K Number
K110947

Validate with FDA (Live)

Device Name
APEX K1 HIP STEM
Date Cleared
2011-06-16

(73 days)

Product Code
Regulation Number
888.3358
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Apex Hip System K1™ Hip Stem is intended for use as the femoral component of a primary or revision total hip replacement when used with the Apex Interface™ Acetabular System. The Apex Interface™ Acetabular System articulates with the Apex Modular Femoral Head (Cobalt Chromium or Ceramic). The femoral hip stem is intended for uncemented fixation and single use implantation. These prosthesis may be used for hip arthroplasty to treat the following conditions, as appropriate:
• Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
• Rheumatoid arthritis;
• Correction of functional deformity;
• Congenital dislocation;
• Revision procedures where other treatments or devices have failed;
• Femoral neck and trochanteric fractures of the proximal femur.

The Apex Hip System K1 Femoral Stem is also intended for use in hemiarthroplasty when used with the Apex Bipolar Head.

The Apex Hip System Bipolar Head is intended for use in combination with an Apex Hip System femoral stem for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:
• Femoral neck and trochanteric fractures of the proximal femur;
• Osteonecrosis of the femoral head;
• Revision procedures where other treatments or devices for these indications have failed.

Device Description

The Apex K1™ Hip Stems are rectangular in cross-section and tapered in the proximal-to-distal direction. The K1 Hip Stem is offered in 12 sizes and two neck offsets for each size. Standard necks have a 135 degree angle and lateralized (Lat) necks have a 130 degree angle. Sizes 7 to 12 also offer a Lat + offset with a 127 degree angle.

AI/ML Overview

The provided text is a 510(k) Summary for the Apex K1™ Hip System, a medical device for hip replacement. It outlines the device's characteristics, intended use, and a comparison to a predicate device. However, it explicitly states: "No clinical studies were performed."

Therefore, it is impossible to provide the requested information about acceptance criteria, device performance, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, or ground truth, as no clinical study was conducted for this device.

The document indicates that the device's substantial equivalence to a legally marketed predicate device (K060072 - Apex K1™ Hip Stem) was established through non-clinical testing.

Here's a summary based on the provided text, highlighting the absence of clinical study data:


Absence of Clinical Study Data:

The provided 510(k) Summary explicitly states: "No clinical studies were performed." Therefore, the device performance against acceptance criteria, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, and ground truth establishment, which are typically derived from clinical studies, cannot be detailed.

Instead, the device's substantial equivalence was demonstrated through non-clinical testing and comparison to an existing predicate device.

Non-Clinical Test Summary:

The following non-clinical tests were conducted:

  • Fatigue Strength Testing: Performed according to ISO 7206-6, ISO-7206-4, ISO 7206-8, and ASTM 2068-09.
  • ROM (Range of Motion) evaluation: Performed per ISO 21535.

Predicate Device Comparison:

The Apex K1™ Hip Stem (subject device, sizes 0, 0 Lat, 1, 1 Lat, and 7 thru 12 Lat +) was found to be substantially equivalent to the predicate device (Apex K1™ Hip Stem, Sz 2 to 12, K060072) based on the following:

CharacteristicApex K1 Hip Stem (Subject Device)Apex K1 Hip Stem (Predicate Device, K060072)
Intended UseYes (for primary and revision total hip replacement)Yes (for primary and revision total hip replacement)
Stem DesignIdentical to K060072Rectangular in cross-section and tapered in the proximal-to-distal direction
Materials (K1 Stem)Identical to K060072Ti6Al4V per ASTMF136
Plasma spray Titanium coatingIdentical to K060072Ti Coating per ASTM F1580
SterilizationIdentical to K060072Ethylene oxide
SAL (Sterility Assurance Level)Identical to K06007210^-6^
PackagingIdentical to K060072Paper Board Box, Double Tyvek inner pouch

Since no clinical studies were performed, the specific details requested for clinical performance acceptance criteria, study design, and ground truth establishment cannot be provided from this document. The basis for acceptance for this device was its substantial equivalence to a previously cleared predicate device, supported by the mentioned non-clinical engineering and material testing.

