K Number
K111748

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-11-23

(154 days)

Product Code
Regulation Number
870.5800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pulsar Scientific device is not a life-supporting or life-sustaining device. It is a new, prescription use only [21 CFR 801 Subpart D] device for use with restricted mobility and non-ambulatory patients, or other patients as deemed appropriate by the physician. It is not intended for over-the-counter use [21 CFR 801 Subpart C]. The device is intended to function as an intermittent, external pneumatic compression device. The intended therapy of the device is to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding in blood flow back to the heart, to increase blood flow to various body areas, as determined by the physician, to aid in the reduction and control of edema and venous stasis ulcers.

Indications For Use:

Compression therapy - Including, but not limited to:

  • Decrease the risk of deep venous thrombosis (DVT); .
  • Aid the blood flow back to the heart: .
  • Treat and assist healing of cutaneous ulceration (wounds). reduce wound healing time, . enhance arterial circulation (blood flow), reduce compartmental pressures, reduce edema (swelling), reduce the need for anticoagulant (blood thinning) medications;
  • Treatment of pain and swelling of acute periarticular processes, treatment of pain and . swelling following mobilization of shoulder stiffness under anesthesia, treatment of pain · and swelling postoperatively for bones, joints and soft tissue, treatment of pain and
  • swelling caused by musculoskeletal contusions and athletic injury;
  • Treatment of disorders associate with vascular or lymphatic insufficiency such as . Chronic Venous Insufficiency (CVI), venous stasis ulcers, post-mastectomy edema and chronic edema;
  • Reduction of edema associated with soft tissue injuries such as ligament sprains, . postoperative edema, and burns.

Thermal therapy - Including but not limited to:

  • Reduction of general edema (swelling); ●
  • Reduction of edema associated with soft tissue injuries such as ligament sprains ● postoperative edema, and burns;
  • Treatment of pain and swelling of acute periarticular processes, treatment of pain and . swelling following mobilization of shoulder stiffness under anesthesia, treatment of pain and swelling postoperatively for bones, joints and soft tissue, treatment of pain and swelling caused by musculoskeletal contusions and athletic injury.
Device Description

The Recovery+ device is comprised of an electronically controlled, compartmentalized reusable pump used with a variety of single-patient use inflatable anatomical cuffs/wraps. The unit is light-weight and portable. The Recovery+ thermal component with standard pressure component and the DVT component (DVTherapy) of the pump are housed in separate compartments so that they can be used together, separately, or removed for portable usage of the DV.Therapy component. For purposes of this application we will refer to the components as the Recovery+ thermal compression unit and the DVTherapy unit.

The Recovery+ device is designed to provide compression therapy and/or thermal therapy (cold or hot). The therapies may be used together or separately, as ordered by the physician. The thermal therapy provides chilled or heated fluid to the cuff affixed on an extremity therapy site and compresses the treatment site for optimum performance and fluid transfer. The compression therapy is accomplished by air pumped through the cuffs, which are affixed on an extremity therapy site, to provide intermittent, sequential pressure. This facilitates venous flow towards the center of the body and simulates the muscle contraction and blood flow of ambulation.

The device utilizes microprocessor control with multiple sensors to ensure patient safety and system functionality, and to provide consistent and repeatable therapy modalities. Indicators are both visual on the unit display and audible. Settings are in place to detect any potentially unsafe situation and to terminate therapy to protect the patient and the system. The unit is light weight and portable. The thermal compression runs only on 115 VAC while the DVT unit can operate on batteries, connected to 115 VAC mains, or while docked in the thermal compression unit. The thermal compression unit has internal power supply while the DVT unit has an external power supply. The unit is protectively housed in a plastic shell, except for the outer membrane switch, the plastic quick-connect locks for tub connection, and an external power supply input jack.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Pulsar Scientific Recovery+ and DVTherapy devices, based on the provided 510(k) summary:

Acceptance Criteria and Device Performance

The 510(k) summary for the Pulsar Scientific Recovery+ and DVTherapy devices does not explicitly list quantitative acceptance criteria for performance metrics in a tabular format, nor does it provide specific measured performance data against such criteria.

Instead, the submission relies on demonstrating substantial equivalence to predicate devices through various tests. The core "acceptance criteria" are implied to be achieving performance characteristics that are substantially equivalent to or superseding the listed predicate devices for the various modalities offered.

