K Number
K110031

Validate with FDA (Live)

Date Cleared
2011-08-24

(231 days)

Product Code
Regulation Number
864.8950
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HemosIL dRVVT Screen and HemosIL dRVVT Confirm assays are qualitative in-vitro diagnostic products to aid in the detection of lupus anticoagulants in human citrated plasma by the diluted Russell's Viper Venom method, on the ACL TOP® Family. The HemosIL dRVVT Screen and HemosIL dRVVT Confirm assays are intended to evaluate patients who have unexplained prolonged APTT test results The HemosIL dRVVT Screen and HemosIL dRVVT Confirm assays should be used in parallel as an integrated test for Lupus Anticoagulant detection.

Device Description

DRVVT Screen and dRVVT Confirm are improved dRVVT reagents, intended to simplify and standardize the detection of Lupus Anticoagulant (LA) disorder in clinical chemistry evaluations. DRVVT Screen is poor in phospholipid, making it sensitive to LA. The additional amount of phospholipid in dRVVT Confirm neutralizes LA to give shorter clotting times. Russell's viper venom, in the presence of calcium, directly activates factor X (in a test sample). DRVVT Screen and dRVVT Confirm are therefore unaffected by contact factor abnormalities, factor VII, VIII and IX deficiencies, or inhibitors. As a result, dRVVT Screen and dRVVT Confirm are more specific tests for the evaluation of LA than APTT.

AI/ML Overview

Acceptance Criteria and Device Performance Study for HemosIL® dRVVT Screen and Confirm Assays

1. Table of Acceptance Criteria and Reported Device Performance

The device (HemosIL® dRVVT Screen and HemosIL® dRVVT Confirm assays) is a qualitative in-vitro diagnostic product. Instead of traditional sensitivity/specificity acceptance criteria for quantitative devices, the performance is demonstrated through comparison with a predicate device and via inter-laboratory validation studies using established cut-offs.

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
Analytical Performance
Precision (Total %CV)Within acceptable ranges for coagulation assays (not explicitly stated, but industry standard for diagnostic tests).LA Negative Control: Lot 1: 2.3%, Lot 2: 3.4%, Lot 3: 2.1% Weakly LA Positive Control: Lot 1: 3.0%, Lot 2: 2.6%, Lot 3: 2.2% LA Positive Control: Lot 1: 5.0%, Lot 2: 3.5%, Lot 3: 3.0% (across three lots and three instruments)
Interferent ToleranceMaximum tolerated concentrations of common interferents should not significantly affect results (defined as base clotting time ± 15%).UFH: ≤ 1.0 IU/mL LMWH: ≤ 1.0 IU/m Hemoglobin: ≤ 200 mg/dL Bilirubin: ≤ 10 mg/dL Triglyceride: ≤ 500 mg/dL (at any LA level tested)
Assay Cut-off DeterminationNormal healthy individual samples should be used to establish a cut-off (Mean + 3SD).Normalized Ratio cut-off determined using 40 normal healthy individuals. ACL TOP: >1.2 ACL TOP 500CTS: >1.2 (Note: Each lab should establish its own cut-off).
Comparison to Predicate DeviceHigh Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) with the predicate device (HemosIL LAC Screen & LAC Confirm).In-house (115 samples): - ACL TOP & ACL TOP 500 CTS: PPA 100.0% (35/35), NPA 100.0% (80/80), Overall 100% (115/115) (CI 95% provided) 3 US Field Sites (100+ samples each): - Site 1: PPA 92.7% (38/41), NPA 98.9% (91/92), Overall 97% (129/133) - Site 2: PPA 90.2% (46/51), NPA 98.9% (91/92), Overall 95.8% (137/143) - Site 3: PPA 98.1% (52/53), NPA 100.0% (80/80), Overall 99.2% (132/133) (CI 95% provided for all sites)
Matrix Comparison (Citrate Type)Normalized Ratio should not be significantly affected by 3.8% versus 3.2% sodium citrate sample tubes, demonstrating high PPA and NPA.ACL TOP: PPA 100% (19/19), NPA 92% (24/26) (CI 95% provided). Results showed the dRVVT NR is not affected by citrate tube type.
Matrix Comparison (Fresh vs. Frozen)Normalized Ratio should not be significantly affected by fresh versus frozen and once-thawed samples, demonstrating high PPA and NPA.ACL TOP: PPA 100% (28/28), NPA 100% (26/26) (CI 95% provided). The method comparison demonstrated the dRVVT NR is not affected by fresh or frozen use.
Specificity to LAThe assay should detect LA specifically and not be significantly affected by other conditions like oral anticoagulants, LMWH, UFH, DIC, or Factor Deficiency.Known LA Positive: 100% (35/35) Oral Anticoagulants: 40% (2/5) LMWH: 0% (0/5) UFH: 20% (1/5) DIC: 0% (0/5) Factor Deficiency: 0% (0/6) (Performance on both ACL TOP and ACL TOP 500CTS were similar).
Reference RangeA normal range study should be performed with a sufficient number of normal healthy individuals to establish reference intervals for the Normalized Ratio (NR).A normal range study (n=120) established reference intervals for dRVVT Screen/Confirm Normalized Ratio: ACL TOP: Lower Limit 0.92 (0.91-0.93), Upper Limit 1.11 (1.10-1.15) ACL TOP 500 CTS: Lower Limit 0.91 (0.89-0.92), Upper Limit 1.13 (1.11-1.16)

