K Number
K101596

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-07-21

(43 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05:

The following drugs had NO breakthrough detected up to 240 minutes:
Bleomycin sulfate (15.0mg/ml), Gemcitabine (38.0mg/ml), Busulfan (6.0mg/ml), Idarubicin (1.0mg/ml), Carboplatin (10.0mg/ml), Ifosfamide (50mg/ml), Cisplatin (1 mg/ml), Irinotecan (20.0mg/ml), Cyclophosphamide (20 mg/ml), Mechlorethamine HCl (1.0mg/ml), Cytarabine HCl (100.0mg/ml), Melphalan (5.0mg/ml), Dacarbazine (10 mg/ml), Methotrexate (25.0mg/ml), Daunorubicin Hcl (5.0mg/ml), Mitomycin (0.5mg/ml), Docetaxel (10.0mg/ml), Mitoxantrone (2 mg/ml), Doxorubicin HCl (Adriamycin) (2 mg/ml), Paclitaxel (Taxol) (6 mg/ml), Ellence (Epirubicin) (2.0mg/ml), Paraplatin (10.0mg/ml), Etoposide (20 mg/ml), Rituximab (10.0mg/ml), Fludarabine (25.0mg/ml), Trisenox (0.1mg/ml), Fluorouracil (adrucil) (50 mg/ml), Vincrinstine Sulfate (1 mg/ml)

The following drugs showed breakthrough detected in less than 30 minutes:
Carmustine (3.3mg/ml): 1.8 minutes, ThioTEPA (10.0mg/ml): 1.7 minutes

Warning - Not for Use with Carmustine and ThioTEPA

Device Description

Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves (Chemotherapy Glove) are 12inch long, non-sterile purple-colored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application. In addition these gloves were tested for use with the drugs listed in the Intended Use(s) section below, per ASTM D6978-05 "Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs."

Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves (Chemotherapy Glove) are 9.5-inch long, non-sterile purple-colored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application. In addition these gloves were tested for use with the drugs listed in the Intended Use(s) section below, per ASTM D6978-05 "Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs."

Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves are 9.5-inch long, non-sterile purplecolored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application.

Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves are 12-inch long, non-sterile purple-colored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves (Chemotherapy), based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The device (Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves) is tested against several FDA-recognized consensus standards. The performance results consistently indicate that the device "Meets ASTM Requirements" or has "NO breakthrough detected" for most chemotherapy drugs up to 240 minutes.

Characteristics / Acceptance Criteria (Standard)Reported Device Performance (Result)
General Glove Requirements
Dimensions (ASTM D6319-00a)Meets ASTM Requirements
Physical Properties (ASTM D6319-00a)Meets ASTM Requirements
Freedom from pinholes (ASTM D6319-00a, D5151-06)Meets ASTM Requirements
Powder Free (Powder Content) (ASTM D6319-00a, D6124-06)Meets ASTM Requirements
Biocompatibility
ISO Skin Irritation Study and Sensitization (ISO 10993, Part 10)Meets ASTM Requirements
ISO Cytotoxicity Study (ISO 10993, Part 5)Meets ASTM Requirements
Chemotherapy Permeation Resistance (ASTM D6978-05 and/or ASTM F739-07)Meets ASTM Requirements
Bleomycin sulfate (15.0mg/ml)NO breakthrough detected up to 240 minutes
Busulfan (6.0mg/ml)NO breakthrough detected up to 240 minutes
Carboplatin (10.0mg/ml)NO breakthrough detected up to 240 minutes
Cisplatin (1 mg/ml)NO breakthrough detected up to 240 minutes
Cyclophosphamide (20 mg/ml)NO breakthrough detected up to 240 minutes
Cytarabine HCl (100.0mg/ml)NO breakthrough detected up to 240 minutes
Dacarbazine (10 mg/ml)NO breakthrough detected up to 240 minutes
Daunorubicin Hcl (5.0mg/ml)NO breakthrough detected up to 240 minutes
Docetaxel (10.0mg/ml)NO breakthrough detected up to 240 minutes
Doxorubicin HCI (Adriamycin) (2 mg/ml)NO breakthrough detected up to 240 minutes
Ellence (Epirubicin) (2.0mg/ml)NO breakthrough detected up to 240 minutes
Etoposide (20 mg/ml)NO breakthrough detected up to 240 minutes
Fludarabine (25.0mg/ml)NO breakthrough detected up to 240 minutes
Fluorouracil (adrucil) (50 mg/ml)NO breakthrough detected up to 240 minutes
Gemcitabine (38.0mg/ml)NO breakthrough detected up to 240 minutes
Idarubicin (1.0mg/ml)NO breakthrough detected up to 240 minutes
Ifosfamide (50mg/ml)NO breakthrough detected up to 240 minutes
Irinotecan (20.0mg/ml)NO breakthrough detected up to 240 minutes
Mechlorethamine HCl (1.0mg/ml)NO breakthrough detected up to 240 minutes
Melphalan (5.0mg/ml)NO breakthrough detected up to 240 minutes
Methotrexate (25.0mg/ml)NO breakthrough detected up to 240 minutes
Mitomycin (0.5mg/ml)NO breakthrough detected up to 240 minutes
Mitoxantrone (2 mg/ml)NO breakthrough detected up to 240 minutes
Paclitaxel (Taxol) (6 mg/ml)NO breakthrough detected up to 240 minutes
Paraplatin (10.0mg/ml)NO breakthrough detected up to 240 minutes
Rituximab (10.0mg/ml)NO breakthrough detected up to 240 minutes
Trisenox (0.1mg/ml)NO breakthrough detected up to 240 minutes
Vincrinstine Sulfate (1 mg/ml)NO breakthrough detected up to 240 minutes
Carmustine (3.3mg/ml)Breakthrough detected in 1.8 minutes (Warning: Not for Use)
ThioTEPA (10.0mg/ml)Breakthrough detected in 1.7 minutes (Warning: Not for Use)

