K Number
K101037

Validate with FDA (Live)

Date Cleared
2011-05-23

(404 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EG V1 Pro Self Monitoring Blood Glucose Test System is intended for the quantitative measurement of glucose in venous whole blood or fresh capillary whole blood from fingertip. Testing is done outside the body (In Vitro diagnostic use). It is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control. The system is only used with single-use lancing devices. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus.

The system consists of the EG V1 Pro meter and the EG Pro Test Strips. The EG V1 Pro meter only is used with the EG Pro Test Strips to quantitatively measure glucose in venous whole blood or fresh capillary whole blood from the fingertip.

EG Glucose Control Solution: For use with the EG V1 Pro Blood Glucose Self Monitoring System as a quality control check to verify the accuracy of blood glucose test results.

The EG V1 (BL) Self Monitoring Blood Glucose Test System is intended for the quantiatative measurement of glucose in fresh capillary whole blood from the fingertip, palm or forearm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter[OTC]) by a single patient with diabetes and should not be shared, as an aid to monitor the effectiveness of diabetes control. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus. Alternative site testing can only be used during steady-state blood glucose conditions.

The system consists of the EG V1 (BL) meter and the EG V1 Test Strips. The EG V1 (BL) meter only is used with the EG V1 Test Strips to quantitatively measure glucose in venous whole blood or fresh capillary whole blood from the fingertip, palm, or forearm.

EG Glucose Control Solution: For use with the EG V1 Pro Blood Glucose Self Monitoring System as a quality control check to verify the accuracy of blood glucose test results.

Device Description

The EG V1 Pro Self-Monitoring Blood Glucose System consists of the EG V1 Pro Blood Glucose Meter, EG Pro Glucose Test Strips, single-use Lancing Device and EG Level 2 Control Solution.

The EG V1 (BL) Self-Monitoring Blood Glucose System consists of the EG V1 (BL) Blood Glucose Meter, EG Glucose Test Strips, Auto-Lancet Device and EG Level 2 Control Solution.

AI/ML Overview

Here's an analysis of the acceptance criteria and study details for the EG V1 Pro Self-Monitoring Glucose Test System and the EG V1 (BL) Self Monitoring Glucose Test System, based on the provided 510(k) Summary.

Key Observation: The document contains two separate 510(k) summaries for two devices:

  1. EG V1 Pro Self Monitoring Glucose Test System: Intended for professional use, measuring glucose in venous or capillary whole blood from the fingertip.
  2. EG V1 (BL) Self Monitoring Glucose Test System: Intended for Over-The-Counter (OTC) / home use, measuring glucose in capillary whole blood from fingertip, palm, or forearm.

Both summaries reference the same performance characteristics data, indicating that the underlying technology and analytical performance studies likely apply to both, with clinical accuracy studies differing based on professional vs. patient use and anatomical sites. I will present the acceptance criteria and study details as they apply to the respective devices and their specific accuracy studies.


EG V1 Pro Self Monitoring Glucose Test System (Professional Use)

1. Table of Acceptance Criteria and Reported Device Performance (Professional Use)

The acceptance criteria for system accuracy are based on ISO 15197:2003.

Performance MetricAcceptance Criteria (ISO 15197:2003)Reported Device Performance (Professional Use)Meets Criteria?
System Accuracy (Professional Use)
For Glucose < 75 mg/dL95% of individual glucose results to be within ±15 mg/dL of the laboratory reference method.100% (19/19) of individual glucose results were within ±15 mg/dL of the YSI 2300D analyzer for glucose concentrations < 75 mg/dL from fingertip samples. (Also reported: 84.21% within ±5mg/dL, 89.47% within ±10mg/dL).Yes
For Glucose ≥ 75 mg/dL95% of individual glucose results to be within ±20% of the laboratory reference method.98.51% (132/134) of individual glucose results were within ±20% of the YSI 2300D analyzer for glucose concentrations ≥ 75 mg/dL from fingertip samples. (Also reported: 76.12% within ±5%, 93.28% within ±10%, 96.27% within ±15%).Yes
Hematocrit InterferenceBias ≤ 15 mg/dL when glucose concentration <75 mg/dL and bias ≤ 15% when glucose concentration ≥75 mg/dL (for hematocrit ranges 20-60%)Max % Bias reported for different hematocrit levels (20%, 30%, 41%, 49%, 60%) across two strip lots: 11.5% and 13.1% (all within 15%). Specific bias for <75 mg/dL not explicitly stated in table but implied to meet criteria as max % bias is given for the overall range.Yes
Reportable RangeNot explicitly stated as a numerical acceptance criterion, but implied to be validated.20-600 mg/dLValidated
Precision (Within-run)Not explicitly stated as a numerical acceptance criterion in the provided text.CV (%) ranged from 2.7% to 6.5% across 6 glucose concentration ranges. SD (mg/dL) ranged from 2.6 to 13.9.N/A
Precision (Day-to-Day)Not explicitly stated as a numerical acceptance criterion in the provided text.CV (%) ranged from 2.1% to 7.5% across 3 control solution levels. SD (mg/dL) ranged from 3.1 to 7.2.N/A
Linearity (R2)Not explicitly stated as a numerical acceptance criterion in the provided text.0.9973 (Linear regression slope 0.9809, Y-intercept 0.8180)N/A

2. Sample Size and Data Provenance (for System Accuracy / Method Comparison Study)

  • Test Set Sample Size: 153 diabetes patients.
  • Data Provenance: Not explicitly stated (e.g., country of origin), but implies a multi-site study ("at three sites"). The study is prospective as it involves taking blood samples from diabetes patients for comparison.

