K Number
K111442

Validate with FDA (Live)

Date Cleared
2011-06-03

(10 days)

Product Code
Regulation Number
878.4493
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

V-Loc™ 180 Absorbable Reload: The V-Loc™ 180 absorbable reloads are intended for use with the Endo Stitch™ suturing device and SILS™ Stitch suturing device. The Endo Stitch™ suturing device and SILS™ Stitch suturing device, when used with the reloads, have application in endoscopic surgery for the placement of running stitches in soft tissue. The V-Loc™ 180 absorbable reloads are indicated for soft tissue approximation where use of an absorbable suture is appropriate.

V-Loc™ PBT Non-Absorbable Reload: The V-Loc™ PBT non-absorbable reloads are intended for use with the Endo Stitch™ suturing device and SILS™ Stitch suturing device. The Endo Stitch™ suturing device and SILS™ Stitch suturing device, when used with the reloads, have application in endoscopic surgery for the placement of running stitches in soft tissue. The V-Loc™ PBT non-absorbable reloads are indicated for soft tissue approximation.

Device Description

Absorbable and non-absorbable synthetic suture reloads for endoscopic suturing devices

AI/ML Overview

The provided text describes a 510(k) premarket notification for surgical suture reloads. It does not contain information about an AI/ML powered medical device, therefore the response will focus on the details that ARE available related to the performance testing of the suturing devices, primarily to establish substantial equivalence with predicate devices.

Here's an analysis of the provided information regarding acceptance criteria and the study:

Acceptance Criteria and Device Performance

The acceptance criteria for the V-Loc™ 180 Absorbable Reload and V-Loc™ PBT Non-Absorbable Reload are based on demonstrating substantial equivalence to predicate devices through various performance tests. The document highlights the following standards and tests:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Performance StandardReported Device Performance and Compliance
USP Monograph for Sutures (33, NF 28, 2010) Physical Test <861> DiameterCompliance with this standard indicates the suture diameter meets specified requirements. The document states "Performance testing has been performed... This testing verifies that the V-Loc™ 180 Absorbable Reload and the V-Loc™ PBT Non-Absorbable Reload are substantially equivalent to the predicate devices."
USP Monograph for Sutures (33, NF 28, 2010) Physical Test <871> Needle AttachmentCompliance with this standard indicates the needle attachment strength meets specified requirements. The document states "Performance testing has been performed... This testing verifies that the V-Loc™ 180 Absorbable Reload and the V-Loc™ PBT Non-Absorbable Reload are substantially equivalent to the predicate devices."
Needle Attachment (Bench Top & In Vivo)Demonstrated performance to establish substantial equivalence. Specific quantitative results are not provided but the successful completion of these tests supports the claim of equivalence.
Insertion / Removal Forces through Port Cannula (Bench Top & In Vivo)Demonstrated performance to establish substantial equivalence. Specific quantitative results are not provided.
Tensile Strength (Bench Top & In Vivo)Demonstrated performance to establish substantial equivalence. Specific quantitative results are not provided.
Barbed Suture Holding Strength (Bench Top & In Vivo)Demonstrated performance to establish substantial equivalence. Specific quantitative results are not provided.
Tissue Passage (Bench Top & In Vivo)Demonstrated performance to establish substantial equivalence. Specific quantitative results are not provided.
Biocompatibility"All component materials... have been evaluated in accordance with ISO 10993-1: Biological Evaluation of Medical Devices, according to their intended use." Also, "These materials have also been previously cleared in predicate devices for the identical intended use."
Technological Characteristics (Indications, Labeling, Materials, Performance)The implantable portions are "identical to the predicate V-Loc™ Wound Closure Devices in terms of indications, labeling, materials, performance, and technology." Modifications (braided leader, ferrule) utilize "previously-cleared" needles and materials.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes used for the benchtop and in vivo performance tests. It mentions "Performance testing has been performed" which implies tests were conducted, but specific numbers of devices, animals, or test conditions are not detailed.

Data provenance is not specified beyond indicating "bench top and in vivo tests." There is no mention of country of origin or whether the studies were retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable to this submission. The submission is for surgical sutures, not an AI/ML powered device that requires expert review for ground truth establishment. The performance tests are largely objective measurements against defined standards (like USP monographs) or comparisons to predicate devices.