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not explicitly state the specific sample sizes for tests such as dimensions, physical properties, pinholes, or powder content. However, the testing for resistance to permeation by chemotherapy drugs was performed per ASTM D6978-05 (and/or ASTM F739-07), which would specify the required sample sizes and methodology for these tests. The data provenance is retrospective, as the tests were conducted by the manufacturer, Kimberly-Clark, to demonstrate substantial equivalence to previously marketed devices. The country of origin of the data is not specified beyond being generated by the applicant (Kimberly-Clark Corporation, Roswell, GA), implying it was likely conducted in the United States or at a facility capable of performing ASTM standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable in this context. The "ground truth" for glove performance (like physical properties, pinholes, or chemical permeation) is established by adhering to widely accepted, standardized laboratory testing procedures outlined in the ASTM and ISO standards (e.g., ASTM D6319-00a, ASTM D5151-06, ASTM D6124-06, ISO 10993 Parts 5 & 10, ASTM D6978-05, ASTM F739-07). These tests do not require human expert consensus for interpreting results, but rather rely on instrumental measurements and defined pass/fail criteria within the standards.

4. Adjudication Method for the Test Set

This is not applicable. As mentioned above, the performance is determined by meeting the objective criteria of the specified ASTM and ISO standards, not through expert adjudication in the traditional sense (e.g., 2+1 radiologist review).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. This submission is for a medical device (exam gloves), not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and AI assistance are not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not applicable. This device is not an algorithm or AI system.

7. The Type of Ground Truth Used

The ground truth used for product evaluation is based on objective measurements and predefined criteria specified within recognized consensus standards (ASTM and ISO standards). For example, for chemical permeation, the ground truth is whether "breakthrough" occurred within a specified timeframe, as measured by precise analytical methods according to ASTM D6978-05. For physical properties, it's defined by measurements falling within the ranges stipulated by ASTM D6319-00a.

8. The Sample Size for the Training Set

This is not applicable. This is a physical medical device, not a machine learning model. There is no "training set" in the context of this regulatory submission.

9. How the Ground Truth for the Training Set was Established

This is not applicable, as there is no training set for this device.

{0}------------------------------------------------

K101596

JUL 2-1 2010

Traditional 510(k) Notification: Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves (Chemotherapy)

Preparation Date:30 June 2010
Applicant:Kimberly Clark Corporation1400 Holcomb Bridge RoadRoswell, GA 30097
Contact Person:Lester F. PadillaTel. No.: 678-352-6766
Trade/Proprietary Name(s):Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free ExamGloves (Chemotherapy Glove)
Common Name(s):Powder-Free Nitrile Patient Examination Chemotherapy Use Gloves
Classification Name:Patient Examination Glove, Specialty (21 CFR Part 880.6250 -Product Code LZC)

Section 5 - 510(k) Summary

Legally Marketed Device(s) to Which Substantial Equivalence is Claimed:

    1. Safeskin Purple Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic Drugs Labeling Claim - K992162
  • Kimberly-Clark STERLING* Nitrile & STERLING* NITRILE-XTRA* Powder-Free Exam Gloves 2. with Chemotherapy Drug Use Claim - K081089
    1. Kimberly-Clark LAVENDER* Nitrile Powder-Free Examination Glove K081260

Device Description(s):

Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves (Chemotherapy Glove) are 12inch long, non-sterile purple-colored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application. In addition these gloves were tested for use with the drugs listed in the Intended Use(s) section below, per ASTM D6978-05 "Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs."