3. Number of Experts and Qualifications for Ground Truth (System Accuracy)

  • Number of Experts: Not explicitly stated, however, the "healthcare professional" at the study sites performed tests using the EG V1 Pro meter and the YSI 2300 analyzer. The YSI 2300D analyzer is the reference method, which is a laboratory instrument.
  • Qualifications of Experts: "Healthcare professional" is mentioned. Additional specific qualifications (e.g., "radiologist with 10 years of experience") are not provided. Given it's a glucose measurement, these would likely be laboratory technicians or nurses/medical staff trained in sample collection and operation of the reference device.

4. Adjudication Method (Test Set)

  • None. The ground truth is established by a single, traceable laboratory reference method (YSI 2300D analyzer) for each sample. Measurements from the EG V1 Pro meter are then compared directly to this reference.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No. This type of study is typically for evaluating the diagnostic performance of a device when interpreted by multiple human readers (e.g., radiology images). For a quantitative self-monitoring blood glucose system, comparing human interpretation of cases is not applicable. The comparison is between the device's numerical output and a laboratory reference.

6. Standalone Performance (Algorithm Only)

  • Yes. The system accuracy study directly compares the results from the EG V1 Pro meter (algorithm only, without human interpretation for the result itself) against the reference method.

7. Type of Ground Truth Used

  • Reference Method (Laboratory Analyzer): The YSI 2300D analyzer served as the ground truth. This is a common and accepted clinical laboratory method for glucose measurement.

8. Sample Size for the Training Set

  • The document does not explicitly mention a "training set" in the context of machine learning. The studies described are performance validation studies for the device itself.
    • Analytical Performance (Precision): 200 tests per range for within-run (total 1200+ for 6 ranges), 400 tests for day-to-day precision.
    • Analytical Performance (Linearity): 630 tests in three test strip lots.
    • Analytical Performance (Hematocrit Interference): 40 tests per hematocrit level (total of 400 for 10 levels across 2 lots).
    • System Accuracy (Clinical Study): 153 diabetes patients.

9. How Ground Truth for Training Set was Established

  • As a traditional quantitative measurement device, it's unlikely to have a "training set" in the machine learning sense. The device's calibration and analytical performance would be established through a series of experiments using known glucose concentrations and reference methods. For the analytical studies (precision, linearity, hematocrit interference), the ground truth was established by:
    • Known glucose concentrations (glucolyzed/spiked venous blood).
    • The YSI 2300D analyzer as the reference method, traceable to NIST 965b.

EG V1 (BL) Self Monitoring Glucose Test System (Over-The-Counter / Home Use)

1. Table of Acceptance Criteria and Reported Device Performance (Patient Use)

The acceptance criteria for system accuracy are based on ISO 15197:2003.

Performance MetricAcceptance Criteria (ISO 15197:2003)Reported Device Performance (Patient Use)Meets Criteria?
System Accuracy (Patient Use)
For Glucose < 75 mg/dL95% of individual glucose results to be within ±15 mg/dL of the laboratory reference method.100% (19/19) of individual glucose results were within ±15 mg/dL of the YSI 2300D analyzer for glucose concentrations < 75 mg/dL from fingertip samples. (Also reported: 73.68% within ±5mg/dL, 100% within ±10mg/dL).Yes
For Glucose ≥ 75 mg/dL95% of individual glucose results to be within ±20% of the laboratory reference method.97.76% (131/134) of individual glucose results were within ±20% of the YSI 2300D analyzer for glucose concentrations ≥ 75 mg/dL from fingertip samples. (Also reported: 61.94% within ±5%, 85.82% within ±10%, 96.27% within ±15%).Yes
Hematocrit InterferenceBias ≤ 15 mg/dL when glucose concentration <75 mg/dL and bias ≤ 15% when glucose concentration ≥75 mg/dL (for hematocrit ranges 20-60%)Max % Bias reported for different hematocrit levels (20%, 30%, 41%, 49%, 60%) across two strip lots: 11.5% and 13.1% (all within 15%). (Same data as EG V1 Pro, implies device uses same fundamental analytical components).Yes
Reportable RangeNot explicitly stated as a numerical acceptance criterion, but implied to be validated.20-600 mg/dLValidated
Precision (Within-run)Not explicitly stated as a numerical acceptance criterion in the provided text.CV (%) ranged from 2.7% to 6.5% across 6 glucose concentration ranges.N/A
Precision (Day-to-Day)Not explicitly stated as a numerical acceptance criterion in the provided text.CV (%) ranged from 2.1% to 7.5% across 3 control solution levels.N/A
Linearity (R2)Not explicitly stated as a numerical acceptance criterion in the provided text.0.9973N/A

2. Sample Size and Data Provenance (for System Accuracy / Method Comparison Study)

  • Test Set Sample Size: 153 diabetes patients.
  • Data Provenance: Not explicitly stated (country of origin), but implies a multi-site study ("at three sites"). The study is prospective as it involves patients performing tests.