4. Adjudication Method for the Test Set

This information is not applicable as there is no mention of human-in-the-loop assessments or expert interpretations that would require an adjudication method. The testing focuses on physical and biological properties.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

This information is not applicable. This submission is for surgical sutures and related devices, not an AI/ML system. There are no "human readers" or "AI assistance" being evaluated in this context.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This information is not applicable. This is not an AI/ML device.

7. The Type of Ground Truth Used

The "ground truth" in this context refers to established scientific and regulatory standards or the documented performance of the predicate devices.

  • USP Monograph for Sutures (33, NF 28, 2010): These serve as the ground truth for physical properties like diameter and needle attachment.
  • Predicate Devices: The V-Loc™ 180 Absorbable Wound Closure Device (K091087 and K082662), V-Loc™ Non-Absorbable Wound Closure Device (PBT) (K103052), Endo Stitch™ Endoscopic Suturing Device (K934738), SILS™ Stitch (K090419), and USSC Non-absorbable Sutures (K955747) serve as the ground truth for establishing "substantial equivalence" in terms of material composition, intended use, indications, labeling, and performance characteristics.
  • ISO 10993-1: Biological Evaluation of Medical Devices: This standard serves as the ground truth for biocompatibility assessments.

8. The Sample Size for the Training Set

This information is not applicable. This is not an AI/ML device, so there is no concept of a "training set" for an algorithm. The development of sutures and devices involves design, material selection, and manufacturing processes, not algorithm training.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness
SUBMITTER:Covidien llc (formerly registered as Tyco Healthcare, LP)60 Middletown AvenueNorth Haven, CT 06473Tel. No .: (203) 492-5000
CONTACT PERSON:Robert ZottProgram Director, Regulatory Affairs60 Middletown AvenueNorth Haven, CT 06473 USAPhone: (203) 492-5493Fax: (203) 492-6602
DATE PREPARED:May 17, 2011
TRADE/PROPRIETARY NAME:V-Loc™ 180 Absorbable Reload
V-Loc™ PBT Non-Absorbable Reload
Endo Stitch™ Endoscopic Suturing Device
SILS™ Stitch Endoscopic Suturing Device
COMMON/USUAL NAME:Synthetic Absorbable and Nonabsorbable Sutures
Endoscopic Suturing Devices
CLASSIFICATION NAME:V-Loc™ 180 Absorbable Reload:Absorbable poly(glycolide/l-lactide) surgical suture21 CFR 878.4493
V-Loc™ PBT Non-Absorbable Reload:Nonabsorbable poly(ethylene terephthalate) surgical suture21 CFR 878.5000
Endo Stitch™ Endoscopic Suturing DeviceEndoscopic Tissue Approximation Device21 CFR 876.1500
SILS™ Stitch Endoscopic Suturing DeviceEndoscopic Tissue Approximation Device21 CFR 876.1500
PREDICATE DEVICE(S):V-Loc™ 180 Absorbable Wound Closure Device,Covidien (K091087 and K082662)
V-Loc™ Non-Absorbable Wound Closure Device (PBT),Covidien (K103052)
Endo Stitch™ Endoscopic Suturing Device, Covidien(K934738)
SILS™ Stitch, Covidien (K090419)

Addendum to Special 510(k) Premarket Notification for V-Loc™ 180 and V-Loc™ PBT Reloads - 06/01/2011

™ Trademark Page 1 of 26

.