However, the document does state performance characteristics that were verified:

Acceptance Criteria (Implied)Reported Device Performance
Safety Certifications- Conformed to UL 60601-1, CSA C22.2 NO 601.1, and CENELEC EN 60601-1.
Electromagnetic Compatibility (EMC)- Verified (emissions, immunity, safety).
Mechanical Integrity- Verified.
Environmental Performance- Verified.
Life Cycle Testing- Verified.
Pressure Delivery- Equivalent to predicate devices.
Cuff Fill Time- Equivalent to predicate devices.
Cycle Time- Equivalent to predicate devices.
Temperature Management- Equivalent to predicate devices. Controlled cooling from 43ºF to 49ºF; heating to 105ºF.
System Operation- Equivalent to predicate devices.

Important Note: The concept of "acceptance criteria" for a 510(k) submission, especially for devices like external pneumatic compression systems, often revolves around demonstrating safety and effectiveness compared to existing legally marketed devices, rather than meeting specific quantitative thresholds for, say, diagnostic accuracy.

Study Information

The provided document describes non-clinical validation testing but explicitly states that clinical testing was not performed. Therefore, several typical questions regarding clinical studies, experts, and ground truth are not applicable in this context.

  1. Sample size used for the test set and the data provenance: Not applicable. No clinical test set was used. The validation was based on non-clinical bench testing.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No clinical ground truth was established, as no clinical study was performed.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. No clinical test set requiring adjudication was used.
  4. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is not an AI-assisted diagnostic tool, and no MRMC study was conducted.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is a medical device, not an algorithm/AI. Its performance is inherent in its electromechanical function.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For non-clinical validation, "ground truth" would correspond to established engineering standards, performance specifications (e.g., target pressure, temperature range), and comparison against the performance of legally marketed predicate devices. The document implies compliance with these engineering and safety standards.
  7. The sample size for the training set: Not applicable. No training set was used, as this is a non-AI medical device that does not involve machine learning.
  8. How the ground truth for the training set was established: Not applicable. No training set was used.

{0}------------------------------------------------

510(k) Summary

K11748

Applicant [21 CFR 807.92 a(1)]:

Pulsar Scientific, LLC 8 Stony Brook Lane Ludlow, MA 01056 (413) 385-0005, ext. 202 - phone (413) 385-0003 - fax

Contact Person:

Joy Long (317) 496-8647 joyslong@gonowmail.com and Fernando Ubidia (413) 385-0005, ext 202 fernando@tandemtech.net

Device Name [21 CFR 807.92 a(2)]:

Regulation No.:CFR Part 870.5800
Classification:Class II
Panel:870 Cardiovascular
Classification Name:Compressible Limb Sleeve
Common/usual name:DeviceIntermittent External Pneumatic Compression
Product Code:JOW

Proprietary Name: Model #'s:

Recovery+ and DVTherapy PS511 and PS511a

Identification of Predicate Devices [21CFR 807.92 a(3)]:

  • DVT Therapy - Doctor's Orders, Inc. DVTcare CA5 K061125
  • . DVT Therapy - BioCompression Systems, Bioarterial Plus K072666
  • . Thermal Therapy - ThermoTek, Inc. NanoTherm/VascuTherm K061866

Name Clarification

The device as a whole will be marketed as the Recovery+, Model PS511. However, the DVT therapy portion of the device is designed to be portable and used as a stand-alone device. Therefore, the therapy portion of the device is named DVTherapy and assigned Model PS511a. Throughout the submission, when the discussion involves the full device it will be referred to as the Recovery+. When referring to the thermal only portion of the device, it will be referred to as the Recovery+ thermal compression unit. When the

{1}------------------------------------------------

discussion involves the portable DVT therapy portion of the device, it will be referred to as DVTherapy.

Device Description [21 CFR 807.92 a(4)]

Intended Use

The Pulsar Scientific device is not a life-supporting or life-sustaining device. It is a new, prescription use only [21 CFR 801 Subpart D] device for use with restricted mobility and non-ambulatory patients, or other patients as deemed appropriate by the physician. It is not intended for over-the-counter use [21 CFR 801 Subpart C]. The device is intended to function as an intermittent, external pneumatic compression device. The intended therapy of the device is to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding in blood flow back to the heart, to increase blood flow to various body areas, as determined by the physician, to aid in the reduction and control of edema and venous stasis ulcers.