2. Sample Size Used for the Test Set and Data Provenance

  • Precision/Reproducibility Study Test Set:
    • Sample Size: N=80 per instrument per lot (Total: 3 lots x 2 instruments x 80 = 480 individual measurements for each control level). Specifically, 20 days, 2 runs/day, 2 replicates/run for each sample level (Negative, Weakly Positive, Positive controls).
    • Data Provenance: Not explicitly stated, but implied to be in-house laboratory testing as part of the manufacturer's analytical performance assessment.
  • Interference Studies Test Set:
    • Sample Size: Not explicitly stated, but different concentrations of interferent were spiked into pooled normal plasma, weak LA positive plasma, and high LA positive plasma. The number of samples for each interferent type is not given.
    • Data Provenance: Not explicitly stated, but implied to be in-house laboratory testing.
  • Assay Cut-off Determination Test Set:
    • Sample Size: 40 normal healthy individual samples.
    • Data Provenance: Not explicitly stated, but likely from a healthy donor pool.
  • "Clinical Sample" Testing (LA Positive, Oral Anticoagulants, etc.) Test Set:
    • Sample Size: Known LA Positive (35 samples), Oral Anticoagulants (5 samples), LMWH (5 samples), UFH (5 samples), DIC (5 samples), Factor Deficiency (6 samples).
    • Data Provenance: Not explicitly stated, but these are patient samples with specific conditions.
  • Comparison Studies (Primary In-house) Test Set:
    • Sample Size: 115 samples (80 Normal / 35 known LA Positive).
    • Data Provenance: Not explicitly stated, but implied to be in-house, possibly from a reference laboratory.
  • Comparison Studies (US Field Sites) Test Set:
    • Sample Size: Site 1: 133 samples (41 LA positive + 92 normal); Site 2: 143 samples (51 LA positive + 92 normal); Site 3: 133 samples (53 LA positive + 80 normal). Over 100 samples per site.
    • Data Provenance: US field sites (presumably clinical laboratories in the US).
  • Matrix Comparison (Citrate Type) Test Set:
    • Sample Size: Plasma from 26 donors. Artificial LA-Positive samples were prepared by spiking this pool.
    • Data Provenance: Not explicitly stated, but implied healthy donors.
  • Matrix Comparison (Fresh vs. Frozen) Test Set:
    • Sample Size: Blood samples from 26 normal healthy donors. LA-Positive samples were prepared by spiking this pool.
    • Data Provenance: Not explicitly stated, but implied healthy donors.
  • Expected Values/Reference Range Study Test Set:
    • Sample Size: 120 normal healthy individuals.
    • Data Provenance: Not explicitly stated, but implied healthy donors.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts or their qualifications used to establish the ground truth for the test sets.