3. Number of Experts and Qualifications for Ground Truth (System Accuracy)

  • Number of Experts: A "healthcare professional" performed the reference tests using the YSI 2300D analyzer.
  • Qualifications of Experts: "Healthcare professional." Specific qualifications are not provided.

4. Adjudication Method (Test Set)

  • None. The ground truth is established by a single, traceable laboratory reference method (YSI 2300D analyzer) for each sample. Measurements from the EG V1 (BL) meter are then compared directly to this reference.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No. Similar to the Pro version, this type of study is not applicable for a quantitative blood glucose monitoring system.

6. Standalone Performance (Algorithm Only)

  • Yes. The system accuracy study compares the results from the EG V1 (BL) meter (algorithm only, with patient applying the sample) directly against the reference method.

7. Type of Ground Truth Used

  • Reference Method (Laboratory Analyzer): The YSI 2300D analyzer served as the ground truth.

8. Sample Size for the Training Set

  • Similar to the Pro version, the document does not explicitly mention a "training set" in the context of machine learning. The studies described are performance validation studies.
    • Analytical Performance (Precision): 200 tests per range for within-run, 400 tests for day-to-day precision.
    • Analytical Performance (Linearity): 630 tests.
    • Analytical Performance (Hematocrit Interference): 40 tests per hematocrit level.
    • System Accuracy (Clinical Study): 153 diabetes patients.

9. How Ground Truth for Training Set was Established

  • For the analytical studies, the ground truth was established by:
    • Known glucose concentrations (glucolyzed/spiked venous blood).
    • The YSI 2300D analyzer as the reference method, traceable to NIST 965b.

{0}------------------------------------------------

MAY 2 3 2011

510(k) SUMMARY for EG V1 Pro Self Monitoring Glucose Test System

A. Submitter's information

EPS Bio Technology Corp. Company: No. 8, R&D RD. III, Hsinchu Science Park Hsinchu City, Taiwan, R.O.C. Address: Mr. Y.C. Lei, General Manager Contact Name: 886-3-6686868 Phone: 886-3-6686866 Fax:

B. Measured:

Glucose

  • C. Type of Test: Quantitative; electrochemical biosensor
  • D. Proprietary and Established Names: EG V1 Pro Self-Monitoring Blood Glucose Test System

E. Common or Usual Name: Glucose Test System

F. Regulatory Information:

    1. Regulation section: 21 CFR 862.1345, Glucose Test System 21 CFR 862.1660, Quality control materials (assayed and unassayed)
    1. Classification: Class II Class I, reserved

3. Product code:

NBW, System Test, Blood Glucose, Over the Counter LFR, Glucose Dehydrogenase, Glucose JJX, Single (Specified) Controls (assayed and Unassayed)

    1. Panel: Chemistry 75

G. Intended Use:

    1. Intended use(s): See Indications for Use below

{1}------------------------------------------------

2. Indication(s) for use:

The EG V1 Pro Self Monitoring Blood Glucose Test System is intended for the quantitative measurement of glucose in venous whole blood or fresh capillary whole blood from fingertip. Testing is done outside the body (In Vitro diagnostic use). It is intended for multiple-patient use in professional healthcare settings, as an aid to monitor the effectiveness of diabetes control. The system is only used with single-use lancing devices. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus.

The system consists of the EG V1 Pro meter and the EG Pro test strips. The EG V1 Pro meter only is used with the EG Pro test strips to quantitatively measure glucose in venous whole blood or fresh capillary whole blood from fingertip.

The EG Glucose Control Solution

For use with the EG V1 Pro Blood Glucose Self Monitoring System as a quality control check to verify the accuracy of blood glucose test results.

H. Device Description:

The EG V1 Pro Self-Monitoring Blood Glucose System consists of the EG V1 Pro Blood Glucose Meter, EG Pro Glucose Test Strips, single-use Lancing Device and EG Level 2 Control Solution. (Please refer to the IFU for the product picture)

l. Substantial Equivalence Information:

    1. Predicate device name(s): ASCENSIA CONTOUR BLOOD GLUCOSE METER, MODEL 7151; REAGENT STRIP, MODEL 7080
    1. Device Company BAYER HEALTHCARE
    1. Predicate 510(k) number(s): K062058

4. Comparison with predicate:

Similarities
ItemDevicePredicate
Detection methodAmperometryAmperometry
EnzymeGlucose Dehydrogenase(FAD)Glucose Dehydrogenase(FAD)
Sample volume$\geq$ 0.6 uL$\geq$ 0.6 uL
Temperature range10-40°C10-40°C
Memory capability480 tests with date and time480 tests with date and time
Hematocrit range20-60%20-60%
Power3V CR2032 batteries3V CR2032 batteries

{2}------------------------------------------------

Differences
ItemDevicePredicate
Test range20-600 mg/dL10-600 mg/dL
Anatomical SitesVenous sample andCapillary samples from thefingertip, palm, forearm,Capillary samples from thefingertip, palm, forearm,abdomen and thigh
Test time5 seconds8 seconds
Size L x W x H (inch)3.5"x 2.1"x 0.97"2.8"x 2.35"x 0.77"
Weight2.05 oz (without batteries)2.0 oz (without batteries)

