K Number
K102430

Validate with FDA (Live)

Date Cleared
2010-09-10

(15 days)

Product Code
Regulation Number
870.2900
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Tyco Electronics Electrocardiograph (ECG) Leadwire Sets are indicated for use in the monitoring of cardiac signals for both diagnostic and monitoring purposes. Use is limited by the indications for use of the connected monitoring or diagnostic equipment.

Device Description

Tyco Electronics Electrocardiograph (ECG) Leadwire Sets are intended for use with ECG monitors manufactured by Philips Medical Systems. Both the Philips and the Tyco Electronics groups of leadwire sets are designed specifically for compatibility with Philips ECG monitors and consist of insulated copper conductors with connectors on each cable end. The term leadwire set includes ECG leadwires as well as the attached proximal and distal connectors. Leadwire sets are used to transfer signals originating from skin-mounted ECG electrodes (distal end) to a patient monitoring device (proximal end). Tyco Electronics ECG Leadwire Sets are single-use devices within the meaning of that term as used by FDA - that is, they cannot be reprocessed and re-used. As described by FDA, a single-use device, also referred to as a disposable device, is not intended to be reprocessed (cleaned, disinfected/sterilized) and used on another patient. All leadwire sets addressed within this 510(k) submission are produced in a configuration that is shielded against electromagnetic interference; they are single-use devices, not sterilizable and not reusable, intended for short-term use only (recommended use duration 7 days or less).

AI/ML Overview

The provided text describes the submission of a 510(k) premarket notification for the Tyco Electronics Electrocardiograph (ECG) Leadwire Set. This document is for a medical device that acts as a conductor of electrical energy and is explicitly stated not to be an AI/ML device. Therefore, many of the requested categories related to AI/ML device studies (e.g., sample size for test set, number of experts, adjudication method, MRMC studies, standalone performance, training set size, ground truth for training set) are not applicable to this submission.

Here's the breakdown of the available information regarding acceptance criteria and the study:

1. Table of Acceptance Criteria and Reported Device Performance:

The document states that compliance with recognized standards will be achieved through verification testing. However, it explicitly mentions that testing against these standards is not complete at the time of 510(k) filing. Therefore, there is no table showing reported device performance against specific acceptance criteria for a completed study.

The acceptance criteria are generally defined by compliance with the following standards:

Acceptance Criteria (Standards to be met)Reported Device Performance
IEC 60601-1:1998 (Medical electrical equipment), including Amendments 1 & 2, and national deviations within UL 60601-1:2003 and ANSI/AAMI ES 60601-1:2005Testing not complete at time of 510(k) filing. (Will be completed prior to market release).
AAMI / ANSI EC13:2002((R)2007 (Cardiac monitors, heart rate meters, and alarms)Testing not complete at time of 510(k) filing. (Will be completed prior to market release).
AAMI / ANSI EC53:1995/(R) 2008 (ECG cables and leadwires, including Amendment 1)Testing not complete at time of 510(k) filing. (Will be completed prior to market release).
AAMI / ANSI / ISO 10993-5:2009 (Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity)Testing not complete at time of 510(k) filing. (Will be completed prior to market release).
ISO 10993-10:2002 (Biological evaluation of medical devices Part 10 and AMENDMENT 1: Tests for irritation and delayed-type hypersensitivity)Testing not complete at time of 510(k) filing. (Will be completed prior to market release).
Product inspection for characteristics like color and intrinsic designCompliance will be achieved through product inspection.
Additional preference testing as specified by Tyco Electronics (not related to safety and effectiveness)Will be performed.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not applicable. (The device is a physical medical electrical equipment component, not an AI/ML diagnostic software).
  • Data Provenance: Not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. (No diagnostic ground truth is established for this device).

4. Adjudication method for the test set:

  • Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. (This is not an AI/ML device).

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. (This is not an AI/ML device).

7. The type of ground truth used:

  • Not applicable. (The study focuses on compliance with electrical and biocompatibility standards for a physical device, not diagnostic ground truth).

8. The sample size for the training set:

  • Not applicable. (No training set is involved for this physical device).

9. How the ground truth for the training set was established:

  • Not applicable.

Summary of the Study and Device:

The device is the Tyco Electronics Electrocardiograph (ECG) Leadwire Set, intended to transmit cardiac signals from electrodes to Philips ECG monitors. Its substantial equivalence is claimed against Philips' ECG Cables (K020531). The primary difference noted is that the Tyco device is single-use, while the predicate is reusable.