  • For the "Known LA Positive" samples and samples from patients with specific conditions (Oral Anticoagulants, LMWH, UFH, DIC, Factor Deficiency), the ground truth is implied to be based on established clinical diagnosis and/or previous laboratory results for those conditions. The method of determining if a sample was "Known LA Positive" (e.g., diagnosis by a hematologist, positive by multiple other LA tests) is not detailed.
  • For the "Normal" samples, ground truth is based on samples from supposedly healthy individuals.

4. Adjudication Method for the Test Set

No explicit adjudication method is described for conflicting results in the test set.

  • For the comparison studies, the device's results (dRVVVT NR) were compared against the predicate device's results (LAC NR) which served as the reference/ground truth for in-house validation.
  • For the field site validations, "LAC Screen/Confirm" is listed as the reference, suggesting comparison against the predicate device or a combination of standard diagnostic practices at those sites.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. This device is a diagnostic reagent for automated analyzers, not an imaging device requiring human reader interpretation. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable. The study focuses on comparing the new reagent's performance against a predicate device and assessing analytical validity.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

Yes, the studies presented are essentially "standalone" performance studies for the reagent and instrument combination. The device (reagent) and the automated analyzer (ACL TOP Family) generate the results without human interpretive input for the final Normalized Ratio. Human involvement is in sample collection, running the assay on the instrument, and interpreting the final numerical ratio against the established cut-off, but the diagnostic determination of the ratio itself is automated.

7. The Type of Ground Truth Used

The ground truth used depends on the specific study:

  • Comparison Studies: The predicate device (HemosIL LAC Screen & LAC Confirm) served as the reference/ground truth.
  • Specificity Studies (LA Positive, Oral Anticoagulants, etc.): Ground truth was based on the "known" status of the plasma samples (e.g., "Known LA Positive," "Oral Anticoagulants"). How these "known" statuses were originally established (e.g., clinical diagnosis, other laboratory methods) is not detailed.
  • Assay Cut-off and Reference Range Studies: Ground truth was based on plasma samples from "normal healthy individuals."

8. The Sample Size for the Training Set

The concept of a "training set" in the context of an AI/machine learning algorithm does not directly apply here, as this is a chemical reagent-based diagnostic assay. Therefore, there is no explicit training set in the AI sense.

However, if "training set" is interpreted as data used to establish device parameters or optimize its performance before formal validation, the following might be considered:

  • The development process of the improved dRVVT reagents (HemosIL dRVVT Screen and dRVVT Confirm) would have involved extensive R&D and internal testing to optimize their composition and function. This internal optimization data is not detailed in the 510(k) summary.
  • The "Assay cut-off" was determined using 40 normal healthy individual samples, which could be seen as an "internal" reference set for parameter setting.
  • The normal range study used 120 normal healthy individuals to establish reference intervals.

9. How the Ground Truth for the Training Set Was Established

As noted in point 8, a formal "training set" as understood in AI/ML is not applicable. For the data used to establish parameters like the assay cut-off or reference ranges, the ground truth was established by using plasma samples from normal healthy individuals, indicating they were free from the condition the test aims to detect (lupus anticoagulants). The method for confirming their "normal healthy" status (e.g., physical examination, medical history review, other lab tests) is not specified.

{0}------------------------------------------------

510K SUMMARY

A.510(k) NumberK110031AUG 2 4 2011
B.Purpose forSubmissionNew product
D.MeasurandLupus anticoagulant
F.Type of TestDiluted Russell's venom clotting assay
H.ApplicantInstrumentation Laboratory Co.
J.Proprietary &EstablishedNamesHemosIL® dRVVT Screen and HemosIL dRVVT Confirm Assays
L.Regulatory Information
1.Regulation section:21CFR §864.8950, Russell's viper venom reagent
2.Classification:Class II
3.Product code:GIR
4.Device classification name:Reagent, Russell's Viper Venom

Intended Use H.

