K Number
K103418

Validate with FDA (Live)

Device Name
DESARA MINI
Date Cleared
2011-11-30

(373 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Desara Mini is intended to be used in females to position a suburethral mesh sling for treatment of Genuine Stress Urinary Incontinence (SUI) and mixed incontinence resulting from urethral hypermobility. The mesh is placed transvaginally using a reusable introducer (sold separately)

Device Description

The Desara Mini is a sterile, single-use pubourethral sling. The device is comprised of monofilament polypropylene yarn, which is knitted into a mesh, two self-fixating tips and two guiding suture loops at each end of the device. The device contains a visible midline marker to assist the surgeon in placement of the device.

The Desara Mini is available in one configuration and a reusable introducer is available separately.

AI/ML Overview

The provided 510(k) summary for the Caldera Medical Desara Mini surgical mesh primarily focuses on establishing substantial equivalence to predicate devices through technical and performance characteristic comparisons, rather than presenting a study to prove acceptance criteria for a new, unique performance claim. As such, the information needed to fully respond to your request regarding acceptance criteria and a study proving device performance in the context of an AI/human reader study is not present in this document.

However, I can extract the available information regarding performance data and testing methods used to demonstrate substantial equivalence, which serves a similar purpose in the medical device approval process:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" in a quantitative, measurable way for novel performance endpoints. Instead, it demonstrates that the Desara Mini meets the established characteristics and performance benchmarks of its predicate devices. The "reported device performance" is essentially that the Desara Mini's characteristics are "substantially equivalent" to the predicate devices.

Here's an interpretation of the performance data in relation to the predicate devices:

Performance Characteristic/TestAcceptance Criteria (Implied by Predicate Device Equivalence)Reported Device Performance
Mesh Characteristics
Mesh ThicknessEquivalent to Desara Mesh (K110169)Substantially equivalent
Mesh Weave CharacteristicsEquivalent to Desara Mesh (K110169)Substantially equivalent
Pore SizeEquivalent to Desara Mesh (K110169)Substantially equivalent
Mesh DensityEquivalent to Desara Mesh (K110169)Substantially equivalent
Tensile StrengthEquivalent to Desara Mesh (K110169)Substantially equivalent
Mesh StiffnessEquivalent to Desara Mesh (K110169)Substantially equivalent
Flexural RigidityEquivalent to Desara Mesh (K110169)Substantially equivalent
Tear ResistanceEquivalent to Desara Mesh (K110169)Substantially equivalent
Burst StrengthEquivalent to Desara Mesh (K110169)Substantially equivalent
Suture PulloutEquivalent to Desara Mesh (K110169)Substantially equivalent
Pyrogen LevelsEquivalent to Desara Mesh (K110169)Substantially equivalent
BiocompatibilityMeets FDA guidance for biocompatibility (as per Desara Mesh)Passed all testing requirements; substantially equivalent
Sterility/PackagingMeets FDA guidance for sterility, aging, shelf life, transportation, sterilization (as per Desara Mesh)Passed all testing requirements; substantially equivalent
Mechanical/Functional
Junction StrengthEquivalent to MiniArc (K073703)Demonstrates equivalent
Device Function (Fixation Tips, Polypropylene Mesh, Suture Loops, Introducer)Equivalent to MiniArc (K073703)Demonstrates equivalent
Device PerformanceEquivalent to MiniArc (K073703)Demonstrates equivalent

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify a "test set" in the context of human data or clinical studies for performance evaluation. The testing described involves bench testing for material properties and mechanical performance, and cadaver lab testing for functional performance. No sample sizes for these tests are provided. Data provenance is not specified beyond being "bench" and "cadaver lab" testing. This is common for a 510(k) submission where substantial equivalence is primarily based on material and design similarities, rather than new clinical effectiveness.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

Not applicable. This document describes a medical device submission based on substantial equivalence through material and functional testing, not a study requiring expert readers to establish ground truth for image interpretation or diagnosis.

4. Adjudication Method:

Not applicable, as no expert consensus or ground truth establishment process is described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC comparative effectiveness study was not done. The submission is for a surgical mesh and its performance relies on physical and mechanical properties, not on diagnostic accuracy enhanced by AI.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

Not applicable. The Desara Mini is a physical surgical device, not a software algorithm.

7. Type of Ground Truth Used:

The "ground truth" for the performance claims in this submission is the established performance characteristics and safety profile of the predicate devices (Desara Mesh and MiniArc), as well as established industry standards and FDA guidance for surgical mesh testing. For instance, the mesh characteristics were assessed against standards, and biocompatibility was assessed against ISO-10993.

8. Sample Size for the Training Set:

Not applicable. This document describes a medical device submission where the relevant "training" would be the historical data and performance of the predicate devices and general knowledge of surgical mesh properties, not a machine learning training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable. (See #8).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for Caldera Medical. The logo consists of a circular graphic on the left, followed by the text "CALDERA MEDICAL" in bold, sans-serif font. The graphic appears to be an abstract representation of a biological or medical element.