J. Standard/Guidance Document Referenced (if applicable):

  • ISO 15197:2003, In Vitro Diagnostic Test Systems Requirements for Blood 1. Glucose Monitoring Test Systems for Self Managing Diabetes Mellitus.
  • ISO 14971:2007, Medical Devices Application of Risk Management to Medical 2. Devices.
  • EN 13640:2002, Stability Testing of In Vitro Diagnostic Reagents. న.
  • IEC 60601-1-2:2004. Medical electrical equipment Part 1: General requirement for 4. safety; Electromagnetic compatibility -Requirements and tests.
  • IEC 61010-2-101:2002, Safety Requirements for Electrical Equipment for 5. Measurement, Control, and Laboratory Use - Part 2-101: Particular Requirements for In Vitro Diagnostic (IVD) Medical Equipment.
  • IEC 60601-1:2005, Medical Electrical Equipment Part 1: General Requirements for ి. Safety.
  • CLSI EP7-2A:2005, Interference Testing in Clinical Chemistry. 7.

K. Test Principle:

The EG V1 Pro Self-Monitoring Blood Glucose System employs a disposable dry reagent strip technology, using glucose dehydrogenase method to quantify glucose. Each test strip features an electrode containing the glucose dehydrogenase (in the presence of the coenzyme: Flavin Adenine Dinucleotide, FAD). A blood sample is applied to the blood collection area at the tip of the strip and is automatically drawn into the reaction zone, where the FAD-binding glucose dehydrogenase catalyzes glucose to dehydrogenate and to produce gluconolactone. During the reaction, a mediator transfers electrons to the electrode surface and generates a current. The amount of the current is proportional to the amount of glucose present in the blood sample. The glucose concentration is calculated in the EG V1 Pro Glucose Meter and displayed on the screen after 5 seconds.

L. Performance Characteristics (if/when applicable):

    1. Analytical performance:
    • a. Precision/Reproducibility:

Within-run Precision – The testing was used by venous blood, which was stored in the heparin blood collection tube. Glucose was glucolyzed/spiked to the blood to prepare 6 different levels of glucose concentration: 20-50 mg/dL, 51-110 mg/dL, 111-150 mg/dL, 151-250 mg/dL, 251-400 mg/dL, and 401~600 mg/dL. For each testing range, 10 meters, 10 pcs of test strips for each meter, and 2 lots of test strips were used. (N=10 Meters x10 tests x 2 lots =200)

{3}------------------------------------------------

EG V1 SMBG system
Range (mg/dL)NMean(mg/dL)SD(mg/dL)CV (%)
20~50 mg/dL20040.72.66.5%
51~110 mg/dL200108.53.53.2%
111~150 mg/dL200137.04.53.3%
151~250 mg/dL200223.56.42.9%
251~400 mg/dL200399.811.02.8%
401~600 mg/dL200512.113.92.7%
Level 2 control solution200118.23.32.8%

Day-to-Day Precision - 10 Meters, 2 lots of test strips, and 3 control solutions (Level 1, 2, and 3) were prepared. Each control was tested twice a day, once in the morning and once in the afternoon consecutively for 10 days.

(N=10 Meter x 2 Lots x 2 tests x10 days =400)

ControlsolutionNmean(mg/dL)SD(mg/dL)CV (%)
Level 140041.83.17.5%
Level 2400120.73.12.6%
Level 3400349.87.22.1%
  • b. Linearity assay reportable range:
    A 25 mL venous whole blood sample was treated with heparin vacuum tube. The blood supplemented with ß -D-glucose to prepare seven different blood glucose levels (20-50 mg/dL, 51-80 mg/dL, 81-120 mg/dL, 121-200 mg/dL, 201-300 mg/dL, 301-400 mg/dL, and 400-600 mg/dL). A total of 630 tests were performed to determine the linear regression in three test strip lots.

For each strip lot, N=3 tests/day x 10 days x 7 glucose ranges=210

The linear regression was as follows:

N=630 (210X3Lot)
Slope0.9809
Y-intercept0.8180
R20.9973
  • Traceability, Stability, Expected values (controls, calibrators, or methods): ﻥ The device was traceable to a laboratory YSI 2300D analyzer. The YSI 2300D analyzer was calibrated before use and the calibrator was manufactured gravimetrically and provided from the instrument supplier. The YSI 2300D analyzer is validated by using the international standard reference material (NIST 965b) every year.
    Stability characteristics of the level 2 and level 3 control solutions were determined by performing real-time stability studies and the shelf life could be up to 19 months

{4}------------------------------------------------

at the room temperature.

  • d. Detection limit:
    Data was provided to support a reportable range of 20-600 mg/dL.