  • Intended use(s): 1.
    The HemosIL dRVVT Screen and HemosIL dRVVT Confirm assays are qualitative in-vitro diagnostic products to aid in the detection of lupus anticoagulants in human citrated plasma by the diluted Russell's Viper Venom method, on the ACL TOP® Family. The HemosIL dRVVT Screen and HemosIL dRVVT Confirm assays are intended to evaluate patients who have unexplained prolonged APTT test results The HemosIL dRVVT Screen and HemosIL dRVVT Confirm assays should be used in parallel as an integrated test for Lupus Anticoagulant detection.

  • Indication(s) for use: 2.
    Same

Special conditions for use statement(s):

For in-vitro diagnostic use only. For prescription use. HemosIL dRVVT Confirm is intended to be used in conjunction with HemosIL dRVVT Screen.

    1. Special instrument requirements:
      ACL TOP Family Analyzers

{1}------------------------------------------------

Device Description

DRVVT Screen and dRVVT Confirm are improved dRVVT reagents, intended to simplify and standardize the detection of Lupus Anticoagulant (LA) disorder in clinical chemistry evaluations. DRVVT Screen is poor in phospholipid, making it sensitive to LA. The additional amount of phospholipid in dRVVT Confirm neutralizes LA to give shorter clotting times.

Russell's viper venom, in the presence of calcium, directly activates factor X (in a test sample). DRVVT Screen and dRVVT Confirm are therefore unaffected by contact factor abnormalities, factor VII, VIII and IX deficiencies, or inhibitors. As a result, dRVVT Screen and dRVVT Confirm are more specific tests for the evaluation of LA than APTT.

Substantial Equivalence Information J.

  • Predicate device name(s): HemosIL LAC Screen & LAC Confirm (self) 1.
  • Predicate 510(k) number(s): K990302 2.
  • Comparison with predicate: 3.

Similarities

The applicants, HemosIL dRVVT Screen and HemosIL dRVVT Confirm (PNs 000200301500 & 00020301600 respectively) are Substantially Equivalent to their predicates, the HemosIL LAC Screen and HemosIL LAC Confirm (K990302).

Table of similarities:
ItemPredicate DeviceApplicant
Device NameHemosIL LAC Screen & LACConfirmHemosIL dRVVT Screen & dRVVTConfirm
K#K990302K110031
Indicationsfor UseHemosiL LAC Screen and HemosILLAC Confirm are in vitro diagnosticproducts for the detection of lupusanticoagulants (a type ofphospholipid interfering antibody)in human citrated plasma on theACL TOP Family.The HemosIL dRVVT Screen andHemosIL dRVVT Confirm assaysare qualitative in-vitro diagnosticproducts to aid in the detection oflupus anticoagulants in humancitrated plasma by the dilutedRussell's Viper Venom method, onthe ACL TOP® Family. The HemosILdRVVT Screen and HemosIL dRVVTConfirm assays are intended toevaluate patients who haveunexplained prolonged APTT testresults The HemosIL dRVVTScreen and HemosIL dRVVTConfirm assays should be used inparallel as an integrated test forLupus Anticoagulant detection.
Sample TypeCitrated plasmaSame
  • l.

{2}------------------------------------------------

ItemPredicate DeviceApplicant
ReagentcompositionRussell's viper venom,phospholipids, calcium and heparininhibitorSame
Test PrincipleLAC Screen and LAC Confirm areimproved dRVVT reagents,intended to simplify andstandardize the detection of LupusAnticoagulant (LA) disorder inclinical chemistry evaluations. LACScreen is poor in phospholipid,making it sensitive to LA. Theadditional amount of phospholipidin LAC Confirm neutralizes LA togive shorter clotting times.Russell's viper venom, in thepresence of calcium, directlyactivates factor X (in a test sample).Same

Differences

The main difference between the applicant, HemosIL dRVVT Screen and dRVVT Confirm, and their predicates HemosIL LAC Screen and LAC Confirm, is that the applicants have improved stability as compared to their predicates.

