K Number
K110332

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-03-24

(49 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ClearCanvas RIS/PACS is an image management system whose intended use is to provide scaleable DICOM compatible PACS solutions for hospitals and related institutions and sites, which will archive, distribute, retrieve and display images and data from all hospital modalities (such as CR, CT, DR, MR, and other devices) and information systems. This also includes the display of structured reports and mammography images that have been created according to DICOM."For Presentation", and will include standard features and other tools for analyzing mammography images.

Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at least 5 mega-pixel resolution and meets other technical specifications approved by the FDA.

Device Description

ClearCanvas RIS/PACS is a Picture Archiving and Communication (PACS) software system for the management and review of medical image data, and other digital images. Such data can be received from all DICOMcompliant imaging modalities or imported directly into the system. This data can then be stored, archived, distributed, processed, enhanced, and displayed for review. Enhancement, processing and analysis of the image data includes but is not limited to compression, magnification, value-of-interest manipulation, orientation changes, image fusion and multiplanar reformat. Additional health data, such as patient demographics, from other information systems can be received from HL7-compliant sources.

The ClearCanvas RIS/PACS is a system of interrelated software components:

  • . ClearCanvas Workstation – a desktop client that is primarily used to retrieve, display, enhance and analyze medical images. Access to patient demographic data is also possible through the Workstation when working with the RIS server.
  • . ClearCanvas Webstation – a browser-based web application that allows the user to retrieve images from the ImageServer for review. Primarily intended for referring physicians and as an image viewer for EMR systems, the images are JPEG-compressed.
  • . ClearCanvas ImageServer – a server application that receives images from imaging modalities, stores and archives them, and distributes them to client applications.
  • . ClearCanvas RIS – a server application that acts as an online transaction processor for the management of patient and workflow information.
  • . ClearCanvas IntegrationServer – a server application composed of optional modules that implement integration and interoperability, such as an inbound and outbound HL7 processor.
  • . ClearCanvas EnterpriseServer -- a server application that provides enterprise-wide facilities such as user authentication, authorization and auditing.

Working in concert or stand-alone, each component plays a valuable role in the management information in the radiology reading room, the clinic, the referring physician's office, or any other patient care setting where access to medical imaging information is important.

ClearCanvas RIS/PACS software is primarily written in C#, employing the most current best practices in objectoriented and component-oriented software architecture resulting in a highly scalable and extensible design. The system is also designed to work on generic PC hardware that meets the minimum system requirements.

AI/ML Overview

This is a 510(k) premarket notification for the ClearCanvas RIS/PACS, which is a Picture Archiving and Communication System (PACS). The submission focuses on demonstrating substantial equivalence to predicate devices rather than providing a study with specific acceptance criteria and performance metrics for a novel AI algorithm. Therefore, much of the requested information regarding AI algorithm performance, sample sizes for training/test sets, expert adjudication, MRMC studies, and ground truth establishment is not present in this document.

The document primarily lists technological characteristics and intended uses to show equivalence to existing PACS systems.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't provide a table of quantitative acceptance criteria for device performance in the typical sense of AI/CADe devices (e.g., sensitivity, specificity, AUC). Instead, it relies on demonstrating that its technological characteristics and intended use are equivalent to legally marketed predicate devices. The "performance" is implicitly accepted if it matches or exceeds the predicate devices' capabilities in areas like image handling, display, and workflow.

Acceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (as stated by ClearCanvas)
Ability to be used as a PACS (Archive & Communication)Y (ClearCanvas RIS/PACS)
Ability to store and distribute imagesY (ClearCanvas RIS/PACS)
Ability to display, analyze, manipulate, and enhance imagesY (ClearCanvas RIS/PACS)
Conformance to DICOM standard for data exchangeY (ClearCanvas RIS/PACS)
Conformance to JPEG standard for image compressionY (ClearCanvas RIS/PACS)
Software device operating on off-the-shelf hardwareY (ClearCanvas RIS/PACS)
System composed of multiple components (client-server)Y (ClearCanvas RIS/PACS)
Key components: Medical image viewer, Image storage/archive serverY (ClearCanvas RIS/PACS)
Functions & Capabilities (e.g., Query/import/send DICOM, W/L, zoom, pan, stack, rotation, flip, measurement, annotation, multi-monitor, image layout, thumbnails, reference lines, user annotations, synchronized stacking, probe tool, shutters, key image marking, MPR, lossy/lossless compression, automatic routing, archiving, patient info/workflow, image streaming, DICOM printing, export to optical media, PET/CT fusion)Y for most (ClearCanvas RIS/PACS), with a few exceptions where ClearCanvas offers a feature not present in one predicate, or vice-versa (e.g., ClearCanvas has Probe tool and Shutters, BRIT PACS does not; ClearCanvas has PET/CT fusion, IntelePACS does not).
Real-time performance requirements not applicableReal-time performance requirements not applicable
Interoperation with other devices based on DICOM standardsInteroperation with other devices based on consensus standards on data exchange (DICOM)
Instructions for search & retrieval, display, manipulation, enhancement, annotation of imagesY (ClearCanvas RIS/PACS)

