K Number
K103384

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-03-16

(118 days)

Product Code
Regulation Number
888.3358
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Escalade Acetabular Cup System is intended for the replacement of the natural articular surface of the hip joint in a total hip replacement surgery. Total hip arthroplasty is intended to provide increased patient mobility and to decrease pain by replacing the damaged hip joint in patients having sufficiently sound bone to support the implants.

The Escalade Acetabular Cup System is indicated for use in total hip arthroplasty procedures where the means of cup fixation is uncemented, biological fixation. The Escalade Acetabular Cup System is compatible with the Ovation Hip Stem System (indicated for uncemented, biological fixation of the stem) and the Encompass Hip Stem System (indicated for both cemented and uncemented, biological fixation of the stem).

Total hip arthroplasty is indicated for the following conditions:

  • Notably impaired hip joints due to osteoarthritis, rheumatoid arthritis and/or post traumatic arthritis.
  • Previously failed hip surgery.
  • Fractures of the femoral neck or head.
  • Avascular necrosis of the femoral head.
  • Congenital dysplasia or other structural abnormalities where sufficient bone stock exists to properly seat the prosthesis.
Device Description

The Escalade Acetabular Cup System is a modular system intended for the replacement of the natural articular surface of the hip joint in a total hip replacement surgery. The system consists of acetabular shells, liners, bone screws, apical screw hole cover, and femoral heads.

Acetabular shells are hemispherical in shape and are designed for press-fit, cementless surgical applications. The shells are machined from titanium alloy (Ti-6Al-4V ELI) and are plasma spray coated (Ti-CP) for biological fixation. Escalade liners are manufactured from extensively cross-linked UHMWPE and are available in neutral face and 10° hooded options. Liners are configured for use with femoral head diameters ranging from 26mm to 40mm. The Ø6.5mm cancellous bone screws and the apical screw hole cover are manufactured from titanium alloy (Ti-6Al-4V ELI). The 36mm and 40mm diameter femoral heads made of cobalt chromium alloy (Co-28Cr-6Mo) are being added to the currently offered femoral head sizes in conjunction with this system.

AI/ML Overview

The provided text does not describe acceptance criteria for a diagnostic device or a study related to its performance in a clinical setting. Instead, it is a 510(k) summary for a medical device called the "Escalade Acetabular Cup System," which is a hip implant.

The "study" referenced in the document is an in-vitro hip simulator study conducted to support a marketing claim about the wear reduction of the Escalade Acetabular Liners, not a diagnostic device's performance study based on expert interpretation or comparison with ground truth.

Therefore, I cannot provide the requested information in the format of acceptance criteria for a diagnostic device's performance because the document describes an orthopedic implant, not a diagnostic tool.

However, I can extract information about the in-vitro wear test if that is helpful:

1. Table of Acceptance Criteria and Reported Device Performance (for wear performance in in-vitro study):

Acceptance Criteria (Implied)Reported Device Performance
Reduced wear compared to predicate device (TriPlus UHMWPE liners)Escalade Acetabular Liners reduce wear by 88% compared to previously cleared TriPlus UHMWPE liners when used with CoCrMo femoral heads. Steady state wear rates: - Escalade 36mm and 40mm ID liners: 1.5 ± 0.6 mg/MC (millions of cycles) - TriPlus control 32mm ID liners: 12.8 ± 1.6 mg/MC
Acceptable absolute wear rateAverage total wear for 5 million cycles: - Escalade 40mm liners: 6.6 mg - Escalade 36mm liners: 6.7 mg - TriPlus 32mm control liners: 67.9 mg

Note: The document explicitly states: "Reduced wear claims are based upon the results of in-vitro hip wear simulator tests which have not been shown to quantitatively predict clinical wear performance. Also, a reduction in total polyethylene wear volume or wear rate alone may not result in an improved clinical outcome as wear particle size and morphology are also critical factors in the evaluation of the potential for wear mediated osteolysis and associated aseptic implant loosening. Particle size and morphology were not evaluated as part of this wear claim." This highlights that the in-vitro wear performance is not directly translatable to clinical outcome or clinical "acceptance criteria."

