K Number
K070080

Validate with FDA (Live)

Device Name
INTELEPACS
Date Cleared
2007-01-24

(15 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

IntelePACS™ is a device that receives digital images and data from various sources (such as, CT scanners, MR scanners, ultrasound systems, R/F units, computer and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or other imaging sources). lmages and data can be communicated, processed, manipulated, enhanced, stored, and displayed within the system and/or across computer networks at distributed locations. Postprocessing of the images can be performed using Multi Planar Reconstruction (MPP),

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mannographic images may only be interpreted using an FDA approved monitor that protateliar manimographio may only be medification specifications reviewed and accepted by FDA.

Typical users of this system are trained professionals, physicians, nurses, and technicians.

Device Description

IntelePACS™ is comprised of software modules that provide image capture, storage, distribution, enhancement, manipulation, and networking of medical images at distributed locations. In cases where DICOM images are not directly available to IntelePACS™, the system can acquire medical imem mugos using a DICOM image gateway, which generates DICOM-type files. For example, and a Broom images from original film and convert them to meet DICOM standards and stored. Stored files are transmitted using a network and can be viewed or manipulated from imaging workstation.

AI/ML Overview

The provided text is a 510(k) summary for IntelePACS™. After a thorough review, it is clear that this document does not contain the acceptance criteria or a study proving that the device meets such criteria.

The 510(k) summary focuses on establishing substantial equivalence to predicate devices, providing a device description, and outlining indications for use. It mentions that the system will be manufactured in accordance with voluntary standards, and a hazard analysis classified potential hazards as "minor," but these are not the detailed performance acceptance criteria and study results typically expected for a medical device.

Therefore, I cannot provide the requested table or answer the specific questions about the study, sample sizes, ground truth, or expert involvement based on the provided text. The document is primarily a regulatory submission demonstrating equivalence, not a scientific study report.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a stylized letter "i" with a unique, handwritten-like appearance. The letter is slightly tilted to the left, and it has a distinct serif at the top. The letter is black and white, and the image is cropped closely around the letter.

510(k) Summary of Safety and Effectiveness K070080

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

Date Prepared: August 14, 2006

Submitter's Information: 21 CFR 807.92(a)(1)

Intelerad Medical Systems Inc. Anibal Jodorcovsky Executive Vice President, Quality Systems 460 Ste-Catherine West, Suite 210 Montreal, QC Canada H3B 1A7 Tel: 514.931.6222 Fax: 514.931.4653

Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)

Trade Name:IntelePACS™
Common Name:Picture Archiving Communications System
Device Classification:892.2050
Name:System, Image Processing, Radiological

Predicate Devices: 21 CFR 807. 92(a)(3)

Device Classification NameSYSTEM, IMAGE PROCESSING, RADIOLOGICAL
Regulation Number892.2050
510(k) NumberK032533
Device NameIntelePACS
ApplicantIntelerad Medical Systems Incorporated
Product CodeLLZ
Date Received08/15/2003
Decision Date10/16/2003
DecisionSUBSTANTIALLY EQUIVALENT (SE)
Classification Advisory CommitteeRadiology
Review Advisory CommitteeRadiology

{1}------------------------------------------------

Device Classification NameSYSTEM, IMAGE PROCESSING, RADIOLOGICAL
Regulation Number892.2050
510(k) NumberK023460
Device NamePACSPLUS
ApplicantMEDICAL STANDARD CO., LTD.
Product CodeLLZ
Date Received10/15/2002
Decision Date01/09/2003
DecisionSUBSTANTIALLY EQUIVALENT (SE)
Classification AdvisoryCommitteeRadiology
Review Advisory CommitteeRadiology
Device Classification NameSystem, Image, Processing, Radiological
Regulation Number892.2050
510(k) NumberK061672
Device NameSynapse 3D MIP/MPR Image Visualization SoftwareOBLIQUUS
ApplicantFUJIFILM Medical Systems USA, Inc.
Product CodeLLZ
Date Received06/14/2006
Decision Date06/29/2006
DecisionSUBSTANTIALLY EQUIVALENT (SE)
Classification AdvisoryCommitteeRadiology
Review Advisory CommitteeRadiology

Device Description: 21 CFR 807 92(a)(4)

IntelePACS™ is comprised of software modules that provide image capture, storage, distribution, enhancement, manipulation, and networking of medical images at distributed locations. In cases where DICOM images are not directly available to IntelePACS™, the system can acquire medical imem mugos using a DICOM image gateway, which generates DICOM-type files. For example, and a Broom images from original film and convert

i

{2}------------------------------------------------

i
z

them to meet DICOM standards and stored. Stored files are transmitted using a network and can be viewed or manipulated from imaging workstation.

