K Number
K110418

Validate with FDA (Live)

Date Cleared
2011-07-13

(149 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Nitrile Examination Gloves is a disposable device intended for medical purposes that is intended to be worn on the hand for medical purpose to provide barrier against potentially infectious materials and other contaminants.

Device Description

Powder Free Nitrile Examination Gloves is substantially equivalent to the Class 1 patient examination glove bearing the product code 80LZA ( 21 CFR 880.6250 ). It meets all the current specifications listed under the ASTM Specification D 6319 - 10 Standard Specification for Nitrile Examination Gloves for Medical Application¹. They are made from nitrile compound(dispersion of butadiene acrylonitrile copolymer. They are powder free gloves.

AI/ML Overview

The provided text describes powdered-free nitrile examination gloves and their compliance with ASTM standards, rather than an AI/ML powered device. Therefore, I cannot extract the requested information about acceptance criteria and study details for an AI/ML device from this document.

However, I can extract the acceptance criteria and performance as described for the gloves.

1. Table of acceptance criteria and the reported device performance

Acceptance Criteria (ASTM D 6319-10)Reported Device Performance
Physical Properties (Dimensions, Tensile Strength, Elongation)Meets all current specifications
Freedom from holes (AQL for pinholes)Meets AQL requirements
Barrier integrityMeets requirements
Biocompatibility (Primary Skin Irritation)No irritation observed in rabbits
Biocompatibility (Delayed Dermal Contact Sensitization)No sensitization observed in guinea pigs

Specific numerical values for tensile strength, elongation, and AQL for pinholes are not explicitly provided in the summary but are stated to meet the ASTM D 6319-10 standard. The document states, "Gloves meet all the current ASTM D 6319-10." and "Non-Clinical laboratory and animal based test data indicate that the powder free product meets all performance and biocompatibility requirements."

2. Sample size used for the test set and the data provenance

The document does not explicitly state the sample size used for the test set for all physical and barrier integrity tests. For biocompatibility tests:

  • Primary skin irritation: Conducted in rabbits (number of rabbits not specified).
  • Delayed dermal contact sensitization: Conducted in guinea pigs (number of guinea pigs not specified).
  • Data provenance: The testing was performed in Thailand by W.A. Rubbermate Co., Ltd. The study appears to be non-clinical (laboratory and animal-based).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable as the device is a medical glove, not an AI/ML device requiring expert ground truth for classification or diagnosis. Performance is based on standardized physical and biocompatibility testing.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable for this type of device and testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable, as this is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable, as this is not an AI/ML device.

7. The type of ground truth used

For the glove's performance:

  • Physical properties: Measured values (e.g., dimensions, tensile strength, elongation) compared against the specifications in ASTM D 6319-10.
  • Freedom from holes: Determined by established inspection methods (e.g., water leak test) and statistical sampling plans (AQL) as defined in ASTM D 6319-10.
  • Biocompatibility: Observed biological responses (irritation, sensitization) in animal models compared against established safety criteria.

8. The sample size for the training set

Not applicable, as this is not an AI/ML device. There is no "training set" in the context of physical glove testing.

9. How the ground truth for the training set was established

Not applicable, as this is not an AI/ML device.

{0}------------------------------------------------

KI 10418

Powder Free Nitrile Examination Gloves

:

. :

. . .

. . . . . . . . .

.

・・・

· .

!

,

.

ः ।

Attachment 13

510(k) SUMMARY

JUL 13 2011

POWDER FREE NITRILE EXAMINATION GLOVES

Submitter's NameW.A RUBBERMATE CO., LTD
Submitter's Address67/1 Bypass Road, Moo 4,Nongmaidaeng, A. Maung,Chonburi 20000, Tailand
Submitter's Phone Number+662-3189442
Submitter's Fax Number+662-3183268
Name of Contact PersonTerence Lim Sin Kooi
Date of Preparation08 February 2011
Name of Device
Trade Name: POWDER FREE NITRILE EXAMINATIONGLOVES
Common Name: Nitrile Examination Gloves
Classification Name: Patient Examination Gloves
Legally Marketed Device to which Equivalency isBeing ClaimedPowder Free Nitrile Examination Gloves asdescribed in this 510 K Notification issubstantially equivalent to the current Class 1Patient Examination glove bearing the productcode 80LZA ( 21 CFR 880.6250 ). It meets all thecurrent specifications listed under the ASTMSpecification D 6319 - 10 Standard Specificationfor Nitrile Examination Gloves for MedicalApplication.
Description of the DevicePowder Free Nitrile Examination Gloves issubstantially equivalent to the Class 1 patientexamination glove bearing the product code
80LZA ( 21 CFR 880.6250 ). It meets all thecurrent specifications listed under the ASTMSpecification D 6319 - 10 Standard Specificationfor Nitrile Examination Gloves for MedicalApplication¹. They are made from nitrilecompound(dispersion of butadiene acrylonitrilecopolymer. They are powder free gloves.
Intended Use of the DevicePowder Free Nitrile Examination Gloves areintended for single use for medical purposes thatis worn on the hand of health care and similarpersonnel to prevent contamination between thehealth care personnel and the patients.
Summary of Technological CharacteristicCompared to the Predicate DeviceThere is no different technological characteristic.Gloves are made from nitrile compound(dispersion of butadiene acrylonitrile copolymer)and the initial products are powder free nitrileexamination gloves.
Brief Description of Non-Clinical TestsTesting performed per ASTM D 6319 - 10Standard Specification for Nitrile ExaminationGloves for Medical Application and 21 CFR800.20. Gloves meet all the current ASTM D6319-10.Primary skin irritation testing in the rabbit anddelayed dermal contact sensitization study in theguinea pigs indicate no irritation or sensitization.
Brief description of Clinical TestsNo new clinical tests were conducted under this510(k).
Conclusions Drawn from the Non-Clinical andClinical TestsNon-Clinical laboratory and animal based testdata indicate that the powder free product meetsall performance and biocompatibilityrequirements.
Other Information Deemed Necessary by FDANot Applicable

.

{1}------------------------------------------------

K110418

Powder Free Nitrile Examination Gloves

Attachment 13

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and features the department's name around the perimeter. In the center is a stylized image of an eagle or bird-like figure, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Terence Lim Sin Kooi Head of Regulatory Affairs W.A Rubbermate Company, Limited 67/1 Bypass Road, Moo 4 Nongmaidaeng A. Maung Chonburi, THAILAND 20000

JUL 13 2011

Re: K110418

Trade/Device Name: Powder Free Nitrile Examination Gloves (Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: June 29, 2011 Received: July 6, 2011

Dear Mr. Lim Sin Kooi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Lim Sin Kooi

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Anthony D. anton

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Powder Free Nitrile Examination Gloves

Attachment 2

INDICATIONS FOR USE

Applicant

W.A RUBBER MATE CO., LTD. 4 Ramkhamhaeng 19 (Chareonploy), Ramkhamhaeng Road, Huamark, Bangkapi, Bangkok 10240, Thailand

510(k) Number (if known)

K110418 *

Device Name

POWDER FREE NITRILE EXAMINATION GLOVES ( BLUE

Indications For Use

Powder Free Nitrile Examination Gloves is a disposable device intended for medical purposes that is intended to be worn on the hand for medical purpose to provide barrier against potentially infectious materials and other contaminants.

Concurrence of CDRH Office of Device Evaluation (ODE )

Prescription Use Per 21 CFR 801.109

× OR Over-The-Counter_

Rosie Pangulung

Vivision of Anesthesiology, General Hospital nfection Control, Dental Devices

510(k) Number: K110418

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.