Regarding the other requested information, it is not applicable or not provided in the document for a diagnostic device context:

  1. Sample size used for the test set and the data provenance:

    • The "test set" in this context refers to the liners tested in the in-vitro hip simulator. The document mentions "Escalade 36mm ID liners for a 52mm OD shell, Escalade 40mm ID liners for a 56mm OD shell, and TriPlus 32mm ID liners for a 52mm OD shell as a control." It does not specify the exact number of each type of liner tested.
    • Data Provenance: The study was an "in-vitro hip simulator study," indicating it was conducted in a laboratory setting, not with human patient data from a specific country.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This was an in-vitro mechanical wear test, not a diagnostic study requiring expert ground truth.

  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is an orthopedic implant, not an AI-powered diagnostic device.

  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for the wear test was the measured mass loss (wear rate) of the liners after a specified number of cycles in the hip simulator, determined by weighing.

  7. The sample size for the training set: Not applicable. There is no "training set" as this is not a machine learning/AI study.

  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a sequence of handwritten characters. The characters are 'K103384'. The characters appear to be written with a dark ink or marker on a white background. The handwriting is somewhat uneven, but the characters are generally legible.

510(k) Summary

MAR 1 6 2011

NAME OF SPONSOR:Ortho Development Corporation12187 South Business Park DriveDraper, Utah 84020
510(k) CONTACT:Tom HaueterRegulatory Affairs ManagerTelephone: (801) 553-9991Facsimile: (801) 553-9993Email: thaueter@orthodevelopment.com
DATE PREPARED:November 15, 2010
PROPRIETARY NAME:Escalade Acetabular Cup System
COMMON NAME:Acetabular Cup Prosthesis
CLASSIFICATION:21 CFR 888.3358, Hip joint metal/polymer/metal semi-constrainedporous-coated uncemented prosthesis, Class II device21 CFR 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis
DEVICE PRODUCT CODE:LPH, LZO
PREDICATE DEVICES:TriPlus Acetabular Cup System, Ortho Development Corp.(K042565)
Pinnacle Acetabular Cup System, DePuy Orthopedics, Inc.(K062148, K000306)

DEVICE DESCRIPTIONS:

The Escalade Acetabular Cup System is a modular system intended for the replacement of the natural articular surface of the hip joint in a total hip replacement surgery. The system consists of acetabular shells, liners, bone screws, apical screw hole cover, and femoral heads.

Acetabular shells are hemispherical in shape and are designed for press-fit, cementless surgical applications. The shells are machined from titanium alloy (Ti-6Al-4V ELI) and are plasma spray coated (Ti-CP) for biological fixation. Escalade liners are manufactured from extensively cross-linked UHMWPE and are available in neutral face and 10° hooded options. Liners are configured for use with femoral head diameters ranging from 26mm to 40mm. The Ø6.5mm cancellous bone screws and the apical screw hole cover are manufactured from titanium alloy (Ti-6Al-4V ELI). The 36mm and 40mm diameter femoral heads made of cobalt chromium alloy (Co-28Cr-6Mo) are being added to the currently offered femoral head sizes in conjunction with this system.

{1}------------------------------------------------

The following marketing claim will be made for Escalade Acetabular Liners:

Ortho Development Corp Escalade™ UHMWPE liners reduce wear by 88% compared to previously cleared TriPlus UHMWPE liners when used with CoCrMo femoral heads. An in-vitro hip simulator study was conducted using the following polyethylene liners: Escalade 36mm ID liners for a 52mm OD shell, Escalade 40mm ID liners for a 56mm OD shell, and TriPlus 32mm ID liners for a 52mm OD shell as a control. The bearing surface thickness was 6.7mm for the Escalade liners and 5.8mm for the TriPlus liners. Test and control liners were manufactured by Ortho Development Corp. The Escalade Liners were made from compression molded GUR 1020 material which was crosslinked, and ETO sterilized prior to the study. TriPlus liners were machined from compression molded GUR 1020 material and sterilized by gamma irradiation in an inert gas package. Test and control liners were tested for 5 million cycles on an AMTI 12 station hip simulator with a standard walking gait cycle as specified by ISO 14242-1 with a 3.0 kN peak load. The femoral head components were CoCrMo heads manufactured by Ortho Development. The lubricant was 94% bovine serum (Sigma-Aldrich) with 10.7mMol EDTA and penicillin-streptomycin solution (Fisher Scientific). The protein concentration of the lubricant was 74 g/L. The steady state wear rates determined from the study were 1.5 ± 0.6 mg/MC for both the 36mm and the 40mm Escalade liners. The steady state wear rate for the TriPlus control liners was 12.8 ±1.6 mg/MC. The average total wear for 5 million cycles was 6.6 mg for the 40mm liners, 6.7 mg for the 36mm liners, and 67.9 mg for the 32mm control liners. Reduced wear claims are based upon the results of in-vitro hip wear simulator tests which have not been shown to quantitatively predict clinical wear performance. Also, a reduction in total polyethylene wear volume or wear rate alone may not result in an improved clinical outcome as wear particle size and morphology are also critical factors in the evaluation of the potential for wear mediated osteolysis and associated aseptic implant loosening. Particle size and morphology were not evaluated as part of this wear claim.

