K Number
K111520

Validate with FDA (Live)

Date Cleared
2011-10-26

(146 days)

Product Code
Regulation Number
882.4310
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IPC® System is indicated for the incision / cutting, removal, drilling, and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.

The IPC® POWEREASE™ System is indicated for drilling, tapping, and driving screws and working end attachments during spinal surgery, including open and minimally invasive procedures. It is also used in the placement of screws, or cutting of screws, posts, and rods.

Device Description

The purpose of this submission is to add a new driver and an accompanying software module to the IPC® . The new system is the IPC® POWEREASE™System. The IPC® POWEREASE™ System provides a powered driver that performs multiple functions and also has nerve stimulation capabilities. The IPC® POWEREASE™ System consists of the Integrated Powered Console (K081475, 10/17/2008), a driver handpiece, and cables to connect the driver, IPC® and NIM-ECLIPSE® System together. The driver is powered by the previously cleared Integrated Powered Console equipped with the appropriate software to operate the driver. The driver is used for drilling, tapping and driving screws in the pedicle, placing screws, and driving working end attachments capable of breaking set screw heads, cutting posts, and cutting rods during spinal surgery. The POWEREASE™ Driver also has nerve stimulation capabilities when connected to the Medtronic NIM -ECLIPSE® System (K061113 - 05/23/2006; K061639 -11/01/2006). An external POWEREASE™NIM Cable that connects the NIM-ECLIPSE® System to the driver enables the driver to have nerve stimulating capability. It should be noted that the IPC® does not provide the stimulation. The stimulation used with the IPC® POWEREASE™ is provided by the Medtronic NIM-ECLIPSE® System.

The driver is equipped with a quick connect feature which allows for easy and secure attachment of different working end attachments compatible with the IPC® POWEREASE™ Driver to perform the various functions mentioned above.

AI/ML Overview

The provided document describes the Medtronic IPC® POWEREASE™ System, a medical device for surgical procedures, and its substantial equivalence to predicate devices. It does not contain information typically found in studies evaluating acceptance criteria and device performance for AI/ML-based medical devices.

Therefore, I cannot extract the requested information such as:

  • A table of acceptance criteria and reported device performance.
  • Sample sizes used for test sets.
  • Number of experts used to establish ground truth or their qualifications.
  • Adjudication methods.
  • MRMC comparative effectiveness study results.
  • Standalone performance.
  • Type of ground truth used.
  • Sample size for the training set.
  • How ground truth for the training set was established.

Instead, the document focuses on:

  • Device Description: The IPC® POWEREASE™ System, a powered driver with nerve stimulation capabilities for spinal surgery.
  • Indications for Use: Detailed uses in various surgical procedures, including drilling, tapping, driving screws, and cutting screws/posts/rods in spinal surgery.
  • Predicate Device Identification: A list of previously cleared devices to which the IPC® POWEREASE™ System is deemed substantially equivalent.
  • Comparison to Predicate Devices: A table comparing features like indications, handpiece compatibility, and nerve monitoring capability.
  • Testing: Stating that "Laboratory bench testing conducted on the POWEREASE™ System demonstrates substantially equivalent performance characteristics to the predicate devices currently on the market. Testing was performed according to IEC 60601 Medical Electrical Equipment."
  • Conclusion: The device is substantially equivalent based on bench tests, intended use, and design control activities.

This document describes a traditional medical device clearance, not an AI/ML diagnostic or prognostic device that would typically involve the types of performance metrics and ground truth establishment you are asking about.

{0}------------------------------------------------

111520

pg 1 of 3

OCT 2 6 2011

5. 510(k) Summary

This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Medtronic Xomed, Inc. Address of Manufacturer: 6743 Southpoint Drive Jacksonville, FL 32216 (817) 788-6685 Phone (817) 788-6222 Facsimile Establishment Registration Number: 1625507 Contact Person: Rishi Sinha Senior Regulatory Affairs Specialist

Date:

Trade or Proprietary Name:

IPC® POWEREASE™ System IPC System with POWEREASE TM POWEREASE™ System

Common usual or Classification Name:

21 CFR 882.4310 - Powered simple cranial drills, burrs, trephines, and their accessories 21CFR 878.4820 - Surgical instrument motors and accessories/attachments 21CFR882.1870 - Evoked Response Electrical Stimulator

Description:

The purpose of this submission is to add a new driver and an accompanying software module to the IPC® . The new system is the IPC® POWEREASE™System. The IPC® POWEREASE™ System provides a powered driver that performs multiple functions and also has nerve stimulation capabilities. The IPC® POWEREASE™ System consists of the Integrated Powered Console (K081475, 10/17/2008), a driver handpiece, and cables to connect the driver, IPC® and NIM-ECLIPSE® System together. The driver is powered by the previously cleared Integrated Powered Console equipped with the appropriate software to operate the driver. The driver is used for drilling, tapping and driving screws in the pedicle, placing screws, and driving working end attachments capable of breaking set screw heads, cutting posts, and cutting rods during spinal surgery. The POWEREASE™ Driver also has nerve stimulation capabilities when connected to the Medtronic NIM -ECLIPSE® System (K061113 - 05/23/2006; K061639 -11/01/2006). An external POWEREASE™NIM Cable that connects the NIM-ECLIPSE® System to the driver enables the driver to have nerve stimulating capability. It should be noted that the IPC® does not provide the stimulation. The stimulation used with the IPC® POWEREASE™ is provided by the Medtronic NIM-ECLIPSE® System.

