K Number
K110413

Validate with FDA (Live)

Date Cleared
2011-05-11

(86 days)

Product Code
Regulation Number
862.1810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The basic Power Processor is an automated sample handling system which processes sample tubes from the pre-centrifugation, pre-sorting step to presentation of centrifuged and decapped samples into Generic or Personality Racks for specific instruments. The Power Processor can be configured with optional software and hardware to allow processing of sample tubes on Generic Connection Instruments. The Power Processor performs all pre-analytical sample tube preparation, and then sorts the sample tubes directly to Generic Connection Modules where the samples are pipetted by the Generic Connection instrument for testing. After the samples are pipetted, the tubes can route to other instruments for additional testing or to Outlet Racks.

The UniCel DxI 600/800 Access Immunoassay System(s) with laboratory automation connection is a microcomputer-controlled, random and continuous access analyzer that includes an external computer. This computer stores the system user interface (U) software and allows the operator to interface with and direct the instrument software. The UniCel DxI 600/800 System(s) uses enzyme immunoassays (utilizing paramagnetic particle solid phase and chemiluminescent detection) for determination of various analytes, such as Vitamin B12, Ferritin, Folate and hTSH along with other various enzyme immunoassays assays that may be adaptable to the analyzer depending on the reagent used to induce the enzyme immunoassay reaction. The UniCel DxI 600/800 System(s) is an in vitro diagnostic device for use in the clinical laboratory.

The Access Ferritin assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of ferritin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism.

The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum, plasma (heparin) and red blood cells using the Access Immunoassay Systems. Folic acid measurements are used in the diagnosis and treatment of megaloblastic anemia.

The Access HYPERsensitive hTSH assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of human thyroid-stimulating hormone (thyrotropin, hTSH) levels in human serum using the Access Immunoassay Systems. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.

The Access Vitamin B12 assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of vitamin B12 in human serum and plasma (heparin) using Access Immunoassay Systems. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.

Device Description

The Power Processor is a scalable laboratory automation system (LAS) designed to streamline pre-analytical processes in the clinical laboratory. A basic Power Processor System is comprised of a Line Control Computer, Preplink™ Computer, Inlet Module, Hematology Module, Centrifugation Module, Decapper Module, and Outlet Module, In the basic configuration, patient sample tubes are loaded onto the Power Processor system to be sorted to a Hematology Module, or to be centrifuged, decapped, and sorted to Personality Racks for further processing on other instruments. Additional modules may be added for aliquot capability, sample capping, and refrigerated storage.

The Power Processor is an open architecture system that can connect to a variety of clinical analyzers, including Beckman Coulter and third party systems. Connection modules are extensions of the Power Processor track system that link with an analyzer's existing LAS interface. Connection modules support one of two types of sample transfer methods: onboard or outboard sampling.

Onboard sampling physically transfers the sample tube/rack from the automation track to the analyzer's sample load and identification area. This method is used to establish connection with Beckman Coulter's Synchron LX® and UniCel® DxC Systems and some third party analyzers.

With outboard sampling, the connection unit performs the sample bar code read function, presents the sample ID to the connected analyzer, and then signals for direct sampling of the open tube by the connected instrument at an aspiration point on the automation track. The Power Processor Generic Connection Module is specifically designed to support the outboard sampling method based on point-in-space pipetting technology aligned with the CSLI guidelines.

AI/ML Overview

The provided text describes a 510(k) submission for a software upgrade to a laboratory automation system, the "Power Processor Sample Processing System with Generic Connection Module to Software Version 5.0." This submission focuses on demonstrating substantial equivalence to a predicate device (the same system with Software Version 3.5). As such, the information provided is not typical for a study demonstrating performance against specific quantitative diagnostic or clinical acceptance criteria as might be seen for a new diagnostic device.

