Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K180985
    Date Cleared
    2018-06-08

    (56 days)

    Product Code
    Regulation Number
    886.5925
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Aveo (omafilcon A) Aspheric Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in not aphakic persons with non-diseased eyes that are myopic or hyperopic and exhibit astigmatism of 1.00D or less that does not interfere with visual acuity.

    Aveo ( omafilcon A) Toric Soft ( Hydrophilic) Contact Lenses are indicated the daily wear for the correction of visual acuity in not aphakic persons with non-diseased eyes that are myopic or hyperopic and exhibit astigmatism of between 0.50D and 2.50D or less that does not interfere with visual acuity.

    Daily wear replacement schedules may vary from patient to patient and should be decided by eyecare practitioners in consultation with their patients.

    Frequent Planned Replacement Wear
    When prescribed for frequent planned replacement wear the lenses are to be cleaned, rinsed and disinfected each time they are removed from the patients' eve and discarded after the recommended wearing period prescribed by the eye care practitioner. The lenses may be disinfected using a chemical disinfection system.

    Disposable Wear
    When prescribed for Daily Disposable Wear the lens is to be discarded after each removal.

    Device Description

    The Aveo (Omafilcon A) Soft (Hydrophilic) Contact Lenses are daily wear soft contact lenses produced from the HEMA-MPC copolymer material. MPC is similar to phospholipids (e.g., phosphatidylcholine) where molecules are found naturally in human cell membranes that improves the lens biocompatibility.

    The contact lenses contain 58% water by weight and are sold in the blister package immersed in phosphate buffered packaging saline. RB 246 pigment conforms to 21 CFR Part 73.3106 is used to provide the handling blue tint for the lens. Reactive ultraviolet absorber RUVA-93 is used to block UV radiation.

    The device is a corneal contact lens having a total diameter less than the visible iris diameter and is designed to be worn in its entirety on the cornea.

    Aveo (Omafilcon A) Aspheric Soft (Hydrophilic) Contact Lenses have an aspheric front curve (external curve) which is tri-curve and spherical base curve (internal curve).

    Aveo (Omafilcon A) Toric Soft (Hydrophilic) Contact Lenses have an aspheric front curve (external curve) which is tri-curve for the sphere power and multi-radius base curve for the cylinder power (internal curve).

    The contact lenses are hydrophilic, soft and it is supplied in sterile state.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device, specifically contact lenses. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving a device meets specific acceptance criteria based on performance metrics like accuracy, sensitivity, or specificity (as would be typical for an AI/ML medical device submission).

    Therefore, the requested information about "acceptance criteria and the study that proves the device meets the acceptance criteria" in terms of performance metrics (like those for an AI/ML diagnostic) is not directly applicable to this document. This submission primarily relies on showing the new device is fundamentally similar to an already approved one.

    However, I can extract the information relevant to the substantial equivalence claim, which acts as the "proof" for this type of device.

    Here's an interpretation based on the document's content:

    1. Table of "Acceptance Criteria" (for Substantial Equivalence) and Reported Device Performance:

    For a 510(k) of a contact lens, the "acceptance criteria" are essentially demonstrating similarity to a predicate device across various characteristics, and the "performance" is the subject device exhibiting these similar characteristics.