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not applicable. This submission does not describe a clinical study with a "test set" in the context of evaluating an AI algorithm's performance on a specific disease detection task. The performance testing conducted was to verify design outputs met input requirements and validate conforming to user needs under simulated use conditions. No specific patient data sample size is mentioned for this testing.
  • Data Provenance: Not applicable. No specific country of origin or retrospective/prospective nature of data is mentioned as the testing was likely functional and system-level, not clinical efficacy evaluation on patient datasets.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: Not applicable. There is no mention of experts or ground truth establishment in the context of evaluating diagnostic accuracy or performance on patient images. The device itself is a PACS, not a diagnostic AI tool.
  • Qualifications of Experts: Not applicable.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Adjudication Method: Not applicable. No adjudication process is described as there's no clinical performance study on patient data.

5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • MRMC Study: No, an MRMC comparative effectiveness study was not done. The device is a PACS, not an AI-assisted diagnostic tool in the sense of a CADe/CADx system. The submission focuses on the PACS system's functionality and equivalence to other PACS systems.
  • Effect Size: Not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Standalone Performance: No. The ClearCanvas RIS/PACS is a "Picture Archiving and Communication (PACS) software system for the management and review of medical image data" and is explicitly described as being "used by trained and qualified professionals affording ample opportunity for competent human intervention in interpreting the images and information presented." It is not an algorithm designed for standalone diagnostic performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Type of Ground Truth: Not applicable. No ground truth is mentioned in the context of clinical diagnostic accuracy. The "performance testing" referenced was likely functional and integration testing against defined system requirements, not against clinical outcomes or expert labels.

8. The sample size for the training set:

  • Training Set Sample Size: Not applicable. This is not an AI/ML algorithm that requires a training set for model development.

9. How the ground truth for the training set was established:

  • Training Set Ground Truth: Not applicable.

{0}------------------------------------------------

K110332

MAR 2 4 2011

.

510(k) Summary of Safety and Effectiveness

Date Prepared

November 22, 2010

Submitter

.

:

ClearCanvas Inc. 439 University Ave., Suite 1920 Toronto, ON M5G 1Y8 " | ways . : が、「私 .. . " -、" Canada

・・

Contact Person '

Clinton Chau Email: clinton@clearcanvas.ca Phone: (647) 547-2987 Fax: (866) 584-5688

Device Name and Classification

Trade NameClearCanvas RIS/PACS
Common NamePicture Archiving and Communications System (PACS)
Device Classification NameSystem, Image Processing, Radiological
Device ClassClass II
Regulation Number892.2050
Product CodeLLZ
Classification PanelRadiology

Predicate Devices

Trade NameBRIT PACS SYSTEMSIntelePACS(tm)
Model Number----
Common NamePicture Archiving andCommunication System (PACS)Picture Archiving andCommunication System (PACS)
510(k) Submitter/HolderBRIT Systems Inc.1909 Hi-Line Dr.Dallas, TX 75207Intelerad Medical Systems Inc.Anibal Jodorcovsky460 Ste-Catherine West, Suite 210Montreal, QC Canada H3B 1A7
510(k) NumberK081168K070080
Regulation Number892.2050892.2050
PanelRadiologyRadiology
Product CodeLLZLLZ

{1}------------------------------------------------

Device Description and Intended Use

ClearCanvas RIS/PACS is a Picture Archiving and Communication (PACS) software system for the management and review of medical image data, and other digital images. Such data can be received from all DICOMcompliant imaging modalities or imported directly into the system. This data can then be stored, archived, distributed, processed, enhanced, and displayed for review. Enhancement, processing and analysis of the image data includes but is not limited to compression, magnification, value-of-interest manipulation, orientation changes, image fusion and multiplanar reformat. Additional health data, such as patient demographics, from other information systems can be received from HL7-compliant sources.