  • e. Analytical specificity:
    Hematocrit interference was evaluated by adjusting the glucose and hematocrit levels of venous blood samples from several donors. The venous blood samples were glucolyzed/spiked to 4 glucose concentrations and the hematocrit levels were adjusted to 20%, 30%, 40%, 50% and 60%. For each hematocrit level, the linear regression was determined by 40 testing results, N=2 times x 5 meters x 4 glucose ranges=40. The maximum percent bias was calculated by comparing to the YSI reading and shown on each linear regression. The acceptance criteria of Hematocrit interference test is: bias ≤15 mg/dL when glucose concentration <75 mg/dL and bias ≤ 15% when glucose concentration ≥75 mg/dL. Results were summarized in the following table. The device was claimed to use blood samples having hematocrit concentrations ranging from 20% to 60%.

Strip lot #Hematocrit %SlopeY-interceptR2% Bias
01207290120%1.1302-0.15490.999411.5
30%1.1020-1.22480.99928.2
41%0.95816.77170.9967-0.3
49%0.8948-0.66860.9984-9.9
60%0.8800-1.90430.9994-11.7
01207270120%1.12570.12200.999211.3
30%1.06207.23610.99949.4
41%1.00091.18450.99770.7
49%0.8875-0.73340.9991-10.3
60%0.8674-1.96110.9991-13.1
  • f. Assay cut-off:
    N/A

    1. System Accuracy studies:
    • a. Accuracy/Method comparison with predicate device at sites 1, 2, and 3:

A total 153 diabetes patients were taken capillary blood from a fingertip, palm, and forearm using the EG V1 Pro system at three sites. A healthcare professional performed the tests using the EG V1 Pro meter and the YSI 2300 analyzer. The qlucose range of testing values among these samples was 50-517 mg/dL. Meter versus YSI at each site met the ISO 15197 requirement of ninety-five percent (95 %) of the individual glucose results falling within ±15 mg/dL of the laboratory test at glucose concentrations <75 mg/dL and within ± 20 % at glucose concentrations ≥75 mg/dL. These results were summarized in the following tables.

{5}------------------------------------------------

System accuracy professional results for glucose concentrations < 75 mg/dL

Within ±5mg/dLWithin ±10mg/dLWithin ±15mg/dL
16/19 (84.21%)17/19 (89.47%)19/19 (100%)

System accuracy professional results for glucose concentrations ≥ 75 mg/dL

Within ±5%Within ±10%Within ±15%Within ±20%
102/134(76.12%)125/134(93.28%)129/134(96.27%)132/134(98.51%)

Method comparison studies were performed by using alternative site testing (AST) samples compared to YSI. The AST samples from the thenar, hypothenar, and forearm taken by a professional were tested. Each linear regression was calculated and shown in the following table:

ComparisonNRange(mg/dL)Slope and Y-interceptR2
Finger(professional) vs. YSI15358.5 - 433.5$Y=1.0190X-2.9299$0.9823
Thenar(professional) vs. YSI15355 - 459$Y=1.0101X-0.6506$0.9719
Hypothenar(professional) vs. YSI15353 - 517$Y=1.0537X-7.1471$0.9673
Forearm(professional) vs. YSI15350 - 514$Y=1.0282 X-3.9814$0.9650
Venous vs. YSI15359 - 432$Y=0.9874 X+1.5835$0.9744

The comparison with the predicate device, the linear regressions was as follows: EG V1 Pro vs. predicate device: Y = 0.9859X +3.5632, R2= 0.9387

  • b. Matrix comparison: N/A
    1. Clinical studies:
    • a. Clinical Sensitivity: N/A
    • b. Clinical specificity: N/A
    • c. Other clinical supportive data (when a. and b. are not applicable): see section 2.a.
    1. Clinical cut-off: N/A
    1. Expected values/Reference range:
    1. The expected blood glucose levels for an adult (referenced from American Diabetes Association. Standards of medical care in diabetes-Table 8. Diabetes care. 2008; S18.):

{6}------------------------------------------------

Time Before meal or fasting Post -meal

Range (mg/dL) 70-130 Less than 180

    1. PDI® SUPER SANI-CLOTH® Germicidal disposable wipes with EPA registration# 9480-4 was validated demonstrating complete inactivation of live virus. It demonstrated that there was no change in performance or in the external materials of the meter and lancing device after 11,000 cleaning and disinfection cycles designed to simulated 3 years for professional use.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The EG V1 Pro Self Monitoring Blood Glucose System has the same intended use and similar technological characteristics as the Bayer Contour Self Monitoring Blood Glucose System (K062058) marketed by BAYER HEALTHCARE. Moreover, bench testing contained in the submission demonstrated that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the EG V1 Pro Self-Monitoring Blood Glucose System is substantially equivalent to the predicate devise.

7

{7}------------------------------------------------

510(k) SUMMARY for EG V1 (BL) Self Monitoring Glucose Test System

A. Submitter's information

Company: EPS Bio Technology Corp. No. 8, R&D RD. III, Hsinchu Science Park Hsinchu City, Taiwan, R.O.C. Address: Contact Name: Mr. Y.C. Lei, General Manager 886-3-6686868 Phone: 886-3-6686866 Fax:

B. Measured:

Glucose

C. Type of Test: Quantitative; electrochemical biosensor

D. Proprietary and Established Names:

EG V1 (BL) Self-Monitoring Blood Glucose Test System

E. Common or Usual Name:

Glucose Test System

F. Regulatory Information:

    1. Regulation section: 21 CFR 862.1345, Glucose Test System 21 CFR 862.1660, Quality control materials (assayed and unassayed)
    1. Classification: Class II Class I, reserved

