K Number
K111463

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-02-17

(267 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Valeo VT-601 Series IR Thermometer, (Model no.: VT-601D, VT-601E, VT-601F) is infrared thermometer intended for the intermittent measurement of human body temperature in people of all ages.

Device Description

The Valeo VT-601 Series IR Thermometer, (Model no.: VT-601D, VT-601E . VT-601F) is hand-held, non-sterile, reusable, battery operated device that can measure human body temperature. The Valeo VT-601D IR Forehead/ Ear Thermometer can measure human body temperature in 2 ways: (1) The temporal artery over forehead. (2) Tympanic temperature via the human ear. The Valeo VT-601E IR Ear Thermometer measures human body temperature natural thermal infrared radiation emitted from the ear tympanic. The Valeo VT-601F IR Forehead Thermometer measure human body temperature by the temporal artery over forehead. Operation is based on the measuring of the natural thermal infrared radiation emitted from the surface of the skin over the temporal artery or from the ear tympanic.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Valeo VT-601 Series IR Thermometer, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Accuracy)Reported Device Performance (Accuracy)
Ear Module:Ear Module:
±0.2°C for 36-39°C±0.2°C for 36-39°C
±0.3°C for the rest of the range±0.3°C for the rest of the range
(±0.4°F for 96.8-102.2°F)(±0.4°F for 96.8-102.2°F)
(±0.5°F for the rest of the range)(±0.5°F for the rest of the range)
Skin Module:Skin Module:
±0.3°C for 22-40.0°C±0.3°C for 22-40.0°C
(±0.5°F for 71.6-104.0°F)(±0.5°F for 71.6-104.0°F)

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not specified directly in the provided text. The document mentions "A Clinical Test Report conducted according to ASTM E 1965: 1998 (2009)" but does not detail the number of subjects or measurements.
  • Data Provenance: The study was "performed by the manufacturer." The country of origin is not explicitly stated for the study data itself, but the manufacturer (VALEO Corporation) is based in Taipei, Taiwan. The study is retrospective in terms of being a report of past activity, but the study itself (the clinical test) would have been prospective to gather the data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the text. As this is a medical device for measuring temperature, the "ground truth" would likely be established by a reference thermometer or highly accurate clinical measurement standards, rather than expert judgment in the human-reader sense.

4. Adjudication Method for the Test Set:

This information is not applicable and therefore not provided. Temperature measurements are typically compared directly to a "gold standard" reference, without the need for human adjudication between multiple readers of the device's output.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. The device is an infrared thermometer, not an AI-powered diagnostic imaging tool that assists human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The device is a standalone algorithm/device. Its performance is evaluated on its own ability to measure temperature. The accuracy criteria listed above define its standalone performance.

7. The Type of Ground Truth Used:

The ground truth used for assessing accuracy was based on the "method recommended in ASTM E1965 standard." This standard specifies methods for determining the accuracy of clinical thermometers, which typically involves comparing the device's readings against a reference thermometer (a highly accurate and calibrated device) in a controlled environment, and/or against a subject's core body temperature measured by an invasive "gold standard" method in clinical trials. Therefore, the ground truth is derived from a standardized clinical measurement method.

8. The Sample Size for the Training Set:

This information is not applicable and therefore not provided. The Valeo VT-601 Series IR Thermometer is a physical device based on infrared sensing technology, not a machine learning model that requires a training set.

9. How the Ground Truth for the Training Set was Established:

This information is not applicable and therefore not provided. As stated above, the device does not use a "training set" in the context of machine learning. The device's design and calibration would be based on engineering principles and physical laws of infrared radiation, validated against established measurement standards like ASTM E1965.

{0}------------------------------------------------

K111463

510(K) SUMMARY

FEB 1 7 2012

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92

Date of Preparation: Dec. 6, 2011

.