510(k) Summary

Date of Summary:

July 12, 2011

Submitted by:

Submitter: Address:

Contact: Phone:

Caldera Medical, Inc. 5171 Clareton Drive Agoura Hills, CA 91301 Vicki Gail, Manager QA/RA (818) 879-6555 x 102

Device Name:

Device Name:Surgical Mesh (878.3300)
Trade Name:Desara Mini
Common Name:Surgical Mesh
Device Class:Class II, Product Code FTL, 21 CFR 878.3300, Surgical Mesh,Plastic and General Surgery
Predicate Devices:Desara mesh, #K101169MiniArc, #K073703

Description of Device:

The Desara Mini is a sterile, single-use pubourethral sling. The device is comprised of monofilament polypropylene yarn, which is knitted into a mesh, two self-fixating tips and two guiding suture loops at each end of the device. The device contains a visible midline marker to assist the surgeon in placement of the device.

The Desara Mini is available in one configuration and a reusable introducer is available separately.

Intended Use of Device:

Desara Mini is intended to be used in females to position a mesh for treatment of Genuine Stress Urinary Incontinence (SUI) and mixed incontinence resulting from urethral hypermobility. The mesh is placed transvaginally using a reusable introducer (sold separately)

Technological Characteristics

The Desara Mini utilizes a similar concept, mesh with self-fixating tips, as that of the predicate device, MiniArc. The Desara Mini is comprised of the same mesh and suture as the predicate device, Desara Mesh. Self-fixating tips are not contained in the predicate device, Desara Mesh.

The Desara Mini is a modification of the predicate devices, MiniArc, and Desara Mesh with the same intended use and does not change the fundamental scientific technology of the predicate device.

Confidential

{1}------------------------------------------------

K103418
PAGE 2/2.

CALDERA MEDICAL

Performance Data Summary

In accordance with the FDA's Guidance for Industry and FDA Staff "Formal for Traditional and Abbreviated 510/k)'s the results of bench, cadaver lab and validation testing has shown the Desara Mini to be substantially equivalent to the predicate devices in its intended use, technological characteristics and performance.

The mesh component of the Desara Mini is comprised of the same mesh as the predicate device. Desara mesh. FDA 510(k) #K110169, also a product of Caldera Medical. In accordance with the FDA's Guidance for the Preparation of a Premarket Notification Application for Surgical Mesh, the following mesh characteristics were assessed: mesh thickness, mesh weave characteristics, pore size, mesh density, tensile strength, mesh stiffness, flexural rigidity, tear resistance, burst strength, suture pullout and pyrogen levels. The Desara Mini mesh demonstrates substantial equivalence to the predicate device, Desara Mesh.

Desara Mini has passed all biocompatibility testing as indicated per the FDA quidance documents, FDA Device Advice, Guidance Documents (Medical Device and Radiation Emitting Product), 3. Biocompatibility and FDA Blue Book Memorandum #G95-1 Entitled "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing". The Desara Mini mesh demonstrates substantial equivalence to the predicate device, Desara Mesh.

In accordance with the FDA Guidance, Updated 510(k) Sterility Review Guidance K90-1; Final Guidance for Industry, FDA, FDA Device Advice, Guidance Documents (Medical Device and Radiation Emitting Product), 4. Labeling and FDA Consensus standard, ASTM F-1980-07, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. Desara Mini has passed all testing requirements in terms of aging. shelf life, transportation and sterilization and has demonstrated substantial equivalence to the predicate device. Desara Mesh.

Results of mechanical bench and functional cadaver testing demonstrate equivalent junction strength, device function and device performance (fixation tips, polypropylene mesh, suture loops and introducer) based upon its intended use to the predicate device, MiniArc.

The performance of the Desara Mini demonstrates that the device is as safe, and as effective, and performs at least as safely and effectively as the predicate devices, Desara Mesh and MiniArc.

Summary of Substantial Equivalence

The Desara Mini sling is safe and effective for its intended use and is substantially equivalent to the predicate devices Desara Mesh, also a product of Caldera Medical, Inc. and MiniArc, a product of American Medical Systems, Inc.

Confidential

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized eagle with its wings spread, symbolizing strength and freedom. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle, indicating the organization's name and location.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Vicki Gail Manager, QA/RA Caldera Medical, Inc. 5171 Clareton Drive AGOURA HILLS CA 91301

SEP 2 8 2012

Re: K103418 Trade/Device Name: Desara Mini Regulation Number: 21 CFR& 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: PAH Dated: November 14, 2011 Received: November 18, 2011

Dear Ms. Gail:

This letter corrects our substantially equivalent letter of November 30, 2011.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 -

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Benjamin R. Perkins

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Caldera Medical. The logo consists of a stylized image to the left of the company name. The company name is written in all capital letters.

Indications For Use

510 (k) Number (if known): K103418

Device Name: Desara Mini

Indications for Use:

Desara Mini is intended to be used in females to position a suburethral mesh sling for treatment of Genuine Stress Urinary Incontinence (SUI) and mixed incontinence resulting from urethral hypermobility. The mesh is placed transvaginally using a reusable introducer (sold separately)

Prescription Use--X--AND/OROver the Counter
Use
(Part 21 CFR 801 Subpart D)(Part 21 CFR 807
Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number: K103418

Desara® Mini Caldera Medical, Inc. Confidential

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.