IMMUKNOW-THE CYLEX IMMUNE CELLFUNCTION ASSAY
K101911 · Cylex, Inc. · NID · Oct 18, 2010 · Hematology
Device Facts
| Record ID | K101911 |
| Device Name | IMMUKNOW-THE CYLEX IMMUNE CELLFUNCTION ASSAY |
| Applicant | Cylex, Inc. |
| Product Code | NID · Hematology |
| Decision Date | Oct 18, 2010 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 864.5220 |
| Device Class | Class 2 |
Intended Use
ImmuKnow®-the Cylex®™Immune Cell Function Assay measures the concentration of ATP from circulating CD4 cells following in vitro stimulation with phytohemagglutinin (PHA) as an indicator of immune cell function. This measurement is made on heparin anti-coagulated whole blood using a luminometer and luciferin/luciferase. The assay is used for the detection of cell mediated immune response in populations undergoing immunosuppressive therapy for organ transplant.
Device Story
The Cylex Immune Cell Function Assay measures ATP concentration in CD4 cells to assess immune function. Input: heparin anti-coagulated whole blood. Process: blood incubated 15-18 hours with phytohemagglutinin (PHA) stimulant vs. non-stimulated control; CD4 cells immunoselected using anti-CD4 monoclonal antibody-coated magnetic particles; cells lysed to release intracellular ATP; luminescence reagent (luciferin/luciferase) added. Output: light emission measured by luminometer (562 nm), proportional to ATP concentration. Result: ATP concentration (ng/mL) calculated via calibration curve. Used in clinical laboratory settings to monitor immune status in transplant recipients on immunosuppressive therapy; helps clinicians evaluate cell-mediated immune response.
Clinical Evidence
Multi-center study of 122 subjects (44 healthy adults, 78 transplant recipients). Evaluated ATP results (ng/mL) against total CD4 counts. Transplant recipients showed lower mean ATP (274 ng/mL) compared to healthy adults (449 ng/mL). Data provided for immune cell function assay results (≤260 vs >260 ng/mL) and CD4 counts (≤350 vs >350) across both populations.
Technological Characteristics
In vitro diagnostic assay using anti-CD4 monoclonal antibody-coated magnetic particles for immunoselection. Detection via luminometer (562 nm emission) using luciferin/luciferase bioluminescence reaction. Requires heparin anti-coagulated whole blood. Calibration curve-based quantification of ATP (ng/mL).
Indications for Use
Indicated for the detection of cell-mediated immune response in patients undergoing immunosuppressive therapy for organ transplantation.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
- Becton Dickinson TriTest™ CD4 FITC/CD8 PE/CD3 PerCP Reagent (K971205)
- Becton Dickinson MultiTest™ CD3 FITC/CD8 PE/CD45 PerCP/CD4 APC Reagent (K974360)
Related Devices
- K013169 — IMMUNE CELL FUNCTION ASSAY · Cylex, Inc. · Apr 2, 2002
- K964618 — CYTO-STAT TRICHROME CD8-FITC/CD4-RDI/CD3-PC5 MONOCLONAL ANTIBODY REAGENT WITH CYTO-STAT TRICHROME MSIGGI-FITC/MSIGGI-RD1 · Coulter Corp. · Dec 23, 1996
- K963263 — TETRAONE SYSTEM FOR EPICS XL FLOW CYTOMETRY SYSTEMS AND CYTO-STAT TERTRACHROME CD45-FITC/CD4-RDI/CD8-ECD/CD3-PC5 · Coulter Corp. · Jan 22, 1997
- K982166 — CYTO-STAT TRICHROME CD45-FITC/CD56-RD1/CD3-PC5 MONOCLONAL ANTIBODY REAGENT WITH CYTO-STAT TRICHROME CD45-FITC/MSIGG1-RD1 · Coulter Corp. · Nov 5, 1998
- K960887 — MOUSE ANTI-HUMAN CD4, T-CELL, HELPER/INDUCER/FITC & CD8, T-CELL, SUPPRESSOR/CYTOTOXIC/RPE · Dako Corp. · May 17, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
January 14, 2002
1101911
OCT 1 8 2010
### 510(K) SUMMARY
#### Cylex Inc. Immune Cell Function Assay
Submitted by:
Cylex Inc.
8980-1 Old Annapolis Road Columbia, MD 21045
Contact:
Dr. Judy Britz
Name of Device:
| Trade Name: | Immune Cell Function Assay |
|----------------------|-------------------------------------|
| Common Name: | CD4 Cell Stimulation Assay |
| Classification Name: | Automated Differential Cell Counter |
Predicate Device:
Becton Dickinson TriTest™ CD4 FITC/CD8 PE/CD3 PerCP Reagent; Becton Dickinson MultiTest™ CD3 FITC/CD8 PE/CD45 PerCP/CD4 APC Reagent
Device Description:
Intended Use: The Cylex Immune Cell Function Assay measures the concentration of ATP from circulating CD4 cells following in vitro stimulation with phytohemagglutinin (PHA) as an indicator of immune cell function. This measurement is made on heparin anti-coagulated whole blood using a luminometer and luciferin/luciferase. The assay is used for the detection of cell mediated immune response in populations undergoing immunosuppressive therapy for organ transplant.
