Study Design: A series of in-vitro and in-vivo preclinical toxicology and biocompatibility tests were performed. These include:
      • Cytotoxicity Test (ISO 10993-5:1999)
      • ISO Ocular Irritation (ISO 10993-10:2002)
      • Systemic Toxicity Study (ISO 10993-11:1996)
      • ISO Contact Lens 22 Day Ocular Irritation Study (ISO 9394)
    • Data Provenance: All tests were conducted in accordance with GLP regulation (21 CFR Part 56) or according to valid scientific protocols. The studies were performed by "Contract laboratories under Good Manufacturing Practice regulations" and "CooperVision" for chemistry, shelf-life, and leachability.
    • Acceptance Criteria & Results: All tests "Pass"ed, confirming that the modified lens "remains non-toxic and biocompatible with the ocular environment."
  • Material Comparison:

    • A detailed table (Table 1) compares various material characteristics between the "Subject Device" and "Predicate Device." These characteristics include Product name, Material USAN Name, FDA Category, Manufacturing method, Curing, Sterilization, Packaging, Visibility tint, Package Saline buffers and surfactant, Water Content, Refractive Index, Oxygen Permeability, Light Transmission, Base Curve, Diameter, and Power.
    • Acceptance Criteria: The conclusion drawn is that "all the material characteristics for the omafilcon A lens are equivalent between the predicate lens and the subject lens."

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: The document does not specify a numerical sample size for the "test set" in terms of number of lenses or subjects. However, the studies mentioned are laboratory-based (physicochemical, in-vitro, in-vivo animal studies).
  • Data Provenance: The studies were performed by "Contract laboratories under Good Manufacturing Practice regulations" and "CooperVision." They were conducted in accordance with GLP regulation (21 CFR Part 56) or valid scientific protocols. The country of origin for the data is not explicitly stated, but the submission is to the FDA (USA). These are prospective laboratory and animal studies, not retrospective human data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This submission relies on non-clinical laboratory and animal studies, not expert consensus on medical images or clinical outcomes with human subjects. The "ground truth" (or acceptance criteria) for these tests are defined by international standards (e.g., ISO, GLP regulations) and direct measurement of physical/chemical properties.

4. Adjudication method for the test set

Not applicable. There was no human adjudication process involved as these were laboratory and animal studies with objective measurements against predefined standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a 510(k) for contact lenses, not an AI/CADe device. Therefore, no MRMC study or AI assistance evaluation was performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is a 510(k) for contact lenses, not an algorithm.

7. The type of ground truth used

The ground truth used for the non-clinical studies was based on:

  • International Standards: ISO 18369-4:2006, ISO 18369-2:2006, ISO 10993-5:1999, ISO 10993-10:2002, ISO 10993-11:1996, ISO 9394.
  • Predicate Device Equivalence: Direct comparison of material characteristics and performance to the previously cleared predicate lenses.
  • Objective Measurements: Physicochemical properties (e.g., water content, Dk, refractive index) and biological responses (e.g., cytotoxicity grade, irritation scores) measured in laboratory settings using established protocols.

8. The sample size for the training set

Not applicable. This is not an AI/machine learning device. The term "training set" doesn't apply here.

9. How the ground truth for the training set was established

Not applicable. As above, this is not an AI/machine learning device.

{0}------------------------------------------------

DEC 2 1 2011

Image /page/0/Picture/2 description: The image shows the CooperVision logo. The logo consists of the word "CooperVision" in a bold, sans-serif font. To the left of the word is a stylized "C" symbol, and to the right is a black square with a white swirl inside. The logo is simple and modern, and the black and white color scheme gives it a clean and professional look.

510(k) Summary K112302

1.Applicant Name andAddressCooperVision, Inc.6150 Stoneridge Mall DriveSuite 370Pleasanton CA 94588
2. ContactLisa HahnGlobal Regulatory Affairs585-385-6810 ext 4221Ihahn@coopervision.com
3. Date PreparedOctober 17, 2011
4. Device IdentificationTrade Name:Proclear
Common Name:Scheduled Replacement Soft ContactLenses for Daily wear or DailyDisposable Soft Contact Lenses
ClassificationName:Daily Wear, Soft (hydrophilic) ContactLenses
DeviceClassification:Class II (21 CFR 886.5925)
FDA MaterialClass:FDA Group II Non-Ionic High WaterContent
Product Code:LPL and MVN

5. Device Description

The Proclear lens is composed of polymer of 2-hydroxy-ethylmethacrylate and 2-metacryloloyoxyethyl phosphorylcholine cross linked with ethylene glycol dimethacrylate. The lenses are tinted blue from edge to edge for visibility purposes. The Proclear (omafilcon A) Soft (hydrophilic) contact lenses are a hemispherical shell. The physical properties and available dimensions are unchanged from predicate 510ks.

