K Number
K111337

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-01-25

(258 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Tornier® Collagen Coated BioFiber Scaffold is intended for use where temporary wound support is required to reinforce soft tissues where weakness exists or for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.

Tornier® Collagen Coated BioFiber Scaffold is also intended for reinforcement of soft tissues, in conjunction with sutures and/or suture anchors during tendon repair surgery; including the reinforcement of rotator cuff, patellar, Achilles, biceps and quadriceps tendons.

Tornier® Collagen Coated BioFiber Scaffold is not intended for use as a replacement for normal body structures or to provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, or quadriceps tendons. Sutures, used to repair the tear, and sutures or bone anchors, used to attach the tissue to the bone, provide mechanical strength for the tendon repair.

Device Description

Tornier® Collagen Coated BioFiber™ Scaffold is a bi-layer, synthetic absorbable reinforced woven fabric made from poly(4-hydroxybutyrate) fibers. The construction permits the mesh to be cut into any desired shape or size without unraveling. The device is supplied sterile, for single-patient use for the reinforcement of soft tissue where weakness exists.

AI/ML Overview

The provided 510(k) summary for the Tornier® Collagen Coated BioFiber™ Scaffold is for a medical device that is not an AI/ML device, but rather an absorbable surgical mesh. Therefore, many of the typical acceptance criteria and study aspects related to AI/ML device performance (like sensitivity, specificity, F1-score, sample sizes for test/training sets, expert consensus for ground truth, MRMC studies, standalone performance, etc.) are not applicable to this submission.

This 510(k) summary focuses on demonstrating substantial equivalence to predicate surgical meshes through physical, mechanical, biocompatibility, and functionality testing, rather than performance metrics for an AI algorithm.

However, I can extract the relevant information presented in the document based on the prompt's categories, even if they don't perfectly align with AI/ML evaluation.


Acceptance Criteria and Device Performance (Not Applicable for AI/ML)

Since this is not an AI/ML device, typical performance metrics like sensitivity, specificity, or F1-score are not relevant. The acceptance criteria for this device revolve around demonstrating substantial equivalence to predicate devices through various tests.

The document states: "In support of the claim of substantial equivalence, Tornier has presented data regarding the physical and mechanical characteristics as outlined in the FDA "Guidance for the Preparation of a Premarket Notification for a Surgical Mesh". In addition, Tornier has conducted studies of collagen purity, biocompatibility testing in accord with ISO 10993, a functionality study, a shelf-life study and a biodegradation study."

This indicates that the acceptance criteria would have been met by passing these specific tests and demonstrating that the device performs equivalently to the listed predicate devices in these aspects. The document does not provide specific performance values (e.g., tensile strength, burst strength, degradation rate) to create a table of "reported device performance." Instead, it asserts that the studies support the claim of substantial equivalence.


Study Details (Adapted for a Non-AI/ML Medical Device)

  1. A table of acceptance criteria and the reported device performance

    • Not applicable in the typical AI/ML sense. The acceptance criteria are implicit in the tests conducted to demonstrate substantial equivalence to predicate devices. The document does not provide a table of performance metrics for the Tornier® Collagen Coated BioFiber™ Scaffold versus specific criteria. It only states that the studies were conducted and support substantial equivalence.
  2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not explicitly stated for specific tests. The document mentions a "functionality study," a "shelf-life study," and a "biodegradation study," along with "biocompatibility testing in accord with ISO 10993." However, the sample sizes, data provenance (e.g., animal studies, in-vitro experiments), or study design (retrospective/prospective) for these tests are not detailed in this 510(k) summary. These are typically included in the full submission, not in this publicly available summary.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. As a physical medical device (surgical mesh), ground truth is established through laboratory testing (e.g., material testing, chemical analysis, animal studies for biocompatibility/biodegradation) and comparison to established standards or predicate device performance, not through expert human review in the context of image interpretation or diagnosis.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving human interpretation or diagnosis, often to resolve discrepancies in expert opinions for ground truth establishment in AI/ML performance evaluation. This is not relevant for a surgical mesh.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. MRMC studies are used to evaluate diagnostic performance, often for AI-assisted workflows. This device is a surgical mesh, not a diagnostic AI tool.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This question pertains to the performance of an AI algorithm in isolation. The Tornier® Collagen Coated BioFiber™ Scaffold is a physical medical device.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Derived from laboratory and material testing. For this device, ground truth would be established through:
      • Chemical analysis (for collagen purity).
      • Standardized physical and mechanical tests (e.g., tensile strength, burst strength, weight, porosity) for the mesh material, often following ISO or ASTM standards.
      • Biocompatibility testing (e.g., cytotoxicity, sensitization, irritation, genotoxicity, implantation tests) according to ISO 10993 standards, typically involving animal models and histopathological evaluation.
      • Biodegradation studies (in vitro and/or in vivo), also involving animal models and material analysis.
    • The "ground truth" here is the objective measurement of the device's properties and its biological interaction.
  8. The sample size for the training set

    • Not applicable. This device is a manufactured product, not an AI model that requires a training set.
  9. How the ground truth for the training set was established

    • Not applicable. As there is no training set for an AI model, this question is not relevant.

