K Number
K111970

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-02-23

(227 days)

Product Code
Regulation Number
888.3540
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The replacement of the patellofemoral joint with the KNEETEC PFJ device is indicated in the case of symptoms and serious signs affecting daily activities involving the joint:

  • Serious degenerative arthritis of the patellofemoral joint,
  • Failure of conservative procedures (realignment, arthroscopy, transfer of the tibial tuberosity) with the persistence of joint pain or dysfunction,
  • Patellofemoral dislocation (dysplastic or other origin),
  • Post-traumatic arthritis (patellar fracture),

The KNEETEC PFJ patellofemoral prosthesis is intended for cemented use only.

When used in association with the KNEETEC PFJ prosthesis, the patellar component of the HLS KNEETEC has the above mentioned indications for use.

Device Description

The Kneetee PFJ is a semiconstrained resecting prosthesis intended for replacement of the patellofemoral joint. The prosthesis consists of two components, both for cemented use only: a femoral implant and a patella. The patella of the knee joint may be resurfaced with the HLS KNEETEC patellar component and a substantially equivalent in K094013) or KNEETEC PFJ patellar implant.

AI/ML Overview

This document is a 510(k) premarket notification for the KNEETEC PFJ and HLS KNEETEC patellar component, which are patellofemoral knee prostheses. A 510(k) submission demonstrates that a new device is substantially equivalent to a legally marketed predicate device. This process does not typically involve clinical studies with human subjects or AI performance evaluations as described in your request. Instead, it relies on demonstrating equivalence through non-clinical (engineering) testing and comparison of technological characteristics.

Therefore, many of the requested categories are not applicable to this type of submission.

Here's an analysis based on the provided text:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Demonstrated Equivalence to Predicate Devices)Reported Device Performance (KNEETEC PFJ)
Non-clinical Testing:
Contact area and trajectories of the patella to compare congruence with the trochlea.No significant difference observed compared to predicate devices.
Contact pressure test.No significant difference observed compared to predicate devices.
Measurement of the subluxation force.No significant difference observed compared to predicate devices.
Material Equivalence:
Femoral component material (CoCr alloy).CoCr alloy (ISO 5832-7 or 5832-12), equivalent to predicate devices.
Patellar component material (UHMWPE).UHMWPE (ISO 5834-2), equivalent to predicate devices.
Method of Fixation:Cemented, equivalent to predicate devices.
Intended Use:Replacement of the patellofemoral joint, similar to predicate devices.
Terminal Sterilization:Gamma, similar to HLS Kneetec (predicate).
General Features & Technological Characteristics Equivalence:Equivalent to predicate devices.
Range of Motion Equivalence:Equivalent to predicate devices.
Prosthetic Dimensions Equivalence:Equivalent to predicate devices.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample size for test set: Not applicable in the context of a 510(k) for this type of device. The "test set" here refers to the actual physical devices undergoing non-clinical engineering tests (e.g., specific components for contact area, pressure, subluxation force). The document states "The following non-clinical testing has been performed on the smallest components of the ranges (worst case)." This implies a limited number of physical prototypes or manufactured components were tested, focusing on the worst-case scenario (smallest size).
  • Data provenance: Not explicitly stated, but these were engineering studies performed by the manufacturer, Tornier (based in France for regulatory affairs contact, and USA for applicant). This is not data from human patients.
  • Retrospective or prospective: Not applicable, as this refers to engineering tests, not clinical data collection.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. Ground truth as typically understood for AI devices (e.g., expert consensus on medical images) is not relevant here. The "truth" in this context is established by engineering standards and measurements.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are typically for clinical studies involving human interpretation or diagnoses, not for engineering performance tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a physical knee prosthesis, not an AI or imaging device that would involve human readers or AI assistance in diagnosis.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device (implant), not an algorithm or software.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the non-clinical testing was based on engineering standards and direct physical measurements. The performance of the KNEETEC PFJ was compared against the measured performance of predicate devices based on parameters like contact area, contact pressure, and subluxation force.

8. The sample size for the training set

Not applicable. This is not an AI/machine learning device.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/machine learning device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the text 'K111970' and 'Page 1' at the top. Below that, the word 'TORNIE' is written in large, bold, sans-serif font. The text appears to be part of a document or label.