Intended Use(s):

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between palient and examiner.

In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05:

{1}------------------------------------------------

Traditional 510(k) Notification: Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves (Chemotherapy)

· The following drugs had NO breakthrough detected up to 240 minutes:

Bleomycin sulfate (15.0mg/ml)Gemcitabine (38.0mg/ml)
Busulfan (6.0mg/ml)Idarubicin (1.0mg/ml)
Carboplatin (10.0mg/ml)Ifosfamide (50mg/ml)
Cisplatin (1 mg/ml)Irinotecan (20.0mg/ml)
Cyclophosphamide (20 mg/ml)Mechlorethamine HCl (1.0mg/ml)
Cytarabine HCl (100.0mg/ml)Melphalan (5.0mg/ml)
Dacarbazine (10 mg/ml)Methotrexate (25.0mg/ml)
Daunorubicin Hcl (5.0mg/ml)Mitomycin (0.5mg/ml)
Docetaxel (10.0mg/ml)Mitoxantrone (2 mg/ml)
Doxorubicin HCI (Adriamycin) (2 mg/ml)Paclitaxel (Taxol) (6 mg/ml)
Ellence (Epirubicin) (2.0mg/ml)Paraplatin (10.0mg/ml)
Etoposide (20 mg/ml)Rituximab (10.0mg/ml)
Fludarabine (25.0mg/ml)Trisenox (0.1mg/ml)
Fluorouracil (adrucil) (50 mg/ml)Vincrinstine Sulfate (1 mg/ml)

The following drugs showed breakthrough detected in less than 30 minutes:

Carmustine (3.3mg/ml): 1.8 minutes ThioTEPA (10.0mg/ml): 1.7 minutesComments of the control a province a desire a desire a de control a de control a province a formation a formation and consideration and consideration of the formation of the
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Warning - Not for Use with Carmustine and ThioTEPA

Summary of Technologies:

The technological characteristics (design, specification, performance) of the Subject Devices and the Predicate Devices are substantially equivalent.

Non-Clinical Testing (Subject Device):

CharacteristicsApplicable FDA-Recognized StandardsPerformance Results
DimensionsASTM D6319-00aMeets ASTM Requirements
Physical PropertiesASTM D6319-00aMeets ASTM Requirements
Freedom from pinholesASTM D6319-00aASTM D5151-06Meets ASTM Requirements
Powder Free (Powder Content)ASTM D6319-00aASTM D6124-06Meets ASTM Requirements
ISO Skin Irritation Study andSensitizationISO 10993, Part 10Meets ASTM Requirements
ISO Cytotoxicity StudyISO 10993, Part 5Meets ASTM Requirements
Resistance to PermeationASTM D6978-05 and/orASTM F739-07Meets ASTM RequirementsSee Intended Use Section forDevice 1

{2}------------------------------------------------

Traditional 510(k) Notification: Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves (Chemotherapy)

K101596

{3}------------------------------------------------

101596

ഇടി Kimberly-Clark Corporation

Traditional 510(k) Notification: Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves (Chemotherapy)

Clinical Testing:

No Clinical testing was required to determine substantial equivalence of these devices.

Conclusion:

The results of the non-dinical testing demonstrate that the gloves meet the FDA-recognized consensus standards and are substantially equivalent to the predicate devices.

{4}------------------------------------------------

101596

ලිපු Kimberly-Clark Corporation

Traditional 510(k) Notification: Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves (Chemotherapy)

Preparation Date:30 June 2010
Applicant:Kimberly Clark Corporation1400 Holcomb Bridge RoadRoswell, GA 30097
Contact Person:Lester F. PadillaTel. No.: 678-352-6766
Trade/Proprietary Name(s):Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves(Chemotherapy Glove)
Common Name(s):Powder-Free Nitrile Patient Examination Chemotherapy Use Gloves
Classification Name:Patient Examination Glove, Specialty (21 CFR Part 880.6250 -Product Code LZC)

Section 5 - 510(k) Summary

Legally Marketed Device(s) to Which Substantial Equivalence is Claimed:

  • Safeskin Purple Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic Drugs 1 . Labeling Claim - K992162
  • Kimberly-Clark STERLING* Nitrile & STERLING* NITRILE-XTRA* Powder-Free Exam Gloves 2. with Chemotherapy Drug Use Claim - K081089
    1. Kimberly-Clark LAVENDER* Nitrile Powder-Free Examination Glove K081260

Device Description(s):

Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves (Chemotherapy Glove) are 9.5-inch long, non-sterile purple-colored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application. In addition these gloves were tested for use with the drugs listed in the Intended Use(s) section below, per ASTM D6978-05 "Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs."