The ClearCanvas RIS/PACS is a system of interrelated software components:

  • . ClearCanvas Workstation – a desktop client that is primarily used to retrieve, display, enhance and analyze medical images. Access to patient demographic data is also possible through the Workstation when working with the RIS server.
  • . ClearCanvas Webstation – a browser-based web application that allows the user to retrieve images from the ImageServer for review. Primarily intended for referring physicians and as an image viewer for EMR systems, the images are JPEG-compressed.
  • . ClearCanvas ImageServer – a server application that receives images from imaging modalities, stores and archives them, and distributes them to client applications.
  • . ClearCanvas RIS – a server application that acts as an online transaction processor for the management of patient and workflow information.
  • . ClearCanvas IntegrationServer – a server application composed of optional modules that implement integration and interoperability, such as an inbound and outbound HL7 processor.
  • . ClearCanvas EnterpriseServer -- a server application that provides enterprise-wide facilities such as user authentication, authorization and auditing.

Working in concert or stand-alone, each component plays a valuable role in the management information in the radiology reading room, the clinic, the referring physician's office, or any other patient care setting where access to medical imaging information is important.

ClearCanvas RIS/PACS software is primarily written in C#, employing the most current best practices in objectoriented and component-oriented software architecture resulting in a highly scalable and extensible design. The system is also designed to work on generic PC hardware that meets the minimum system requirements.

Indications for Use

The ClearCanvas RIS/PACS is an image management system whose intended use is to provide scaleable DICOM compatible PACS solutions for hospitals and related institutions and sites, which will archive, retrieve and display images and data from all hospital modalities (such as CR, CT, DR, MR, and other devices) and information systems. This also includes the display of structured reports and mammography images that have been created according to DICOM "For Presentation", and will include standard features and other tools for analyzing mammography images.

Page 2 of 8

{2}------------------------------------------------

Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at ieast 5 mega-pixel resolution and meets other technical specifications approved by the FDA.

Technological Characteristics

The ClearCanvas RIS/PACS has no patient contact and does not control any life sustaining devices. It is a software PACS device used in the management of medical imaging information. This device is used by trained and qualified professionals affording ample opportunity for competent human intervention in interpreting the images and information presented. These technological characteristics are identical to those in the predicate devices. Furthermore, the following comparison table summarizes the specific characteristics of the ClearCanvas RIS/PACS and its predicate devices that demonstrate substantial equivalence:

Trade NameSubject DeviceClearCanvas RIS/PACSPredicate DeviceBRIT PACS SYSTEMPredicate DeviceIntelePACS
Intended Use
1To be used as a Picture Archive andCommunication System (PACS)YYY
2To be used for storing anddistributing imagesYYY
3To be used for displaying,analyzing, manipulating andenhancing of imagesYYY
Indications For Use Statement
4Statement uses the term, “picturearchive and communicationsystem” or “PACS”YYY
5Statement references the displayof medical imagesYYY
6Statement references receivingimages from image sourcesYYY
Subject DevicePredicate DevicePredicate Device
7Statement references multi-planar reformatNNY
8Statement references mammographic imagesYYY
Standards Met
9Conforms to the essential requirements of the DICOM standard for data exchange (PS 3.3-PS 3.12)YYY
10Conforms to the JPEG standard for image compression of digital medical imagesYYY
Design
11Software device that operates on off-the-shelf hardwareYYY
12Device is a system composed of multiple components that work stand-alone or in concertYYY
13Device components are organized in a client-server architectureYYY
14Medical image viewer, the client application, is a key component of the device systemYYY
15Image storage and archive server, the server application, is a key component of the device systemYYY
Subject DevicePredicate DevicePredicate Device
Functions and Capabilities
16Query, import, send, receiveDICOM imagesYYY
17Image analysis tools W/L, zoom,pan, stack, rotation, flipYYY
18Measurement and annotationYYY
19Multi-monitor awarenessYYY
20Image layoutYYY
21Image ThumbnailsYYY
22Reference lines and spatial locatorfor tomographic imagesYYY
23User annotations of imagesYYY
24Synchronized stackingYYY
25Probe toolYNY
26ShuttersYNY
27Key Image markingYYY
28Multiplanar Reformat (MPR)YYY
29Lossy and lossless imagecompressionYYY
30Automatic routing of imagesYYY
31Image archivingYYY
Subject DevicePredicate DevicePredicate Device
32Patient information and workflow management componentsYPartialY
33Image streaming from serverYYN
34Operating System/PlatformWindowsLinux and WindowsLinux and Windows
35DICOM image printingYYY
36Export DICOM images to optical mediaYYY
37PET/CT fusionYYN
Other Areas of Comparison
38Target populationNo restrictionsNo restrictionsNo restrictions
39Anatomical sitesNo restrictionsNo restrictionsNo restrictions
40Environment for useNo restrictions; to be used by trained professionals onlyNo restrictions; to be used by trained professionals onlyNo restrictions; to be used by trained professionals only
41Energy used and/or deliveredElectric power supply to computer hardware onlyElectric power supply to computer hardware onlyElectric power supply to computer hardware only
42Human factorsStandard windows-based or web-based graphical user interfaces, with standard input devicesStandard windows-based or web-based graphical user interfaces, with standard input devicesStandard windows-based or web-based graphical user interfaces, with standard input devices
Subject DevicePredicate DevicePredicate Device
43PerformanceReal-time performance requirements not applicableReal-time performance requirements not applicableReal-time performance requirements not applicable
44MaterialsNot applicableNot applicableNot applicable
45BiocompatibilityNot applicableNot applicableNot applicable
46Compatibility with environment and other devicesInteroperation with other devices based on consensus standards on data exchange (DICOM)Interoperation with other devices based on consensus standards on data exchange (DICOM)Interoperation with other devices based on consensus standards on data exchange (DICOM)
47SterilityNot applicableNot applicableNot applicable
48Electrical safetyNot applicableNot applicableNot applicable
49Mechanical safetyNot applicableNot applicableNot applicable
50Chemical safetyNot applicableNot applicableNot applicable
51Thermal safetyNot applicableNot applicableNot applicable
52Radiation safetyNot applicableNot applicableNot applicable
Product Labelling
53Instructions for searching and retrieval of imagesYYY
54Instructions for displaying of imagesYYY
Subject DevicePredicate DevicePredicate Device
55Instructions for manipulatingorientation of imagesYYY
56Instructions for enhancing imagesYYY
57Instructions for annotating imagesYYY