Summary of Safety and Effectiveness information

510(k) Premarket Notification - KNEETEC PFJ

Date prepared: November 4th, 2011

Regulatory authority: Safe Medical Devices Act of 1990, 21 CRF 807.92

1) Device name

KNEETEC PFJ & HLS KNEETEC patellar component Trade name: Common name: Patellofemoral knee prosthesis Classification name: 888.3540 Knee joint patellofemoral polymer/metal semi-constrained cemented prosthesis

2) Submitter

Tornier Rue Doyen Gosse 38330 Saint Ismier - France

3) Appliquant

Tornier 7701 France avenue South Suite 600 Edina MN 55345 - USA

4) Company contact

Tomier Mrs Stephanie BERNARD Regulatory Affairs Specialist 161, rue Lavoisier - Montbonnot 38334 Saint Ismier Cedex - France Tel: + 33 (0)4 76 61 35 00 Fax: + 33 (0)4 76 61 35 59 e-mail: stephanie.bernard(@)tornier.fr

5) Classification

Device class:Class II
Classification panel:Orthopedic
Product code:87 KRR

6) Equivalent / Predicate device

Avon Patello-Femoral Joint Prosthesis, Howmedica Osteonics Corp, K010100, K020841, K041160, K051948 Patello-Femoral Knee Implant, Journey, Smith & Nephew, Inc, K051086 HLS Kneetec, Tornier, K094013

Image of a logo with the letters T and P inside a hexagon.

Tél. : 33 (0)4 76 61 35 00 TORNIER S.A.S. Fax : 33 (0)4 76 61 35 33 161, rue Lavoisier - Montbonnot 38334 SAINT-ISMIER CEDEX FRANCE

Section 5 - Page 1/ page 3

S.A.S. au capital de 288 000 € SIRET : 070 501 275 000 13 R.C.S. : 070 501 275 CODE APE : 331 B

SIEGE SOCIAL : rue du Doyen Gosse - 38330 SAINT-ISMIER - FRANCE

{1}------------------------------------------------

K11/1970 Page

TORNIER

7) Device description

The Kneetee PFJ is a semiconstrained resecting prosthesis intended for replacement of the patellofemoral joint. The prosthesis consists of two components, both for cemented use only: a femoral implant and a patella. The patella of the knee joint may be resurfaced with the HLS KNEETEC patellar component and a substantially equivalent in K094013) or KNEETEC PFJ patellar implant.

8) Materials

The femoral implant is manufactured from chromium cobalt alloy (CoCr) according to ISO standard 5832-7 or 5832-12. The patella is made of ultra high molecular weight polyethylene (UHMWE) according to ISO standard 5834-2.

9) Indications for use

The replacement of the patellofemoral joint with the KNEETEC PFJ device is indicated in the case of symptoms and serious signs affecting daily activities involving the joint:

  • Serious degenerative arthritis of the patellofemoral joint,
  • Failure of conservative procedures (realignment, arthroscopy, transfer of the tibial tuberosity) with the persistence of joint pain or dysfunction,
  • Patellofemoral dislocation (dysplastic or other origin),
  • Post-traumatic arthritis (patellar fracture),

The KNEETEC PFJ patellofemoral prosthesis is intended for cemented use only.

When used in association with the KNEETEC PFJ prosthesis, the patellar component of the HLS KNEETEC has the above mentioned indications for use.

Main features orsystem characteristicsKneetecPFJHLS KneetecAvonJourney
MaterialsFemoral partCoCr alloyISO 5832-7orISO 5832-12CoCr alloyISO 5832-7orISO 5832-12CoCr alloyOxinium
MaterialsPatellarcomponentUHMWPEISO 5834-2UHMWPEISO 5834-2UHMWPEUHMWPEASTM F 648
Method of fixationCementedCementedCementedCemented
Intended useReplacement of thepatellofemoral jointTotal knee prosthesisReplacement of thepatellofemoral jointReplacement of thepatellofemoral joint
Terminal sterilizationGammaGammaN/KN/K
ManufacturerTORNIERTORNIERStrykerSmith & Nephew
K-numberK111970K094013K010100K020841K041160K051948K051086K951987

10) Summary of technological characteristics

Image /page/1/Picture/16 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is bold and appears to be constructed from geometric shapes. The logo is black and white, with the 'T' and the outline of the hexagon being black against a white background.

TORNIER S.A.S. 161, rue Lavoisier - Montbonnot 38334 SAINT-ISMIER CEDEX FRANCE

Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33

S.A.S. au capital de 288 000 € SIRET : 070 501 275 000 13 R.C.S. : 070 501 275 CODE APE : 331 B

Section 5 - Page 2/ page 3

SIEGE SOCIAL : rue du Doyen Gosse - 38330 SAINT-ISMIER - FRANCE

{2}------------------------------------------------

K111970 Page

Image /page/2/Picture/1 description: The image shows the word "page" followed by the numbers "3/3". The word "page" is written in cursive. The numbers "3/3" are written in a simple, clear font. The image is black and white.

TORNIER

11) Non-clinical testing

The following non-clinical testing has been performed on the smallest components of the ranges (worst case):

  • Contact area and trajectories of the patella to compare the congruence of the patella with the । trochlea,
  • Contact pressure test, -
  • Measurement of the subluxation force. r

Results: there is no significant difference between the results for Kneetec PFJ prosthesis and the predicate devices.

The results of engineering studies referenced in this 510(k) submission demonstrate the Kneetec PFJ is substantially equivalent to the predicate devices.