1. Submitter's Name:VALEO Corporation
Address:4F-3, No. 161, Sung-Teh Road Taipei, Taiwan 110
Phone:+886-2-23463113
Fax:+886-2-23463115
Contact:James Jean/ Presidente-mail: valeo@valeo.com.tw , daisy@valeo.com.tw
2. Device Name :
Trade Name:Valeo VT-601 Series IR Thermometer,Model no.: VT-601D, VT-601E, VT-601F
Common Name:IR Thermometer
Classification namethermometer, electronic, clinical
3. DEVICE CLASSThe Valeo VT-601 Series IR Thermometer, (Model no.: VT-601D, VT-601E, VT-601F) has been classified asRegulatory Class: IIPanel: 80Product Code: FLLRegulation Number: 21CFR 880.2910
4. Predicate Device:The predicate devices are the• BRAUN Thermoscan IRT 3020 Thermometer (K983295) marketed by BRAUN AG.&• EXERGEN Temporal - Scanner (K011291) marketed by EXERGEN CORP.
5. Intended Use:The Valeo VT-601 Series IR Thermometer, (Model no.: VT-601D, VT-601E, VT-601F) is infrared thermometers intended for theintermittent measurement of human body temperature in people of all ages.

Product:VT-601 Series IR Thermometer (Model no.: VT-601D , VT-601E , VT-601F ) Section 4 - 510(k) Summary Page 1 of 4 REV. 【B】

{1}------------------------------------------------

6. Device Description:

Equivalence Discussion

The Valeo VT-601 Series IR Thermometer, (Model no.: VT-601D,

VT-601E . VT-601F) is hand-held, non-sterile, reusable, battery operated device that can measure human body temperature.

The Valeo VT-601D IR Forehead/ Ear Thermometer can measure

human body temperature in 2 ways:

  • (1) The temporal artery over forehead.
  • (2) Tympanic temperature via the human ear.

The Valeo VT-601E IR Ear Thermometer measures human body temperature natural thermal infrared radiation emitted from the ear tympanic.

The Valeo VT-601F IR Forehead Thermometer measure human body temperature by the temporal artery over forehead.

Operation is based on the measuring of the natural thermal infrared radiation emitted from the surface of the skin over the temporal artery or from the ear tympanic.

The Valeo VT-601 Series IR Thermometer, (Model no.: VT-601D, 7. Comparison to 510(k) Predicate VT-601E, VT-601F) have the same intended use, principles of operation, and similar technological characteristics as predicate devices. Please find Devices & Substantial Technological Characteristics as follows

VT-601 Series IRThermometerPredicate DeviceBRAUN ThermoscanIRT 3020Thermometer(K983295)Predicate DeviceEXERGEN Temporal- Scanner(K011291)
RegulationNumber:21CFR 80.2910Item21CFR 880.2910
DisplayedTemperatureRangeEar: 3243°CSkin: 22 42.2°C(Ear:89.6109.5°FSkin:71.6107.9°F)34°C42.2°C(93.2°F108°F)15.5°C~42°C
AmbientTemperatureEnvironment1. Operating:10°C40°C2. Storage:-20°C50°C1. Operating:10°C40°C2. Storage:-20°C50°C1. Operating:15.5°C40°C2. Storage:-20°C50°C

Product:VT-601 Series IR Thermometer (Model no.: VT-601D , VT-601E , VT-601F ) Section 4 - 510(k) Summary REV. (B) Page 2 of 4

{2}------------------------------------------------

AccuracyEar: ±0.2°C,3639°C±0.3°C, the restSkin: ±0.3°C,2240.0°C(Ear: ±0.4°F,96.8102.2°F±0.5°F, the restSkin: ±0.5°F,71.6104.0°F)1. 36.0°C(96.8 °F) to39.0°C(102.2°F):± 0.2°C (0.4°F)2. Outside thisrange:± 0.3°C (0.5°F)36.0°C to 39.0°C:± 0.2°COutside this range:± 0.3 °C
Resolution0.1°C / 0.1°F0.1°C / 0.1°F0.1°C
Display4 digits LCD3 and 1/2 digits LCD3 digits LCD
ComponentsMaincomponents:1. Chip IC (µP)2. IR Sensor3. LCD(LiquidCrystalDisplay)Main components:1. Chip IC (µP)2. IR Sensor3. LCD(LiquidCrystalDisplay)Main components:1. Chip IC (µP)2. IR Sensor3. LCD(LiquidCrystalDisplay)
PowerrequirementsTwo 1.5VdcAlkalineManganesebattery (AAA)Two 3V CR 2032batteriesOne Alkaline 9Vbatteries
(PowerConsumption)Standby mode:14 miliwattsMeasurementmode:8 miliwatts1. Standby mode:16.8 miliwatts2. Measurementmode:12 miliwatts1. Standby mode:112.68miliwatts2. Measurementmode: 112.68miliwatts
VoiceFunctionWithNoNo