The "study" in this context refers to verification testing against established medical device standards for electrical safety (IEC 60601-1, AAMI EC13, AAMI EC53) and biocompatibility (ISO 10993-5, ISO 10993-10). At the time of the 510(k) filing, these tests were not yet complete, but the manufacturer committed to completing them and meeting the specified acceptance criteria prior to market release. This approach demonstrates a plan to prove the device meets acceptance criteria, but the results are not included in this particular document.

{0}------------------------------------------------

4. 510(k) Summary

4.1 General Information

Submitter's Name:Tyco Electronics Corporation
Address:10025 S.W. Freeman CourtWilsonville, OR 97070
Establishment Registration:3026961
Telephone Number:(503) 673-5027
Fax Number:(503) 685-9305
Contact Person:Dennis M. Gilkey
Submission Preparation Date:July 21, 2010

SEP 1 0 2010

4.2 Device

Trade/Proprietary Name: Common Name: Classification Name: Product Code: DSA CFR Reference: Class: -

Tyco Electronics Electrocardiograph (ECG) Leadwire Set ECG Leadwire Set Patient transducer and electrode cable (including connector) 21 CFR 870.2900

4.3 Predicate Device

Common Name:ECG Cables
Proprietary Name:ECG Cables
Product Code:DSA
510(k) Owner:Philips Medical Systems (Andover, MA)
510(k) Number:#K020531 (cleared March 20, 2002)

4.4 Device Description

Tyco Electronics Electrocardiograph (ECG) Leadwire Sets are intended for use with ECG monitors manufactured by Philips Medical Systems. Both the Philips and the Tyco Electronics groups of leadwire sets are designed specifically for compatibility with Philips ECG monitors and consist of insulated copper conductors with connectors on each cable end.

The term leadwire set includes ECG leadwires as well as the attached proximal and distal connectors. Leadwire sets are used to transfer signals originating from skin-mounted ECG electrodes (distal end) to a patient monitoring device (proximal end).

Tvco Electronics ECG Leadwire Sets are single-use devices within the meaning of that term as used by FDA - that is, they cannot be reprocessed and re-used. As described by FDA', a single-use device, also referred to as a disposable device, is not intended to be reprocessed (cleaned, disinfected/sterilized) and used on another patient.

All leadwire sets addressed within this 510(k) submission are produced in a configuration that is shielded against electromagnetic interference; they are single-use devices, not sterilizable and not reusable, intended for short-term use only (recommended use duration 7 days or less).

Tyco Electronics Electrocardiograph (ECG) Leadwire Set

1 Enforcement Priorities for Single-Use Devices Reprocessed by Third Parties and Hospitals; August 14, 2000 Page 7 of 18 Premarket Notification:

Tyco Electronics Corporation

{1}------------------------------------------------

4.5 Intended Use

The Tyco Electronics ECG Leadwire Set is used to transmit signals from patient-connected electrodes or transducers to Philips electrocardiograph recorders/monitors for both diagnostic and monitoring purposes. Use is limited by the indications for use of the connected monitoring or diagnostic equipment.

Tyco Electronics Electrocardiograph (ECG) Leadwire Sets are single-use devices; they are supplied non-sterile and are not sterilizable or reusable. Tyco Electrocardiograph (ECG) Leadwire Sets are intended for short-term use only (recommended use duration 7 days or less), by trained operators in a medical environment.

4.6 Product Comparison

The table below provides a comparison between Tyco Electronics ECG Leadwire Set cables and the predicate ECG cables manufactured by Philips Medical Systems. The primary difference between the two products is that Tyco Electronics ECG Leadwire Sets are single-use devices whereas Philips leadwire sets are reusable.

