K Number
K061125

Validate with FDA (Live)

Device Name
DVTCARE CA5
Manufacturer
Date Cleared
2006-05-23

(29 days)

Product Code
Regulation Number
870.5800
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DVTcare™ CA5 is intended to be an easy to use, portable system that is prescribed by healthcare professionals, to help prevent the onset of DVT in patients, by stimulating blood flow in the legs (simulating muscle contractions). Furthermore, the unit can be used as an aid in the prophylaxis for DVT by persons traveling, or those expecting to be stationary for long periods of time (> 4 hours). This device can also be used to: aid in the prevention of DVT, enhance blood circulation, diminish post-operative pain and swelling, reduce wound healing time, and aid in the treatment and healing of stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency, and reduction of edema in the lower limbs.

Device Description

The DVTcare™ CA5 system is a light weight, portable, prescriptive device that helps stimulate blood flow in the legs through the use of pneumatically controlled, single chamber leg cuff(s), actuated by an electronically controlled pump unit and solenoid valves. All pump control unit components are protectively housed in a plastic shell except the outer membrane switch (needed for user interface), 2 plastic quick disconnects for air tube connection, and an external power supply input jack. The option exists for the unit to be used in single leg or double leg modes.

During operation, the pump unit provides air to one cuff through flexible plastic tubing, inflating it to a specified pressure, to compress the lower limb, thus aiding venous return. Air pressure and delivery are monitored by a pressure transducer and integrated system software contained in the plastic control unit. Immediately after the pressure transducer detects that the cuff has achieved the current set pressure, the cuff deflates to ambient pressure. This allows the blood flow to return to the limb. To provide maximum flexibility, the CA5 also provides adjustable cycle times. System software prevents users from entering a cycle time, for any one limb, to be less than 60 seconds. This is done to prevent excessive stimulation of the limb.

In single leg mode, this cycle repeats until the unit is turned off. In double leg mode the first and second leg cuffs do not fill simultaneously. The second leg begins its fill once the first leg has reached the midpoint of the overall cycle time for both legs. In double leg mode, air flow is directed to fill the second bladder cuff using the same cycle profile as the initial leg. After completing the fill, exhaust and rest period of leg 1, the control unit directs air flow to the second leg and fills that bladder cuff until the set pressure is reached and exhausted. The cycle repeats, reverting air flow to the initial leg.

The control unit is supplied with a non-serviceable, rechargeable battery, to allow user portability, and a power supply transformer for mains connection.

AI/ML Overview

The provided document is a 510(k) summary for the DVTcare™ CA5 device, an intermittent pneumatic compression device. It details the device's substantial equivalence to predicate devices but does not contain typical acceptance criteria and study information for AI/ML-based diagnostic devices.

Specifically, the document states:

  • "No clinical testing was performed on the DVTCare CA5" (page 5).
  • The device is an electromechanical pump and sleeve system, not an AI/ML diagnostic.

Therefore, many of the requested categories for AI/ML device evaluation (such as sample sizes for test and training sets, number of experts, adjudication methods, MRMC studies, standalone performance, and ground truth for AI models) are not applicable to this submission.

However, based on the information provided, I can construct a table for the non-clinical testing performed and describe how "equivalence" was established, which serves a similar purpose to acceptance criteria in this context for non-AI devices.


Acceptance Criteria and Device Performance for DVTcare™ CA5 (Non-AI Device)