Intended Use(s):

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05:

{5}------------------------------------------------

Traditional 510(k) Notification: Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves (Chemotherapy)

The following drugs had NO breakthrough detected up to 240 minutes:

Bleomycin sulfate (15.0mg/ml)Gemcitabine (38.0mg/ml)
Busulfan (6.0mg/ml)Idarubicin (1.0mg/ml)
Carboplatin (10.0mg/ml)Ifosfamide (50mg/ml)
Cisplatin (1 mg/ml)Irinotecan (20.0mg/ml)
Cyclophosphamide (20 mg/ml)Mechlorethamine HCl (1.0mg/ml)
Cytarabine HCl (100.0mg/ml)Melphalan (5.0mg/ml)
Dacarbazine (10 mg/ml)Methotrexate (25.0mg/ml)
Daunorubicin Hcl (5.0mg/ml)Mitomycin (0.5mg/ml)
Docetaxel (10.0mg/ml)Mitoxantrone (2 mg/ml)
Doxorubicin HCI (Adriamycin) (2 mg/ml)Paclitaxel (Taxol) (6 mg/ml)
Ellence (Epirubicin) (2.0mg/ml)Paraplatin (10.0mg/ml)
Etoposide (20 mg/ml)Rituximab (10.0mg/ml)
Fludarabine (25.0mg/ml)Trisenox (0.1mg/ml)
Fluorouracil (adrucil) (50 mg/ml)Vincrinstine Sulfate (1 mg/ml)

The following drugs showed breakthrough detected in less than 30 minutes:

Carmustine (3.3mg/ml): 1.8 minutesThioTEPA (10.0mg/ml): 1.7 minutes
-----------------------------------------------------------------------

Warning - Not for Use with Carmustine and ThioTEPA

Summary of Technologies:

The technological characteristics (design, specification, performance) of the Subject Devices and the Predicate Devices are substantially equivalent.

Non-Clinical Testing (Subject Device):

CharacteristicsApplicable FDA-Recognized StandardsPerformance Results
DimensionsASTM D6319-00aMeets ASTM Requirements
Physical PropertiesASTM D6319-00aMeets ASTM Requirements
Freedom from pinholesASTM D6319-00aASTM D5151-06Meets ASTM Requirements
Powder Free (Powder Content)ASTM D6319-00aASTM D6124-06Meets ASTM Requirements
ISO Skin Irritation Study andSensitizationISO 10993, Part 10Meets ASTM Requirements
ISO Cytotoxicity StudyISO 10993, Part 5
Resistance to PermeationASTM D6978-05 and/orASTM F 739-07Meets ASTM RequirementsSee Intended Use Section

{6}------------------------------------------------

K101596

്‍ട്ടി Kimberly-Clark Corporation

Traditional 510(k) Notification: Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves (Chemotherapy)

Clinical Testing:

No Clinical testing was required to determine substantial equivalence of these devices.

Conclusion:

The results of the non-dinical testing demonstrate that the FDA-recognized consensus standards and are substantially equivalent to the predicate devices.

{7}------------------------------------------------

Traditional 510(k) Notification: Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves

Preparation Date:30 June 2010
Applicant:Kimberly Clark Corporation1400 Holcomb Bridge RoadRoswell, GA 30097
Contact Person:Lester F. PadillaTel. No.: 678-352-6766
Trade/Proprietary Name(s):Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves
Common Name(s):Powder-Free Nitrile Patient Examination Gloves
Classification Name:Polymer Patient Examination Glove (21 CFR Part 880.6250 -Product Code LZA)

Section 5 - 510(k) Summary

Legally Marketed Device(s) to Which Substantial Equivalence is Claimed:

    1. Safeskin Purple Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic Drugs Labeling Claim - K992162
    1. Kimberly-Clark STERLING* Nitrile & STERLING* NITRILE-XTRA* Powder-Free Exam Gloves with Chemotherapy Drug Use Claim - K081089
    1. Kimberly-Clark LAVENDER* Nitrile Powder-Free Examination Glove - K081260

Device Description(s):

Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves are 9.5-inch long, non-sterile purplecolored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application.

