K Number
K062058

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-09-11

(53 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ascensia® CONTOUR®Blood Glucose Monitoring System is used for the measurement of glucose in whole blood. The Ascensia CONTOUR® Blood Glucose Monitoring System is an over-the-counter (OTC) device used by persons with diabetes and by healthcare professionals in home settings and in healthcare facilities.

The Ascensia CONTOUR® Blood Glucose Monitoring System is indicated for use with capillary, venous, and arterial whole blood samples and neonatal blood samples. Capillary samples may be drawn from the fingertip, palm, forearm, and in the case of neonates, the heel.

The frequent monitoring of blood glucose is an adjunct to the care of persons with diabetes.

Device Description

The Ascensia® CONTOUR® Blood Glucose Monitoring System is used for the measurement of glucose in whole blood. The system contains a blood glucose meter, a bottle of strips, a bottle of normal control solution, a lancing device and lancets and instructions for use.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Ascensia CONTOUR® Blood Glucose Monitoring System, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state numerical acceptance criteria in a table format for accuracy. Instead, it focuses on demonstrating "equivalent performance" to a predicate device. The performance assessment section describes the studies conducted.

Acceptance Criteria (Implied)Reported Device Performance
Equivalent performance to predicate deviceStudies showed equivalent performance with the original Ascensia® CONTOUR® Blood Glucose Monitoring System
Substantially equivalent to predicate deviceDemonstrated that the device produces blood glucose results that are substantially equivalent to results obtained on the predicate device.
(Specific accuracy criteria - e.g., % within +/- threshold)Not specified in the provided text.

2. Sample Size Used for the Test Set and Data Provenance

The document mentions that the device was "studied in the laboratory and in a clinical setting by persons with diabetes." However, it does not specify the sample size for the test set or the country of origin. It indicates that the clinical setting involved "persons with diabetes," suggesting a prospective clinical study component.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not provide information on the number of experts used or their qualifications for establishing ground truth specifically for the test set. It mentions comparison to "a laboratory method," which implies a reference method, but details about the personnel executing this reference method are absent.

4. Adjudication Method for the Test Set

The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study. There is no information about human readers or the effect size of AI assistance. This type of study is typically relevant for interpretative devices (e.g., imaging) where expert readers provide interpretations. For a blood glucose meter, the primary comparison would be against a reference method rather than different human interpretations.

6. Standalone (Algorithm Only) Performance

The device itself is a standalone algorithm (blood glucose meter and its internal software). Its performance against the predicate and laboratory method is its standalone performance. There isn't a separate "algorithm only" study distinct from the device's overall performance.

7. Type of Ground Truth Used

The ground truth for evaluating the device's performance was established by:

  • Comparison to the original Ascensia® CONTOUR® Blood Glucose Monitoring System (predicate device).
  • Comparison to a "laboratory method." The specific type of laboratory method (e.g., YSI analyzer, enzymatic assay) is not detailed, but it serves as the reference standard.

8. Sample Size for the Training Set

The document does not provide information on the sample size for the training set. Blood glucose meters are typically calibrated during manufacturing and would have undergone extensive internal testing, but the specifics of a "training set" for algorithm development are not disclosed here.

9. How Ground Truth for the Training Set Was Established

The document does not describe how ground truth for any potential "training set" was established. Given the nature of a blood glucose meter, the "training" would likely involve robust manufacturing calibration and quality control procedures using known glucose concentrations.

{0}------------------------------------------------

Bayer HealthCare Diabetes Care Division

Image /page/0/Picture/1 description: The image shows the Bayer company logo. The logo consists of the name "BAYER" written twice, once horizontally and once vertically, forming a cross shape. The entire logo is enclosed within a circle.

SEP 1 1 2006

510(k) SUMMARY

Ascensia Contovr® Blood Glucose Monitoring System

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is K062058.

Prepared:August 4, 2006
Submitter:Bayer HealthCare, Diabetes Care Division
Address:430 South Beiger StreetMishawaka, IN 46544Phone (574) 262-7152; FAX (574) 262 6945
Contact:Roger Sonnenburg, Regulatory Affairs Manager
Device:Trade/Proprietary Name: Ascensia® CONTOUR® BloodGlucose Monitoring System
Common/Usual Name:Blood Glucose Meter
Classification:Division of Clinical Laboratory DevicesPanel - Clinical Chemistry and ToxicologyClassification Code – 75 LFR, Glucose Dehydrogenase,Glucose
Predicate Device:Ascensia® CONTOUR® Diabetes Care System, K023657
Device Description:The Ascensia® CONTOUR® Blood Glucose Monitoring Systemis used for the measurement of glucose in whole blood. Thesystem contains a blood glucose meter, a bottle of strips, abottle of normal control solution, a lancing device and lancetsand instructions for use.

{1}------------------------------------------------

Ascensia Page 2 of 2

.

Intended Use:The Shogun Blood Glucose Monitoring System is used for the measurement of glucose in whole blood. The Shogun Blood Glucose Monitoring System is an over-the-counter (OTC) device used by persons with diabetes and by healthcare professionals in home settings and in healthcare facilities. The Shogun Blood Glucose Monitoring System is indicated for use with capillary, venous, and arterial whole blood samples and neonatal blood samples. Capillary samples may be drawn from the fingertip, palm, forearm, and in the case of neonates, the heel. The frequent monitoring of blood glucose is an adjunct to the care of persons with diabetes.
Technological Characteristics:There were no changes to the fundamental scientific technology.
Comparison to Predicate device:The modifications to the device encompass meter design changes, software changes, and labeling changes. There has been no change to the intended use, operating principle, or functionality of the device.
Assessment of Performance:An evaluation of the Ascensia ® CONTOUR ® Blood Glucose Monitoring System was studied in the laboratory and in a clinical setting by persons with diabetes. The results were compared to results from the original Ascensia ® CONTOUR ® Blood Glucose Monitoring System and to a laboratory method. The studies showed equivalent performance with the original Ascensia ® CONTOUR ® Blood Glucose Monitoring System.
Conclusion:The results of the laboratory and clinical evaluations of the Shogun Blood Glucose Monitoring System demonstrated that the device produces blood glucose results that are substantially equivalent to results obtained on the predicate device. Therefore, the system with the modified meter is as safe and effective as the original system.

:

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services, USA. The seal features a stylized eagle with its wings spread, clutching a staff with a snake wrapped around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

SEP 1 1 2006

Mr. Roger Sonnenburg Bayer Healthcare LLC Diabetes Care Division 430 South Beiger Street Mishawaka, IN 46544

Re: K062058

Trade/Device Name: Ascensia® CONTOUR ® Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW, LFR Dated: July 19, 2006 Received: July 20, 2006

Dear Mr. Roger Sonnenburg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Alberto Gu

Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K062058

Device Name: Ascensia CONTOUR® Blood Glucose Monitoring System

Indications For Use: The Ascensia® CONTOUR®Blood Glucose Monitoring System is used for the measurement of glucose in whole blood. The Ascensia CONTOUR® Blood Glucose Monitoring System is an over-the-counter (OTC) device used by persons with diabetes and by healthcare professionals in home settings and in healthcare facilities.

The Ascensia CONTOUR® Blood Glucose Monitoring System is indicated for use with capillary, venous, and arterial whole blood samples and neonatal blood samples. Capillary samples may be drawn from the fingertip, palm, forearm, and in the case of neonates, the heel.

The frequent monitoring of blood glucose is an adjunct to the care of persons with diabetes.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)

Concurrence of CDRH, Office of Device Evaluation (OIVD)

Carol Benson

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

Page 1 of

K062058

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.