K Number
K110146

Validate with FDA (Live)

Date Cleared
2011-05-13

(115 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nitrile Powder Free Patient Examination Gloves, Coloured (Blue) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Nitrile Powder Free Patient Examination Gloves that meets all of the requirements of ASTM standard D 6139-05

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance Criteria (Standard)Device Performance
DimensionASTM standard D 6139-05Meets
Physical PropertiesASTM standard D 6139-05Meets
Freedom from pinholes21 CFR 800.20Meets (specifically, waterleak test on pinhole AQL)
Powder ResidualASTM standard D 5250-06 and D6124-06<2mg/glove
Biocompatibility: Primary Skin IrritationPrimary Skin Irritation in rabbitsPasses
Biocompatibility: Dermal SensitizationDermal sensitization in the guinea pigPasses (Not a Dermal sensitization)

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes used for each test (e.g., how many gloves were tested for pinholes, or how many animals were used for biocompatibility tests). It only refers to compliance with the standards, which would implicitly define sample sizes within those standards.

The provenance of the data is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. It is implied to be data collected to demonstrate compliance with the referenced ASTM standards and CFR regulations.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

This information is not applicable and not provided in the document. The "ground truth" for glove testing is established by objective measurements against predefined criteria in the referenced standards (e.g., tensile strength, elongation, water leak test for pinholes, chemical analysis for powder residue, animal studies for biocompatibility). These are not expert interpretations requiring adjudication in the way medical imaging diagnosis would.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in human interpretation studies (e.g., radiology reads). For physical or chemical testing of medical devices like gloves, the assessment is based on direct measurement and meeting specified thresholds in standards, not subjective expert opinion.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Readers Improvement with AI vs. without AI Assistance

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-assisted diagnostic devices where human reader performance is being evaluated. This document pertains to a physical medical device (examination gloves), not a diagnostic AI system.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

This is not applicable. This document describes the performance of a physical device (examination gloves), not an algorithm or AI system.

7. The Type of Ground Truth Used

The "ground truth" for the device's performance is established by objective measurements and chemical/biological tests against the specific criteria and methodologies outlined in the referenced ASTM standards (D 6139-05, D 5250-06, D6124-06) and FDA regulations (21 CFR 800.20, ISO10993-10). Examples include:

  • Dimensional measurements (e.g., length, width, thickness).
  • Physical property tests (e.g., tensile strength, elongation at break).
  • Water leak tests for freedom from pinholes.
  • Chemical analysis to quantify powder residue.
  • Animal studies (rabbits for primary skin irritation, guinea pigs for dermal sensitization) for biocompatibility.

8. The Sample Size for the Training Set

This is not applicable. This device is a physical product, not a software or AI model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This is not applicable. As stated above, this device does not involve a training set.

{0}------------------------------------------------

Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

MAY 1 3 2011

"The assigned 510(k) number is: K110146 _ ." (applicant leave blank)

Premarket Notification [510(k)] Summary

[(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

Submitter's name :TANGSHAN KAIYUE GLOVE CO., LTD.
Submitter's address :TANGLE ROAD, LUANNAN COUNTY,TANGSHAN CITY,HEBEI PROVINCE,063500,CHINA
Phone number :(86)315-4169201
Fax number :(86)315-4169311
Name of contact person:Mr. ZHANG Liang
Date the summary was prepared:Jan.5th. 2010

[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Nitrile Powder Free Patient Examination Gloves, Coloured (Blue)
Proprietary/Trade name:Nitrile Powder Free Patient Examination Gloves, Coloured (Blue)
Common Name:Patient examination glove
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LZA

[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* Nitrile Powder Free Patient Examination Gloves that meets all of the requirements of ASTM standard D 6139-05

Predicate device : Nitrile powder-free patient examination glove, JDA International Inc. K993247 .

[(a)(4)] A description of the device

Device Description : Nitrile Powder Free Patient Examination Gloves that meets all of the requirements of ASTM standard D 6139-05

{1}------------------------------------------------

[(a)(5)] The summary describes the intended use of the device

Device Intended Use: Nitrile Powder Free Patient Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 6139-05Meets
Physical PropertiesASTM standard D 6139-05Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06Meets
Biocompatabilityand D6124-06<2mg/glove
Primary Skin Irritation in rabbitsPasses
Dermal sensitization in the guinea pigPassesNot a Dermal sensitization

[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Nitrile powder-free patient examination glove meet requirements per ASTM D6319-05, per ASTM D6124-06, per 21 CFR 800.20 and ISO10993-10.

[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

[(b)(3)| The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Nitrile powder-free patient examination glove meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle in flight, composed of three curved lines that suggest the bird's wings and body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, with the text oriented to follow the curve of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAY 1 3 2011

Tangshan Kaiyue Glove Company, Limited C/O Mr. Chu Xiaoan Room 1606 Building 1, Jianxiang Yuan No. 209, Bei Si Huan Zhong Road Haidian District, Beijing CHINA 100083

Re: K110146

Trade/Device Name: Nitrile Powder Free Patient Examination Gloves, Coloured (Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: April 25, 2011 Received: April 25, 2011

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

hr for

Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

Applicant: TANGSHAN KAIYUE GLOVE CO., LTD.

510(k) Number (if known):_* KIIO I 46

Device Name: Nitrile Powder Free Patient Examination Gloves, Coloured (Blue)

Indications For Use:

Nitrile Powder Free Patient Examination Gloves, Coloured (Blue) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Edith B. Clavie- Will

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

11 014 510(k) Number:

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.