K Number
K082597

Validate with FDA (Live)

Date Cleared
2008-10-24

(46 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the "Powder Free Vinyl Patient Examination Gloves, Clear(non-colored)" device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance CriteriaDevice Performance
DimensionASTM standard D 5250-06Meets
Physical PropertiesASTM standard D 5250-06Meets
Freedom from pinholes21 CFR 800.20 (Water leak test on pinhole AQL)Meets
Powder ResidualASTM standard D 5250-06 and D6124-01 (< 2mg/glove implied from predicate)<2mg/glove
Biocompatibility (Primary Skin Irritation)Primary Skin Irritation in rabbits (ISO10993-10)Passes (Not a Primary Skin Irritation)
Biocompatibility (Dermal Sensitization)Dermal sensitization in the guinea pig (ISO10993-10)Passes (Not a Dermal sensitization)

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample sizes used for testing each characteristic. The data provenance is not explicitly stated in terms of country of origin for the individual tests, but the device manufacturer is based in China. The studies appear to be prospective as they are conducted to demonstrate the device meets specific standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

This information is not provided in the document. For device testing (like these gloves), the "ground truth" is typically defined by the objective measurement adhering to the specified ASTM standards and CFR regulations, rather than expert consensus on subjective interpretations. For biocompatibility, the interpretation of results would be by qualified laboratory personnel following established protocols.

4. Adjudication Method for the Test Set

This information is not applicable and therefore not provided. Adjudication methods like "2+1" or "3+1" are typically used in scenarios where multiple human readers are assessing subjective data (e.g., medical images) and their disagreements need to be resolved. For objective tests on physical properties of gloves, the results are typically directly measured against a standard, not adjudicated by a panel of experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance, which is not applicable to the assessment of physical properties or biocompatibility of examination gloves.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, a standalone algorithm performance study was not done. This device is a physical product (gloves), not an algorithm or AI system.

7. The Type of Ground Truth Used

The ground truth used for these tests is based on objective measurements against established industry standards and regulatory requirements.

  • ASTM standards (D 5250-06, D6124-01): These standards define the acceptable range for dimensions, physical properties, and powder residue. The ground truth is whether the measured values fall within these defined ranges.
  • 21 CFR 800.20: This regulation defines the acceptable level of freedom from pinholes based on a water leak test. The ground truth is the pass/fail outcome of this test.
  • ISO10993-10: This standard governs biocompatibility testing. The ground truth is the biological response observed (e.g., presence or absence of irritation/sensitization) as interpreted by standard toxicological protocols.

8. The Sample Size for the Training Set

This information is not applicable and therefore not provided. The development and testing of these gloves do not involve a "training set" in the context of machine learning or AI models. The gloves are manufactured and then tested against specified standards.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no "training set" for this type of device.

{0}------------------------------------------------

Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: ." (applicant leave blank)

Premarket Notification [510(k)} Summary

.

[{a)(1}]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

Submitter's name :TANGSHAN LINYUAN PLASTIC PRODUCTSCO.,LTD.
Submitter's address :SIGEZHUANG, LUANNAN COUNTY,TANGSHAN CITY, HEBEI PROVINCE,063503, CHINA
Phone number :(86) 315-4169201
Fax number :(86) 315-4169311
Name of contact person:Mr. Zhang Liang
Date the summary was prepared:Sep. 1st,2008

[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Powder Free Vinyl Patient Examination Gloves, Clear(non-colored)
Proprietary/Trade name:Powder Free Vinyl Patient Examination GlovesOther clients private labeling
Common Name:Patient examination glove
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LYZ

[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* powder free vinyl patient examination gloves , Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.

Predicate device : FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908 .

[(a)(4)] A description of the device

{1}------------------------------------------------

Device Description ; powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.

[(a)(5)] The summary describes the intended use of the device

Device Intended Use: powder free vinyl patient examination glove, Clear(non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 5250-06Meets
Physical PropertiesASTM standard D 5250-06Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06and D6124-01<2mg/glove
BiocompatabilityPrimary Skin Irritation inrabbitsPassesNot a Primary Skin Irritation
Dermal sensitization in theguinea pigPassesNot a Dermal sensitization

[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the ((0)(1)) A Brief Choussission for a determination of substantial equivalence .

Powder free vinyl patient examination gloves , Clear(non-colored) meet requirements per ASTM D5250-06, per ASTM D6124-01, per 21 CFR 800.20 and ISO10993-10.

[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

[(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that ((0)(-)) The concludions effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims .

{2}------------------------------------------------

Image /page/2/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird-like figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Tangshan Linyuan Plastic Products Company, Limited C/O Mr. Chu Xiaoan Chu Xiaoan Room 1606, Building 1, Jianxiang Yuan No. 209 Bei Si Huan Zhong Road Haidian District, Beijing CHINA 100083

OCT 2 4 2008

Re: K082597

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (Non-Colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: October 14, 2008 Received: October 14, 2008

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

signature

Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

Applicant: TANGSHAN LINYUAN PLASTIC PRODUCTS CO.,LTD.

1082597 510(k) Number (if known): *

Device Name: Powder Free Vinyl Patient Examination Gloves, Clear(Non-colored)

Indications For Use:

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Shirlen A. Murphy, MD

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K082547

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.