12) Substantial equivalence conclusion

Substantial equivalence of the Kneeter PFJ to the already cleared Avon, Journey and HLS Kneeter prostheses can be demonstrated on the following grounds, according to the FDA's Guidelines for Substantial Equivalence Decision Making Process:

  • The components of the Kneetec PFJ have been compared to the predicate devices.
  • The components of the Kneeter PFJ have the same intended use as the Avon and Journey prostheses and very similar indications for use.
  • Major technological characteristics are equivalent between the Kneetec PFJ and the predicate devices: -
    • Equivalence of general features and material, י
    • Equivalence of non-clinical testing results, .
    • -Equivalence of range of motion,
    • Equivalent means of fixation, .
    • Equivalent prosthetic dimensions. -

Therefore, in light of the above information, the company believes that the components of the Kneeter PFJ and the patellar component of the HLS Kneetec may be cleared via the 510(k) notification process for use as a Patellofemoral Knee Prosthesis.

Image /page/2/Picture/21 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is bold and has a unique design, with the top part of the 'T' appearing to be a smaller, inverted hexagon. The logo is simple and geometric, with a focus on the letter 'T' as the central element.

TORNIER S.A.S. 161, rue Lavoisier - Montbonnot 38334 SAINT-ISMIER CEDEX FRANCE

Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33

Section 5 - Page 3/ page 3

S.A.S. au capital de 288 000 € SIRET : 070 501 275 000 13 R.C.S. : 070 501 275 CODE APE : 331 B

SIEGE SOCIAL : rue du Doyen Gosse - 38330 SAINT-ISMIER - FRANCE

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. The text is written in all capital letters.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

FEB 2 3 2012

Tornier, Incoporated % Ms. Stéphanie Bernard Regulatory Affairs Specialist 161, rue Lavoiser - Montbonnot 38334 SAINT-ISMIER CEDEX FRANCE

Re: K111970

Trade/Device Name: Kneetec PFJ & HLS Kneetec Patellar Component Regulation Number: 21 CFR 888.3540 Regulation Name: Knee joint patellofemoral polymer/metal semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: KRR Dated: February 10, 2012 Received: February 13, 2012

Dear Ms. Bernard:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roved above and have determined the device is substantially equivalent (for the indications ferenced above and har sure) to legally marketed predicate devices marketed in interstate for use stated in the encreate) to ttg_segactment date of the Medical Device Amendments, or to conniner of the 11th) 2011-11-11 in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic (10. (110.) has the device, subject to the general controls provisions of the Act. The 1 ou may, morelore, manot of the Act include requirements for annual registration, listing of general voltrold providering practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (boo above) insting major regulations affecting your device can be may be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean i lease oc advised that I Dri bristian that your device complies with other requirements of the Act that I Drinas made a cond regulations administered by other Federal agencies. You must of any I catal statutes and regulations and limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2 - Ms. Stéphanie Bernard

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic rood ct radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sincerely yours,

Mark M. Millman

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K111970 Page 1/2

Indications for Use

K111970 510(k) Number (if known): K111970

Device Name: KNEETEC PFJ

Indications For Use:

The replacement of the patellofemoral joint with the KNEETEC PFJ device is indicated in the case of symptoms and serious signs affecting daily activities involving the joint:

  • Serious degenerative arthritis of the patellofemoral joint,

  • Failure of conservative procedures (realignment, arthroscopy, transfer of the tibial tuberosity) with the persistence of joint pain or dysfunction,

  • Patellofemoral dislocation (dysplastic or other origin),

  • Post-traumatic arthritis (patellar fracture).

The KNEETEC PFJ patellofemoral prosthesis is intended for cemented use only.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (2) CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) (Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number

Page 1 of 1

Section 4

Tomier

{6}------------------------------------------------

K111970 Page 2/2

Indications for Use

K111970 510(k) Number (if known): K094013

Device Name: HLS KNEETEC patellar implant

Indications For Use:

The HLS KNEETEC is indicated for use as a total knee replacement for the relief of pain and significant disability following the effects of primary or secondary osteoarthritis and rheumatoid arthritis. This device is also intended for the revision of knee prostheses.

The HLS KNEETEC is intended for cemented use only.

When used in association with the KNEETEC PFJ prosthesis, the patellar component of the HLS KNEETEC has the following indications for use:

The replacement of the patellofemoral joint with the KNEETEC PFI device is indicated in the case of symptoms and serious signs affecting daily activities involving the joint:

  • Serious degenerative arthritis of the patellofemoral joint,

  • Failure of conservative procedures (realignment, arthroscopy, transfer of the tibial tuberosity) with the persistence of joint pain or dysfunction.

  • Patellofemoral dislocation (dysplastic or other origin),

· Post-traumatic arthritis (patellar fracture).

The KNEETEC PFJ patellofemoral prosthesis is intended for cemented use only.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative DevicesPage 1 of 1
Tornier510(k) NumberSection 4
K111970

N/A