Standard/Guidance Document Referenced (if applicable) K.

No performance standard or FDA guidance has been established for these reagents.

L. Test Principle

In dRVVT screening assays, a low, rate-limiting concentration of phospholipid is used to give a clotting time which is sensitive to the presence of lupus anticoagulants, since anti-phospholipid antibodies interfere with the clot-promoting role of phospholipid in vitro. A prolonged clotting time of a patient sample that does not correct with the addition of an equal volume of normal plasma suggests the presence of a lupus anticoagulant.

An abnormal result for the initial dRVVT screening assay should be followed by a dRVVT confirmatory test. In this test, the inhibitory effect of lupus anticoagulants on phospholipids in the dRVVT can be overcome by adding an excess of phospholipid to the assay.

The clotting times of both the initial dRVVT assay and confirmatory test are subsequently normalized and then used to determine a ratio: the time without phospholipid excess to time with phospholipid excess, the so called "normalized ratio". A ratio greater than the laboratory established cut-off is considered a positive result and implies that the patient may have anti-phospholipid antibodies.

{3}------------------------------------------------

Performance Characteristics

  • Analytical performance 1.
    • Precision/Reproducibility a.

Precision was assessed utilizing 3 lots of reagent on 3 representative members of the ACL TOP Family (ACL TOP and ACL TOP 500 CTS) by 3 independent operators. Precision was evaluated in accordance with CLSI EP05-A213, for 20 days, with 2 runs per day and 2 replicates per run for each sample level (n=80/ instrument/ lot), with the following results:

ACL TOP FamilydRVVT NR
LA Negative Control1.00
Weakly LA Positive Control1.35
LA Positive Control1.77
HemosIL dRVVT NRWithin Run (%CV)Between Run (%CV)Total (%CV)
Lot 1Lot 2Lot 3Lot 1Lot 2Lot 3Lot 1Lot 2Lot 3
LA Negative Control1.22.00.81.72.81.92.33.42.1
Weakly LA Positive1.10.90.62.72.12.03.02.62.2
LA Positive Control1.50.91.14.43.42.55.03.53.0
  • Linearity/assay reportable range: NA, qualitative assay. b.
  • C. Traceability, Stability, Expected values (controls, calibrators, or methods):

Unopened reagents are stable until the expiration date shown on the vial when stored at 2-8°C. Stability after reconstitution: 15 days at 2-8°C in the closed original vial or 3 days at 15°C in the original vials on the ACL TOP Family. dRVVT Screen/ Confirm may be used with either fresh or frozen samples. For optimal stability remove reagents from the system and store them, closed, at 2-8°C, in their original vials. Based on the results of the accelerated stability study, a shelf-life of at least 2 years is claimed for the products when stored at 2- 8°C. Real-time stability testing is ongoing, and will be used to update the shelf life as more data becomes available.

  • Detection limit: NA d.
  • Analytical specificity: e.

Interference studies were conducted on a representative member of the ACL TOP Family. Different concentrations of interferent were spiked into pooled normal plasma, weak LA Positive plasma and high LA Positive plasma The maximum concentration tolerated in the assay was defined as the highest concentration of interferent relative to the recovered value of the base clotting time ± 15%. The maximum tolerated concentrations not causing interference at any LA level tested were:

Possible InterferantNot affected by concentrations
Unfractionated Heparin (UFH)≤ 1.0 IU/mL
Low Molecular Weight Heparin (LMWH)≤ 1.0 IU/m
Hemoglobin≤ 200 mg/dL
Bilirubin≤ 10 mg/dL
Triglyceride≤ 500 mg/dL

M.