Intended Use(s):

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Summary of Technologies:

The technological characteristics (design, specification, performance) of the Subject Devices and the Predicate Devices are substantially equivalent.

{8}------------------------------------------------

Traditional 510(k) Notification: Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves

Non-Clinical Testing (Subject Device):

CharacteristicsApplicable FDA-Recognized StandardsPerformance Results
DimensionsASTM D 6319-00aMeets ASTM Requirements
Physical PropertiesASTM D 6319-00aMeets ASTM Requirements
Freedom from pinholesASTM D 6319-00aASTM D 5151-06Meets ASTM Requirements
Powder Free (Powder Content)ASTM D 6319-00aASTM D 6124-06Meets ASTM Requirements
ISO Skin Irritation Study andSensitizationISO 10993, Part 10Meets ASTM Requirements
ISO Cytotoxicity StudyISO 10993, Part 5Meets ASTM Requirements

Clinical Testing:

No Clinical testing was required to determine substantial equivalence of these devices.

Conclusion:

The results of the non-clinical testing demonstrate that the FDA-recognized consensus standards and are substantially equivalent to the predicate devices.

{9}------------------------------------------------

Traditional 510(k) Notification: Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves

Preparation Date:30 June 2010
Applicant:Kimberly Clark Corporation1400 Holcomb Bridge RoadRoswell, GA 30097
Contact Person:Lester F. PadillaTel. No.: 678-352-6766
Trade/Proprietary Name(s):Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free ExamGloves
Common Name(s):Powder-Free Nitrile Patient Examination Gloves
Classification Name:Polymer Patient Examination Glove (21 CFR Part 880.6250 -Product Code LZA)

Section 5 - 510(k) Summary

Legally Marketed Device(s) to Which Substantial Equivalence is Claimed:

  • Safeskin Purple Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic Drugs 1. Labeling Claim - K992162
  • Kimberly-Clark STERLING* Nitrile & STERLING* NITRILE-XTRA* Powder-Free Exam Gloves 2. with Chemotherapy Drug Use Claim - K081089
    1. Kimberly-Clark LAVENDER* Nitrile Powder-Free Examination Glove K081260

Device Description(s):

Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves are 12-inch long, non-sterile purple-colored nitrile, powder-free, ambidextrous patient examination glove that meets all of the requirements of ASTM D6319-00a, Standard Specification for Nitrile Examination Gloves for Medical Application.

Intended Use(s):

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Summary of Technologies:

The technological characteristics (design, specification, performance) of the Subject Devices and the Predicate Devices are substantially equivalent.

{10}------------------------------------------------

Traditional 510(k) Notification: Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves

Non-Clinical Testing (Subject Device):

CharacteristicsApplicable FDA-Recognized StandardsPerformance Results
DimensionsASTM D6319-00aMeets ASTM Requirements
Physical PropertiesASTM D6319-00aMeets ASTM Requirements
Freedom from pinholesASTM D6319-00aASTM D5151-06Meets ASTM Requirements
Powder Free (Powder Content)ASTM D6319-00aASTM D6124-06Meets ASTM Requirements
ISO Skin Irritation Study andSensitizationISO Cytotoxicity StudyISO 10993, Part 10ISO 10993, Part 5Meets ASTM Requirements

Clinical Testing:

No Clinical testing was required to determine substantial equivalence of these devices.

Conclusion:

The results of the non-clinical testing demonstrate that the gloves meet the FDA-recognized consensus standards and are substantially equivalent to the predicate devices.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows a partial view of the seal for the Department of Health & Human Services. The seal includes the department's name around the perimeter and a stylized caduceus symbol in the center. The caduceus is depicted with a staff and intertwined snakes, representing medicine and healing. The image is in black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Kimberly-Clark C/O Mr. Ned Devine Responsible Third Party Official Underwriters Laboratories Incorporated 333 Pfingsten Road Northbrook, Illinois 60062

JUL 21 2010

Re: K101596

Trade/Device Name: Kimberly -Clark PURPLE NITRILE Powder-Free Exam Glove (Chemotherapy Gloves)

Kimberly -Clark PURPLE NITRILE -XTRA Powder-Free Exam Glove (Chemotherapy Gloves)