{1}------------------------------------------------

K111520

pg 2 of 3

The driver is equipped with a quick connect feature which allows for easy and secure attachment of different working end attachments compatible with the IPC® POWEREASE™ Driver to perform the various functions mentioned above.

Indications for Use:

The IPC® is indicated for the incision / cutting, removal, drilling, and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.

The IPC® POWEREASE™ System is indicated for drilling, tapping and driving screws and working end attachments during spinal surgery, including open and minimally invasive procedures. The IPC® POWEREASE™ System is also used in the placement of screws, or cutting of screws, posts, and rods.

Predicate Device Identification:

The Medtronic IPC® POWEREASE™ System is substantially equivalent to the following predicate devices:

DeviceManufacturer510(k)NumberClearanceDate
Integrated Powered Console (ElectricDrill System)Medtronic Xomed, Inc.K08147510/17/2008
Stimulating Bur GuardMedtronic Xomed, Inc.K06330511/29/2006
Stryker Consolidated Operating RoomEquipment (CORE) SystemStryker InstrumentsK0323031/16/2004
OrthomonAxon Systems, Inc.K06111305/23/2006
Eclipse TCD Neurvascular WorkstationAxon Systems, Inc.K06163911/01/2006
XPS 4000 System, Midas Rex LegendEHS System, Integrated Power Console(IPC)Medtronic Xomed, Inc.K08127709/05/2008

Table 1: Predicate Devices for the IPC® POWEREASE™ System

Comparison to Predicate Device:

The IPC® POWEREASE™ System is similar in device design, function, intended use and fundamental scientific technology to the previously cleared devices listed above.

FeaturesIPC® POWEREASE™ SystemIPC® System (K081475)
IndicationsThe IPC® is indicated for the incision /cutting, removal, drilling, and sawing ofsoft and hard tissue and bone, andbiomaterials in Neurosurgical (Cranial,Craniofacial), Orthopedic, Arthroscopic,Spinal, Sternotomy, and Generalsurgical procedures.The IPC® POWEREASE™ System isindicated for drilling, tapping and drivingThe IPC® is indicated for the incision /cutting, removal, drilling, and sawing ofsoft and hard tissue and bone, andbiomaterials in Neurosurgical (Cranial,Craniofacial) Orthopedic, Arthroscopic,Spinal, Sternotomy, and General surgicalprocedures

Table 2: Comparison of Subject Device to Predicate Device

{2}------------------------------------------------

K111520

screws and working end attachmentsduring spinal surgery, including openand minimally invasive procedures.The IPC® POWEREASE™ System isalso used in the placement of screws,or cutting of screws, posts, and rods.
HandpieceDriverCompatibilityDrills including the POWEREASE™Driver, Saws, Microdebriders. All ClassII devices in the IPC® System otherthan the POWEREASE™ Driver arecleared via K081475.Drills, Saws, Microdebriders - All Class IIdevices cleared via K081475
NerveMonitoringCapabilityYes - allows for connection to nervemonitoring systemsYes - allows for connection to nervemonitoring systems

Testing:

Laboratory bench testing conducted on the POWEREASE™ System demonstrates substantially equivalent performance characteristics to the predicate devices currently on the market. Testing was performed according to IEC 60601 Medical Electrical Equipment.

Conclusion/Summary:

Based upon the laboratory bench test summaries, intended use, and the successful completion of design control activities; the IPC® POWEREASE™ System has been shown to be substantially equivalent to currently marketed predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

OCT 2 6 2011

Medtronic Xomed, Inc. % Mr. Rishi Sinha Senior Regulatory Affairs Specialist 6743 Southpoint Drive North Jacksonville, Florida 32216

Re: K111520

Trade/Device Name: IPC POWEREASE System Regulation Number: 21 CFR 882.4310 Regulation Name: Powered simple cranial drills, burrs, trephines, and their accessories Regulatory Class: Class II Product Code: HBE, HWE, GWF Dated: September 29, 2011 Received: October 03, 2011

  • Dear Mr. Sinha:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Trug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act . The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract and warranties. We remind you; however, that device labeling must be truthful and not muclery,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

Page 2 – Mr. Rishi Sinha

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,
Erine Keith

Jos Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

  1. Indications for Use Statement

pg i of l

K111520

510(k) Number (if known):

® 2024 Experian

Device Name: IPC® POWEREASE™ System

Indications for Use:

The IPC® System is indicated for the incision / cutting, removal, drilling, and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.

The IPC® POWEREASE™ System is indicated for drilling, tapping, and driving screws and working end attachments during spinal surgery, including open and minimally invasive procedures. It is also used in the placement of screws, or cutting of screws, posts, and rods.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nail RPE Ogden for uxm
(Division Sign Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K111520

Medtronic IPC® POWEREASE™ System

§ 882.4310 Powered simple cranial drills, burrs, trephines, and their accessories.

(a)
Identification. Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.(b)
Classification. Class II (performance standards).