{4}------------------------------------------------

Normalized dRVVT ratio higher than the internal-study cut-off (NR > 1.2) was found in the following plasma samples using dRVVT Screen/dRVVVT Confirm:

SampleACL TOP BaseACL TOP 500CTS
Known LA Positive100% (35/35)100% (35/35)
Oral Anticoagulants40% (2/5)40% (2/5)
LMWH0% (0/5)0% (0/5)
UFH20% (1/5)0% (0/5)
DIC0% (0/5)0% (0/5)
Factor Deficiency0% (0/6)0% (0/6)

f. Assay cut-off:

The Normalized Ratio cut off was determined as recommended using 40 normal healthy individual samples and calculating the Mean + 3SD. The results were obtained using a specific lot of reagent. Due to many variables which may affect results, each laboratory should establish its own NR cut off.

SystemNormalized dRVVT Ratio Cut Off
ACL TOP>1.2
ACL TOP 500CTS>1.2

Comparison studies: 2.

  • a. An in-house method comparison was performed in accordance with EP09-A244, on 115 samples (80 Normal/ 35 known LA Positive), on a representative member of the ACL TOP Family (ACL TOP Base & ACL TOP 500 CTS). The Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) were calculated with the following result(s):
Cut-Off(Mean+3SD)LACNRdRVVTNRPPACI 95%NPACI 95%OverallReference
ACL TOP> 1.2> 1.2100.0%(35/35)90.1-100.0%100.0%(80/80)95.4-100.0%100%(115/115)LAC Screen/Confirm
ACL TOP500 CTS> 1.2> 1.2100.0%(35/35)90.1-100.0%100.0%(80/80)95.4-100.0%100%(115/115)

Results were subsequently validated by 3 US field sites. Each site established its own cut-off, and validated that cut-off with 100+ samples, with the following result(s):

Cut-Off(Mean+3SD)LACNRdRVVTNRPPACI 95%NPACI 95%OverallReference
Site 1> 1.2> 1.292.7%(38/41)80.6-97.5%98.9%(91/92)94.1- 99.8%97%(129/133)LAC Screen,Confirm
Site 2> 1.3> 1.390.2%(46/51)79.0-95.7%98.9%(91/92)94.1- 99.8%95.8%137/143
Site 3> 1.3> 1.298.1%(52/53)90.1-99.7%100.0%(80/80)95.4-100.0%99.2%(132/133)

{5}------------------------------------------------

Matrix Comparison b.

A citrate study was performed to assess the effect on the assays of collecting the blood samples in 3.8% versus 3.2% sodium citrate sample tubes. Plasma from 26 donors was collected, in parallel, in both tube types. Artificial LA-Positive samples were prepared by spiking with different amounts of B2gPl antibodies to produce a range of concentrations. Using the previously established cut-off, the dRVVT Normalized Ratios for both 3.8% and 3.2% sodium citrate sample tubes were calculated. The two NRs were compared for their Positive and Negative Percent Agreement.

Results showed that the dRVVT Normalized Ratio on the ACL TOP is not affected by the type of citrate tubes used to draw blood samples.

3.8 v. 3.2% Na CitratePPACI 95%NPACI 95%
ACL TOP19/19 (100%)83.2-100%24/26 (92%)75.9-97.9%

A fresh v. frozen study was conducted to demonstrate that the results of fresh and frozen and once thawed samples are equivalent. Blood samples were drawn from 26 normal healthy donors. LA-Positive samples were prepared by spiking this pool with different amounts of ß2gPl antibodies. Fresh samples were kept at room temperature. Frozen samples were stored at -65℃ for 24 hr, prior to being thawed and analyzed at room temperature. Using the previously established cut-off, the dRVVT Normalized Ratios for both fresh and frozen (normal and LA-antibodies-spiked) samples were calculated. The two NRs were compared for their Positive and Negative Percent Agreement. The method comparison demonstrated that the dRVVT Normalized Ratio on the ACL TOP Family is not affected by whether the analysis is performed on fresh or frozen samples.

