K Number
K111300

Validate with FDA (Live)

Date Cleared
2011-08-12

(95 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Green, Orange, Violet and White Nitrile examination gloves, Powder Free, Non - Sterile are disposable device and made of Synthetic Polymer that exhibits rubber like characteristics intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powder Free Green, Orange, Violet and White Nitrile Examination Gloves, Non Sterile is a disposable device intended for medical purposes that is worn on the examiner's to prevent contamination between patient and examiner.

AI/ML Overview

The provided document describes the performance testing and acceptance criteria for "Senstouch Nitrile Examination Gloves, Powder Free, Non Sterile". This device is a Class I medical device and the submission focuses on its physical and chemical properties rather than an AI/ML algorithm. Therefore, many of the requested categories (e.g., number of experts, adjudication methods, MRMC studies, standalone algorithm performance, training set details) are not applicable.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

RequirementStandardAcceptance CriteriaReported Device Performance
Physical DimensionD 6319-00aE3Length ≥ 230 mmLength = 246.77 mm
Width = 95 ± 10 mmWidth = 94.46 mm
Thickness: ≥ 0.05 mm (Finger, Palm, Cuff)Thickness: Finger = 0.150 mm, Palm = 0.095 mm, Cuff = 0.082 mm
Physical Properties (Unaged)D 6319-00aE3TS = 14 MPaTS = 15.9 MPa
UE = 500 %UE = 583.4 %
Physical Properties (Aged)D 6319-00aE3TS = 14 MPaTS = 22.3 MPa
UE = 400 %UE = 569.2 %
Freedom from PinholesD 6319-00aE3, FDA 21 CFR 800.20Acc / Rej = 3 / 40 piece found
Moisture ContentIn-house< 0.8%0.67%
Powder ResidueD 6319-00aE3, D6124 - 01< 2.0 mg/glove0.84 mg/glove
Biocompatibility: Primary Skin Irritation in Rabbits"Consumer Product Safety Commission, Title 16, Chapter II, Part 1500 and ASTM F720-86"PassPass (Negative)
Biocompatibility: Dermal Sensitization"Consumer Product Safety Commission, Title 16, Chapter II, Part 1500 and ASTM F720-86"PassPass (Negative)

2. Sample Size Used for the Test Set and Data Provenance

The document specifies some sample sizes for specific tests:

  • Residual Powder: N = 5
  • Moisture Content: N = 8
  • Visual Inspection: Not explicitly stated as a count, but AQL levels (0.65 for critical, 2.5 for major, 4.0 for minor defects) imply sampling plans.
  • Water Leak Test: G-1, AQL 1.5
  • Physical Properties (General): S-2, AQL 4.0 and S-2, AQL 2.5 (Single Sampling)

The data provenance is from Indonesia, where the manufacturer (PT MAHAKARYA INTI BUANA) is located. The study is prospective in the sense that these are laboratory tests conducted on the manufactured product to demonstrate compliance with standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

Not applicable. This device is a patient examination glove, and its performance is assessed through physical and chemical laboratory tests, not by human expert interpretation of data.

4. Adjudication Method for the Test Set:

Not applicable. Performance is determined by objective laboratory measurements against defined standards.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC study was not done. This is not an AI/ML diagnostic or assistive device where human reader performance would be a factor.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

No, a standalone algorithm performance study was not done. This device is a physical product (nitrile gloves), not a software algorithm.

7. The Type of Ground Truth Used:

The "ground truth" for the device's performance is established by international and national standards (e.g., ASTM D 6319-00aE3, FDA 21 CFR 800.20, D6124 - 01, ISO standards for biocompatibility implicitly via "Consumer Product Safety Commission... and ASTM F720-86"). The performance measurements are directly compared against the numerical criteria specified in these standards.

8. The Sample Size for the Training Set:

Not applicable. This is not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set was Established:

Not applicable. As there is no training set for this device.

{0}------------------------------------------------

K111300

AUG 1 2 2011

Image /page/0/Picture/2 description: The image shows a stylized logo with the letters 'MB' in a bold, gothic-style font. The letters are interconnected and have a textured appearance, giving them a three-dimensional effect. A horizontal line runs through the middle of the letters, adding a graphic element to the design. The overall impression is one of a strong, established brand or organization.

PT. MAHAKARYA INTI BUANA

Jl .Desa Dalu X-A, Dalu 10
Deli Serdang – T. Morawa
Sumatera Utara – 20362
Indonesia