The DVTcare™ CA5 is an intermittent pneumatic compression device, not an AI/ML diagnostic algorithm. Therefore, the "acceptance criteria" and "study" are focused on demonstrating substantial equivalence to predicate devices through non-clinical validation and comparison to published clinical studies of predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (based on predicate device characteristics/standards)Reported Device Performance (DVTcare™ CA5)
Electrical Safety Standards ComplianceValidated to UL 60601-1, CSA C22.2 NO 601.1, and CENELEC EN 60601-1. Confirmed by an accredited test lab.
Electromagnetic Compatibility (EMC)Demonstrated substantially equivalent or superseding performance to listed predicate devices.
Mechanical IntegrityDemonstrated substantially equivalent or superseding performance to listed predicate devices.
Environmental & Life Cycle TestingDemonstrated substantially equivalent or superseding performance to listed predicate devices.
Vibrational Integrity (for portability)Tested to MIL-STD 810D, section 514.3-1.
Pressure DeliveryBench tested to verify equivalent pressure delivery as predicate devices. Adjustable pressure set point (20-65 mmHg) with default similar to predicate (40mmHg).
Cuff (Garment) Fill TimeBench tested to verify equivalent cuff fill time as predicate devices. Default fill time approximately 10-12 seconds, similar to predicate (approximately 10 seconds).
Cycle TimeBench tested to verify equivalent cycle time as predicate devices. Adjustable cycle time (60-75 seconds) with default similar to predicate (approximately 60 seconds). System software prevents cycle time less than 60 seconds.
System Operation (overall functionality)Bench tested to verify equivalent system operation as predicate devices. Combines features of predicate devices: portability, adjustable pressure, and cycle time capabilities. Uses single bladder PVC chambers in soft Nylon material, similar to Flowtron Excel. Microprocessor control, internal battery, and mains power similar to WizAir DVT and TravelAir. Alarms for system monitoring and fault recognition similar to Flowtron Excel.
Hemodynamic Effect (Indirect, via predicate device studies)"Significant equivalence is concluded" based on comparison to published clinical studies of predicate devices (e.g., Venodyne model producing a mean blood flow velocity of 76.2 ± 23.77 using similar parameters).

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: Not applicable in the context of an AI/ML diagnostic test set. For the non-clinical bench testing, no specific sample sizes are mentioned, but it implies multiple cycles and components were tested to verify performance.
  • Data Provenance: The "clinical testing" referenced for comparison was from published studies of predicate devices. These included two studies:
    • One study used five healthy individuals aged 21-35 years old.
    • Another study used ten patients about to undergo primary unilateral knee arthroplasty, with a mean age of 68 years and mean weight of 85kg.
    • These data are retrospective in relation to the DVTcare™ CA5 device, as they are from previously published studies on different, but similar, devices. The country of origin is not specified but is implied to be academic/research settings.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable as this is not an AI/ML diagnostic device requiring expert-established ground truth. The "ground truth" for the device's operational performance was against engineering specifications and the performance of established predicate devices.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable as this is not an AI/ML diagnostic test set requiring human adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No MRMC study was done, as this is not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable as this is not an AI/ML algorithm. The device's "standalone" performance refers to its electromechanical operation as described in the non-clinical testing section.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For the non-clinical validation of the DVTcare™ CA5 device itself, the "ground truth" was engineering specifications, compliance with recognized safety and performance standards (UL, CSA, CENELEC, MIL-STD), and the established performance parameters of the predicate devices.
  • For the clinical effectiveness claims, the "ground truth" for efficacy was based on the outcomes and blood flow velocity measurements reported in the published clinical studies of the predicate devices. These studies reported outcomes such as "decreasing the risk of postoperative DVT" and measured "mean blood flow velocity."

8. The sample size for the training set:

  • Not applicable as this is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable as this is not an AI/ML device that requires a training set with established ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for "DVTCare, LLC". To the left of the company name is a graphic of a grid with four black dots. Below the company name is the address "731 Construction Ct., Suite B, Zeeland, MI 49464".

731 Construction Ct. Suite

510(k) Summary DVTcare™ CA5

510(k) Number K 061125

MAY 2 3 2006

Summary Prepared: Date 12/19/05

Submitter:

Doctor's Orders, Inc. 731 Construction Court, Suite B Zeeland, MI 49464 Phone: 616-931-9939 Fax: 616- 582-6200

Contact Person:

Jerry Kulas Development Engineer

3rd Party Reviewer

Intertek - ETL SEMKO 1365 Adams Court Menlo Park, CA 94025 (650) 463-2946

Name of the Device:

Trade Name: DVTcare™ CA5 Model #: CA5 Common or usual name: Compressible Limb Sleeve Device Classification Name: Compressible Limb Sleeve Class II Product Code: JOW Regulation No. CFR Part 870.5800 Establishment Registration Number: 3004739895

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for "DVTCare, LLC". The logo includes a grid-like design with four black circles on the left side. Below the logo is the address "731 Construction Ct. Suite B Zeeland, MI 49464".

