K Number
K113191

Validate with FDA (Live)

Date Cleared
2011-11-17

(17 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Vinyl Patient Examination Gloves, Yellow Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powder Free Vinyl Patient Examination Gloves, Yellow Color are disposable devices which made of PVC material ,intended for medical purpose that worn on examiner's hand or finger to prevent contamination between patient and examiner and they meets all of the requirements of ASTM standard D 5250-06 e1.

AI/ML Overview

The provided text describes the acceptance criteria and performance of "Powder Free Vinyl Patient Examination Gloves, Yellow Color" but does not describe a study involving a device that uses AI or requires a human-in-the-loop for its operation.

Therefore, most of the requested information (like sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth types for AI/ML models) is not applicable to this medical glove.

Here's the relevant information that can be extracted:

Acceptance Criteria and Device Performance

Acceptance CriteriaStandardReported Device Performance
DimensionASTM standard D 5250-06 e1.Meets
Physical PropertiesASTM standard D 5250-06 e1.Meets
Freedom from Pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06 e1 and D6124-06Meets (<2mg/glove)
Biocompatability:
- Primary Skin IrritationPrimary Skin Irritation in rabbits (ISO 10993-10)Passes (Not a Primary Skin Irritation)
- Dermal SensitizationDermal sensitization in the guinea pig (ISO 10993-10)Passes (Not a Dermal Sensitization)

Additional Information based on the provided text:

  1. Sample sizes used for the test set and the data provenance: Not applicable. Performance is based on compliance with ASTM and FDA standards for medical gloves.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Performance is assessed against established technical standards, not expert consensus on medical imaging or diagnoses.
  3. Adjudication method for the test set: Not applicable.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a medical glove, not an AI-powered diagnostic device.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" here is compliance with defined ASTM standards (D 5250-06 e1, D6124-06) and FDA regulations (21 CFR 800.20, ISO 10993-10) for medical gloves.
  7. The sample size for the training set: Not applicable. This is not an AI/ML device requiring a training set.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Section C 510(k) Summary (21 CFR 807.92)

NOV 1 7 2011

:

KI 3 9

510(K) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is:

Premarket Notification [510(k)] Summary

Submitter's name :Suqian Ruijiang Medical Supplies Co., Ltd.
Submitter's address :No.119 Hongzhehu Road, Suqian, Jiangsu,223800, China
Phone number :86-527-84391929
Fax number :86-527-84391929
Name of contact person:Mr.Donmei Yu
Date the summary was prepared:2011-10-23
Device Name:Powder Free Vinyl Patient Examination Gloves,Yellow Color
Proprietary/Trade name:Powder Free Vinyl Patient Examination Gloves,Yellow ColorOther clients private labeling
Common Name:Exam gloves
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LYZ

Class I* Powder Free Vinyl Patient Examination Gloves, Yellow Color that meets all of the requirements of ASTM D 5250-06 e1.

网站_365bet官网_365bet官网官网 狗万体育 Predicate device: Powder-Free Vinyl Patient Examination Glove, Yellow Color, PPP; i Medical and Safety Products Limited. K110972.

Device Description: Powder Free Vinyl Patient Examination Gloves, Yellow Color are disposable devices which made of PVC material ,intended for medical purpose that worn on examiner's hand or finger to prevent contamination between patient and examiner and they meets all of the requirements of ASTM standard D 5250-06 e1.

{1}------------------------------------------------

Device Intended Use(Indication for use): Powder Free Viny) Patient Examination Gloves, Yellow Color is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

A summary of the technological characteristics of new device compared to the predicate device.

The Powder Free Vinyl Patient Examination Gloves, Yellow Color, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 5250-06 el.Meets
Physical PropertiesASTM standard D 5250-06 el.Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06 eland D6124-06Meets<2mg/glove
BiocompatabilityPrimary Skin Irritation in rabbitsISO 10993-10Dermal sensitization in the guinea pigISO 10993-10PassesNot a Primary Skin IrritationPassesNot a Dermal sensitization

A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Powder Free Vinyl Patient Examination Gloves, Yellow Color, meet requirements per ASTM D5250-06 e1, per ASTM D6124-06, per 21 CFR 800.20 and ISO 10993-10: 2002/Amd. 1:2006.

The performance test data of the non clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above.

A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

The conclusions

... ..

It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Yellow Color meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims.

It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Yellow Color is as safe, as effective, and performs as well as the predicate device, Powder-Free Vinyl Patient Examination Glove, Yellow Color PPP Medical and Safety Products Limited. K110972.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the perimeter. Inside the seal, there is a stylized representation of a human figure with three overlapping arms or lines, which is a common symbol associated with the department.

DEPARTMENT OF HEALTH & HUMAN SERVICES


Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Sugian Ruijang Medical Supplies Company, Limited C/O Mr. Chu Xianoan Official Correspondent Beijing Easy-Link Company, Limited 1 Juan Xiang Yuan No. 209 Bei Si Huam Zhong Road Haidian District, Beijing 100083 CHINA

NOV 1 7 2011

Re: K113191

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Yellow Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: October 23, 2011 Received: October 31, 2011

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Anthony V. a. alar

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K113191

INDICATIONS FOR USE

Applicant: Suqian Ruijiang Medical Supplies Co., Ltd.

510(k) Number (if known): *

Device Name: Powder Free Vinyl Patient Examination Gloves, Yellow Color

Indications For Use:

Powder Free Vinyl Patient Examination Gloves, Yellow Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth P. Clarence Wells

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K113191

...

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.