K Number
K971883

Validate with FDA (Live)

Manufacturer
Date Cleared
1997-10-16

(147 days)

Product Code
Regulation Number
888.3060
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Small Stature Anterior Cervical Vertebrae Plate System (as part of the Anterior Cervical Vertebrae Plate System) is intended for anterior screw fixation to the cervical spine (C2-C7) for the following indications:

  1. Spondylolisthesis
  2. Fracture
  3. Spinal stenosis
  4. Tumor
Device Description

The Synthes Small Stature Anterior Cervical Vertebrae Plate consists of a plate with at least 4 screws and is manufactured from CP titanium. The plate has 6° of angulation, is available in lengths from 20mm is 2mm thick and is 16.49mm wide. The 20mm - 26mm plates have four holes: the 34mm - 47mm plates have six holes. 30mm -59mm have eight holes. and 62mm has 10 holes . The plate accepts a 3.5mm cancellous screw. a 3.5mm corex screw, or a 4.0mm or 4.35mm expansion screw with 1.8mm locking screw. The expansion screws provide a locking mechanism to reduce the risk of the screw backing out with subsequent loss of fixation. It is inserted (screwed) into the bone through the plate, and tightened until the screw head is fully seated in the plate. The 1.8mm locking screw is then inserted into the expansion screw by way of the mating threads (1.8mm). As the locking screw is tightened, the conical head of this screw engages the conical bore in the head of the expansion screw. In addition to the conical bore, the head of the expansion screw is slotted so that further tightening of the locking screw serves to expand the four sections of the expansion screw head outward. This outward (radial) expansion of the expansion screw head results in a locking force created between the screw and the plate. Implant removal is accomplished in the reverse order. The device functions as follows: The placed on the anterior aspect of the vertebral body of the cervical spine by inserting a 3.5mm cancellous screw, a 3.5mm cortex screw, or a 4.0mm or 4.35mm expansion screw with 1.8mm locking screw through the cephalad holes and into the vertebral body. The screws, or expansion screws with locking screws are then inserted into the caudal holes in the plate and inserted into the vertebral bodies of the cervical spine. It is manufactured from CP titanium, which conforms to ASTM Standard F67. This system is provided non-sterile; moist heat sterilization is recommended.

AI/ML Overview

The provided text is a 510(k) summary for a medical device: the SYNTHES Spine Small Stature Anterior Cervical Vertebrae Plate System. It details the device's indications for use and compares it to a predicate device for substantial equivalence.

However, the document does not contain information about specific acceptance criteria or a study proving the device meets those criteria, as typically found in performance studies for AI/software devices. This document is a regulatory submission focused on substantial equivalence of a physical implant (a spinal plate system), not an AI or software device. As such, many of the requested categories (like sample size for test/training sets, expert ground truth adjudication, MRMC studies, or standalone algorithm performance) are not applicable or present in this type of submission.

Therefore, I cannot provide a table of acceptance criteria and reported device performance from this text, nor can I fill in the details for a study that proves it meets those criteria in the context of an AI/software device.

The information that can be extracted relates to the device's intended use and the regulatory finding of substantial equivalence:

Device: SYNTHES Spine Small Stature Anterior Cervical Vertebrae Plate System

Intended Use/Indications:
Anterior screw fixation to the cervical spine (C2-C7) for:

  1. Spondylolisthesis
  2. Fracture
  3. Spinal stenosis
  4. Tumor

Regulatory Finding: Substantially equivalent to the Synthes Anterior Cervical Vertebrae Plate System (K945700 & K926453).

Regarding the specific questions posed:

  1. A table of acceptance criteria and the reported device performance: Not applicable/not present in this type of document. The regulatory pathway here is substantial equivalence to a predicate, not performance against specific, quantifiable acceptance criteria in the manner of an AI/software device.
  2. Sample size used for the test set and the data provenance: Not applicable. This is a physical device, not an AI model evaluated on a data set.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth, in the context of AI, doesn't apply to this physical device submission.
  4. Adjudication method: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI device.
  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is not an AI device.
  7. The type of ground truth used: Not applicable. The "ground truth" for this device revolves around its material properties (conforming to ASTM Standard F67 for CP titanium), mechanical design, and clinical indications being similar to the predicate device.
  8. The sample size for the training set: Not applicable. This is not an AI device.
  9. How the ground truth for the training set was established: Not applicable.

This document is a 510(k) Pre-market Notification, which focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing detailed performance data from a clinical trial or AI model validation study. The "study" here is implied by the comparison to the predicate device and adherence to material standards.

{0}------------------------------------------------

Summary of Jafety and Effectiveness Information 1971883 [510(k) Summary]

OCT 1 6 1997

SYNTHES (U.S.A.) 1690 Russell Road Paoli, PA 19301

(610) 647-9700 Contact: Barry E. Sands 2/8/97

Device: SYNTHES Spine Small Stanre Anterior Cervical Vertebrae Plate System (as part of the Anterior Cervical Vertebrae Plate System) compared to the Synthes Anterior Cervical Vertebrae Plate System (K945700 & K926453).