Test Description The Cylex Immune Cell Function Assay detects cell-mediated immunity in whole blood after a 15-18 hour incubation with stimulant. During incubation, increased ATP synthesis occurs within the cells that respond to the stimulant phytohemagglutinin (PHA). Concurrently, whole blood is incubated in the absence of stimulant for the purpose of assessing basal ATP activity. Anti-CD4 monoclonal antibody coated magnetic particles are added to immunoselect CD4 cells from both the stimulated and non-stimulated wells. After washing the selected CD4 cells on a magnet tray, Lysis Reagent is added to release intracellular ATP. Addition of Luminescence Reagent (luciferin/luciferase) to the released ATP produces light according to the following equation:
Mg24 Luciferin +ATP +O2 Luciferase Oxyluciferin +AMP +Pyrophosphate +CO2 + Light
The amount of light measured by a luminometer (emission maximum 562 nm) is proportional to the concentration of ATP. The concentration of ATP (ng/mL) is calculated from a calibration curve and compared to ATP level ranges to characterize the cellular immune function of the sample.
510(k) Summary - 1
81
{1}------------------------------------------------
## 510(K) SUMMARY
### Cvlex Inc. Immune Cell Function Assay (cont.)
# Substantial Equivalence:
The Cylex Inc. Immune Cell Function Assay has been found to be substantially equivalent to the Becton Dickinson TriTest™ CD4 FITC/CD8 PE/CD3 PerCP Reagent (K971205) and MultiTest™ CD3 FITC/CD8 PE/CD45 PerCP/CD4 ACP Reagent (K974360). All assays differentiate CD4 cells; the Cylex assay determines the responsiveness of those cells and the Becton Dickinson assays count the number of those cells.
A multi-center study was conducted on freshly drawn blood collected from 44 apparently healthy adults and 78 transplant recipients (17 at discharge from the hospital and 61 post-discharge follow-up). The samples were evaluated with the Cylex Immune Cell Function Assay. The apparently healthy adult population consisted of 11% (5) females, 86% (38) males and 3% (1) unknown, with an age range of 20 - 60 years. The ethnicity of the population was 80% (35) African American, 16% Caucasian (7), and 4% (2) other or unknown. The transplant population consisted of 33% (26) females and 67% (52) males, with an age range of 20 - 64 years. The ethnicity of the population was 15% (12) African American, 74% Caucasian (58), 10% (8) other or unknown. The organs transplanted were 55% (43) liver, 36% (28) kidney, 4% (3) pancreas, and 5% (4) multiple organs.
The means of the two populations were found to be statically different, the results are summarized in the following table.
| Population | Apparently Healthy | Transplant |
|-------------------------------------------------------|--------------------|------------|
| Cylex Inc. Immune Cell Function Assay<br>ATP Results | n = 44 | n = 78 |
| Mean / SD (ng/mL) | 449 / 152 | 274 / 166 |
| Median (ng/mL) | 430 | 251 |
| Range (ng/mL) | 205 - 967 | 21 - 759 |
| Becton Dickinson Total CD4 count by<br>Flow Cytometry | n = 44 | n = 78 |
| Mean / SD (ng/mL) | 786 / 490 | 487 / 403 |
| Median (ng/mL) | 654 | 389 |
| Range (ng/mL) | 130 - 2659 | <68 - 1904 |
| Patient Status | Transplant<br>Subjects | Normal<br>Subjects | Total |
|-----------------------------------------------|------------------------|--------------------|-------|
| Immune cell<br>functions assay<br>result ≤260 | 41 | 4 | 45 |
| Immune cell<br>function assay<br>result >260 | 37 | 40 | 77 |
| Total | 78 | 44 | 122 |
| Patient Status | Transplant<br>Subjects | Normal<br>Subjects | Total |
|-------------------------|------------------------|--------------------|-------|
| Total CD4<br>count ≤350 | 33 | 5 | 38 |
| Total CD4<br>count >350 | 45 | 39 | 84 |
| Total | 78 | 44 | 122 |
*U.S. Patent No. 5,773,232
510(k) Summary - 2
82
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. The logo is in black and white.
Food & Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993
Cylex Inc. c/o Ms. Judi Smith President, Judi Smith LLC 8980 I Old Annapolis Road Columbia, MD 21045
0CT 1-8 2010
Re: k101911
Trade/Device Name: Immuknow® Immune Cell Function Assay Regulation Number: 21 CFR§864.5220 Regulation Name: Automated Differential cell counter Regulatory Class: Class II Product Code: NID Dated: August.20, 2010 Received: September 1, 2010
Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice
{3}------------------------------------------------
Page 2 - Ms. Judi Smith
requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
maria m chan
Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indication for Use
510(k) Number (if known): K101911
Device Name: Immune Cell Function Assay
Indication For Use:
ImmuKnow®-the Cylex®™Immune Cell Function Assay measures the concentration of ATP from circulating CD4 cells following in vitro stimulation with phytohemagglutinin (PHA) as an indicator of immune cell function. This measurement is made on heparin anti-coagulated whole blood using a luminometer and luciferin/luciferase. The assay is used for the detection of cell mediated immune response in populations undergoing immunosuppressive therapy for organ transplant.
Prescription Use × (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
1/10/9/1
OCT 1 8 2010
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Reena Philip
Division Sign-Off Office of Device Evaluation
510(k) k101911
Page 1 of 1