{0}------------------------------------------------

K111337

1 of 3

510(k) SUMMARY (Per 21 CFR 807.92)

JAN 2 5 2012

స్టే

General Company Information

Name:Tornier, Inc.
Contact:Howard SchrayerRegulatory Affairs Consultant

7701 France Avenue South - Suite 600 Address: Edina, MN 55435

Telephone:952-426-7641
Fax:952-426-7601
  • Date Prepared January 11, 2012

General Device Information

Product Name:Tornier® Collagen Coated BioFiber™ Scaffold
------------------------------------------------------------

Classification: "Absorbable Surgical Mesh" Product code: FTL - Class II

Predicate Devices

Tepha, Inc., Inc.TephaFLEXTM Surgical Mesh[510(k) Number K070894]
Tornier, Inc.Tornier® BioFiberTM Scaffold - (Surgical Mesh)[510(k) K102788 and K093799]
Sofradim Production(Covidien)PREVADHTM Surgical Mesh[510(k) K073287]
Artimplant ABArtelon® SportMeshTM Tissue Reinforcement Mesh[510(k) K071887]

{1}------------------------------------------------

11/ 1337 20F3

Description

Tornier® Collagen Coated BioFiber™ Scaffold is a bi-layer, synthetic absorbable reinforced woven fabric made from poly(4-hydroxybutyrate) fibers. The construction permits the mesh to be cut into any desired shape or size without unraveling. The device is supplied sterile, for single-patient use for the reinforcement of soft tissue where weakness exists.

Intended Use (Indications)

Tornier® Collagen Coated BioFiber™ Scaffold is intended for use where temporary wound support is required to reinforce soft tissues where weakness exists or for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.

Tornier® Collagen Coated BioFiber™ Scaffold is also intended for reinforcement of soft tissues, in coniunction with sutures and/or suture anchors during tendon repair surgery; including the reinforcement of rotator cuff, patellar, Achilles, biceps and quadriceps tendons.

Tornier® Collagen Coated BioFiber™ Scaffold is not intended for use as a replacement for normal body structures or to provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, or quadriceps tendons. Sutures, used to repair the tear, and sutures or bone anchors, used to attach the tissue to the bone, provide mechanical strength for the tendon repair.

Substantial Equivalence

The mesh structure of the Collagen Coated BioFiber™ Scaffold is the same material used to form the mesh used in the predicate TephaFLEX™ Surgical Mesh and both the material and mesh structure are the same as that of the predicate Tornier® BioFiber™ Scaffold. The use of the collagen coating as a component of the Collagen Coated BioFiber™ Scaffold is equivalent to the structure used for the predicate PREVADH™ Surgical Mesh and the predicate Artelon® SportMesh™ Tissue Reinforcement Mesh.

In support of the claim of substantial equivalence, Tornier has presented data regarding the phvsical and mechanical characteristics as outlined in the FDA "Guidance for the Preparation of a Premarket Notification for a Surgical Mesh". In addition, Tornier has conducted studies of collagen purity, biocompatibility testing in accord with ISO 10993, a functionality study, a shelf-life study and a biodegradation study.

This submission supports the position that the Tornier® Collagen Coated BioFiber™ Scaffold is substantially equivalent to previously cleared devices, including those listed above. A number of predicate devices list the same range of clinical uses.

{2}------------------------------------------------

ר331 ווא 30f3

Conclusions

Tornier, Inc. believes that the information provided establishes that similar legally marketed devices have been used for the same clinical applications as the Tornier® Collagen Coated BioFiber™ Scaffold. The materials from which the Tornier device is fabricated have an established history of use, and the devices have been tested in accordance with applicable FDA guidelines.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image is a logo for the Department of Health & Human Services - USA. The logo features a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of curved lines. The logo is presented in black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JAN 2 5 2012

Tornier, Inc. % Mr. Howard Schrayer 7701 France Avenue South - Suite 600 Edina, Minnesota 55435

Re: K111337

Trade/Device Name: Tornier® Collagen Coated BioFiber Scaffold Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL Dated: January11, 2012 Received: January 12, 2012

Dear Mr. Schrayer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

Page 2 - Mr. Howard Schrayer .

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

cerely yours,

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): 〈〈〈 | | | 33フ

Device Name: Tomier® Collagen Coated BioFiber Scaffold

Indications For Use:

Tornier® Collagen Coated BioFiber Scaffold is intended for use where temporary wound support is required to reinforce soft tissues where weakness exists or for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.

Tornier® Collagen Coated BioFiber Scaffold is also intended for reinforcement of soft tissues, in conjunction with sutures and/or suture anchors during tendon repair surgery; including the reinforcement of rotator cuff, patellar, Achilles, biceps and quadriceps tendons.

Tornier® Collagen Coated BioFiber Scaffold is not intended for use as a replacement for normal body structures or to provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, or quadriceps tendons. Sutures, used to repair the tear, and sutures or bone anchors, used to attach the tissue to the bone, provide mechanical strength for the tendon repair.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David Krane for MXIM

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K111337

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.