The Synthes Small Stature Anterior Cervical Vertebrae Plate consists of a plate with at least 4 screws and is manufactured from CP ritanium. The Small Stature Anterior Cervical Vertebrae Plate System (as part of the Anterior Cervical Vertebrae Plate System) is intended for anterior screw fixation to the cervical spine (C2-C7) for the following indications:

    1. Spondvlolisthesis
    1. Fracture
    1. Spinal stenosis
    1. Tumor

The plate has 6° of angulation, is available in lengths from 20mm is 2mm thick and is 16.49mm wide. The 20mm - 26mm plates have four holes: the 34mm - 47mm plates have six holes. 30mm -59mm have eight holes. and 62mm has 10 holes . The plate accepts a 3.5mm cancellous screw. a 3.5mm corex screw, or a 4.0mm or 4.35mm expansion screw with 1.8mm locking screw

The expansion screws provide a locking mechanism to reduce the risk of the screw backing out with subsequent loss of fixation. It is inserted (screwed) into the bone through the plate, and tightened until the screw head is fully seated in the plate. The 1.8mm locking screw is then inserted into the expansion screw by way of the mating threads (1.8mm). As the locking screw is tightened, the conical head of this screw engages the conical bore in the head of the expansion screw. In addition to the conical bore, the head of the expansion screw is slotted so that further tightening of the locking screw serves to expand the four sections of the expansion screw head outward. This outward (radial) expansion of the expansion screw head results in a locking force created between the screw and the plate. Implant removal is accomplished in the reverse order.

The device functions as follows: The placed on the anterior aspect of the vertebral body of the cervical spine by inserting a 3.5mm cancellous screw, a 3.5mm cortex screw, or a 4.0mm or 4.35mm expansion screw with 1.8mm locking screw through the cephalad holes and into the vertebral body. The screws, or expansion screws with locking screws are then inserted into the caudal holes in the plate and inserted into the vertebral bodies of the cervical spine.

It is manufactured from CP titanium, which conforms to ASTM Standard F67.

The Synthes Anterior Cervical Vertebrae Plate is indicated for the same clinical indications. Material composition is identical to the Small Stature Anterior Cervical Vertebrae Plate System. The surgical technique and instrumentation to implant this system is also the same.

This system is provided non-sterile; moist heat sterilization is recommended.

Small Stature ACVPS

10/8/97

{1}------------------------------------------------

Based on the above, the SYNTHES Spine Small Stature Anterior Cervical Vertebrae Plate System (as part of the Anterior Cervical Vertebrae Plate System) is substantially equivalent to the Synthes Anterior Cervical Vertebrae Plate System.

Small Stature ACVPS

10/8/97

ర్

{2}------------------------------------------------

Image /page/2/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human form.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 16 1997

Mr. Barry Sands Manager, Regulatory Affairs Synthes (USA) P.O. Box 0548 1690 Russell Road Paoli, Pennsylvania 19301

K971883 Re: Synthes Spine Small Stature Anterior Cervical Vertebrae Plate System (as part of Synthes Anterior Cervical Vertebrae Plate System, K945700) Regulatory Class: II Product Code: KWQ Dated: Auqust 19, 1997 Received: August 20, 1997

Dear Mr. Sands:

We have reviewed your Section 510(k) notification of intent to market the device system referenced above and we have determined the device system is substantially equivalent (for the indications for use stated in the enclosure) to device systems marketed in interstate commerce prior to May 28, 1976 or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act You may, therefore, market your device system subject (Act) . . to the general controls provisions of the Act and the limitations identified below.

The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Note that labeling or otherwise promoting this device system for pedicular screw fixation/attachment would cause the device system to be adulterated under 501(f)(1) of the Act.

FDA identifies that any device system, if intended for use in pedicular screw fixation/attachment, except for some limited indications, would be found not substantially equivalent and would be a class III device under Section 513 (f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing. Accordingly:

{3}------------------------------------------------

Page 2 - Mr. Barry Sands

  • You may not label or in any way promote this device 1. system for pedicular screw attachment to, or fixation of the cervical, thoracic or lumbar vertebral column. Therefore, in order to prevent off-label promotion, the package insert must include the following statement, This device system is not approved for screw "WARNING: attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.";
  • All labeling for this device system, including the 2. package label, must state that there are labeling limitations. The package insert must prominently state that the device system is intended for the specific use(s) described in the enclosure only; and
  • Pedicular screw fixation/attachment to the cervical, 3 . thoracic or lumbar vertebral column, except for limited indications, of any device system is considered investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulations under 21 CFR, Part 812. All users of the device system for pedicular screw fixation/attachment must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) prior to conducting an investigation.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, · through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

{4}------------------------------------------------

Page 3 - Mr. Barry Sands

FDA advises that the use of the subject device system and/or device components with any other device components but those identified in this 510(k) would require submission of a new 510(k) providing documentation of design, material, and labeling compatibility between the device components. Mechanical testing of a spinal system consisting of the subject device components and other device components, whether vours or other manufacturers', may also be required.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

to cselee

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

CONFIDENTIAL: The contents of this document are proprietary to Synthes and should be considered contidential

Page 1 of 1

510(k) Number (if known): K971883

Device Name: Small Stature Anterior Cervical Vertebrae Plate System (as part of the Anterior Cervical Vertebrae Plate System)

Indications for Use:

The Small Stature Anterior Cervical Vertebrae Plate System (as part of the Anterior Cervical Vertebrae Plate System) is intended for anterior screw fixation to the cervical spine (C2-C7) for the following indications:

    1. Spondylolisthesis
    1. Fracture
    1. Spinal stenosis
    1. Tumor

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)
OR
Over-the-Counter Use

(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K97883

§ 888.3060 Spinal intervertebral body fixation orthosis.

(a)
Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.(b)
Classification. Class II.