Statement of Substantial Equivalence:

The DVTcare™ CA5 is substantially equivalent to the following commercially available devices:

WizAir DVT (510k number: K023573) in that the basis of operation of both devices uses intermittent, pneumatic compression to simulate muscle contractions in the legs aiding the return of venous flow. Compression is achieved by air delivery, through flexible plastic tubing, to inflatable bladders, that are wrapped around the limb(s), to transmit the pneumatic force to the leg. The air delivery is controlled by a microprocessor controlled pump and valve system. Both devices are light weight, portable and can operate on rechargeable batteries alone or while being connected to 110 VAC mains. Indications for use are similar. Cycle operating parameters are similar in pressure delivery and cycle time.

TravelAir Portable Compression System (510k number: K022340) in that the basis of operation of both devices uses intermittent, pneumatic compression to simulate muscle contractions in the legs aiding the return of venous flow. Compression is achieved by air delivery, through flexible plastic tubing, that are terminated with CPC quick lock connectors, to inflatable bladders, that are wrapped around the limb(s), to transmit the pneumatic force to the leg. Bladders are composed of PVC. The air delivery is controlled by a microprocessor controlled pump and valve system. Both devices are light weight, portable and can operate on rechargeable batteries alone or while being connected to 110 VAC mains. Indications for use are similar. Cycle operating parameters are similar in pressure delivery and cycle time.

Venodyne Advantage Plus (510k number: K011318) in that the basis of operation of both devices uses intermittent, pneumatic compression to simulate muscle contractions in the legs aiding the return of venous flow. Compression is achieved by air delivery, through flexible plastic tubing, to single chamber inflatable bladders, that are wrapped around the lower limb(s), to transmit the pneumatic force to the leg. The air delivery is controlled by a microprocessor controlled pump and valve system that are encased in a molded plastic enclosure. Indications for use are similar. Cycle operating parameters are similar in pressure delivery and cycle time.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for "DVTCare, LLC". The logo includes a grid-like design with three black circles on the left side. Below the logo is the address "731 Construction Ct. Suite B Zeeland, MI 49464".

Flowtron Excel AC550 (510k number: K961166) is similar in operation and intended use to the DVTCare CA5. Both incorporate adjustable pressures (between 20mmHg and 65mmHg), single and double leg modes, audible and visual alarms for system monitoring and fault recognition, soft Nylon covered, single chambered garments (cuffs) that wrap around the calf and held in place with hook and loop fasteners.

Device Description:

The DVTcare™ CA5 system is a light weight, portable, prescriptive device that helps stimulate blood flow in the legs through the use of pneumatically controlled, single chamber leg cuff(s), actuated by an electronically controlled pump unit and solenoid valves. All pump control unit components are protectively housed in a plastic shell except the outer membrane switch (needed for user interface), 2 plastic quick disconnects for air tube connection, and an external power supply input jack. The option exists for the unit to be used in single leg or double leg modes.

During operation, the pump unit provides air to one cuff through flexible plastic tubing, inflating it to a specified pressure, to compress the lower limb, thus aiding venous return. Air pressure and delivery are monitored by a pressure transducer and integrated system software contained in the plastic control unit. Immediately after the pressure transducer detects that the cuff has achieved the current set pressure, the cuff deflates to ambient pressure. This allows the blood flow to return to the limb. To provide maximum flexibility, the CA5 also provides adjustable cycle times. System software prevents users from entering a cycle time, for any one limb, to be less than 60 seconds. This is done to prevent excessive stimulation of the limb.