3. Product code:

NBW, System Test, Blood Glucose, Over the Counter LFR, Glucose Dehydrogenase, Glucose JJX, Single (Specified) Controls (assayed and Unassayed)

    1. Panel: Chemistry 75

G. Intended Use:

    1. Intended use(s): See Indications for Use below

{8}------------------------------------------------

2. Indication(s) for use:

The EG V1 (BL) Self Monitoring Blood Glucose Test System is intended for the quantitative measurement of glucose in fresh capillary whole blood from fingertip, palm, or forearm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by a single patient with diabetes and should not be shared, as an aid to monitor the effectiveness of diabetes control. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus. Alternative site testing can be only used during steady-state blood qlucose conditions.

The system consists of the EG V1 (BL) meter and the EG test strips. The EG V1 (BL) meter only is used with the EG V1 test strips to quantitatively measure glucose in fresh capillary whole blood from fingertip, palm, or forearm.

The EG Glucose Control Solution

For use with EG V1(BL) Blood Glucose Self Monitoring System as a quality control check to verify the accuracy of blood glucose test results.

    1. Special conditions for use statement(s):

EG V1 (BL) Self Monitoring Blood Glucose Test System:

For Over-the-Counter use. Not for use with newborns.

Alternative site testing (AST) can only be used during steady-state blood glucose conditions. AST (forearm) should only be performed under the following conditions:

  • ◆ Testing before a meal.
  • ♦ In a fasting state.
  • ♦ Two hours or more after a meal.
  • ♦ Two hours or more after insulin dosing.
  • ♦ Two hours after physical activity.

Measurements from alternative site testing should never be used to calibrate a continuous glucose monitor (CGM) or entered into insulin dose calculators for insulin dosing recommendations.

    1. Special instrument requirements:
      EG V1(BL) Blood Glucose Meter.

H. Device Description:

The EG V1 (BL) Self-Monitoring Blood Glucose System consists of the EG V1 (BL) Blood Glucose Meter, EG Glucose Test Strips, Auto-Lancet Device and EG Level 2 Control Solution. (Please refer to the IFU for the product picture)

l. Substantial Equivalence Information:

    1. Predicate device name(s):
      ASCENSIA CONTOUR BLOOD GLUCOSE METER, MODEL 7151; REAGENT STRIP, MODEL 7080

{9}------------------------------------------------

    1. Device Company BAYER HEALTHCARE
    1. Predicate 510(k) number(s): K062058
    1. Comparison with predicate:
Similarities
ItemDevicePredicate
Detection methodAmperometryAmperometry
EnzymeGlucose Dehydrogenase(FAD)Glucose Dehydrogenase(FAD)
Sample volume> 0.6 uL> 0.6 uL
Temperature range10-40°C10-40°C
Memory capability480 tests with date andtime480 tests with date andtime
Hematocrit range20-60%20-60%
Power3V CR2032 batteries3V CR2032 batteries
Differences
ItemDevicePredicate
Test range20-600 mg/dL10-600 mg/dL
Anatomical SitesVenous sample andCapillary samples from thefingertip, palm, forearm,Capillary samples from thefingertip, palm, forearm,abdomen and thigh
Test time5 seconds8 seconds
Size L x W x H (inch)3.5"x 2.1"x 0.97"2.8"x 2.35"x 0.77"
Weight2.05 oz (without batteries)2.0 oz (without batteries)

J. Standard/Guidance Document Referenced (if applicable):

  • ISO 15197:2003, In Vitro Diagnostic Test Systems Requirements for Blood 1. Glucose Monitoring Test Systems for Self Managing Diabetes Mellitus.
  • ISO 14971:2007, Medical Devices Application of Risk Management to Medical 2. Devices.
  • EN 13640:2002, Stability Testing of In Vitro Diagnostic Reagents. 3.
  • IEC 60601-1-2:2004. Medical electrical equipment Part 1: General requirement for 4. safety: Electromagnetic compatibility -Requirements and tests.
  • IEC 61010-2-101:2002, Safety Requirements for Electrical Equipment for 5. Measurement, Control, and Laboratory Use - Part 2-101: Particular Requirements for In Vitro Diagnostic (IVD) Medical Equipment.
  • IEC 60601-1:2005, Medical Electrical Equipment Part 1: General Requirements for 6. Safety.
  • CLSI EP7-2A:2005, Interference Testing in Clinical Chemistry. 7.

{10}------------------------------------------------

K. Test Principle:

The EG V1 (BL) Self-Monitoring Blood Glucose System employs a disposable dry reagent strip technology, using glucose dehydrogenase method to quantify glucose. Each test strip features an electrode containing the glucose dehydrogenase (in the presence of the coenzyme: Flavin Adenine Dinucleotide, FAD). A blood sample is applied to the blood collection area at the tip of the strip and is automatically drawn into the reaction zone, where the FAD-binding glucose dehydrogenase catalyzes glucose to dehydrogenate and to produce qluconolactone. During the reaction, a mediator transfers electrons to the electrode surface and generates a current. The amount of the current is proportional to the amount of glucose present in the blood sample. The glucose concentration is calculated in the EG V1 (BL) Glucose Meter and displayed on the screen after 5 seconds.