Physical Description

The Recovery+ device is comprised of an electronically controlled, compartmentalized reusable pump used with a variety of single-patient use inflatable anatomical cuffs/wraps. The unit is light-weight and portable. The Recovery+ thermal component with standard pressure component and the DVT component (DVTherapy) of the pump are housed in separate compartments so that they can be used together, separately, or removed for portable usage of the DV.Therapy component. For purposes of this application we will refer to the components as the Recovery+ thermal compression unit and the DVTherapy unit.

Therapy Modalities Used

The Recovery+ device is designed to provide compression therapy and/or thermal therapy (cold or hot). The therapies may be used together or separately, as ordered by the physician. The thermal therapy provides chilled or heated fluid to the cuff affixed on an extremity therapy site and compresses the treatment site for optimum performance and fluid transfer. The compression therapy is accomplished by air pumped through the cuffs, which are affixed on an extremity therapy site, to provide intermittent, sequential pressure. This facilitates venous flow towards the center of the body and simulates the muscle contraction and blood flow of ambulation.

Safety Features

The device utilizes microprocessor control with multiple sensors to ensure patient safety and system functionality, and to provide consistent and repeatable therapy modalities. Indicators are both visual on the unit display and audible. Settings are in place to detect

{2}------------------------------------------------

any potentially unsafe situation and to terminate therapy to protect the patient and the system. The unit is light weight and portable. The thermal compression runs only on 115 VAC while the DVT unit can operate on batteries, connected to 115 VAC mains, or while docked in the thermal compression unit. The thermal compression unit has internal power supply while the DVT unit has an external power supply. The unit is protectively housed in a plastic shell, except for the outer membrane switch, the plastic quick-connect locks for tub connection, and an external power supply input jack.

Statement of Intended Use [21 CFR 807.92 a(5)]

The Recovery+ device is used on patient populations for which the Recovery+ device is applicable, based on prescription by a physician.

Indications

Including, but not limited to:

  • Decrease the risk of deep venous thrombosis (DVT); .
  • . Aid the blood flow back to the heart;
  • Treat and assist healing of cutaneous ulceration (wounds), reduce wound healing . time, enhance arterial circulation (blood flow), reduce compartmental pressures, reduce edema (swelling), reduce the need for anticoagulant (blood thinning) medications;
  • . Treatment of pain and swelling of acute periarticular processes, treatment of pain and swelling following mobilization of shoulder stiffness under anesthesia. treatment of pain and swelling postoperatively for bones, joints and soft tissue, treatment of pain and swelling caused by musculoskeletal contusions and athletic injury:
  • Treatment of disorders associated with vascular or lymphatic insufficiency such . as Chronic Venous Insufficiency (CVI), venous stasis ulcers, post-mastectomy edema and chronic edema:
  • Reduction of edema associated with soft tissue injuries such as ligament sprains, . postoperative edema, and burns.

Contraindications

As with all predicate devices, the device should not be used in the following instances:

  • . Presumptive evidence of congestive heart failure (CHF)
  • 트 Suspected/observed pre-existing deep vein thrombosis (DVT) or pulmonary embolism
  • Suspected/observed deep acute venal thrombosis (phlebothrombosis)
  • l Suspected/observed inflammatory phlebitis process
  • l Suspected/observed pulmonary edema
  • Suspected/observed acute inflammations of the veins (thrombophlebitis)
  • 트 Suspected/observed decompensated cardiac insufficiency
  • Suspected/observed arterial dysregulation
  • l Suspected/observed erysipelas

{3}------------------------------------------------

  • 1 Suspected/observed carcinoma and carcinoma metastasis in the affected extremity
  • 다 Suspected/observed decompensated hypertonia
  • 포 Suspected/observed acute inflammatory skin diseases or infection
  • I Suspected/observed venous or arterial occlusive disease
  • I Determine venous and lymphatic return is undesirable
  • Suspected/observed patient has Raynaud's Disease
  • 미 Suspected/observed poor peripheral circulation
  • I Suspected/observed hypersensitivity to cold
  • l Patient therapy contact on extremity containing a fracture
  • 트 Extremities that are not sensitive to pain

It is also not recommended that the Recovery+ device be used on a limb where the cuffs would interfere with open wounds, vein ligation, gangrene, dermatitis, recent skin graft, massive edema, or extreme deformity of the leg. They should not be used where increased venous or lymphatic return is undesirable.

