K Number
K032303

Validate with FDA (Live)

Date Cleared
2004-01-16

(175 days)

Product Code
Regulation Number
882.4310
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stryker Consolidated Operating Room Equipment (CORE) System is intended for use in the cutting, drilling, reaming, decorticating, and smoothing of bone and bone cement in a variety of surgical procedures, including but not limited to Neuro and Endoscopic. It is also usable in the placement or cutting of screws, metal, wires, pins, and other fixation devices.
The endoscopic applications with CORE include use with the SE5 Small Joint Shaver in the wrist for any need for morselization of tissue within the joint. Cutters will be used to debride synovitis, articular cartilage flaps, or torn ligaments when surgeons deem resection appropriate. Burs are indicated for management of osseous lesions such as eburnated articular surfaces or osteophytes.

Device Description

The device description of the Stryker System includes drills, shavers, shields, guards, motors, attachments, saws, wire drivers, collets, console, irrigation pump, cords, footswitch, handswitch, clips, tubing, cutting accessories, and sterilization cases.
The scope of this modification is limited to the console of the system.

AI/ML Overview

The provided text is a 510(k) summary for the Stryker Consolidated Operating Room Equipment (CORE) System. This document focuses on establishing substantial equivalence to previously cleared predicate devices, rather than an independent study proving the device meets specific acceptance criteria in terms of performance metrics like accuracy, sensitivity, or specificity.

Therefore, many of the requested details about acceptance criteria and study design are not present in this type of regulatory submission. The 510(k) process primarily relies on demonstrating that a new device is as safe and effective as a legally marketed predicate device, not necessarily on new clinical performance studies with predefined acceptance criteria.

However, I can extract information related to the device description and the comparison to predicate devices, which implicitly serves as the "study" for a 510(k) application in the sense of demonstrating equivalence.

Here's a breakdown of the information that can be extracted and an explanation of why other requested information is not available:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implicit)Reported Device Performance (Summary)
Safety and Effectiveness Equivalence to Predicate DevicesThe Stryker Consolidated Operating Room Equipment (CORE) System is deemed substantially equivalent to the following legally marketed predicate devices: - Stryker TPS (Dental-K943540, ENT-K943569, Neuro- 943541) - TPS Hermes (K991696) - Stryker Navigation System (K012380) - Dyonics (K771218) The technological characteristics of the CORE Console are stated to be the same as previously cleared predicate devices. This implies that the performance characteristics (e.g., power output, rotational speed, torque, vibration, material compatibility of cutting accessories, sterilization compatibility) meet or exceed those of the predicate devices.

Explanation: In a 510(k) submission, the primary "acceptance criterion" is often demonstrating substantial equivalence to a predicate device. This means the new device is as safe and effective as the predicate device(s). The "reported device performance" is essentially that its technological characteristics are the same as or comparable to the cleared predicates for the stated intended uses. Specific quantitative performance metrics (e.g., "95% accuracy in tissue identification") are typically not required or provided unless the device has novel technology or a different intended use compared to its predicates, which would necessitate a more rigorous performance study.

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided in the 510(k) summary. A 510(k) for a system like the CORE Console, which is a surgical tool based on established technology, generally relies on engineering testing, performance bench testing, and comparisons to predicate devices rather than a "test set" of patient data as you might see for an AI/diagnostic device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided and is not applicable in the context of this 510(k) submission. The "ground truth" for a mechanical surgical system like this would be established through engineering specifications, material science, and clinical experience with the predicate devices, not typically by expert consensus on a test set of data.

4. Adjudication Method for the Test Set

This information is not provided and is not applicable.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not provided and is not applicable. The Stryker CORE System is a surgical tool, not an AI-assisted diagnostic or interpretative device that would involve human "readers."

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not provided and is not applicable. The CORE System is a human-operated surgical console, not an "algorithm only" device.

7. The Type of Ground Truth Used

The "ground truth" for a device of this nature is implicitly established through:

  • Engineering specifications and standards: The console and its accessories are designed to meet certain power outputs, safety features, ergonomic requirements, and material compatibility.
  • Performance of predicate devices: The functionality and safety profile of the predicate devices (Stryker TPS, TPS Hermes, Stryker Navigation System, Dyonics) serve as the benchmark or "ground truth" for what is considered safe and effective for the stated intended uses.
  • Clinical experience and literature: The acceptance of the intended uses (cutting, drilling, reaming, etc., in neuro and endoscopic procedures) is based on established surgical practices and the proven utility of similar devices.

8. The Sample Size for the Training Set

This information is not provided and is not applicable. This device is not an AI/machine learning model that requires a "training set."

9. How the Ground Truth for the Training Set was Established

This information is not provided and is not applicable. This device is not an AI/machine learning model.

In summary:

The provided document is a 510(k) premarket notification, which seeks to demonstrate substantial equivalence to previously cleared devices. It does not present a clinical performance study with specific acceptance criteria, sample sizes, expert ground truth establishment, or AI-related metrics (like MRMC or standalone algorithm performance). The "study" in this context is the comparison of technical characteristics and intended use to the predicate devices to argue for substantial equivalence.