Proclear Toric and Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are back surface toric.

Proclear Multifocal and Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are available as a multifocal lens with an aspherical front surface and spherical back surface for the correction of

CooperVision 510(k) Summary - Proclear omafilcon A

{1}------------------------------------------------

visual acuity in presbyopic persons who are myopic or hyperopic. The multifocal lens has two multifocal zones, as well as the edge shape being optimized to provide comfort without sacrificing tensile strength.

Proclear Multifocal Toric (omafilcon A) Soft (Hydrophilic) Contact Lens front surface is aspherical, with the anterior surface having a toric generated surface for the purpose of correcting vision in an eve that is astigmatic.

Proclear Sphere and Asphere: (omafilcon A) Soft (hydrophilic) Contact Lenses. The sphere lenses have spherical optical zone and asphere lens have an aspherical optical zone. This aspheric optical zone design (front surface) controls and limits the amount of longitudinal spherical aberration generated by the lens across the power range

6. Intended Use

Proclear Multifocal Toric (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with nondiseased eyes that are myopic or hyperopic which, possess astigmatism to -5.75 diopters or less, and are presbyopic.

Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with nondiseased eyes that are myopic or hyperopic and are presbyopic. The lenses may be worn by persons who exhibit astigmatism of 0.75 diopters or less that does not interfere with visual acuity.

Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic which possess astigmatism corrections to -5.75 diopters.

Proclear Sphere and Asphere: (omafilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in non-aphakic persons with nondiseased eyes that are myopic or hyperopic and exhibit astigmatism of 2.00D or less that does not interfere with visual acuity.

Proclear Toric: (omafilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in non-aphakic persons with non-diseased eyes that are myopic or hyperopic The lens may be worn by persons who have astigmatism of 5.00D or less.

Proclear Multifocal: (omafilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, and astigmatism) and presbyopia in non-aphakic persons with non-diseased eyes. The lens may be worn by persons who have astigmatism of 2.00D or less that does not interfere with visual acuity.

Proclear (omafilcon A) Soft (hydrophilic) Contact Lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only).

Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients.

{2}------------------------------------------------

Page 3 of 6

FREQUENT PLANNED REPLACEMENT WEAR

When prescribed for frequent planned replacement wear the lenses are to be cleaned, rinsed and disinfected each time they are removed from the patients' eye and discarded after the recommended wearing period prescribed by the eye care practitioner. The lenses may be disinfected using a chemical disinfection system

DISPOSABLE WEAR

When prescribed for Dally Disposable Wear the lens is to be discarded after each removal

7. Predicate Device(s)

Proclear Multifocal, Toric and Multifocal Toric (Omafilcon A) Soft Contact Lenses (K110099)

Proclear Sphere, Asphere, Toric, Multifocal and Multifocal Toric (Omafilcon A) Soft Contact Lenses (K061948)

8. Characteristics of Substantial Equivalence

The soft contact lenses have the following similarities to the predicate lenses which previously received 510(k) concurrence:

  • have the same indicated use. .
  • incorporate the same design, .
  • incorporate the same materials.
  • . have the same shelf life, and
  • . are packaged and sterilized using the same materials and processes.

The modifications to the hydration and inspection do not change the lens specifications or characteristics.

In summary, the omafilcon A soft contact lenses described in this submission are, substantially equivalent to the predicate device.

9. Physiochemical Studies

Results from physical, optical and chemical properties were performed and indicate no change to physicochemical properties of the lenses. Change does not affect physicochemical properties of the lenses.