L. Performance Characteristics (if/when applicable):

    1. Analytical performance:
    • a. Precision/Reproducibility:

Within-run Precision – The testing was used by venous blood, which was stored in the heparin blood collection tube. Glucose was glucolyzed/spiked to the blood to prepare 6 different levels of glucose concentration: 20-50 mg/dL, 51-110 mg/dL, 111-150 mg/dL, 151-250 mg/dL, 251-400 mg/dL, and 401~600 mg/dL. For each testing range, 10 meters, 10 pcs of test strips for each meter, and 2 lots of test strips were used. (N=10 Meters x10 tests x 2 lots =200)

EG V1 SMBG system
Range (mg/dL)NMean(mg/dL)SD(mg/dL)CV (%)
20~50 mg/dL20040.72.66.5%
51~110 mg/dL200108.53.53.2%
111~150 mg/dL200137.04.53.3%
151~250 mg/dL200223.56.42.9%
251~400 mg/dL200399.811.02.8%
401~600 mg/dL200512.113.92.7%
Level 2 control solution200118.23.32.8%

Dav-to-Day Precision - 10 Meters, 2 lots of test strips, and 3 control solutions (Level 1, 2, and 3) were prepared. Each control was tested twice a day, once in the morning and once in the afternoon consecutively for 10 days.

(N=10 Meter x 2 Lots x 2 tests x10 days =400) .

ControlsolutionNmean(mg/dL)SD(mg/dL)CV (%)
Level 140041.83.17.5%
Level 2400120.73.12.6%
Level 3400349.87.22.1%
  • b. Linearity assay reportable range:
    A 25 mL venous whole blood sample was treated with heparin vacuum tube. The

{11}------------------------------------------------

blood supplemented with { -D-glucose to prepare seven different blood glucose levels (20-50 mg/dL, 51-80 mg/dL, 81-120 mg/dL, 121-200 mg/dL, 201-300 mg/dL, 301-400 mg/dL, and 400-600 mg/dL). A total of 630 tests were performed to determine the linear regression in three test strip lots.

For each strip lot, N=3 tests/day x 10 days x 7 glucose ranges=210

The linear regression was as follows:

N=630 (210X3Lot)
Slope0.9809
Y-intercept0.8180
R20.9973
  • c. Traceability, Stability, Expected values (controls, calibrators, or methods): The device was traceable to a laboratory YSI 2300D analyzer. The YSI 2300D analyzer was calibrated before use and the calibrator was manufactured gravimetrically and provided from the instrument supplier. The YSI 2300D analyzer is validated by using the international standard reference material (NIST 965b) every year.
    Stability characteristics of the Level 2 and Level 3 control solutions were determined by performing real-time stability studies and the shelf life could be up to 19 months at the room temperature.

  • d. Detection limit:
    Data was provided to support a reportable range of 20-600 mg/dL.

  • e. Analytical specificity:
    Hematocrit interference was evaluated by adjusting the glucose and hematocrit levels of venous blood samples from several donors. The venous blood samples were glucolyzed/spiked to 4 glucose concentrations and the hematocrit levels were adjusted to 20%, 30%, 40%, 50% and 60%. For each hematocrit level, the linear regression was determined by 40 testing results, N=2 times x 5 meters x 4 glucose ranges=40. The maximum percent bias was calculated by comparing to the YSI reading and shown on each linear regression. The acceptance criteria of Hematocrit interference test is: bias ≤ 15 mg/dL when glucose concentration <75 mg/dL and bias ≤ 15% when glucose concentration ≥75 mg/dL. Results were summarized in the following table. The device was claimed to use blood samples having hematocrit concentrations ranging from 20% to 60%.

Strip lot #Hematocrit %SlopeY-interceptR2% Bias
01207290120%1.1302-0.15490.999411.5
30%1.1020-1.22480.99928.2
41%0.95816.77170.9967-0.3
49%0.8948-0.66860.9984-9.9
60%0.8800-1.90430.9994-11.7

{12}------------------------------------------------

01207270120%1.12570.12200.999211.3
30%1.06207.23610.99949.4
41%1.00091.18450.99770.7
49%0.8875-0.73340.9991-10.3
60%0.8674-1.96110.9991-13.1
  • f. Assay cut-off:
    N/A

2. System Accuracy studies:

a. Accuracy/Method comparison with predicate device at sites 1, 2, and 3:

A total 153 diabetes patients performed a fingerstick, palm and forearm using the EG V1 (BL) system at three sites. A patient performed the testing using the EG V1 (BL) meter. A healthcare professional performed the test using the YSI 2300 analyzer. The glucose range of testing values among these samples was 50-517 ma/dL.

Meter versus YSI at each site met the ISO 15197 requirement of ninety-five percent (95 %) of the individual glucose results falling within ±15 mg/dL of the laboratory test at glucose concentrations <75 mg/dL and within ± 20 % at glucose concentrations ≥75 mg/dL. These results were summarized in the following tables.