{0}------------------------------------------------

JUN 1 6 2011

510(K) Summary JUN 16 2011
16 April, 2011
SubmitterContact
OMNIlife science, Inc.50 O'Connell WayE. Taunton MA 02718Radhika PondicherryRegulatory Affairs774-226-1852(508) 822-6030 (fax)
Preparation Date16 April, 2011
Device NameApex Hip System
Trade NameApex K1 ™ Hip Stem,
SizesApex K1™ Hip Stem, Size 0Apex K1™ Hip Stem, Size 0 LatApex K1™ Hip Stem, Size 1Apex K1™ Hip Stem, Size 1 LatApex K1™ Hip Stem, Size 7 Lat +Apex K1™ Hip Stem, Size 8 Lat +Apex K1™ Hip Stem, Size 9 Lat +Apex K1™ Hip Stem, Size 10 Lat +Apex K1™ Hip Stem, Size 11 Lat +Apex K1™ Hip Stem, Size 12 Lat +
Common name/ClassificationHip joint metal/polymer/metal semi-constrained porous-coated uncementedprosthesisHip joint metal/ceramic/polymer semi-constrained cemented or nonporousuncemented prosthesis.Hip joint femoral (hemi-hip) metal/polymer cemented or uncementedprosthesis.
Regulatory ClassClass II per 21 CFR §888.3358, §888.3353, §888.3390
Product CodeLPH, LZO, KWY
Legally MarketedPredicate Device(s)K060072- Apex K1™ Hip Stem
Device DescriptionThe Apex K1™ Hip Stems are rectangular in cross-section and tapered in theproximal-to-distal direction. The K1 Hip Stem is offered in 12 sizes and twoneck offsets for each size. Standard necks have a 135 degree angle andlateralized (Lat) necks have a 130 degree angle. Sizes 7 to 12 also offer a Lat +offset with a 127 degree angle.
Indications for UseThe Apex Hip System K1™ Hip Stem is intended for use as the femoralcomponent of a primary or revision total hip replacement when used with theApex Interface™ Acetabular System. The Apex Interface™ Acetabular Systemarticulates with the Apex Modular Femoral Head (Cobalt Chromium orCeramic). The femoral hip stem is intended for uncemented fixation andsingle use implantation. These prosthesis may be used for hip arthroplasty totreat the following conditions, as appropriate:• Non-inflammatory degenerative joint disease, includingosteoarthritis and avascular necrosis;• Rheumatoid arthritis;• Correction of functional deformity;

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・

·

{1}------------------------------------------------

K110947

p. 2/2

  • · Congenital dislocation;
  • · Revision procedures where other treatments or devices have failed;
  • · Femoral neck and trochanteric fractures of the proximal femur.

The Apex Hip System K1 Femoral Stem is also intended for use in hemiarthroplasty when used with the Apex Bipolar Head.

The Apex Hip System Bipolar Head is intended for use in combination with an Apex Hip System femoral stem for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:

  • · Femoral neck and trochanteric fractures of the proximal femur;
  • · Osteonecrosis of the femoral head;
  • · Revision procedures where other treatments or devices for these indications have failed.
Apex K1 Hip Stem,size 0, 0 Lat, 1, 1 Latand 7 thru 12 Lat +(subject device)Apex K1 Hip Stem,Sz 2 to 12(K060072)
Intended UseYesYes
Primary and revisiontotal hip replacementYesYes
Design
Stem DesignIdentical to K060072Rectangular in cross-section and tapered in theproximal-to-distaldirection
Materials
K1 StemIdentical to K060072Ti6Al4V per ASTMF136
Plasma sprayTitanium coatingIdentical to K060072Ti Coating per ASTM F1580
PACKAGING ANDSTERILIZATION
SterilizationIdentical to K060072Ethylene oxide
SalIdentical to K06007210-6
PackagingIdentical to K060072Paper Board Box, DoubleTyvek inner pouch

Non-Clinical Test Summary

The following tests were conducted:

  • . Fatigue Strength Testing per ISO 7206-6, ISO-7206-4, ISO 7206-8 and ASTM 2068-09
  • . ROM evaluation per ISO 21535

Clinical Test Summary No clinical studies were performed.

Conclusions The Apex K1 Stems, sizes 0, 0 Lat, 1, 1 Lat, Lat + stems sizes 7,8,9,10,11 and 12 are substantially equivalent to the predicate devices.

Predicate Device Comparison

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a bird-like figure with three lines forming the body and wings.

Food.and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

OMNIlife Science, Inc. % Ms. Radhika Pondicherry Regulatory Affairs Specialist 50 O'Connell Way East Taunton, Massachusetts 02718

JUN 1 6 2011

Re: K110947 Trade/Device Name: Apex K 1 Hip System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO, KWY Dated: May 17, 2011 Received: May 18, 2011

Dear Ms. Pondicherry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Ms. Radhika Pondicherry

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number: K110947

Device Name: Apex K1™ Hip System

The Apex Hip System K1™ Hip Stem is intended for use as the femoral component of a primary or revision total hip replacement when used with the Apex Interface™ Acetabular System. The Apex Interface™ Acetabular System articulates with the Apex Modular Femoral Head (Cobalt Chromium or Ceramic). The femoral hip stem is intended for uncemented fixation and single use implantation. These prosthesis may be used for hip arthroplasty to treat the following conditions, as appropriate:

  • · Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
  • Rheumatoid arthritis;
  • · Correction of functional deformity;
  • · Congenital dislocation;
  • · Revision procedures where other treatments or devices have failed;
  • · Femoral neck and trochanteric fractures of the proximal femur.

The Apex Hip System K1 Femoral Stem is also intended for use in hemiarthroplasty when used with the Apex Bipolar Head.

The Apex Hip System Bipolar Head is intended for use in combination with an Apex Hip System femoral stem for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate:

  • Femoral neck and trochanteric fractures of the proximal femur;
  • · Osteonecrosis of the femoral head;
  • Revision procedures where other treatments or devices for these indications have failed.

AND/OR Over-The-Counter Use Prescription Use x (Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

Page 1 of 1

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Emil Reith

f Surgical, Orthopedic,

510(k) Number K110947

§ 888.3358 Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis.

(a)
Identification. A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.(b)
Classification. Class II.