SUBSTANTIAL EQUIVALENCE DISCUSSION:

Valeo VT-601 Series IR Thermometer, (Model no.: VT-601D , VT-601E , VT-601F) has the same general design with the predicate devices. It has the following similarities to the predicate devices in:

  • · having the same intended use
  • · using similar operating principle
  • using similar technological characteristics

In summary, the Valeo VT-601 Series IR Thermometer, (Model no.: VT-601D , VT-601E , VT-601F) described in this submission is substantially equivalent to the predicate devices.

Product:VT-601 Series IR Thermometer (Model no.: VT-601D , VT-601E , VT-601F ) Section 4 - 510(k) Summary Page 3 of 4

{3}------------------------------------------------

    1. Non-Clinical Discussion of Non-Clinical Tests Verification Activities Performed to Tests Verification Determine the Safety and Performance of Valeo VT-601 Series IR
      Thermometer (Model no .: VT-601D, VT-601E, VT-601F ) is as the followings:
    1. Performance Compliance Test is according to ASTM E1965: 1998(2009) conducted by manufacturer.
    1. Electrical Safety Compliance Test is according to IEC 60601-1 by accredited laboratory.
    1. EMC Compliance Test is according to IEC 60601-1-2 by accredited laboratory.
  • Discussion of Clinical Test Validation Activities Performed to Determine the 9. Clinical Test for Effectiveness of Device is as the followings: Measurement Accuracy

    • A Clinical Test Report conducted according to ASTM E 1965: 1998 (2009) performed by the manufacturer was included as Clinical Investigation report. This report was carried out in such a way that compared the accuracy performance between Valeo VT-601 Series IR Thermometer, (Model no.: VT-601D , VT-601E , VT-601F ) and the predicate devices according to the method recommended in ASTM E1965 standard.

The results of the Clinical Test Report could positively support the claim of Substantial Equivalence for Valeo VT-601 Series IR Thermometer, ( Model no.: VT-601D , VT-601E , VT-601F ) against the chosen 510(k) predicate devices.

10. Conclusions:

The Valeo VT-601 Series IR Thermometer (Model no.: VT-601D , VT-601E , VT-601F ) has the same intended use and similar technological characteristics as the BRAUN Thermoscan IRT 3020 Thermometer (K983295) marketed by BRAUN AG. & EXERGEN Temporal – Scanner (K011291) marketed by EXERGEN CORP. Moreover, bench testing contained in this submission demonstrates that any differences in their technological characteristics DO NOT RAISE ANY NEW QUESTIONS OF SAFETY OR EFFECTIVENESS. Thus, the Valeo VT-601 Series IR Thermometer, (Model no.: VT-601D , VT-601E , VT-601F ) is substantially equivalent to the predicate devices.

Product:VT-601 Series IR Thermometer (Model no.: VT-601D , VT-601E , VT-601F ) REV. 【B】 Page 4 of 4 Section 4 - 510(k) Summary

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, a traditional symbol of medicine, with three snakes intertwined around a staff. The seal is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

VALEO Corporation C/O Ms. Jennifer Reich Senior Consultant Harvest Consulting Corporation 2904 N. Boldt Drive Flagstaff, Arizona 86001

FEB 1: 7 2012

Re: K111463

Trade/Device Name: Valeo VT-601 Series IR Thermometer, Model no.: VT-601D, VT-601E, VT-601F, VALEO Corporation Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: February 1, 2012 Received: February 6, 2012

Dear Ms. Reich: .

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Reich

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to Compilance ification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the I vision of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and

Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Valeo VT-601 Series IR Thermometer, Device Name: Model no.: VT-601D , VT-601E , VT-601F VALEO Corporation

Indications for Use:

Valeo VT-601 Series IR Thermometer, (Model no.: VT-601D, VT-601E, VT-601F) is infrared thermometer intended for the intermittent measurement of human body temperature in people of all ages.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

V Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ald C. Chyn 2/17/12

(Division Sign-Üff) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K111463

Page 1 of 1

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.