Kimberly -Clark PURPLE NITRILE Powder-Free Exam Glove Kimberly -Clark PURPLE NITRILE -XTRA * Powder-Free Exam Glove

Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZC, LZA Dated: July 2, 2010 Received: July 6, 2010

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your

{12}------------------------------------------------

Page 2-Mr. Devine

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{13}------------------------------------------------

යි හි Kimberly-Clark Corporation

Indications for Use

510(k) Number (if known):

Device Name(s):

Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves (Chemotherapy Gloves)

Indications for Use:

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05:

The following drugs had NO breakthrough detected up to 240 minutes:

Bleomycin sulfate (15.0mg/ml)Gemcitabine (38.0mg/ml)
Busulfan (6.0mg/ml)Idarubicin (1.0mg/ml)
Carboplatin (10.0mg/ml)Ifosfamide (50mg/ml)
Cisplatin (1 mg/ml)Irinotecan (20.0mg/ml)
Cyclophosphamide (20 mg/ml)Mechlorethamine HCl (1.0mg/ml)
Cytarabine HCl (100.0mg/ml)Melphalan (5.0mg/ml)
Dacarbazine (10 mg/ml)Methotrexate (25.0mg/ml)
Daunorubicin Hcl (5.0mg/ml)Mitomycin (0.5mg/ml)
Docetaxel (10.0mg/ml)Mitoxantrone (2 mg/ml)
Doxorubicin HCl (Adriamycin) (2 mg/ml)Paclitaxel (Taxol) (6 mg/ml)
Ellence (Epirubicin) (2.0mg/ml)Paraplatin (10.0mg/ml)
Etoposide (20 mg/ml)Rituximab (10.0mg/ml)
Fludarabine (25.0mg/ml)Trisenox (0.1mg/ml)
Fluorouracil (adrucil) (50 mg/ml)Vincrinstine Sulfate (1 mg/ml)

The following drugs showed breakthrough detected in less than 30 minutes:

Carmustine (3.3mg/ml): 1.8 minutes : ThioTEPA (10.0mg/ml): 1.7 minutes

Warning - Not for Use with Carmustine and ThioTEPA

Page 1 of 2

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: __

{14}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name(s):

Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves (Chemotherapy Gloves)

Indications for Use:

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

In addition, these chemotherapy gloves were tested for use with the following drug concentrations per ASTM D6978-05:

The following drugs had NO breakthrough detected up to 240 minutes.
Bleomycin sulfate (15.0mg/ml)Gemcitabine (38.0mg/ml)
Busulfan (6.0mg/ml)Idarubicin (1.0mg/ml)
Carboplatin (10.0mg/ml)Ifosfamide (50mg/ml)
Cisplatin (1 mg/ml)Irinotecan (20.0mg/ml)
Cyclophosphamide (20 mg/ml)Mechlorethamine HCl (1.0mg/ml)
Cytarabine HCl (100.0mg/ml)Melphalan (5.0mg/ml)
Dacarbazine (10 mg/ml)Methotrexate (25.0mg/ml)
Daunorubicin Hcl (5.0mg/ml)Mitomycin (0.5mg/ml)
Docetaxel (10.0mg/ml)Mitoxantrone (2 mg/ml)
Doxorubicin HCl (Adriamycin) (2 mg/ml)Paclitaxel (Taxol) (6 mg/ml)
Ellence (Epirubicin) (2.0mg/ml)Paraplatin (10.0mg/ml)
Etoposide (20 mg/ml)Rituximab (10.0mg/ml)
Fludarabine (25.0mg/ml)Trisenox (0.1mg/ml)
Fluorouracil (adrucil) (50 mg/ml)Vincrinstine Sulfate (1 mg/ml)

The following drugs had NO breakthrough detected up to 240 minutes:

The following drugs showed breakthrough detected in less than 30 minutes:

Carmustine (3.3mg/ml): 1.8 minutes ThioTEPA (10.0mg/ml): 1.7 minutes

Warning - Not for Use with Carmustine and ThioTEPA

Page 1 of 2

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

{15}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name(s):

Kimberly-Clark PURPLE NITRILE* Powder-Free Exam Gloves

Indications for Use:

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

signature

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

{16}------------------------------------------------

Indications for Use

510(k) Number (if known): K 101596

Device Name(s):

Kimberly-Clark PURPLE NITRILE-XTRA* Powder-Free Exam Gloves

Indications for Use:

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

(Division Sign-Uff) Division of Anesthesiology. General Hospital Infection Control, Dental Os neas

510(k) Number: |<101596

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.