INTENDED USE:

The Escalade Acetabular Cup System is intended for the replacement of the natural articular surface of the hip joint in a total hip replacement surgery. Total hip arthroplasty is intended to provide increased patient mobility and to decrease pain by replacing the damaged hip joint in patients having sufficiently sound bone to support the implants.

INDICATIONS FOR USE:

The Escalade Acetabular Cup System is indicated for use in total hip arthroplasty procedures where the means of cup fixation is uncemented, biological fixation. The Escalade Acetabular Cup System is compatible with the Ovation Hip Stem System (indicated for uncemented, biological fixation of the stem) and the Encompass Hip Stem System (indicated for both cemented and uncemented, biological fixation of the stem).

Total hip arthroplasty is indicated for the following conditions:

  • Notably impaired hip joints due to osteoarthritis, rheumatoid arthritis and/or post traumatic 1. arthritis.
  • Previously failed hip surgery. 2.

{2}------------------------------------------------

    1. Fractures of the femoral neck or head.
    1. Avascular necrosis of the femoral head.
    1. Congenital dysplasia or other structural abnormalities where sufficient bone stock exists to properly seat the prosthesis.

BASIS OF SUBSTANTIAL EQUIVALENCE:

Ortho Development believes that the Escalade Acetabular Cup System is substantially equivalent to the previously cleared predicate devices based on similarities in intended use, design, materials, size range, manufacturing methods, packaging, liner material and mechanical performance.

Testing that supports the substantial equivalence of the devices includes construct wear, liner/shell interconnection strength testing (assembly, push-out, torque-out), static and dynamic impingement, range of motion, shell deformation testing, and offset liner rim fatigue testing. In addition, porous plasma spray coating and polyethylene liner material characterization was performed.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings and tail feathers. The eagle is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ortho Development Corporation % Mr. Thomas D. Haueter Regulatory Affairs Manager 12187 South Business Park Drive Draper, Utah 84020

MAR 1 6 2511

Re: K103384

Trade/Device Name: Escalade Acetabular Cup System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO Dated: March 08, 2011 Received: March 11, 2011

Dear Mr. Haueter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

Page 2 - Mr. Thomas D. Haueter

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97): For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Aj. VS. n 2
fer

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K10384

Indication for Use Form Ortho Development Escalade 510(k)

510(k) Number: K103384

Device Name: Escalade Acetabular Cup System

Indications for Use:

The Escalade Acetabular Cup System is indicated for use in total hip arthroplasty procedures where the means of cup fixation is uncemented, biological fixation. The Escalade Acetabular Cup System is compatible with the Ovation Hip Stem System (indicated for uncemented, biological fixation of the stem) and the Encompass Hip Stem System (indicated for both cemented and uncemented, biological fixation of the stem).

Total hip arthroplasty is indicated for the following conditions:

  • Notably impaired hip joints due to osteoarthritis, rheumatoid arthritis and/or post traumatic 1. arthritis.
  • Previously failed hip surgery. 2.
    1. Fractures of the femoral neck or head.
  • Avascular necrosis of the femoral head. ধ ..
    বাংলাদেশে পারে বিশ্ববিদ্যালয়ের প্রতিষ্ঠান করে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পা
    1. Congenital dysplasia or other structural abnormalities where sufficient bone stock exists to properly seat the prosthesis.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

M.melkersm

(Division Sign-Off) (Division Sign pical. Orthopedic, Divisionative Devices

510(k) Number K103384

§ 888.3358 Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis.

(a)
Identification. A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.(b)
Classification. Class II.