Tested Characteristics SummaryAcceptance CriteriaResults
Total Extractables , Water content, Dk,Light Transmittance, refractive indextested per ISO 18369-4:2006Ophthalmic optics - Contact lenses -Part 4: Physicochemical properties ofcontact lens materials: section 4.2Equivalent to predicatelens and meettolerances in ISO18369-2:2006Pass

{3}------------------------------------------------

Material Comparison
CharacteristicSubject DevicePredicate Device
Product nameProclear ToricProclear Toric XRProclear MultifocalProclear Multifocal XRProclear Multifocal ToricProclear SphereProclear AsphereProclear ToricProclear Toric XRProclear MultifocalProclear Multifocal XRProclear Multifocal ToricProclear SphereProclear Asphere
Material USAN NameOmafilcon AOmafilcon A
510(k) numberThis submissionK110099 for material
FDA Category (Group)Group II Non-IonicHigh WaterGroup II Non-IonicHigh Water
Manufacturing methodFinished InsidePolymerization System IIFinished InsidePolymerization System II
CuringThermal CureThermal Cure
SterilizationMoist Heat (steam) invalidated AutoclaveMoist Heat (steam) invalidated Autoclave
PackagingBlister PackBlister Pack
Visibility tintVAT Blue 6VAT Blue 6
Package Saline buffersand surfactantPhosphate buffersPEG200 and Tween 80Phosphate buffersTween 80
Water Content59% ± 2%59% ± 2%
Refractive Index1.395 ± 0.0051.395 ± 0.005
Oxygen Permeability x 101121.0521.05
Light Transmission>90%>90%
Base Curve8.0 to 9.3 mm8.0 to 9.3 mm
Diameter13.6 to 15.2 mm13.6 to 15.2 mm
Power-20.00 to +20.00-20.00 to +20.00

Table 1: Material characteristics of the omafilcon A lens before and after modification.

Comparison Conclusion: all the material characteristics for the omafilcon A lens are equivalent between the predicate lens and the subject lens is omafilcon A made using FIPSII process with modified wet process).

10. Toxicology Studies

Results from in-vivo and in-vitro studies were performed as support for this modification and indicate no issues of toxicity or biocompatibility with the modified lens. Change will not affect lenses ability to remain non-toxic and biocompatible with the ocular environment.

A series of in-vitro and in-vivo preclincal toxicology and biocompatibility tests were performed to assess the safety and effectiveness of the contact lens. All tests were conducted in accordance with the GLP regulation (21 CFR Part 56) or according to valid scientific protocols.

CooperVision 510(k) Summary - Proclear omafilcon A

{4}------------------------------------------------

TestAcceptance CriteriaResult
Cytotoxicity Test ISO 10993 - 5:1999: Biological Evaluation ofMedical Devices - Part 5: Testsfor In Vitro CytotoxicityAll 3 monolayers exposedto the test article show nograde greater than grade 2(reactivity mild)Pass
ISO Ocular IrritationISO 10993 - 10:2002: BiologicalEvaluation of Medical Devices -Part 10: Tests for Irritation andDelayed HypersensitivityTest extract shows nosignificant irritation over thereagent control during theobservation period.Pass
Systemic Toxicity StudyISO 10993 - 11: 1996: BiologicalEvaluation of Medical Devices -Part 11: Tests for SystemicToxicity.During the observationperiod, none of the animalstreated with the individualtest extract exhibit asignificantly greaterreaction that the controlanimals.Pass
ISO Contact Lens 22 Day OcularIrritation StudyISO 9394 Ophthalmic Optics -Contact Lenses and Contact ·Lens Care Products -The determination ofBiocompatibility by Ocular Studywith Rabbit Eyes.Scores from macroscopicand microscopic ocularexaminations equivalentbetween test and controleyes.Pass

11. Conclusions of Non-Clinical Tests Performed:

Physiochemical: .

The physical, optical and chemical properties of this lens remain unchanged from the unmodified device, and are within established specifications for the lenses.

Toxicology: ■

Results from in-vivo and in-vitro studies conducted verify that the lenses remain non-toxic and are biocompatible with the ocular environment.

12. Clinical Studies

The technical characteristics, formulation, manufacturing, and sterilization processes of this lens are not changing and are equivalent to omafilcon A soft contact lenses currently marketed by CooperVision, therefore no clinical data is required.