い、

Differences in Indications

The indications for the Pulsar Scientific Recovery+ device are generally the same as those published for the predicate devices.

Technological Characteristics [21 CFR 807.92a(6)1]

The technological characteristics of this Recovery+ device are the same in overall design function, materials, energy source, and mode of operation as the predicate units. The material in contact with the patient's skin is medical grade, non-latex, non-woven nylon and polyester, designed for patient comfort and convenience. The inflation deflation tubes are polyurethane.

The Pulsar Scientific Recovery+ device has the same performance characteristics as the predicate devices. The pneumatic control circuitry is a microprocessor-controlled system. Multiple safety redundancies are built into the system including: high and low temperature indicators, indicators for unit malfunctions, and system malfunction overpressure safety via a pressure vent valve. Power is supplied via 115 VAC line current, or battery. The surface contact temperature range is microprocessor controlled in cooling from 43ºF to 49ºF and to 105ºF in heating mode.

Testing Conducted

Non-clinical validation, including electromagnetic compatibility (emissions, immunity and safety), mechanical integrity, environmental and life cycle testing, have shown that the Recovery+ has performance characteristics substantially equivalent to or superseding the listed predicate devices. The Recovery+ device has been validated by Pulsar Scientific at a design validation level based on requirement of UL 60601-1, CSA C22.2 NO 601.1 and CENELEC EN 60601-1 and confirmed by an accredited test lab.

{4}------------------------------------------------

Additional bench testing has verified equivalent pressure delivery, cuff fill time, cycle time, temperature management and system operation as the predicate devices listed.

Clinical testing is not appropriate to this device and was therefore not done.

Per the requirements of 21 CR 807, non-clinical validation testing and the information provided in this submission, Pulsar Scientific concludes that the Recovery+ system is safe, effective, and performs in a manner that is substantially equivalent to the predicate devices listed in this submission.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird symbol, with three overlapping wing-like shapes, representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Pulsar Scientific c/o Ms. Joy Long 8 Stony Brook Lane Ludlow, MA 01056

NOV 2 3 2011

Re: K111748

Trade/Device Name: Recovery+/DVTherapy (Models PS511 and PS511a) Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible limb sleeve Regulatory Class: II Product Code: JOW Dated: November 20, 2011 Received: November 23, 2011

Dear Ms. Long:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{6}------------------------------------------------

Page 2 - Ms. Joy Long

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

M.A. Hielm

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

Pulsar Scientific - 510(k) Number: K111748

Device Name: Recovery+

Indications For Use:

Compression therapy - Including, but not limited to:

  • Decrease the risk of deep venous thrombosis (DVT); .
  • Aid the blood flow back to the heart: .
  • Treat and assist healing of cutaneous ulceration (wounds). reduce wound healing time, . enhance arterial circulation (blood flow), reduce compartmental pressures, reduce edema (swelling), reduce the need for anticoagulant (blood thinning) medications;
  • Treatment of pain and swelling of acute periarticular processes, treatment of pain and . swelling following mobilization of shoulder stiffness under anesthesia, treatment of pain · and swelling postoperatively for bones, joints and soft tissue, treatment of pain and
  • swelling caused by musculoskeletal contusions and athletic injury;
  • Treatment of disorders associate with vascular or lymphatic insufficiency such as . Chronic Venous Insufficiency (CVI), venous stasis ulcers, post-mastectomy edema and chronic edema;
  • Reduction of edema associated with soft tissue injuries such as ligament sprains, . postoperative edema, and burns.

Thermal therapy - Including but not limited to:

  • Reduction of general edema (swelling); ●
  • Reduction of edema associated with soft tissue injuries such as ligament sprains ● postoperative edema, and burns;
  • Treatment of pain and swelling of acute periarticular processes, treatment of pain and . swelling following mobilization of shoulder stiffness under anesthesia, treatment of pain and swelling postoperatively for bones, joints and soft tissue, treatment of pain and swelling caused by musculoskeletal contusions and athletic injury.

Prescription Use × AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices

510(k) Number K111748

Page 1 of

§ 870.5800 Compressible limb sleeve.

(a)
Identification. A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.(b)
Classification. Class II (performance standards).