(51 days)
The U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer are intended for the intermittent measurement and monitoring of human body temperature from ear canal. The device is indicated for use by people of all ages in the home.
The U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer are characterized by measuring human body temperature in the ear canal. It utilizes infrared technology to measure either infrared energy emitted from the eardrum and surrounding tissues when making a temperature measurement.
Here's a breakdown of the acceptance criteria and study information for the U-RIGHT TD-1115 and TD-1118 ear thermometers, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document does not explicitly state numerical acceptance criteria in a table format for the device's accuracy. Instead, it refers to "system accuracy performance" and concludes substantial equivalence based on this.
| Metric | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| System Accuracy Performance (Ear Thermometer) | Not explicitly stated as a numerical value, but implied to be equivalent to the predicate device. | "demonstrated that the U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer and the currently marketed Fora ComfortScan Ear Thermometer, TD-1261B (cleared under K081445) are substantially equivalent." |
2. Sample Size Used for the Test Set and Data Provenance
The document does not provide details on the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It only mentions a "comparison of system accuracy performance."
3. Number of Experts Used to Establish Ground Truth and Their Qualifications
This information is not available in the provided text. The study described focuses on comparison to a predicate device, not on establishing ground truth via expert consensus.
4. Adjudication Method for the Test Set
The document does not mention any adjudication method for a test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No multi-reader multi-case (MRMC) comparative effectiveness study is mentioned. The study performed aims to demonstrate substantial equivalence to a predicate device.
6. Standalone Performance Study
A standalone performance study was done, but the details are minimal. The document states:
- "A comparison of system accuracy performance demonstrated that the U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer and the currently marketed Fora ComfortScan Ear Thermometer, TD-1261B (cleared under K081445) are substantially equivalent."
- "Software verification and validation, performance and safety tests confirmed that the performance, safety and effectiveness of the U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer are equivalent to the predicate device."
This indicates that internal performance and safety tests were conducted on the device itself, likely comparing its output against a reference or the predicate.
7. Type of Ground Truth Used
The document does not explicitly state the type of ground truth used. However, given the nature of a thermometer and the comparison to a predicate device (Fora ComfortScan Ear Thermometer, model TD-1261B), the "ground truth" for the performance comparison would likely be the temperature readings from the predicate device or a clinical standard reference thermometer during the testing. This is inferred, not explicitly stated.
8. Sample Size for the Training Set
The document does not mention a training set, as the device is a medical measurement instrument (thermometer), not an AI/machine learning model that typically requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no mention of a training set for this device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for TaiDoc Technology Corp. The logo features the company name in bold, black font, with three heart symbols underneath the "TaiDoc" portion. Below the company name, there is a tagline that reads "A Good Future of Health for People" in a smaller, italicized font.
http://www.taidoc.com 6F, No. 127, Wugong 2nd Rd., Wugu, Taipei County 248, tel: +886-2-66258188 fax: +886-2-66
Attachment 1. 510(k) Summary
510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
MAY 2,6 2010
The Assigned 510(k) number is:
- l. Submitter's Identification: TaiDoc Technology Corporation 3F, 5F, No.127, Wugong 2nd Rd., Wugu Township, Taipei County, 248, Taiwan
Correspondence: Nicky Pan Regulatory Affairs Specialist Tel: +886-2-6625-8188 #1196 Fax: +886-2-6625-0288 Email: nicky@taidoc.com.tw
Date of submission: April 02, 2010
-
- Device name:
Proprietary name: U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer
- Device name:
Regulatory information:
- A. Regulation section: 21 CFR 880.2910
- B. Classification: Class II
- FLL, Clinical electronic thermometer C. Product Code:
- Panel: General Hospital (80) D.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for TaiDoc Technology Corp. The logo features the company name in bold, black font, with three heart symbols underneath the first word, TaiDoc. Below the company name and heart symbols is the tagline, "A Good Future of Health for People," written in a cursive font.
-
- Intended Use:
The U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer are intended for the intermittent measurement and monitoring of human body temperature from ear canal. The device is indicated for use by people of all ages in the home.
- Intended Use:
-
- Device Description:
The U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer are characterized by measuring human body temperature in the ear canal. It utilizes infrared technology to measure either infrared energy emitted from the eardrum and surrounding tissues when making a temperature measurement.
- Device Description:
Substantial Equivalence Information:
A. Predicate device name:
Fora ComfortScan Ear Thermometer, model TD-1261B
- B. Predicate K number: K081445
C. Comparison with predicate:
The modified U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer have the following similarities to the predicate device:
- I same operating principle,
- same fundamental scientific technology.
- . incorporate the same basic circuit design,
- 트 incorporate the same materials,
- . same shelf life
- packaged using the same materials, and
- Manufactured by the same process. .
{2}------------------------------------------------
Image /page/2/Figure/1 description: The image shows the logo for TaiDoc Technology Corp. The logo features the company name in bold, black font. Below the name are three black heart symbols, followed by the tagline "A Good Future of Health for People" in a cursive font.
The modifications encompass:
- a modification in the user interface
- modification in the physical appearance
- . labeling change due to the modifications
- . Removed the data transmission function
-
- Test Principle:
The ear thermometer measures temperature by reading infrared radiation emitting from eardrum tissue. The small con-shape end of the thermometer is inserted into the ear canal, where the eardrum (tympanic membrane) and surrounding tissues give off heat. The thermometer converts it into a temperature value.
-
- Performance Characteristics:
U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer have the same performance characteristics as the predicate device.
- Performance Characteristics:
A comparison of system accuracy performance demonstrated that the U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer and the currently marketed Fora ComfortScan Ear Thermometer, TD-1261B (cleared under K081445) are substantially equivalent.
Software verification and validation, performance and safety tests confirmed that the performance, safety and effectiveness of the U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer are equivalent to the predicate device.
-
- Conclusion:
Based on the information provided in this submission, the U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer are substantially equivalent to the predicate FORA ComfortScan Ear Thermometer, model TD-1261B.
- Conclusion:
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is a stylized image of a bird or abstract human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Nicky Pan Specialist of Regulatory Affairs Taidoc Technology Corporation 3F, 5F, No. 127 Wugong 2nd Road, Wugu Township Taipei County China (Taiwan) 24888
MAY 2 6 2010
Re: K100942
Trade/Device Name: U-RIGHT TD-1115 Ear Thermometer and U-RIGHT TD-1118 Ear Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: April 27, 2010 Received: April 29, 2010
Dear Mr. Pan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page Mr. Pan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Susan Runyon
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Section 8. Indications for Use
Indications for Use
510(k) Number:
Device Name: U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer
Indications for Use:
.
The U-RIGHT TD-1115 ear thermometer and U-RIGHT TD-1118 ear thermometer are intended for the intermittent measurement and monitoring of human body temperature from ear canal. The devices are indicated for use by people of all ages in the home.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart D)
Over the Counter Use _________________________________________________________________________________________________________________________________________________________ And/Or (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) uns WARN
Division Sign-Off Office of Device Evaluation (ODE) 510(k)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
§ 880.2910 Clinical electronic thermometer.
(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.