Indications for Use: 21 CFR 807 92(a)(5) Technological Characteristics: 21 CFR 807 92(a)(6)

The device is medical device image software that is used with computer hardware in a picture archiving and communications system user environment. The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human interprets images and information being displayed and printed.

Conclusion: 21 CFR 807 92(b)(1)

The 510(k) Pre-Market Notification for IntelePACS™ contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate devices.

IntelePACS™ system will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The submission contains the results of a hazard analysis and the "Level of Concern" for potential hazards has been classified as "minor".

{3}------------------------------------------------

Third Party Review Quality Assessment

Section 1 – Submission Information

510(k) No.:K070080
Third Party Organization:Intertek Testing Services
Third Party's Primary Reviewer(s):Jay Y. Kogoma
ODE/OIVD Division:DRARD
Branch/Team:RDB

Section 2 - 510(k) Decision

Third party recommendation:SE x ___ NSE ___ Other (specify): ___
ODE/OIVD final decision:SE x ___ NSE ___ Other (specify): ___

Section 3 - Assessment of Third Party Review

Review ElementRating (check one)
AdequateMinor Issue(s)Major Issue(s)
a. Determination of device eligibility for third party reviewX
b. Extent of pre-submission consultation with ODE/OIVD division
c. Organization and format of review documentationX
d. Determination of 510(k) administrative completeness (screening review)X
e. Summary of device characteristics, intended use, and performance (including accessories, if applicable) and reason for 510(k) submissionX
f. Comparison to legally marketed devices—identification and analysis of key similarities and differencesX
g. Rationale for conclusions and recommendationX
h. Use of guidance documents and standardsX
i. Resolution of 510(k) deficiencies and FDA requests for additional informationX
j. Scope of reviewer expertise and use of consulting reviewers
k. Other (specify):

Comments (explanation of ratings/issues): ___________________________________________________________________________________________________________________________________

Section 4 – ODE/OIVD Assessor Information

Sunder Rajan Date: 1/22/07 Tel. No.: (240) 276 3968 Assessed by:

Routing: Division--Clip completed assessment (this page only) to inside front cover of 510(k).

DMC-Forward this page only to Eric Rechen, POS/ODE, Rm. 120J, Corp. Blvd. (HFZ-402).

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Intelerad Medical Systems, Inc. % Mr. Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Services NA, Inc. 2307 East Aurora Rd., Unit B7 TWINSBURG OH 44067

JAN 2 4 2007

Re: K070080

K070080
Trade/Device Name: IntelePACS™ from Intelerad Medical Systems, Inc. Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 8, 2007 Received: January 9, 2007

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device wo nate reviewed you beceremined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that provision in the provision accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general androis, mainton are as no requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/4/Picture/10 description: The image is a circular logo or emblem. The text "1906-2006" is at the top of the logo. The letters "FDA" are in the center of the logo in a bold, stylized font. The word "Centennial" is below the letters "FDA". Three stars are below the word "Centennial".

Protecting and Promoting Public Health

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)240-276-0115
21 CFR 894.xxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): _ KO 70080

Device Name:

IntelePACS™ from Intelerad Medical Systems Inc.

Indications for Use:

IntelePACS™ is a device that receives digital images and data from various sources (such as, CT scanners, MR scanners, ultrasound systems, R/F units, computer and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or other imaging sources). lmages and data can be communicated, processed, manipulated, enhanced, stored, and displayed within the system and/or across computer networks at distributed locations. Postprocessing of the images can be performed using Multi Planar Reconstruction (MPP),

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mannographic images may only be interpreted using an FDA approved monitor that protateliar manimographio may only be medification specifications reviewed and accepted by FDA.

Typical users of this system are trained professionals, physicians, nurses, and technicians.

Prescription Use _ V __ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David R. Segner

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _

Page 1 of 1

N/A