Intended Use

A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Classified by FDA's General Hospital and Personal Use Device panel as Class I, 21 CFR 880.6250, Power Free Vinyl Patient Examination Glove, 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1.

AI/ML Overview

The provided text describes the submission of a 510(k) premarket notification for Powder Free Vinyl Patient Examination Gloves. This information focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of a device's performance against specific acceptance criteria for diagnostic or algorithmic tasks. Therefore, many of the requested categories (like number of experts, adjudication methods, multi-reader multi-case studies, standalone performance, and training set information) are not applicable to this type of device submission.

Here's the available information presented in the requested format, with "Not Applicable" for categories that don't fit this context:

Acceptance Criteria and Device Performance Study

The submission demonstrates substantial equivalence to a predicate device by meeting established ASTM standards and FDA requirements for patient examination gloves.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
Physical and Dimensions TestingMeets ASTM D-5250-06e1 requirementsAll testing meets requirements for physical and dimensions
Water Fill Test (Pinhole)FDA 1000 ml. Water Fill Test, AQL 2.5, Inspection Level IMeets these requirements
Primary Skin IrritationNo primary skin irritant reactionsNo primary skin irritant reactions
Skin SensitizationNo sensitization (allergic contact dermatitis) reactionsNo sensitization reactions
Residual Powder TestMeets ASTM D-6124-06: no more than 2 mg powder per gloveMeets "powder-free" claims (contains ≤ 2 mg powder)

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: For the Water Fill Test, inspection levels are S-2 (for Physical and Dimensions) and I (for Water Fill Test), and AQL is 4.0 and 2.5 respectively. The exact number of samples tested is not explicitly stated but is implied by these AQL and inspection level standards.
  • Data Provenance: The data is based on non-clinical tests performed by the manufacturer, Hong Di Plastic Products Co., Ltd. (China). It is retrospective in the sense that the tests were conducted prior to the 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not Applicable. The ground truth for this device (e.g., whether a glove has a pinhole, its dimensions, or causes irritation) is determined by objective physical and chemical tests against established standards, not by expert consensus on subjective observations.

4. Adjudication method for the test set

Not Applicable. Adjudication methods like 2+1 or 3+1 are used for expert review of human-interpreted data (e.g., medical images). For physical and chemical tests, the results are typically direct measurements or observations against a standard, not requiring expert adjudication in this manner.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not Applicable. This is a medical device (patient examination glove), not an AI-powered diagnostic or assistive technology.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not Applicable. This is a physical medical device.

7. The type of ground truth used

The ground truth is based on objective measurements and laboratory test results against established national and international standards (ASTM D-5250-06e1, ASTM D-6124-06) and FDA requirements (e.g., 1000 mL Water Fill Test).

8. The sample size for the training set

Not Applicable. This device does not involve a "training set" in the context of machine learning or AI models.

9. How the ground truth for the training set was established

Not Applicable. As there is no training set for this device.

{0}------------------------------------------------

JUL = 1 2010

EXHIBIT #1 Page 1 of 2

510(K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(K) number is: Κιο|132

l. Owner's Identification:

. ... .

Mr. Zhiqiang Qiao Hong Di Plastic Products Co., Ltd. Donggao Industrial Zone Zanhuang, Hebei, China 050000 Date Summary Prepared: February 8, 2010

  1. Name of the Device: Hong Di Plastic Products Co., Ltd.

Powder Free Vinyl Patient Examination Gloves, (Blue, White, Pink)

3. Predicate Device Information:

Shijiazhuang Great Eagle Plastic Products Co., Ltd Powder-Free (yellow) Vinyl Patient Examination Gloves (K992861).

4. Device Description:

Classified by FDA's General Hospital and Personal Use Device panel as Class I, 21 CFR 880.6250, Power Free Vinyl Patient Examination Glove, 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1.

5. Intended Use:

A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

6. Comparison to Predicate Devices:

Hong Di Plastic Products Co., Ltd.'s Powder Free Vinyl Patient Examination Gloves, (Blue, White, Pink) is substantially equivalent in safety and effectiveness to the Shijiazhuang Great Eagle Plastic Products Co., Ltd.'s Powder-Free (yellow) Vinyl Patient Examination Gloves.

{1}------------------------------------------------

K101132

EXHIBIT #1 Page 2 of 2

7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows:

The standards used for Hong Di Plastic Products Co., Ltd.'s glove production are based on ASTM D-5250-06e1. All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AQL 4.0.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level I, meeting these requirements.

Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

A Residual Powder Test that based on ASTM D-6124-06 for powder at finished inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).

8. Labeling:

There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.

9. Discussion of Clinical Tests Performed:

Not Applicable - There is no hypoallergenic Claim.

10. Conclusions:

Hong Di Plastic Products Co., Ltd.'s Powder Free Vinyl Patient Examination Gloves, (Blue, White, Pink) conform fully to ASTM D-5250-06e1 standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited. The device herein mentioned is as safe, as effective, and performs as well as or better than the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The text is in all caps and is evenly spaced around the circle.

JUL -1 2010

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WQ66-G609 Silver Spring, MD 20993-0002

Hong Di Plastic Products Company, Limited C/O Ms. Jie Liu Project Manager Surprotect Incorporated 3973 Schaefer Avenue Chino, California 91710

Re: K101132

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, (Blue, White, Pink)

Regulation Number: 21 CFR 880.6250

Regulation Name: Patient Examination Glove Regulatory Class: I

Product Code: LYZ

Dated: February 8, 2010

Received: April 22, 2010

Dear Ms. Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Liu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Susan Suave

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Attachment A

INDICATION FOR USE

KIOII32 510 (k) NUMBER (IF KNOW):

APPLICANT: DEVICE NAME:

Hong Di Plastic Products Co., Ltd. Powder Free Vinyl Patient Examination Gloves, (Blue, White, Pink)

INDICATIONS FOR USE:

A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

S

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K101132 .

Prescription Use (Part 21 CFR 801 Subpart D)

AND/ OR

.

Over-The-Counter-Use __ J (21CFR 801Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrent of CDRH, Office of Device Evaluation (ODE)

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.