K Number
K103673

Validate with FDA (Live)

Date Cleared
2011-08-18

(245 days)

Product Code
Regulation Number
866.3220
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the simultaneous qualitative detection and differentiation of Giardia cyst antiqen and Cryptosporidium oocyst antigen in a single test device. It is intended for use with human fecal specimens from patients with gastrointestinal symptoms to aid in the diagnosis of Giardia and/or Cryptosporidium gastrointestinal infection. The test results should be considered in conjunction with the patient history.

Device Description

The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test is a rapid membrane immunoassay for the simultaneous detection of Giardia cyst antigen and Cryptosporidium occyst antigen in a single test device. It is performed with a 25 to 30-minute total incubation time. The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test uses monoclonal and polyclonal antibodies to cell-surface antigens of the device contains a Reaction Window with three vertical lines of immobilized antibodies. The Giardia test line ("Giar") contains mouse monoclonal antibodies against Giardia. The Crypto test line ("Cryp") contains mouse monoclonal antibodies against Cryptosporidium. The control line ("C") is a dotted line that contains anti-horseradish peroxidase (HRP) antibodies. The Conjugate consists of polyclonal antibodies coupled to horseradish peroxidase. To perform the test, the sample is added to a tube containing a mixture of Diluent and Conjugate. The diluted sample-conjugate mixture is added to the Sample Well and the device is allowed to incubate at room temperature for 15 minutes. During the incubation, cyst antigens in the sample bind the antibody-peroxidase conjuqates. The antigen-antibody-conjugate complexes migrate through a filter pad to a membrane where they are captured by the immobilized Giardia and/or Cryptosporidium-specific antibodies in the test lines. The Reaction Window is subsequently washed with Wash Buffer, followed by the addition of Substrate. After a 10 minute incubation period, the reaction is examined visually for the appearance of a vertical blue line on either side of the Reaction Window. A blue line indicates a positive "control" reaction, indicated by a vertical dotted blue line under the "C" portion of the Reaction Window. confirms that the test is working properly and the results are valid.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ device, based on the provided text:

Acceptance Criteria and Reported Device Performance

The acceptance criteria for a diagnostic test like this are typically defined by performance metrics such as sensitivity, specificity, and agreement with a predicate device or gold standard. The reported performance for the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test against the Microscopy - IFA gold standard is as follows:

Performance MetricAcceptance Criteria (Implicit from reported results and regulatory acceptance)Reported Device Performance (Giardia)Reported Device Performance (Cryptosporidium)
SensitivityHigh (e.g., >95%)98.9% (95% CI: 95.7 - 99.8%)100% (95% CI: 96.7 – 100%)
SpecificityHigh (e.g., >95%)100% (95% CI: 99.2 - 100%)99.8% (95% CI: 99.0 – 100%)
Overall CorrelationHigh (e.g., >95%)99.7% (95% CI: 99.7 - 99.7%)99.9% (95% CI: 100 – 100%)

The reported performance against Commercial ELISA Predicate Devices is as follows:

Performance MetricAcceptance Criteria (Implicit from reported results and regulatory acceptance)Reported Device Performance (Giardia)Reported Device Performance (Cryptosporidium)
Percent Positive AgreementHigh (e.g., >95%)99.1% (95% CI: 96.3 - 99.8%)99.2% (95% CI: 95.2 - 100%)
Percent Negative AgreementHigh (e.g., >95%)99.7% (95% CI: 98.7 - 99.9%)99.6% (95% CI: 98.7 - 99.9%)
Overall Percent AgreementHigh (e.g., >95%)99.5% (95% CI: 99.5 - 99.5%)99.5% (95% CI: 99.5 - 99.5%)

The regulatory acceptance of the device (K103673) by the FDA suggests that these reported performance metrics met their criteria for substantial equivalence to legally marketed predicate devices.