{0}------------------------------------------------

:

510(k) Summary

. -

大032303

Trade Name:Stryker Consolidated Operating Room Equipment (CORE) System
Common Name:Console
Classification Names:Bone Cutting Instruments and Accessories. (per 21 CFR section872.4120)Ear, Nose and Throat Electric or Pneumatic Surgical Drill. (per 21CFR section 874.4250)Powered Simple Cranial Drill, Burrs, Trephines and their Accessories(per 21 CFR section 882.4310)
Equivalent to:Stryker TPS (Dental-K943540, ENT-K943569, Neuro- 943541), TPSHermes (991696), Stryker Navigation System (K012380) and Dyonics(K771218).
Device Description:The device description of the Stryker System includes drills, shavers,shields, guards, motors, attachments, saws, wire drivers, collets,console, irrigation pump, cords, footswitch, handswitch, clips, tubing,cutting accessories, and sterilization cases.The scope of this modification is limited to the console of the system.
Intended Use:The Stryker Consolidated Operating Room Equipment (CORE) Systemis intended for use in the cutting, drilling, reaming, decorticating, andsmoothing of bone and bone cement in a variety of surgicalprocedures, including but not limited to Neuro and Endoscopic. It isalso usable in the placement or cutting of screws, metal, wires, pins,and other fixation devices.
The endoscopic applications with CORE include use with the SE5Small Joint Shaver in the wrist for any need for morselization of tissuewithin the joint. Cutters will be used to debride synovitis, articularcartilage flaps, or torn ligaments when surgeons deem resectionappropriate. Burs are indicated for management of osseous lesionssuch as eburnated articular surfaces or osteophytes.
TechnologicalComparison:Technological characteristics are the same as previously clearedfor the Stryker TPS System (K943540, K943569, K943541, andK991696), Stryker Navigation System (K012380) and Dyonics(K771218).
Submitted by:Jean W. SheppardSr. Regulatory Affairs RepresentativeStryker InstrumentsJuly 24, 2003

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/12 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The seal is black and white and has a simple, clean design.

JAN 1 6 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Jean W. Sheppard Sr. Regulatory Affairs Representative Stryker Instruments 4100 East Milham Avenue Kalamazoo, Michigan 49001

Re: K032303

Trade/Device Name: Stryker Consolidated Operating Room Equipment (CORE) Console Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, and their accessories Regulatory Class: II Product Codc: HBE Dated: October 20, 2003 Received: October 21, 2003

Dear Ms. Sheppard:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your been worked the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreative to the enactment date of the Medical Device Amendments, or to conniner of the rial) 2011-11-3) in accordance with the provisions of the Federal Food, Drug, devices that have been recuire approval of a premarket approval application (PMA). alle Costiette Act (71ct) that to hot requestly of the general controls provisions of the Act. The r ou may, merelere, mainer of the Act include requirements for annual registration, listing of general controls provisions of are tice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (500 abor of the existing major regulations affecting your device can may be subject to such additions, Title 21, Parts 800 to 898. In addition, FDA may be found in the Court of Counts concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I DA 3 issuated of a basenia.
that FDA has made a determination that your device complies with other requirements of the Act that FDA has made a dolorimiation administered by other Federal agencies. You must or any Federal statutes and regulations and annualing, but not limited to: registration and listing (21 comply with an the Act s requirements, me; good manufacturing practice requirements as set CFK Part 807), labeling (21 CFR Part 820); and 1f applicable, the clectronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Jean W. Sheppard

This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin maneting your antial equivalence of your deversity your develop premarket nothcation. THE PDA Intellig of babbankal via revice and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific atvice for your do rice of the same of the regulation entitled, and comact the Office of Compliance an (21CFR Part 807.97). You may obtain "Whisoranding by relevelee to premail.org.ibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

Colia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K032303

The Stryker Consolidated Operating Room Equipment (CORE) Device Name System

  • The Stryker Consolidated Operating Room Equipment (CORE)System is Indications intended for use in the cutting, drilling, reaming, decorticating, and smoothing of bone and bone cement in a variety of surgical procedures, including but not limited to Neuro and Endoscopic. It is also usable in the placement or cutting of screws, metal, wires, pins, and other fixation devices.
    The endoscopic applications with CORE include use with the SE5 Small Joint Shaver in the wrist for any need for morselization of tissue within the joint. Cutters will be used to debride synovitis, articular cartilage flaps, or torn ligaments when surgeons deem resection appropriate. Burs are indicated for management of osseous lesions such as eburnated articular surfaces or osteophytes.

Prescription Use _ X (Per 21 CFR 801 Subpart D) Over-The-Counter Use _ (21 CFR 807 Subpart C)

(IPLEASE DO NOT WRITE BELOW THISS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Muriam C. Provost

Sivision Sign-Off) Division of General, Restorative and Neurological Devices

K632303 (1) Number ...

Page __ 1 of __ 1_

§ 882.4310 Powered simple cranial drills, burrs, trephines, and their accessories.

(a)
Identification. Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.(b)
Classification. Class II (performance standards).