K Number
K111135

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-08-17

(117 days)

Product Code
Regulation Number
888.3060
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pyrenees and K2M Cervical Plate Systems are indicated for use in anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (DDD), spondylolisthesis, trauma (including fractures), spinal stenosis and tumors (primary and metastatic), failed previous fusions (pseudarthrosis) and deformity (defined as scoliosis, kyphosis or lordosis).

Device Description

The Pyrenees Cervical Plate System is a spinal fixation system which consists of cervical screws and plates. All of the components are available in a variety of sizes to match more closely the patient's anatomy. Materials: The devices are manufactured from Commercially Pure titanium alloy per ASTM and ISO standards. Function: The system functions as an adjunct to fusion to provide immobilization of cervical segments of the spine.

AI/ML Overview

The Pyrenees Cervical Plate System, Translational Plate, is a spinal fixation system consisting of cervical screws and plates made from Commercially Pure titanium alloy. It's intended for anterior screw fixation to the cervical spine (C2-C7) as an adjunct to fusion.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Mechanical Performance (per ASTM F1717):The Pyrenees Cervical Plate System performed equally to or better than predicate systems in the following tests:
- Static Torsion- Met or exceeded predicate performance
- Dynamic Compression- Met or exceeded predicate performance
- Static Tension- Met or exceeded predicate performance
Material:Manufactured from Commercially Pure titanium alloy per ASTM and ISO standards, same as predicate devices.
Design Features and Sizing:Substantially the same as predicate systems.
Intended Use:Same intended use as predicate systems.

2. Sample Size Used for the Test Set and Data Provenance

The document describes mechanical testing and comparison to predicate devices rather than a clinical study with a test set of human subjects or data. Therefore, the concepts of "sample size for the test set" and "data provenance" (country of origin, retrospective/prospective) are not applicable in the context of this submission. The "test set" here refers to the physical devices subjected to mechanical loads.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as the "ground truth" for mechanical testing is established by the specifications of ASTM F1717 and the performance of predicate devices, not expert consensus.

4. Adjudication Method for the Test Set

This information is not applicable as the assessment is based on objective mechanical measurements against a standard (ASTM F1717) and direct comparison to predicate device performance. No human adjudication is mentioned or implied for this type of test.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of an MRMC comparative effectiveness study, AI assistance, or human readers in the provided document. This submission focuses on the mechanical and functional equivalence of a spinal fixation device, not a diagnostic or interpretative AI-based system.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical spinal implant, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for this device's acceptance is based on:

  • Established mechanical testing standards: Specifically, ASTM F1717.
  • Performance of legally marketed predicate devices: The Pyrenees system's performance was compared directly to that of predicate devices in static torsion, dynamic compression, and static tension.
  • Material and design specifications: Adherence to ASTM and ISO standards for materials, and equivalence in design features and sizing to predicate devices.

8. The Sample Size for the Training Set

This information is not applicable. The device is an orthopedic implant, not a machine learning model that undergoes training on a dataset.

9. How the Ground Truth for the Training Set was Established

This is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

AUG 1 7 2011

510(k) Summary for the

Pyrenees Cervical Plate System, Translational Plate

This safety and effectiveness summary for the System is provided as required per Section 513(i(i(i) of the Food, Drug and Cosmetic Ast.

1. Submitter :K2M, Inc.751 Miller Drive SE,Suite F1Leesburg, VA 20175Date Prepared: April 21, 2011Contact Person :Nancy GiezenK2M, Inc.751 Miller Drive SELeesburg, VA 20175Telephone: 703-777-315
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
2. Tradename:Pyrenees Cervical Plate System 510(k), Translational Plate
Common Name:Spinal Fixation System
Classification Name:Spinal Intervertebral Body Fixation Orthosis (888.3060)
Product Code:KWQ
Device Class:II

3. Predicate or legally marketed devices which are substantially equivalent :

  • Pyrenees Cervical Plate System ( K2M, Inc. ) K063544 .
  • . Cervical Spine Locking Plate (Synthes) K971883, K000536
  • . Vectra-T (Synthes) K051665

4. Description of the device :

The Pyrenees Cervical Plate System is a spinal fixation system which consists of cervical screws and plates. All of the components are available in a variety of sizes to match more closely the patient's anatomy.

Materials: The devices are manufactured from Commercially Pure titanium alloy per ASTM and ISO standards.

Function: The system functions as an adjunct to fusion to provide immobilization of cervical segments of the spine.

5. Intended Use:

The Pyrenees and K2M Cervical Plate Systems are indicated for use in anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (DDD), spondylolisthesis, trauma (including fractures), spinal stenosis and tumors (primary and metastatic), failed previous fusions (pseudarthrosis) and deformity (defined as scollosis, kyphosis or lordosis).

6. Comparison of the technological characteristics of the device to predicate and legally marketed

devices :

The Pyrenees Cervical Plate System was mechanically tested and compared to the predicate systems and other currently marketed systems. The Pyrenees system performed equally to or better than these systems in static torsion, dynamic compression, and static tension testing per ASTM F1717. The design features and sizing of the components were also compared and the Pyrenees Cervical Plate System found to be substantially the same as these systems. It is manufactured from the same materials and is indicated for the same intended uses as these systems.

There are no significant differences between the Pyrenes Cervical Plate Systems currently being marketed. It is substantially equivalent to these other devices in design, function, material and intended use.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are arranged in a circular fashion around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

K2M, Inc. % Ms. Nancy Giezen Manager, Regulatory Affairs 751 Miller Drive SE, Suite F1 Leesburg, Virginia 20175

AUG 17 2011

Re: K111135

Trade/Device Name: Pyrenees Cervical Plate System, Translational Plate Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: July 19, 2011 Received: July 20, 2011

Dear Ms. Giezen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{2}------------------------------------------------

Page 2 - Ms. Nancy Giezen

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours

Mark N. Mellema

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

KITI 113 S

Indications for Use

510(k) Number: Pending

Device Name: Pyrenees Cervical Plate System, Translational Plates

Indications for Use:

The Pyrenees and K2M Cervical Plate Systems are indicated for use in anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (DDD), spondylolisthesis, trauma (including fractures), spinal stenosis and tumors (primary and metastatic), failed previous fusions (pseudarthrosis) and deformity (defined as scoliosis, kyphosis or lordosis).

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-the-counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS-LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign C

(Division Sign-() Division of Surgical, Orthopedic, and Restorative ﺗﺤﺘﺎﺭ ﻴﻨﺎ ﻋ

510(k) Number_ KIIII 35

N/A