Fresh vs. FrozenPPACI 95%NPACI 95%
ACL TOP28/28 (100%)87.9%-100%26/26 (100%)87.1%-100%
    1. Clinical Studies:
    • a. Clinical Sensitivity: NA
    • b. Clinical Specificity: NA
    • C. Other clinical supportive data (when a. and b. are not applicable): NA
  • Clinical cut-off: NA 4.
    1. Expected values/Reference range:

A normal range study (n=120) was performed, in accordance with CLSI C28-A32, using dRVVT Screen/dRVVT Confirm on representative members of the ACL TOP Family. The following Reference intervals were established for dRVVT Screen, and for the dRVVT Screen/ Confirm Normal Ration (NR):

Normal Ratio Reference Interval (NR)
SystemLower LimitUpper Limit
ACL TOP0.92 (0.91-0.93)1.11 (1.10-1.15)
ACL TOP 500 CTS0.91 (0.89-0.92)1.13 (1.11-1.16)

{6}------------------------------------------------

N. Proposed Labeling

The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10.

o. Conclusion

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

P. Administrative Information

Applicant Contact Information

Name of applicant:Instrumentation Laboratory Co.
Mailing address:180 Hartwell RoadBedford, MA 01730, USA
Phone #:781-861-4350
Fax #:781-861-4207
E-mail address:jemery@ilww.com
Contact:Jacqueline Emery, BSEERegulatory Affairs Manager
Date PreparedJuly 30, 2011

Reference(s):

    1. CLSI C28-A3: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory, 30 edition.
    1. Pengo V et al. Update of the Guidelines for Lupus Anticoagulant Detection. J. Thromb. Haem. 2009; 7:1737-1740.
    1. Clinical and Laboratory Standards/CLSI. Establishment of Quantitative Measurement Procedures; Approved Guideline. Document EP5-A3: Vol. 0 No. 0.
    1. Clinical and Laboratory Standards/CLSI. Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline. Document EP9-A2: Vol. 22 No.19.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized image of a human figure, represented by flowing lines, with its arms outstretched.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Instrumentation Laboratory Co. c/o Ms. Jacqueline Emery Regulatory Affairs Manager 180 Hartwell Road Bedford. MA 01730

AUG 2 4 2011

Re: K110031

Trade/Device Name: HemosIL® dRVVT Screen and dRVVT Confirm Regulation Number: 21 CFR 864.8950 Regulation Name: Russell Viper Venom Test Regulatory Class: Class II Product Code: GIR, GGC Dated: August 16, 2011 Received: August 19, 2011

Dear Ms. Emery:

(

We have reviewed your Section 510(k) premarket notification of intent to market the device w & nave and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, do nots may been revire approval of a premarket approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, additional come of Enting may 2007 - 10 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I toase of a rised a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must of any I out all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket

{8}------------------------------------------------

Page 2 - Ms. Jacqueline Emery

requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirents as set form in the quality of think (<=) (<=) (<= ) == ============================================================================================================ will anow you to begin marketing your antial equivalence of your device to a legally marketed nonication. The I Drice results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and II you desire specific advice of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-807), prease condor the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to under the MDK regulation (27 Of RT LEV 807) products good gefault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may oount only generarers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

ie m. elan

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K 11003|

Device Name: HemosIL® dRVVT Screen and HemosIL® dRVVT Confirm

Indications for Use:

The HemosiL dRVVT Screen and HemoslL dRVVT Confirm assays are qualitative in-vitro diagnostic products to aid in the detection of lupus anticoagulants in human citrated plasma by the diluted Russell's Viper Venom method, on the ACL TOP® Family. The HemostL dRVVT Screen and HemosIL dRVVT Confirm assays are intended to evaluate patients who have unexplained prolonged APT test results The HemoslL dRVVT Screen and HemosIL dRVVT Confirm assays should be used in parallel as an integrated test for Lupus Anticoagulant detection.

Prescription Use __ V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

mchan

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K110031

§ 864.8950 Russell viper venom reagent.

(a)
Identification. Russell viper venom reagent is a device used to determine the cause of an increase in the prothrombin time.(b)
Classification. Class I (general controls).