Within ±5mg/dLWithin ±10mg/dLWithin ±15mg/dL
14/19 (73.68%)19/19 (100%)19/19 (100%)

System accuracy patient results for glucose concentrations < 75 mg/dL

System accuracy patient results for glucose concentrations ≥ 75 mg/dL

Within ±5%Within ±10%Within ±15%Within ±20%
83/134(61.94%)115/134(85.82%)129/134(96.27%)131/134(97.76%)

Method comparison studies were performed by using alternative site testing (AST) samples compared to YSI. The AST samples from the thenar, hypothenar, and forearm taken by a patient were tested. Each linear regression was calculated and shown in the following table:

ComparisonNRange(mg/dL)Slope and Y-interceptR2
Finger(a patient) vs. YSI15358.5 - 433.5Y=1.0100X-0.96790.9696
Thenar(a patient) vs. YSI15355 - 459Y=0.9688X+4.82410.9481
Hypothenar(a patient) vs. YSI15353 - 517Y=1.0896 X-11.75960.9584
Forearm(a patient) vs. YSI15350 - 514Y=1.0263X-4.64250.9503

{13}------------------------------------------------

The comparison with the predicate device, the linear regressions was as follows: EG V1 (BL) vs. predicate device: Y = 0.9859X +3.5632, R2= 0.9387

  • b. Matrix comparison: N/A
    1. Clinical studies:
    • a. Clinical Sensitivity: N/A
    • b. Clinical specificity: N/A
    • c. Other clinical supportive data (when a. and b. are not applicable): see section 2.a.
    1. Clinical cut-off:

N/A

    1. Expected values/Reference range:
      The expected blood glucose levels for an adult (referenced from American Diabetes Association. Standards of medical care in diabetes-Table 8. Diabetes care. 2008; S18.):
TimeRange (mg/dL)
Before meals or fasting70-130
Post -mealsLess than 180
    1. PDI® SUPER SANI-CLOTH® Germicidal disposable wipes with EPA registration# 9480-4 was validated demonstrating complete inactivation of live virus. It demonstrated that there was no change in performance or in the external materials of the meter and lancing device after 11,000 cleaning and disinfection cycles designed to simulated 4 years for home use.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The EG V1 (BL) Self Monitoring Blood Glucose System has the same intended use and similar technological characteristics as the Bayer Contour Self Monitoring Blood Glucose System (K062058) marketed by BAYER HEALTHCARE. Moreover, bench testing contained in the submission demonstrated that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the EG V1 (BL) Self-Monitoring Blood Glucose System is substantially equivalent to the predicate devise.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a sans-serif font. The text is centered on the image and is the only element present.

EPS Bio Technology Corp. c/o Y.C. Lei No. 8 R&D. Road III, Hsinchu Science Park Hsinchu City, ROC 30077

Public Health Service

Food and Druq Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

MAY 2 3 2011

Re: K101037

Taiwan

Trade/Device Name: EG V1(BL) Self Monitoring Blood Glucose Test System, EG V1 Pro Self Monitoring Blood Glucose Test System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW, LFR, JJX Dated: 18 May 2011 Received: 20 May 2011

Dear: Mr. Lei,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{15}------------------------------------------------

Page 2 -

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

C.C.

Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

: 173 :

Enclosure

{16}------------------------------------------------

Indications for Use

510(k) Number (if known): K101037

Device Name: EG V1 Pro Self Monitoring Blood Glucose Test system

Indication For Use:

The EG V1 Pro Self Monitoring Blood Glucose Test System is intended for the quantitative measurement of glucose in venous whole blood or fresh capillary whole blood from the fingertip. Testing is done outside the body (In Vitro diagnostic use). It is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control. The system is only used with single-use lancing devices. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus.

The system consists of the EG V1 Pro meter and the EG Pro Test Strips. The EG V1 Pro meter only is used with the EG Pro Test Strips to quantitatively measure glucose in venous whole blood or fresh capillary whole blood from the fingertip.

EG Glucose Control Solution

For use with the EG V1 Pro Blood Glucose Self Monitoring System as a quality control check to verify the accuracy of blood glucose test results.

Prescription Use V (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use V . (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Signature

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) k/010103

{17}------------------------------------------------

Indications for Use

510(k) Number (if known): K101037

Device Name: EG V1(BL) Self Monitoring Blood Glucose Test system

Indication For Use:

The EG V1 (BL) Self Monitoring Blood Glucose Test System is intended for the quantiatative measurement of glucose in fresh capillary whole blood from the fingertip, palm or forearm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter[OTC]) by a single patient with diabetes and should not be shared, as an aid to monitor the effectiveness of diabetes control. The system is not to be used on neonates, nor for the diagnosis of, or screening for diabetes mellitus. Alternative site testing can only be used during steady-state blood glucose conditions.

The system consists of the EG V1 (BL) meter and the EG V1 Test Strips. The EG V1 (BL) meter only is used with the EG V1 Test Strips to quantitatively measure glucose in venous whole blood or fresh capillary whole blood from the fingertip, palm, or forearm.

EG Glucose Control Solution

For use with the EG V1 Pro Blood Glucose Self Monitoring System as a quality control check to verify the accuracy of blood glucose test results.

Prescription UseAnd/OrOver the Counter Use V.
(21 CFR Part 801 Subpart D)(21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)


Rights Sign Off

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K10103

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.