Conclusion Drawn from Studies:

Validity ofContract laboratories under Good Manufacturing Practice regulations
Scientific Dataconducted toxicological and microbiology studies.
Chemistry, shelf-life and leachability studies were conducted by

CooperVision 510(k) Summary - Proclear omafilcon A

{5}------------------------------------------------

CooperVision and followed scientific protocols. The data were determined to be scientifically valid under 21 CFR 860.7

Information presented in this Premarket Notification establishes that the Substantial CooperVision (omafilcon A) Proclear Toric XR, Proclear Multifocal and Equivalence Proclear Multifocal XR contact lens is as safe and effective as the predicate devices when used in accordance with the labeled directions for use and for the requested indications.

Risk and The risks of the subject lens are the same as those normally attributed to Benefits the wearing of soft (hydrophilic) contact lenses on a daily wear basis. The benefits to the patient are the same as those for other soft (hydrophilic) contact lenses.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three overlapping profiles forming the body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DEC 2 1 2011

CooperVision, Inc. C/O Lisa Hahn Global Regulatory Affairs 6150 Stoneridge Mall Road Pleasanton, CA 94588

Re: K112302

Trade/Device Name: Proclear (omafilcon A) Soft Contact Lenses Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL and MVN Dated: November 17, 2011 Received: November 18, 2011

Dear Ms. Hahn:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your becermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encroours) (c regirly and ment date of the Medical Device Amendments, or to connine.co province may 20, 1978, the eccordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, merelore, mainet the act include requirements for annual registration, listing of general volulors provision practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability addition. Thease noter u, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean 1 that FDA has made a determination that your device complies with other requirements of the Act that I DA has intatutes and regulations administered by other Federal agencies. You must or any I each other of the Act's requirements, including, but not limited to: registration and listing (21

{7}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical CFK Part 807), labeling (21 CFR 803); good manufacturing practice requirements as set device-related adverse events) (21 CFR Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your dovices/CDRH/CDRHOffices/ucm115809.htm for go to midth www.lua.gowritoour Drical Health's (CDRH's) Office of Compliance. Also, please the Center for Devices and Radions by reference to premarket notification" (21CFR Part note and regulations regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may obtain of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Debra Falls

Image /page/7/Picture/7 description: The image shows a signature and the words "Mal" and "Dire". The signature is illegible, but it appears to be written in black ink. The words "Mal" and "Dire" are written in a simple, sans-serif font. The image is cropped, so it is not possible to see the full context of the signature and words.

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices

Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use

510(k) Number (if known): K112302

Device Name: Proclear (omafilcon A) Soft (hydrophilic) Contact Lenses

Indications for Use:

Proclear Multifocal Toric (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with nondiseased eyes that are myopic or hyperopic which, possess astigmatism to -5.75 diopters or less, and are presbyopic.

Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with nondiseased eyes that are myopic or hyperopic and are presbyopic. The lenses may be worn by persons who exhibit astigmatism of 0.75 diopters or less that does not interfere with visual acuity.

Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic which possess astigmatism corrections to -5.75 diopters.

Proclear Sphere and Asphere: (omafilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in non-aphakic persons with non-diseased eyes that are myopic or hyperopic and exhibit astigmatism of 2.00D or less that does not interfere with visual acuity.

Proclear Toric: (omafilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in non-aphakic persons with non-diseased eyes that are myopic or hyperopic The lens may be worn by persons who have astigmatism of 5.00D or less.

Proclear Multifocal: (omafilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, and astigmatism) and presbyopia in non-aphakic persons with non-diseased eyes. The lens may be worn by persons who have astigmatism of 2.00D or less that does not interfere with visual acuity.

Proclear (omafilcon A) Soft (hydrophilic) Contact Lenses may provide improved comfort for I round. (only arers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only).

Page 1 of 2

{9}------------------------------------------------

Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients.

FREQUENT PLANNED REPLACEMENT WEAR

When prescribed for frequent planned replacement wear the lenses are to be cleaned, rinsed and disinfected each time they are removed from the patients' eye and discarded after the recommended wearing period prescribed by the eye care practitioner. The lenses may be disinfected using a chemical disinfection system

DISPOSABLE WEAR

When prescribed for Dally Disposable Wear the lens is to be discarded after each removal

Over-The-Counter Use Prescription Use X__ AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PI.EASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

J. Angelo Green

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K112302

Page 2 of 2

N/A