Instead, the "acceptance criteria" here are implicitly related to the successful validation of the software upgrade and its functionality, ensuring it performs as intended and maintains the safety and effectiveness of the existing device. The "study" that proves these criteria are met is the internal validation testing of the software.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Implied)Reported Device Performance
Increased Generic Connection Capacity: Support up to 12 generic connections (from 4).Software Version 5.0 allows the system to connect up to 12 generic connections.
Increased Primary Sample ID Length: Support 13 characters for primary ID + 2 secondary.Software Version 5.0 provides an option to increase the primary sample ID to 13 characters, plus 2 characters for the secondary sample ID.
Increased Stockyard Capacity: Support up to three 3K or three 5K tube stockyards.Software Version 5.0 provides an option to connect up to three 3K or three 5K tube stockyards.
Improved Load Balancing: Option for load balancing by tubes per rack instead of round robin.Software Version 5.0 provides an option for load balancing number of tubes per rack instead of round robin.
Data Migration Utility: Provision for upgrading from V3.5/V3.6 to V5.0.Software Version 5.0 provides Data Migration Utility (DMU) for V3.5/V3.6 to V5.0.
Unassigned Test Tube Routing: Option to route primary tubes with no assigned tests to stockyard.Software Version 5.0 provides an option to route a primary tube with no assigned tests in its test order to the stockyard.
Data Backup/Restore: Option to backup/restore data files on USB drives.Software Version 5.0 provides an option to backup/restore the data files on USB drives.
Overall Software Functionality and Equivalence: All software design, development, and verification activities completed and passed, supporting equivalency to V3.5."Performance data from validation testing shows that all software design, development and verification activities have been completed and passed, and supports equivalency of Power Processor V5.0 to Power Processor V3.5."

2. Sample Size Used for the Test Set and Data Provenance:

The document explicitly mentions "validation testing" but does not provide details on the sample size for this testing (e.g., number of test cases, tubes processed, or specific scenarios simulated).

The data provenance is internal to Beckman Coulter, Inc. as it refers to their internal "software design, development and verification activities." The data would be prospective in the sense that the testing was conducted specifically for this software validation. There is no mention of country of origin for the data itself, but the manufacturer is based in the USA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

This type of information (experts, qualifications) is typically relevant for diagnostic devices where human interpretation or clinical judgment is part of establishing ground truth. For a software upgrade to an automation system, the "ground truth" is defined by the functional specifications and expected behavior of the software. Therefore, no external experts or clinicians are mentioned as being involved in establishing ground truth in the context of this software validation. The "ground truth" for the software's performance is its adherence to its design specifications, verified by the internal testing team.

4. Adjudication Method for the Test Set:

Not applicable. Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among human readers or experts when establishing a ground truth for diagnostic accuracy. Since this is a software upgrade to an automation system, and not a diagnostic device involving human interpretation of results, no adjudication method would be relevant or mentioned. The "adjudication" in a software validation context would be the resolution of any bugs or deviations from specifications identified during testing by the software development and quality assurance teams.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study was not done. MRMC studies are used to assess the impact of a device (often AI-assisted) on human reader performance, typically in diagnostic imaging or similar fields. This submission is for a software upgrade to a laboratory automation system, which does not involve human readers in interpreting clinical data to make a diagnosis in the same way. The software's role is to streamline pre-analytical processes, not to assist humans in diagnostic interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The device itself (Power Processor Sample Processing System with Generic Connection Module V5.0) operates as a standalone automation system that processes samples. The "performance data from validation testing" is essentially a standalone performance assessment of the software and system. There is no mention of human-in-the-loop performance measurement as part of this validation, confirming that the assessment was of the system's inherent functional capabilities.

7. The Type of Ground Truth Used:

The ground truth used for this software validation is based on functional specifications and expected system behavior. The "performance data from validation testing shows that all software design, development and verification activities have been completed and passed" implies that the software's outputs and actions were compared against its predefined and expected operational characteristics. It is not expert consensus, pathology, or outcomes data, as these are related to clinical diagnostic accuracy rather than automation system functionality.

8. The Sample Size for the Training Set:

This document describes a software upgrade and its validation, not the development of a predictive algorithm or AI model that typically requires a "training set." Therefore, no information on a training set sample size is provided or relevant in this context.

9. How the Ground Truth for the Training Set was Established:

As explained in point 8, there is no mention or indication of a "training set" for an algorithm. The software's functionality is developed based on engineering requirements and validated against those requirements.