{1}------------------------------------------------

USSC Non-absorbable Sutures, Covidien (K955747)
DEVICE DESCRIPTION:Absorbable and non-absorbable synthetic suture reloads for endoscopic suturing devices
INTENDED USE:V-Loc™ 180 Absorbable Reload:The V-Loc™ 180 absorbable reloads are intended for use with the Endo Stitch™ suturing device and SILS™ Stitch suturing device. The Endo Stitch™ suturing device and SILS™ Stitch suturing device, when used with the reloads, have application in endoscopic surgery for the placement of running stitches in soft tissue. The V-Loc™ 180 absorbable reloads are indicated for soft tissue approximation where use of an absorbable suture is appropriate.V-Loc™ PBT Non-Absorbable Reload:The V-Loc™ PBT non-absorbable reloads are intended for use with the Endo Stitch™ suturing device and SILS™ Stitch suturing device. The Endo Stitch™ suturing device and SILS™ Stitch suturing device, when used with the reloads, have application in endoscopic surgery for the placement of running stitches in soft tissue. The V-Loc™ PBT non-absorbable reloads are indicated for soft tissue approximation.
TECHNOLOGICAL CHARACTERISTICS:The implantable portions of the V-Loc™ Reloads are identical to the predicate V-Loc™ Wound Closure Devices in terms of indications, labeling, materials, performance, and technology. The only differences are the manner in which they are attached to their respective needles and the lengths in which they are supplied. The modifications include the addition of a braided leader and ferrule in order to connect the V-Loc™ suture to a needle specifically designed to be captured and passed between the jaws of an endoscopic suturing device. The needles and materials used for the leader and ferrule have been previously-cleared in the cited predicate devices. The Endo Stitch™ and SILS Stitch™ Endoscopic Suturing Devices have not been changed.
MATERIALS:All component materials of the V-Loc™ 180 Absorbable Reload and the V-Loc™ PBT Non-Absorbable Reload have been evaluated in accordance with ISO 10993-1: Biological Evaluation of Medical Devices, according to their intended use. These materials have also been previously cleared in predicate devices for the identical intended use.
PERFORMANCE DATA:Performance testing has been performed in support of the intended use of these devices. This testing verifies that the V-Loc™ 180 Absorbable Reload and the V-Loc™ PBT Non-Absorbable Reload are substantially equivalent to the predicate devices. The following performance standards were used:

{2}------------------------------------------------

  • · USP Monograph for Sutures (33, NF 28, 2010) Physical Test <861> Diameter
  • · USP Monograph for Sutures (33, NF 28, 2010) Physical Test <871> Needle Attachment

The subject devices were evaluated in the following bench top and in vivo tests:

  • Needle Attachment
  • · Insertion / Removal Forces through Port Cannula
  • · Tensile Strength
  • · Barbed Suture Holding Strength
  • · Tissue Passage

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like strokes. The bird is positioned to the right of the text, which is arranged in a circular pattern around the bird. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Covidien % Mr. Robert Zott Program Director, Regulatory Affairs 60 Middletown Avenue North Haven, Connecticut 06473

JUN - 3 2011

Re: K111442

Trade/Device Name: V-Loc™ 180 Absorbable Reload, V-Loc™ PBT Non-Absorbable Reload Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly(glycolide/L-lactide) surgical suture Regulatory Class: II Product Code: GAM, GAT, OCW Dated: May 17, 2011 Received: May 24, 2011

Dear Mr. Zott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

Page 2 - Mr. Robert Zott

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.him for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

fn Dothan,
Oct.

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known).

V-Loc™ 180 Absorbable Reload Device Name:

Indications for Use:

The V-Loc™ 180 absorbable reloads are intended for use with the Endo Stitch™ suturing device and SILS™ Stitch suturing device. The Endo Stitch™ suturing device and SILS™ Stitch suturing device, when used with the reloads, have application in endoscopic surgery for the placement of running stitches in soft tissue. The V-Loc™ 180 absorbable reloads are indicated for soft tissue approximation where use of an absorbable suture is appropriate.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K111442

{6}------------------------------------------------

Covidien IIc (formerly registered as Tyco Healthcare, LP)

Indications for Use

510(k) Number (if known):

Device Name: V-Loc™ PBT Non-Absorbable Reload

Indications for Use:

The V-Loc™ PBT non-absorbable reloads are intended for use with the Endo Stitch™ suturing device and SILS™ Stitch suturing device. The Endo Stitch™ suturing device and SILS™ Stitch suturing device, when used with the reloads, have application in endoscopic surgery for the placement of running stitches in soft tissue. The V-Loc™ PBT non-absorbable reloads are indicated for soft tissue approximation.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Kuneber Mxm

rgical, Orthopedic, and Restorative Devices

510(k) Number K111442

§ 878.4493 Absorbable poly(glycolide/l-lactide) surgical suture.

(a)
Identification. An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.