    CharacteristicPredicate Device (Proclear Asphere)Subject Device (Aveo)"Acceptance Criteria" (Similarity)Reported Device Performance (Aveo)
    Intended UseSimilar scopeSimilar scopeMetDaily wear for vision correction in non-aphakic, non-diseased eyes with myopia/hyperopia and limited astigmatism.
    Material NameOmafilcon AOmafilcon AMetOmafilcon A
    FDA Group (Material)Group II (Non-ionic, High water)Group II (Non-ionic, High water)MetGroup II (Non-ionic, High water)
    ModalityDaily WearDaily WearMetDaily Wear
    Lens DesignAsphericalAsphericalMetAspherical
    Manufacturing MethodFinished Inside Polymerization System IICast MoldedMinor difference, justified.Cast Molded
    CuringThermal CureThermal CureMetThermal Cure
    SterilizationMoist Heat (Steam)Moist Heat (Steam)MetMoist Heat (Steam)
    PackagingBlister PackBlister PackMetBlister Pack
    Visibility TintVAT Blue 6Reactive Blue Dye 246Minor difference, justified.Reactive Blue Dye 246
    Water Content59% ± 2%59% ± 2%Met59% ± 2%
    Package SalinePhosphate Buffers, PEG200, Tween 80Phosphate Buffered SalineMinor difference, justified.Phosphate Buffered Saline
    Refractive Index1.395 ± 0.0051.4002Within range/similar.1.4002
    Oxygen Permeability (Dk)21.0525.68Higher Dk is generally an improvement, not a concern for safety.25.68
    Light Transmission>90%98%Met/Exceeded98%
    Base Curve8.0mm to 9.3mm8.4mm to 8.8mmOverlapping range.8.4mm to 8.8mm
    Diameter Øт13.6mm to 15.2mm14.0mm to 14.7mmOverlapping range.14.0mm to 14.7mm
    Power-20.00 to +20.00-10.00 to +6.00Within a clinically accepted range for specified indications.-10.00 to +6.00

    Note: The "acceptance criteria" for a 510(k) submission are not rigid pass/fail thresholds like for a performance study, but rather demonstrating that the new device is as safe and effective as a legally marketed predicate device. Differences are acceptable if they do not raise new questions of safety or effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set (Non-Clinical Performance Data): The document refers to "Physiochemical Studies" and "Toxicology Studies."
      • Physiochemical Studies: Conducted according to ISO 18369-4:2006 and ISO 18369-3:2006. The sample sizes are not explicitly stated but would adhere to the standards' requirements (e.g., number of lenses tested for Dk, water content, etc.).
      • Toxicology Studies:
        • Cytotoxicity Test: Sample size not specified, typically in-vitro (cell cultures).
        • Ocular Irritation Study: "New Zealand White Rabbit" - not specified, but typically a small number of animals (e.g., 3-6).
        • Skin Sensitization Study: "Guinea Pigs" - not specified, typically a small number of animals (e.g., 10-20).
        • Acute Systemic Toxicity Study: "Swiss Albino Mice" - not specified, typically a small number of animals (e.g., 5-10 per group).
    • Data Provenance: Not explicitly stated (e.g., country of origin for the non-clinical tests). These are prospective tests conducted specifically for the device submission.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Not applicable in the context of this 510(k) submission. Ground truth for an AI/ML device (e.g., presence/absence of disease based on expert consensus) is not relevant here. The "ground truth" for contact lens substantial equivalence relies on established international standards (ISO) for physical and chemical properties, and standard toxicology testing protocols.

    4. Adjudication Method for the Test Set:

    • Not applicable. This isn't a human-reader or AI performance study requiring adjudication. The results of the physiochemical and toxicology tests are directly measured/observed by the testing facility.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. This type of study (comparing human readers with and without AI assistance) is not conducted for contact lens submissions. The submission states, "no clinical studies were required to demonstrate the safety or effectiveness of the subject device" due to its substantial equivalence to the predicate.

    6. Standalone (Algorithm Only) Performance:

    • Not applicable. This submission is for a physical medical device (contact lenses), not an algorithm or AI.

    7. Type of Ground Truth Used:

    • Standardized Test Results and Predicate Device Characteristics. The "ground truth" for this 510(k) relies on:
      • Results from physiochemical tests (e.g., water content, Dk, refractive index) measured according to ISO standards.
      • Results from toxicology tests (cytotoxicity, irritation, sensitization, systemic toxicity) measured according to ISO 10993 standards.
      • The established characteristics and safety profile of the legally marketed predicate device (Proclear, K112302, and also Aveo K162223 as another similar device).

    8. Sample Size for the Training Set:

    • Not applicable. Contact lenses are physical medical devices, not AI/ML algorithms that require training sets. Their manufacturing process is well-defined, and their properties are measured after production.

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable. As there's no training set for an AI/ML algorithm, this question doesn't apply.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1