{0}------------------------------------------------

K110413

Summary of Safety & Effectiveness

Power Processor Sample Processing System with Generic Connection Module V5.0

MAY 1 1 2011

1.0 Submitted By:

Marine Boyajian Senior Regulatory Affairs Specialist Beckman Coulter, Inc. 250 S. Kraemer Blvd., Mail Code E2.SE.08 Brea, California 92822-8000 Telephone: (714) 961-6536 FAX: (714) 961-4234

2.0 Date Submitted:

April 13, 2011

3.0 Device Name(s):

3.1 Proprietary Names

Power Processor Sample Processing System with Generic Connection Module

3.2 Common Name

Laboratory Automation System

3.2 Classification Name

Discrete photometric chemistry analyzer for clinical use (862.2160)

4.0 Predicate Device:

The Power Processor Sample Processing System with Generic Connection Module Software Version 5.0 claims substantial equivalence to the Power Processor Sample Processing System with Generic Connection Module currently in commercial distribution, FDA 510(k) Number K090588.

Candidate(s)PredicateManufacturerDocketNumber
Power Processor SampleProcessing System withGeneric Connection ModuleSoftware Version 5.0Power Processor SampleProcessing System withGeneric Connection ModuleSoftware Version 3.5BeckmanCoulter, IncK090588

5.0 Description:

The Power Processor is a scalable laboratory automation system (LAS) designed to streamline pre-analytical processes in the clinical laboratory. A basic Power Processor System is comprised of a Line Control Computer, Preplink™ Computer, Inlet Module, Hematology Module, Centrifugation Module, Decapper Module, and Outlet Module, In the basic configuration, patient sample tubes are loaded onto the Power Processor system to be sorted to a Hematology Module, or to be centrifuged, decapped, and sorted to Personality Racks for further processing on other instruments. Additional modules may be added for aliquot capability, sample capping, and refrigerated storage.

The Power Processor is an open architecture system that can connect to a variety of clinical analyzers, including Beckman Coulter and third party systems. Connection modules are extensions of the Power Processor track system that link with an analyzer's existing LAS interface. Connection modules support one of two types of sample transfer methods: onboard or outboard sampling.

{1}------------------------------------------------

Onboard sampling physically transfers the sample tube/rack from the automation track to the analyzer's sample load and identification area. This method is used to establish connection with Beckman Coulter's Synchron LX® and UniCel® DxC Systems and some third party analyzers.

With outboard sampling, the connection unit performs the sample bar code read function, presents the sample ID to the connected analyzer, and then signals for direct sampling of the open tube by the connected instrument at an aspiration point on the automation track. The Power Processor Generic Connection Module is specifically designed to support the outboard sampling method based on point-in-space pipetting technology aligned with the CSLI guidelines.

6.0 Intended Use:

The basic Power Processor is an automated sample handling system which processes sample tubes from the pre-centrifugation, pre-sorting step to presentation of centrifuged and decapped samples into Generic or Personality Racks for specific instruments. The Power Processor can be configured with optional software and hardware to allow processing of sample tubes on Generic Connection Instruments. The Power Processor performs all pre-analytical sample tube preparation, and then sorts the sample tubes directly to Generic Connection Modules where the samples are pipetted by the Generic Connection instrument for testing. After the samples are pipetted, the tubes can route to other instruments for additional testing or to Outlet Racks

The UniCel Dxl 600/800 Access Immunoassay System(s) with laboratory automation connection is a microcomputer-controlled, random and continuous access analyzer that includes an external computer stores the system user interface (UI) software and allows the operator to interface with and direct the instrument software. The UniCel Dxl 600/800 System(s) uses enzyme immunoassays (utilizing paramagnetic particle solid phase and chemiluminescent defection) for determination of various analytes, such as Vitamin B12, Ferritin, Folate and hTSH along with other various enzyme immunoassays assays that may be adaptable to the analyzer depending on the reagent used to induce the enzyme immunoassay reaction. The UniCel Dx1 600/800 System(s) is an in vitro diagnostic device for use in the clinical laboratory.

The Access Ferritin assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of ferritin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism.

The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum, plasma (heparin) and red blood cells using the Access Immunoassay Systems. Folic acid measurements are used in the diagnosis and treatment of megaloblastic anemia.

The Access HYPERsensitive hTSH assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of human thyroid-stimulating hormone (thyrotropin, hTSH) levels in human serum using the Access Immunoassay Systems. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.

The Access Vitamin B12 assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of vitamin B12 in human serum and plasma (heparin) using Access Immunoassay Systems. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.

{2}------------------------------------------------

7.0 Comparison to the Predicate:

(Description of the Modification to the Legally Marketed Device)

The modification to the Power Processor Sample Processing System with Generic Connection Module involves upgrading the PrepLink Software from Version 3.5 to Version 5.0.