Study Information

  1. Sample size used for the test set and the data provenance:

    • Test set for comparison to Microscopy (IFA): N = 791 (combined from Study Sites #1 and #3). This included 220 fresh, 140 frozen, 216 preserved-formalin, and 215 preserved-SAF fecal specimens.
    • Test set for comparison to Commercial ELISAs (predicate devices): N = 849 (combined from Study Sites #1 and #2). This included 349 fresh, 322 frozen, 36 preserved-formalin, and 142 preserved-SAF fecal specimens.
    • Data Provenance: The studies were conducted at "3 geographically diverse sites" within the US. The data is prospective in the sense that the test samples were collected and then tested with the new device, but the text doesn't specify if these were newly collected for the study or a collection of existing samples. Given the various preservation methods, it's likely a mix of prospective collection and retrospective analysis of stored samples.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth for the clinical specimens. It refers to "Microscopy - IFA (considered the gold standard)" and "two commercially available ELISAs (predicate devices)" as the comparators, implying the expertise lies in the performance of those established methods.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • The document does not specify an adjudication method like 2+1 or 3+1 for resolving discrepancies in the test set. The comparisons are presented as direct comparisons between the new device's results and the results from the gold standard/predicate devices.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not reported. This device is a rapid membrane enzyme immunoassay (a lab test), not an AI-assisted imaging device that would involve human readers.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, this was a standalone performance study of the diagnostic device itself. The "device performance" metrics (sensitivity, specificity, agreement) directly reflect the algorithm's (immunoassay's) ability to detect the antigens in the samples. There is no human-in-the-loop scenario described for the device's main function, as it's a qualitative visual membrane assay. The reading of the lines by a technician would be part of the standard operation of this type of IVD, but the fundamental detection is by the assay itself.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For clinical performance, the primary ground truth reference was Microscopy - IFA (Immunofluorescence Assay), which is explicitly stated as "considered the gold standard" for detecting Giardia cysts and Cryptosporidium oocysts in fecal specimens.
    • Additionally, two commercially available ELISA predicate devices were used as comparators for agreement in other parts of the study.
    • For analytical sensitivity (LOD) and cross-reactivity studies, the ground truth involved purified Giardia cysts or Cryptosporidium oocysts quantified by immunofluorescent antibody microscopy (IFA) or documented positive specimens for other organisms by microscopy.
  7. The sample size for the training set:

    • The document does not explicitly mention a "training set" in the context of machine learning or AI. This is a traditional immunoassay. Therefore, there's no training set as understood in AI/ML development. The development of the assay (e.g., antibody selection, reagent concentrations) would be an iterative process, but not in the "training set" paradigm.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no specific "training set" mentioned in the context of an AI/ML algorithm for this immunoassay device. The assay development would rely on internal validation and optimization against known positive and negative samples, similar to how the analytical studies for sensitivity and cross-reactivity were performed.

{0}------------------------------------------------

K103673

510(k) SUMMARY

Contact InformationDonna T. LinkDirector QA, RA and ComplianceTECHLAB®, Inc.2001 Kraft DriveCorporate Research CenterBlacksburg, VA 24060Phone: 540-953-1664FAX: 540-953-1665Email: dlink@techlab.com
Date Prepared09 August 2011
Product and Trade NameGIARDIA/CRYPTOSPORIDIUM QUIK CHEK™
Classification21 CFR 866.3220

Predicate Devices

  • Merifluor™ Cryptosporidium / Giardia Kit .
  • GIARDIA II .
  • . CRYPTOSPORIDIUM II

Intended Use

The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the simultaneous qualitative detection and differentiation of Giardia cyst antiqen and Cryptosporidium oocyst antigen in a single test device. It is intended for use with human fecal specimens from patients with gastrointestinal symptoms to aid in the diagnosis of Giardia and/or Cryptosporidium gastrointestinal infection. The test results should be considered in conjunction with the patient history.

Device Description

The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test is a rapid membrane immunoassay for the simultaneous detection of Giardia cyst antigen and Cryptosporidium occyst antigen in a single test device. It is performed with a 25 to 30-minute total incubation time. The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test uses monoclonal and polyclonal antibodies to cell-surface antigens of the device contains a Reaction Window with three vertical lines of immobilized antibodies. The Giardia test line ("Giar") contains mouse monoclonal antibodies against Giardia. The Crypto test line ("Cryp") contains mouse monoclonal antibodies against Cryptosporidium. The control line ("C") is a dotted line that contains anti-horseradish peroxidase (HRP) antibodies. The Conjugate consists of polyclonal antibodies coupled to horseradish peroxidase. To perform the test, the sample is added to a tube containing a mixture of Diluent and Conjugate. The diluted sample-conjugate mixture is added to the Sample Well and the device is allowed to incubate at room temperature for 15 minutes. During the incubation, cyst antigens in the sample bind the antibody-peroxidase conjuqates. The antigen-antibody-conjugate complexes migrate through a filter pad to a membrane where they are captured by the immobilized Giardia and/or Cryptosporidium-specific antibodies in the test lines. The Reaction Window is subsequently washed with Wash Buffer, followed by the addition of Substrate. After a 10 minute incubation period, the reaction is examined visually for the appearance of a vertical blue line on either side of the Reaction Window. A blue line indicates a positive "control" reaction, indicated by a vertical dotted blue line under the "C" portion of the Reaction Window. confirms that the test is working properly and the results are valid.