In single leg mode, this cycle repeats until the unit is turned off. In double leg mode the first and second leg cuffs do not fill simultaneously. The second leg begins its fill once the first leg has reached the midpoint of the overall cycle time for both legs. In double leg mode, air flow is directed to fill the second bladder cuff using the same cycle profile as the initial leg. After completing the fill, exhaust and rest period of leg 1, the control unit directs air flow to the second leg and fills that bladder cuff until the set pressure is reached and exhausted. The cycle repeats, reverting air flow to the initial leg.

The control unit is supplied with a non-serviceable, rechargeable battery, to allow user portability, and a power supply transformer for mains connection.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "DVTCare, LLC" along with their address. The logo is black and white and features a stylized design to the left of the company name. The address is "731 Construction Ct., Suite B, Zeeland, MI 49464".

Construction Ct.

Intended Use:

The DVTcare™ CA5 is intended to be an easy to use, portable system that is prescribed by healthcare professionals, to help prevent the onset of DVT in patients, by stimulating blood flow in the legs (simulating muscle contractions). Furthermore, the unit can be used as an aid in the prophylaxis for DVT by persons traveling, or those expecting to be stationary for long periods of time (> 4 hours). This device can also be used to: aid in the prevention of DVT, enhance blood circulation, diminish post-operative pain and swelling, reduce wound healing time, and aid in the treatment and healing of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency, and reduction of edema in the lower limbs.

Contraindications:

The DVTcare™ CA5 should not be used to treat the following conditions:

  • Persons with suspected, active or untreated: deep vein thrombosis, . ischemic vascular disease, severe arteriosclerosis, pulmonary edema, congestive heart failure, thrombophlebitis or an active infection.
  • · It is not recommended for use on a leg where the cuffs would interfere-with the following conditions: vein ligation, gangrene, dermatitis, open wounds, a recent skin graft, massive edema or extreme deformity of the leg.
  • · Not for use with patients with neuroapathy.
  • Do not use on extremities that are insensitive to pain. .
  • Do not use where increased venous or lymphatic return is undesirable. .
  • · Leg cuffs are not to be used in direct contact with skin.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for "DVTCare, LLC". The logo has a grid-like design with four black circles on the left side. Below the logo is the address "731 Construction Ct. Suite B Zeeland, MI 49464".

Technological Characteristics:

The DVTcare™ CA5 is similar to the predicate devices listed above in function and operating principles to achieve identical results however, the CA5 possesses a combination of features that each of the four predicate devices incorporate individually. The CA5 combines these features to offer greater treatment flexibility due to portability, adjustable pressure, and cycle time capabilities. All systems use microprocessor controlled pumps to deliver pressurized air (approximately 40mmHg) to bladders that are attached to the patient's lower limbs, using a cycle of approximately 60 seconds / leg as prescribed by a healthcare provider. Each cycle consists of one inflation (fill time approximately 10-12 seconds) followed by deflation. The CA5 has the capability of allowing a healthcare professional the following feature adjustment options: pressure set point (20-65 mmHg), hold time (0 seconds, not adjustable), cycle time (60-75 seconds). The cycle time is the length of time for one complete cycle on one leg including fill time, exhaust, and idle time (relaxation).

Patient access to these pressure and cycle time adjustments is prevented by requiring the healthcare provider to enter an access code for modification. The unit's default settings are similar to predicate devices in fill time (approximately 10 seconds), pressure (40mmHg), and cycle time (approximately 60 seconds) as noted above.

The CA5 uses similar means for pressure delivery as the predicate devices. Pressurized air is delivered by the pump to the leg cuffs via flexible, plastic air tubes connected to the plastic pump / control unit by locking, quick disconnect couplings. Like the Flowtron Excel, the CA5 leg cuffs are comprised of single bladder PVC chambers encased in a soft Nylon material to increase patient comfort and compliance.

Like the WizAir DVT and TravelAir the microprocessor and pump units are powered by internal battery supplies and can be connected to the mains power supply for operation and recharging.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for "DVTcare, LLC". The logo consists of a grid of lines and circles on the left, followed by the text "DVTcare, LLC" in a bold, sans-serif font. Below the logo is the address "731 Construction Ct. Suite B Zeeland, MI 49464".