Tel +62-61- 30007150, 51, 52, 53 Fax +62-61-30007156

Date : May 3, 2011

510 (K) SUMMARY

1.0Submitter:
Name:PT MAHAKARYA INTI BUANA
Address:Jl. Desa Dalu X-A, Dalu 10Deli Serdang - T. MorawaSumatera Utara - 20362
Phone No.:+62-61-30007150, 51
Fax No.:+62-61-30007156
FDA Registration No:3005813621
2.0Contact Person:
Name:(Ms) Lina Ng
Phone:+62-61-30007150,51
Fax No.:+62-61-30007156
E-mail:qa@mhk-ia.com
3.0Name or The Device:
Trade Name:1) Senstouch and2) Multiple or Customers' Trade Name
Device Name:Nitrile Examination Gloves, Powder Free, Non Sterile(Green, Orange, Violet and White)
Common Name:Patient Examination Gloves
Classification Name:Nitrile Examination Gloves
4.0Identification of The Legally Marketed Device:
Polymer:Nitrile Latex
Device Class:Class I
Substantial Equivalent
Device Description:Patient Examination Gloves, 21 CFR 880.6250
Product Code:Nitrile -- 80LZA
Standard:ASTM D 6319-00a (2005)
5.0Performance Testing Standard
Water Leak Test:G-1, AQL 1.5
Physical Properties:S-2, AQL 4.0
Physical Properties:S-2, AQL 2.5, Single Sampling
Residual Powder:N = 5
Moisture Content:N = 8
Visual Inspection:Critical Defect AQL 0.65Major Defect AQL 2.5Minor Defect AQL 4.0
REFERENCEDEVICESTANDARD
REQUIREMENTSTANDARDSPERFORMANCESPECIFICATION
Physical DimensionD 6319-00aE3Length = 246.77 mmWidth = 94.46 mmThickness:- Finger = 0.150 mm- Palm = 0.095 mm- Cuff = 0.082 mmLength ≥ 230 mmWidth = 95 ± 10Thickness ≥ 0.05
Physical PropertiesD 6319-00aE3Unaged:TS = 15.9 MPaUE = 583.4%Aged:TS = 22.3 MPaUE = 569.2 %Unaged:TS = 14 MPaUE = 500 %Aged:TS = 14 MPaUE = 400 %
Freedom from PinholesD 6319-00aE3FDA 21 CFR800.200 piece foundAcc / Rej = 3 / 4
Moisture ContentIn-house0.67%0.8%
Powder ResidueD 6319-00aE3D6124 - 010.84 mg/glove< 2.0 mg/glove
BiocompatibilityAccordance with "ConsumerProduct Safety Commission,Title 16, Chapter II, Part1500 and ASTM F720-86Primary SkinIrritation inRabbitsDermalSensitizationPass(Negative)Pass(Negative)PassPass

. .

{1}------------------------------------------------

K111300

PT. MAHAKARYA INTI BUANA

J1 .Desa Dalu X-A, Dalu 10 Deli Serdang - T. Morawa Sumatera Utara - 20362 Indonesia_

Tel +62-61- 30007150, 51, 52, 53 Fax +62-61-30007156

6.0 Intended Use of The Device

Powder Free Green, Orange, Violet and White Nitrile Examination Gloves, Non Sterile is a disposable device intended for medical purposes that is worn on the examiner's to prevent contamination between patient and examiner.

7.0 Summary of The Technological Characteristics of The Device (Performance and Conformance Test Data) :

Technological Characteristics of White Nitrile Examination Gloves, Powder Free, Non Sterile are summarized as below :-

Table 7.0 Performance and Conformance Data of White Nitrile Examination Glove

{2}------------------------------------------------

K 111300

PT. MAHAKARYA INTI BUANA

Image /page/2/Picture/2 description: The image shows a stylized logo with the letters 'MB' in a bold, gothic-style font. A horizontal line runs through the middle of the letters, adding a graphic element to the design. The logo is rendered in black against a white background, creating a high-contrast image.

Jl .Desa Dalu X-A, Dalu 10 Deli Serdang - T. Morawa Sumatera Utara – 20362 Indonesia ____________________________________________________________________________________________________________________________________________________________________

Tel +62-61- 30007150, 51, 52, 53 Fax +62-61-30007156

8.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data The performance test data of the non-clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above.

Based on the above data and information, the device is substantially equivalent to its predicate device approved for distribution in the United States. Part 4 of this submission discusses further on substantial equivalent comparison

9.0 Substantial Equivalent Based on Assessment of Clinical Performance Data Clinical data is not required for gloves for this submission.

10.0 Conclusion

It can be concluded that the Green, Orange, Violet and White Nitrile Examination Gloves, Non Sterile perform according to the gloves performance standards referenced in Section (5) and (7) above and hence meet ASTM standards and FDA requirements.

Conclusively, we therefore claim that this device is substantially equivalent to its predicate device approved by FDA and is safe and effective for its intended for purposes.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the emblem.

Food and Drug Administration 19903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Pt. Mahakarya Inti Buana Ms. Lina Ng OSP Manager Dalu 10, Deli Serdang T. Morawa Sumatera Utara Indonesia 20362

AUG 12 2011

Re: Ki11300

Trade/Device Name: Senstouch Nitrile Examination Gloves, Powder Free, Non Sterile (Green, Orange. Violet and White) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: May 3, 2011 Received: August 3, 2011

Dear Ms. Ng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

ff your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Ng

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Cintron D. Martin

Anthony D. Watson. B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devic Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name:

Indications For use:

KII1300

GREEN, ORANGE, VIOLET AND WHITE NITRILE EXAMINATION GLOVES POWDER FREE, NON-STERILE

Green, Orange, Violet and White Nitrile examination gloves, Powder Free, Non - Sterile are disposable device and made of Synthetic Polymer that exhibits rubber like characteristics intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth F. Clausing-Williams

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: __KIJI 300

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.