Comparative Information of Equivalent Devices

{1}------------------------------------------------

Kit Name510(k)NumbersIntended UseFormatMaterialsTargetPopulation
MicroscopyN/ADirect detectionof Giardia cystsandCryptosporidiumoocysts in fecalspecimensMicroscopyVarious StainsPersonssuspected ofhaving Giardia orCryptosporidiuminfection
Merifluor™Cryptosporidium/Giardia KitK912408Direct detectionof Giardia cystsandCryptosporidiumoocysts in fecalspecimensMicroscopywith DirectFluorescentAntibody –Immuno-fluorescence(IFA)Highly specificantibodiesagainst GiardiaandCryptosporidiumPersonssuspected ofhaving Giardia orCryptosporidiuminfection
GIARDIA IIK033274Detection ofGiardia cystantigen in fecalspecimensELISAHighly specificantibodiesagainst GiardiaPersonssuspected ofhaving Giardiainfection
CRYPTOSPORIDIUM IIK052932Detection ofCryptosporidiumoocyst antigen infecal specimensELISAHighly specificantibodiesagainstCryptosporidiumPersonssuspected ofhavingCryptosporidiuminfection

Summary of Performance Data Clinical Performance

The performance of the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test was evaluated at 3 geographically diverse sites. At Site #1 and Site #3 the performance of the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test was compared to Microscopy (IFA) and included 220 fresh, 140 frozen, 216 preserved-formalin, and 215 preserved-SAF. At Site #1 and #2 the performance of the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test was compared to two commercially available ELISAs (predicate devices for Giardia and Cryptosporidium) and included 349 fresh, 322 frozen, 36 preserved-formalin, and 142 preserved-SAF.

Performance as compared to Microscopy (IFA) - Combined Results for Study Sites #1 and #3:

For Giardia spp.

The following table shows a summary of the clinical performance of the Giardia portion of the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test. The results show that the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test exhibited a sensitivity of 98.9%, a specificity of 100%, and an overall correlation of 99.7% with Microscopy - IFA (considered the gold standard).

{2}------------------------------------------------

Combined Results - Clinical Performance Comparing the Giardia Line of the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ Test to Microscopy - IFA

N = 791
Microscopy - IFA Giardia positiveMicroscopy - IFA Giardia negative
GIARDIA/CRYPTOSPORIDIUM QUIK CHEK TMGiardia Line Positive1810
GIARDIA/CRYPTOSPORIDIUM QUIK CHEK TMGiardia Line Negative2608
95% Confidence Limits
Sensitivity98.9%95.7 - 99.8%
Specificity100%99.2 - 100%
Correlation99.7%99.7 - 99.7%

For Cryptosporidium spp.

The following table shows a summary of the clinical performance of the Cryptosporidium portion of the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test. The results show that the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test exhibited a sensitivity of 100%, a spectlicity of 99.8%, and an overall correlation of 99.9% with Microscopy - IFA (considered the gold standard).

Combined Results - Clinical Performance Comparing the Cryptosporidium Line of the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ Test to Microscopy - IFA

N = 791Microscopy - IFA Crypto. positiveMicroscopy - IFA Crypto. negative
GIARDIA/CRYPTOSPORIDIUM QUIK CHEKTMCryptosporidium Line Positive1401
GIARDIA/CRYPTOSPORIDIUM QUIK CHEKTMCryptosporidium Line Negative0650
95% Confidence Limits
Sensitivity100%96.7 – 100%
Specificity99.8%99.0 – 100%
Correlation99.9%100 – 100%

{3}------------------------------------------------

Performance as Compared to Two Predicate Devices (commercially available devices for Giardia and Cryptosporidium) - Combined Results for Study Sites #1 and #2:

The combined results of our performance evaluations at Study Site #1 and #2 as compared to two commercially available ELISAs (predicate devices for Giardia and Cryptosporidium) exhibited a 99.1% agreement for Giardia positive specimens, a 99.7% agreement for Giardia negative specimens, with an overall agreement of 99.5%. The test exhibited a 99.2% agreement for Cryptosporidium positive specimens, 99.6% agreement for Cryptosporidium negative specimens, and an overall agreement of 99.5%.

GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test Giardia Line versus the Commercial ELISA for Giardia Detection

N = 849Commercial ELISA -Giardia positiveCommercial ELISA -Giardia negative
GIARDIA/CRYPTOSPORIDIUM QUIK CHEK TMGiardia Line Positive2132
GIARDIA/CRYPTOSPORIDIUM QUIK CHEK TMGiardia Line Negative2632
95% Confidence Limits
Percent Positive Agreement99.1%96.3 - 99.8%
Percent Negative Agreement99.7%98.7 - 99.9%
Overall Percent Agreement99.5%99.5 - 99.5%

GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test Cryptosporidium Line versus the Commercial ELISA for Cryptosporidium Detection

N = 849Commercial ELISA -Cryptosporidium positiveCommercial ELISA -Cryptosporidium negative
GIARDIA/CRYPTOSPORIDIUM QUIK CHEK ™Cryptosporidium Line Positive1303
GIARDIA/CRYPTOSPORIDIUM QUIK CHEK ™Cryptosporidium Line Negative1715
95% Confidence Limits
Percent Positive Agreement99.2%95.2 - 100%
Percent Negative Agreement99.6%98.7 - 99.9%
Overall Percent Agreement99.5%99.5 - 99.5%

{4}------------------------------------------------

Analytical Sensitivity

The analytical sensitivity of the device was determined by spiking purified Giardia cysts or Cryptosporidium oocysts quantified by immunofluorescent antibody microscopy (IFA) into negative human fecal specimens. The concentration of Giardia cysts and Cryptosporidium occysts in fecal matrix where specimens were positive by the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test 95% of the time were utilized to describe the assay limitof-detection (LOD). Test results determined the LOD for the assay to be 6000 cysts/mL of feces for Giardia (equivalent to 133 cysts detected per test) and 6000 occysts/mL feces for Cryptosporidium (equivalent to 133 oocysts detected per test). Because the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK test detects soluble antigen in fecal specimens in addition to cysts and occysts, this LOD study represents an estimate of analytical sensitivity based on purified Giardia cysts and Cryptosporidium occysts. Clinical specimens contain varying amounts of free antigen per Giardia cyst or Cryptosporidium oocyst.

Cross-Reactivity

The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test was evaluated for cross-reactivity with the bacterial and viral strains listed below. None of the strains were shown to cross-react with the GIARDIA/CRYPTOSPORIDIUM QUIK CHEKTM test.

Aeromonas hydrophilaClostridium difficile (strain 630)Salmonella typhimurium
Bacillus cereusEnterococcus faecalisShigella dysenteriae
Bacillus subtilisEscherichia coliShigella flexneri
Bacteroides fragilisEscherichia coli 0157:H7Shigella sonnei
Campylobacter coliEscherichia coli ETEC (enterotoxic)Staphylococcus aureus
Campylobacter fetusEscherichia coli EPECStaphylococcus aureus (Cowan's)
Campylobacter jejuni(enteropathogenic)Staphylococcus epidermidis
Candida albicansEscherichia coli EIEC (enteroinvasive)Vibrio parahaemolyticus
Clostridium bifermentansKlebsiella pneumoniaeYersinia enterocolitica
Human Adenovirus 1 and 3Coxsackievirus B5Human paraechovirus 1 (Echovirus
Adenovirus, Type 2, 5, 40 and 41Human CoronavirusEnterovirus 68, 69
Human Coxsackievirus B2, B3,Echovirus 11, 18, 33Human Enterovirus 70, 71

Human Echovirus 9

n paraechovirus 1 (Echovirus 22) ovirus 68, 69 Human Enterovirus 70, 71 Human rotavirus

Additionally, the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test was run on fecal specimens documented to be positive for other parasites by microscopy. No cross-reactivity was seen with the following organisms for either the Giardia-portion or the Cryptosporidium-portion of the test. Cross-reactivity to Astrovirus and Caliciviruses has not been established.