Page 3 of 8

.

{3}------------------------------------------------

. ·

Page 4 of 8

{4}------------------------------------------------

.

.

.

.

Page 5 of 8

.

.

{5}------------------------------------------------

·

: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :

.

. .

{6}------------------------------------------------

.

.

.

: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :

.

Page 7 of 8

{7}------------------------------------------------

Testing

The ClearCanvas RIS/PACS complies with the voluntary DICOM and JPEG standards for device performance and is designed and manufactured according to the process and management system standards set in ISO 13485:2003 and ISO 62304, as discussed in the 510(k) submission.

Performance testing was conducted to verify the design output met the design input requirements and to validate the device conforms to the defined user needs and intended uses. Nonclinical software testing, as well as human factors testing, was conducted under simulated use conditions. Predefined acceptance criteria were met and demonstrated that the device is as safe and as effective as the predicate devices.

Substantial Equivalence Conclusions

ClearCanvas concludes that the intended use for the ClearCanvas RIS/PACS is the same as that of the predicate devices, and that the technological characteristics demonstrate that they are equivalent to the predicate devices. A comparison of the technological characteristics of the predicate and legally marketed devices available has been performed.

Conclusion

The 510(k) Pre-Market Notification for ClearCanvas RIS/PACS contains adequate information and data to determine that ClearCanvas RIS/PACS is as safe and effective as the legally marketed predicate devices

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

ClearCanvas. Inc. % Ms. Paula Wilkerson Sr. Staff Engineer-Medical Devices Intertek Testing Services 2307 East Aurora Rd., Unit B7 TWINSBURG OH 44087

MAR 2 4 2011

Re: K110332

Trade/Device Name: ClearCanyas RIS/PACS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 15, 2011 Received: March 16, 2011

Dear Ms. Wilkerson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{9}------------------------------------------------

Page 2

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k), remarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, perceits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (01) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of advertal Reents under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolli the (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours.

Mary Pastel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): _K110332

Device Name: ClearCanvas RIS/PACS

Indications for Use:

The ClearCanvas RIS/PACS is an image management system whose intended use is to provide scaleable DICOM compatible PACS solutions for hospitals and related institutions and sites, which will archive, distribute, retrieve and display images and data from all hospital modalities (such as CR, CT, DR, MR, and other devices) and information systems. This also includes the display of structured reports and mammography images that have been created according to DICOM."For Presentation", and will include standard features and other tools for analyzing mammography images.

Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammography images may only be interpreted using an FDA approved monitor that offers at least 5 mega-pixel resolution and meets other technical specifications approved by the FDA.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Mary S. Patel

Division Sign-Öff Office of In Vitro Diagnostic Device Evaluation and Safety 10332

. ...
... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... .

Page 1 of 1

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).