PrepLink Version 5.0 provides the following key software features:

  • . Allow the system to connect up to 12 generic connections (increase from current 4), and up to 4 Beckman Coulter (SYNCHRON LX/UniCel DxC) analyzer connections (remains unchanged).
  • . Provide an option to increase the primary sample ID to 13 characters, plus 2 characters for the secondary sample ID.
  • . Provide an option to connect up to three 3K or three 5K tube stockyards.
  • Provide an option for load balancing number of tubes per rack instead of round robin . of one tube for each analyzer. Round robin (RR) is one of the simplest scheduling algorithms for processes in an operating system, which assigns time slices to each process in equal portions and in circular order, handling all processes without priority.
  • . Provide Data Migration Utility (DMU) for V3.5/V3.6 to V5.0.
  • Provide an option to route a primary tube with no assigned tests in its test order to the . stockvard.
  • Provide an option to backup/restore the data files on USB drives. .

8.0 Summary of Performance Data:

Performance data from validation testing shows that all software design, development and verification activities have been completed and passed, and supports equivalency of Power Processor V5.0 to Power Processor V3.5.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized human profiles facing to the right.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

MAY 11 2011

Beckman Coulter, Inc. c/o Marine Boyajian Senior Regulatory Affairs Specialist 250 South Kraemer Blvd, Brea, CA 92822

Re: K110413

Trade/Device Name: Power Processor Sample Processing System with Generic Connections Regulation Number: 21CFR Sec .- 862.1810 Regulation Name: Vitamin B12 test system. Regulatory Class: II Product Code: CDD, DBF, CGN, JLW, JJE Dated: April 13, 2011 Received: April, 14, 2011

Dear: Ms. Boyajian:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

・・

{4}------------------------------------------------

Page 2 -

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (30) 796 5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolli-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

C.C.

Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

:

Enclosure

{5}------------------------------------------------

Indication for Use

510(k) Number (if known): K110413

Device Name:

Power Processor Sample Processing System with Generic Connection Module and Access Immunoassay System Reagents

Indication For Use:

The basic Power Processor is an automated sample handling system which processes sample tubes from the pre-centrifugation, pre-sorting step to presentation of centrifuged and decapped samples into Generic or Personality Racks for specific instruments. The Power Processor can be configured with optional software and hardware to allow processing of sample tubes on Generic Connection Instruments. The Power Processor performs all pre-analytical sample tube preparation, and then sorts the sample tubes directly to Generic Connection Modules where the samples are pipetted by the Generic Connection instrument for testing. After the samples are pipetted, the tubes can route to other instruments for additional testing or to Outlet Racks.

The UniCel DxI 600/800 Access Immunoassay System(s) with laboratory automation connection is a microcomputer-controlled, random and continuous access analyzer that includes an external computer. This computer stores the system user interface (U) software and allows the operator to interface with and direct the instrument software. The UniCel DxI 600/800 System(s) uses enzyme immunoassays (utilizing paramagnetic particle solid phase and chemiluminescent detection) for determination of various analytes, such as Vitamin B12, Ferritin, Folate and hTSH along with other various enzyme immunoassays assays that may be adaptable to the analyzer depending on the reagent used to induce the enzyme immunoassay reaction. The UniCel DxI 600/800 System(s) is an in vitro diagnostic device for use in the clinical laboratory.

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE, CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Signature

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) == K110413

{6}------------------------------------------------

The Access Ferritin assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of ferritin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism.

The Access Folate assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of folic acid levels in human serum, plasma (heparin) and red blood cells using the Access Immunoassay Systems. Folic acid measurements are used in the diagnosis and treatment of megaloblastic anemia.

The Access HYPERsensitive hTSH assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of human thyroid-stimulating hormone (thyrotropin, hTSH) levels in human serum using the Access Immunoassay Systems. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.

The Access Vitamin B12 assay is a paramagnetic particle, chemiluminescent assay for the quantitative determination of vitamin B12 in human serum and plasma (heparin) using Access Immunoassay Systems. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.

Prescription Use X 1 (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

CARS

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K110413

§ 862.1810 Vitamin B

12 test system.(a)
Identification. A vitamin B12 test system is a device intended to measure vitamin B12 in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.(b)
Classification. Class II.