Ascaris lumbricoides eggs Blastocystis hominis Chilomastix mesnili Cyclospora cayetanensis

and B4

Dientamoeba fragilis Diphyllobothrium latum eggs Endolimax nana Entamoeba coli

Entamoeba hartmanni Entamoeba histolytica/E. dispar Hookworm egas lodamoeba bütschlii Trichuris trichiura eggs

{5}------------------------------------------------

INTERFERING SUBSTANCES (U.S. Formulations)

The following substances had no effect on positive or negative test results analyzed at the concentrations indicated: Hog gastric mucin (3.5% w/v), Human blood (40% v/v), Barium sulfate (5% w/v), Kaopectate® (5% v/v), Pepto-Bismol® (5% v/v), Steric/Palmitic Acid (40% w/v), Metronidazole (0.25% w/v), Vancomycin (0.25% w/v).

REPRODUCIBILITY

A total of 22 fecal specimens were pre-characterized by commercially available predicate devices. The samples included 6 Giardia-positive specimens (3 mid-range positives), 6 Cryptosporidium-positive specimens (3 low positives), and 4 Giardia/Cryptosporidium-positive specimens (2 of which were low Giardia-positives and 2 of which were low Cryptosporidium-positives), and 6 specimens negative for both parasites. All specimens were coded to prevent their identification during testing. Testing was performed at 3 sites. The samples were tested, twice a day over a 5-day period by multiple technicians at each site using 2 different kit lots. A positive and negative control was run with each panel of the masked samples. The results from each laboratory were subsequently submitted to TECHLAB®, Inc. and compared with in-house results. The results were consistent among the different locations, and exhibited a correlation of 100%. The positive specimens consistently tested positive and the negative specimens consistently tested negative at all sites using the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test.

PRECISION - INTRA-ASSAY

For the determination of intra-assay performance, 6 positive fecal specimens (two positive for Giardia, two positive for Cryptosporidium, two positive for both Giardia and Cryptosporidium) and six negalive fecal specimens were analyzed. Each specimen was assayed on 5 cassettes. All positives remained positive and all negatives remained negative.

PRECISION - INTER-ASSAY

For the determination of inter-assay performance, 16 positive fecal specimens (six positive for Giardia, six positive for Cryptosporidium, and four positive for both Giardia and Cryptosporidium) and six neqative fecal specimens were assayed twice a day over a four-day period using the GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ kit. All positives remained positive and all negatives remained negative.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized bird-like figure with three curved lines representing its wings or body.

` Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

TechLab, Inc. c/o Ms. Donna T. Link Director, Quality Assurance, Regulatory & Compliance 2001 Kraft Drive Blacksburg, VA 24060-6358

AUG 1 8 2011

Re: K103673

Trade/Device Name: The GIARDIA/CRYTOSPORIDIUM QUIK CHEK™ Regulation Number: 21 CFR§ 866.3220 Regulation Name: Qualitative membrane enzyme immunoassay Regulatory Class: Class II Product Code: MHJ, MHJ, MHI Dated: August 8, 2011 Received: August 8, 2011

Dear Ms. Link:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter

{7}------------------------------------------------

Page 2 – Donna T. Link

will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Fally acton

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

2. INDICATIONS FOR USE

510(k) Number: K103673

Device Name: GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™

Indications For Use:

The GIARDIA/CRYPTOSPORIDIUM QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the simultaneous qualitative detection and differentiation of Giardia cyst antigen and Cryptosporidium oocyst antigen in a single test device. It is intended for use with human fecal specimens from patients with gastrointestinal symptoms to aid in the diagnosis of Giardia and/or Cryptosporidium gastrointestinal infection. The test results should be considered in conjunction with the patient history.

FOR IN VITRO DIAGNOSTIC USE.

Prescription Use __ V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

Luddi te. Pode

vision Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K103673

§ 866.3220

Entamoeba histolytica serological reagents.(a)
Identification. Entamoeba histolytica serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies toEntamoeba histolytica in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identifyEntamoeba histolytica directly from clinical specimens. The identification aids in the diagnosis of amebiasis caused by the microscopic protozoan parasiteEntamoeba histolytica and provides epidemiological information on diseases caused by this parasite. The parasite may invade the skin, liver, intestines, lungs, and diaphragm, causing disease conditions such as indolent ulcers, an amebic hepatitis, amebic dysentery, and pulmonary lesions.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.