K Number
K111134

Validate with FDA (Live)

Date Cleared
2012-01-26

(279 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

Non sterile, Powder free Nitrile Examination Glove, Blue.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

CharacteristicAcceptance Criteria (Standards)Reported Device Performance
Dimensions: overall length; width, palm and finger thicknessASTM D 6319-10Meets
Tensile strength: before and after agingASTM D 6319-10Meets
Ultimate elongation: before and after agingASTM D 6319-10Meets
Freedom from holes: pinholes AQL 2.5ASTM D 6319-10Meets
Powder Free ResidueASTM D 6319-10Meets
BiocompatibilityPrimary Skin Irritation in Rabbits, Guinea Pig SensitizationPasses

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not specified in the provided text. The document states "The performance test data of the non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above," referring to the table of characteristics, but does not detail the sample sizes used for each test.
  • Data Provenance: Not explicitly stated as retrospective or prospective, or country of origin for the data collection itself. However, the tests cited are ASTM standards, which are internationally recognized. The manufacturer is Siam Sempermed Corp., Ltd from Thailand.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. This device is a nitrile examination glove. The "ground truth" for its performance is established through standardized physical and biocompatibility testing (e.g., tensile strength, freedom from holes, skin irritation), not through expert interpretation of medical images or clinical outcomes that would require adjudicators.

4. Adjudication method for the test set:

  • Not Applicable. As mentioned above, the evaluation of this device relies on standardized physical and chemical tests against ASTM standards, not on subjective assessment or consensus among experts.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This device is a patient examination glove and does not involve AI or human "readers" in its intended use or evaluation.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This device is a physical medical device, not an algorithm.

7. The type of ground truth used:

  • The "ground truth" is based on objective measurements and observations according to established industry standards (ASTM D 6319-10) for physical properties (dimensions, tensile strength, elongation, freedom from holes, powder residue) and standardized biological assays for biocompatibility (Primary Skin Irritation in Rabbits, Guinea Pig Sensitization).

8. The sample size for the training set:

  • Not Applicable. This device concerns a physical product tested against standards, not a machine learning model requiring a training set.

9. How the ground truth for the training set was established:

  • Not Applicable. There is no training set for this type of device.

{0}------------------------------------------------

K111134

510 (k) Summary

As Required by 21 section 807.92 (c)

JAN 2 6 2012

    1. Submitter Name: Siam Sempermed Corp., Ltd
      352 Kanjanavanit Road. Pahtong Hatyai 2. Address: Songkhla. Thailand 90230
  • (+66) 74 291 471 3. Phone:
    . .},

    1. Fax: (+66) 74 471 111
  1. Contract Person: Mrs. Parawan Paiyasan (Quality System Manager)
  • Date summary prepared: November , 20 2010 6.
  1. Official Correspondent: Sempermed USA Inc.
  • 13900 49th Street North 8. Address:
    Clearwater, USA , FL 33762

    1. Phone: 727 787 7250
  • 727 787 7558 10. Fax:

    1. Contact person: Mr. William E. Harris
  1. Device Trade or Proprietary Name: Non sterile, Powder free Nitrile Examination Glove, Blue.
    1. Device Common or usual name: Examination glove
  1. Device Classification Name: Nitrile Patient Examination Glove (Powder free, Blue color)
    1. Description of the Device:
    • Non sterile, Powder free Nitrile Examination Glove, Blue.
  1. Indications for use of the device: Based on 21 CFR Part 880.6250: "Patient examination glove." A patient examination glove is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

17. Summary of The Technological Characteristics of The devices:

Non sterile, Powder free Nitrile Examination Glove, Blue are summarized with the following technological characteristics compared to ASTM or equivalent standards.

CHARACTERISTICSSTANDARDSDEVICE PERFORMANCE
Dimensions: overall length; width, palmand finger thicknessASTM D 6319-10Meets
Tensile strength: before and after agingASTM D 6319-10Meets
Ultimate elongation: before and afteragingASTM D 6319-10Meets
Freedom from holes: pinholes AQL 2.5ASTM D 6319-10Meets
Powder Free ResidueASTM D 6319-10Meets
BiocompatabilityPrimary Skin Irritation inRabbitsGuinea Pig SensitizationPasses
Dimensions: overall length; width, palmand finger thicknessASTM D 6319-10Meets

{1}------------------------------------------------

18. Substantial Equivalents Based on Assessment of Non-Clinical Performance Data

The performance test data of the non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above.

19. Conclusion

、、。

ﻪ ﻣﻌﻪ

It can be concluded that the Non sterile, Powder free Nitrile Examination Glove, Blue will perform of according to the glove performance standards referenced in section 17 above and meet ASTM standards, and EDA requirements. Consequently, this device is substantially equivalent to currently marketed devices. This device is as safe and effective as the predicate device Siam Sempermed Non-sterile, Powder-free Nitrile Examination Glove, Blue with Polymer Coating, Tested for Use with Chemotherapy Drugs. Indeed, it is equivalent. This is better expressed in the tabulated comparison as below.

FDA file reference number ofpredicate device510k number : K080520 (Nitrile Examination glove,Powder Free with polymer coating, Blue Color,Non-sterile, Glove were tested chemotherapydrugs)
TECHNOLOGICALCHARACTERISTICSSee additional details in main submission, Section4, Subpart H
Indications for useIdentical
Target populationIdentical
DesignSimilar; new device does not have inner coating
MaterialsSimilar; new device does not have inner coating
PerformanceIdentical
SterilityIdentical (Not applicable)
BiocompatibilityIdentical
Mechanical safetyIdentical
Chemical safetyIdentical
Anatomical sitesIdentical
Human factorsIdentical
Energy used and/or deliveredIdentical (Not applicable)
Compatibility with environmentand other devicesIdentical
Where usedIdentical
Standards metIdentical
Electrical safetyIdentical (Not applicable)
Dimensional standardsidentical
Physical propertiesIdentical
Thermal safetyIdentical (Not applicable)
Radiation safetyIdentical (Not applicable)
Production processSimilar; new device is online chlorinated. Seefollowing page for details.
LabelingSimilar: new device lacks a chemotherapy labelingclaim and associated statements

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the human services aspect of the department. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Siam Sempermed Corporation, Limited C/O Mr. William E. Harris, III · President & CEO Sempermed USA, Incorporated 13900 49th Street Clearwater, Florida 33762

JAN 2 6 2012

Re: K11134

Trade/Device Name: Non-Sterile, Powder-Free Nitrile Examination Glove, Blue Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: January 5, 2012 Received: January 6, 2012

Dear Mr. Harris:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

{3}------------------------------------------------

Page 2 - Mr. Harris

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the I ou him) of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Susan Kummer

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K111134

Device Name: Non-sterile, Powder-free Nitrile Examination Glove, Blue

Indications For Use: A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices

510(k) Number:K111134
Page 1 of

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.