ParameterTyco ElectronicsPhilips Medical Systems
New DevicePredicate Device
Intended UseTransmit signals from patient-connectedsurface electrodes to Philipselectrocardiograph recorders/monitorsfor both diagnostic and monitoringpurposes.Same
SterilitySupplied non-sterile; cannot be sterilizedor otherwise reprocessedSupplied non-sterile; can be sterilized(no autoclave)
ReusabilityNot reusableReusable
Anatomical SitesAttached to electrodes placed atstandard specified locations on chestwall and extremitiesSame
Design/AppearanceCables with "snap" configuration of ECGelectrode connector (distal connector)and common "header" connection(proximal connector)Same
Type of ConstructionFlexible, shielded multi conductorelectrical cableSame
Distal ConnectorDesign1. Incorporates proprietary design tofacilitate connection with Philips ECGmonitors/recorders;2. "Snap" electrode connectors are colorcoded - e.g., red, white, green, black.brown3. Connector designations (LL, RL etc.)molded into plastic1. Same2. Same3. Connector designations (LL, RL etc.)indicated by adhesive labels
Cable Length1.0 m1.0 m; 1.6 m
Wire ColorsWhiteWires are color coded - e.g., red, white,green, black, brown
Wire MaterialShielded copper lead wire with polymeriacketSame

Premarket Notification: Tyco Electronics Electrocardiograph (ECG) Leadwire Set Tyco Electronics Corporation

Page 8 of 18

{2}------------------------------------------------

ParameterTyco ElectronicsNew DevicePhilips Medical SystemsPredicate Device
Proximal ConnectorDesignAll-in-one common connector, fits onlyPhilips ECG monitor/recorders; color-coded for use with ECG systemsSame

4.7 Performance Data

The Company has chosen to perform testing including but not necessarily limited to the external and/or recognized standards identified below:

  • · Medical electrical equipment IEC 60601-1:1998, including Amendments 1 (1991) and 2 (1995) and the national deviations described within UL 60601-1:2003 and ANSI/AAMI ES 60601-1:2005
  • · AAMI / ANSI EC13:2002((R)2007, Cardiac monitors, heart rate meters, and alarms.
  • · AAMI / ANSI EC53:1995/(R) 2008, ECG cables and leadwires, including Amendment 1,
  • · AAMI / ANSI / ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity.
  • · ISO 10993-10:2002, Biological evaluation of medical devices Part 10 and AMENDMENT 1: Tests for irritation and delayed-type hypersensitivity (including sensitization).

Compliance with the requirements of these standards will be achieved through verfilcation testing. except in cases such as color and intrinsic design, where compliance will be achieved through product inspection. Testing against these standards is not complete at time of 510(k) filing. Testing will be conducted and will meet specified acceptance criteria prior to market release of the associated medical device.

Additional preference testing of product characteristics not related to safety and effectiveness and as specified by Tyco Electronics also will be performed.

4.8 Conclusions

Tyco Electronics ECG Leadwire Sets (the new device) serve as a conductor of electrical energy. The new device is substantially equivalent to electrocardiograph cable sets manufactured by Philips Medical Systems (Andover, MA), cleared as components of the Philips M1275B Component Compact Monitor (#K020531, clearance date March 20, 2002).

Product design and testing will be in conformance with FDA-recognized standards. Conformance with recognized standards ensures product design and function will raise no new issues related to safety and effectiveness.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

Electronics Corporation c/o Ms. Dawn Tibodeau Responsible Third Party Official TUV SUD America, Inc 1775 Old Hwy 8 NW, Ste 104 Wilsonville, OR 97070

SEP 1 0 2010

Re: K102430 Trade/Device Name: Tyco Electronics Electrocardiograph (ECG) Leadwire Set Regulatory Number: 21 CFR 870.2900 Regulation Name: Transducer and Electrode Patient Cable (including connector) Regulatory Class: II (two) Product Code: DSA Dated: August 21, 2010 Received: August 26, 2010

Dear Ms. Tibodeau:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 – Ms. Dawn Tibodeau

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

3. Indications for Use Statement

510(k) Number (if known):

K102430

Device Name: Tyco Electronics Electrocardiograph (ECG) Leadwire Set

SEP 1 0 2010

Indications for Use:

Tyco Electronics Electrocardiograph (ECG) Leadwire Sets are indicated for use in the monitoring of cardiac signals for both diagnostic and monitoring purposes. Use is limited by the indications for use of the connected monitoring or diagnostic equipment.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of

(Division Sign-Off)

Division of Cardiovascular Devices

510(k) Number K162430

Premarket Notification: Tyco Electronics Electrocardiograph (ECG) Leadwire Set Tyco Electronics Corporation

Page 6 of 18

§ 870.2900 Patient transducer and electrode cable (including connector).

(a)
Identification. A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.(b)
Classification. Class II (performance standards).