731 Construction Ct. Suite B Zeeland, MI 494

Non-Clinical Testing

Nonclinical validation including electrical safety, EMC, mechanical integrity, environmental and life cycle testing have shown that the DVTcare™ CA5 has performance characteristics substantially equivalent to or superseding the listed predicate devices. The CA5 has been validated by DVTCare at a design validation level based on the requirements of UL 60601-1, CSA C22.2 NO 601.1 and CENELEC EN 60601-1 and confirmed by an accredited test lab. Being portable, CA5 samples were also tested to MIL-STD 810D, section 514.3-1 for vibrational integrity. Additional bench testing has verified equivalent pressure delivery, cuff (garment) fill time, cycle time and system operation as the predicate devices listed.

Clinical Testing

No clinical testing was performed on the DVTCare CA5, however test results of the predicate devices have been compared in the following published clinical studies : Evaluation of Intermittent Pneumatic Compression Devices (Orthopedics 24(3):257-261, 2001), and Venous hemodynamics after total knee arthroplasty: Evaluation of active dorsal to plantar flexion and several mechanical compression devices (Journal of Bone and Joint Surgery, Nov 1998) . Test subjects for the former study consisted of five healthy individuals aged 21-35 years old and the later consisted of ten patients who were about to have a primary unilateral knee arthroplasty for osteoarthritis. This patient group's mean age was 68 years old with mean weight of 85kg. In summary, the conclusions of each study state that the use of intermittent pneumatic compression devices is useful in decreasing the risk of postoperative DVT. The Venodyne model produced a mean blood flow velocity of 76.2 +/-23.77 using an average pressure of 43mmHg, a compression time of 11.6 seconds and a total cycle time of 60 seconds. These values are extremely similar to corresponding default parameters used by the DVTCare CA5. Since the methods of pressure delivery and operation are identical in both systems, significant equivalence is concluded.

Summary Conclusion:

Per the requirements of 21CFR807, surrogate clinical data, non-clinical validation testing and the information provided in the accompanying 510k pre-market submission, Doctor's Orders cocludes that the DVTcare™ CA5 is safe, effective and performs in a manner that is substantially equivalent to the predicate devices listed above.

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is enclosed within a circle, with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" surrounding the upper half of the circle. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 3 2006

Doctor's Orders, Inc. c/o Mr. Daniel W. Lehtonen Intertek Testing Services 2307 East Aurora Road Twinsburg, Ohio 44087

Re: K061125

DVTcare™ CA5 Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II (two) Product Code: JOW Dated: April 21, 2006 Received: April 24, 2006

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device w for a reveal above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for asso surved in the enactment date of the enactment date of the Medical Device Amendments, or to conimer of the 112) 2011-03-120, in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2 - Mr. Daniel W. Lehtonen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Duna R. Wehner

Image /page/7/Picture/5 description: The image shows a black and white drawing of a signature. The signature appears to be stylized and illegible. The lines are thick and bold, and the signature is compact.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: DVTCare CA5

Indications for Use: The DVTcare™ CA5 is intended to be an easy to use, portable system that is prescribed by healthcare professionals, to help prevent the onset of DVT in patients, by stimulating blood flow in the legs (simulating muscle contractions). Furthermore, the unit can be used as an aid in the prophylaxis for DVT by persons traveling, or those expecting to be stationary for long periods of time (> 4 hours). This device can also be used to: aid in the prevention of DVT, enhance blood circulation, diminish post-operative pain and swelling, reduce wound healing time, and aid in the treatment and healing of stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency, and reduction of edema in the lower limbs.

Prescription Use ﺮﺍ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

DMR R.V. Jones

(Division Sign-Off) Division of Cardiovascular Devices

Page 1 of 1

510(k) Number_k 66 1125

§ 870.5800 Compressible limb sleeve.

(a)
Identification. A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.(b)
Classification. Class II (performance standards).