K Number
K100500

Validate with FDA (Live)

Date Cleared
2010-03-09

(15 days)

Product Code
Regulation Number
886.1850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Keeler PSL Classic Slit Lamp is an AC-powered Slit lamp biomicroscope. It is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior eye segment.

Device Description

The Keeler Portable Slit Lamp comprises a rechargeable hand held portable illuminated biomicroscope system and a desk mounted base charger unit that is powered from a low voltage (12V) power supply.

The hand held unit incorporates a built-in rechargeable battery powering the illumination system. The illumination system and fixation targets are activated using a single/double click trigger located on the front of the grip/handle. To increase or reduce the light intensity there is a rheostat located below the eyepieces on the rear of the grip /handle.

The 10 x and 16 x magnifications optical system is controlled using the lever located under the adjustable eyepieces.

AI/ML Overview

The Keeler PSL Classic Slit Lamp is a medical device designed for non-invasive illumination, magnification, and observation of the human eye. It is used to examine the anterior and posterior segments of the eye, aiding in the diagnosis of various eye conditions.

Here's an analysis of its acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The provided 510(k) summary focuses on demonstrating substantial equivalence to a predicate device rather than detailing specific numerical acceptance criteria for performance metrics. However, based on the Summary of Performance Testing and the regulatory context of a slit lamp, the implicit acceptance criteria are related to safety, effectiveness, and functional equivalence to the predicate device.

Acceptance Criteria CategorySpecific Criteria (Implicitly based on document)Reported Device Performance (Summary)
SafetyCompliance with Optical Radiation Safety StandardsMet acceptance criteria
Compliance with Electrical Safety StandardsMet acceptance criteria
Compliance with Electromagnetic Compatibility (EMC) StandardsMet acceptance criteria
Reduced risk of physical injuryRisk reduced as much as practicably possible
Effectiveness/FunctionalityOperating range and usabilityMet acceptance criteria
Performance in comparative study with predicate deviceJudged to be as safe, as effective, and performs as well as predicate devices
Clinical PerformancePerformance in field trial in clinical environmentMet acceptance criteria and usability
Quality System ComplianceISO 9001:2000 and BS EN 13485:2003 complianceKeeler is a compliant company

Note: The 510(k) summary does not provide specific quantitative metrics or thresholds for "acceptance criteria" (e.g., minimum luminescence, maximum scatter, specific dimensions of the slit). Instead, it relies on demonstrating compliance with recognized safety standards and functional equivalence to a legally marketed predicate device.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for any specific "test set" in the context of device performance. It mentions "units under test" for the safety evaluations and "field trial in clinical environment" for performance, but no numbers are provided for either.

The data provenance is not specified regarding country of origin, nor is it explicitly stated whether the clinical data was retrospective or prospective. Given the nature of a 510(k) summary for a slit lamp, any clinical "field trial" would likely be a limited prospective observation to confirm usability and basic functionality rather than a large-scale, long-term clinical trial.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided. The relevant "ground truth" for a slit lamp device would typically involve clinical observations and diagnoses made by healthcare professionals. While a "field trial in clinical environment" is mentioned, there's no detail on how many experts were involved or their qualifications.

4. Adjudication Method for the Test Set

The document does not describe any adjudication method. This type of detail is typically not included in a 510(k) for a device like a slit lamp, especially when demonstrating substantial equivalence. The focus is more on demonstrating that the device functions correctly and safely, rather than on diagnostic accuracy studies requiring expert adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned or indicated in the provided 510(k) summary. These types of studies are more common for diagnostic imaging devices where human interpretation interacts with AI algorithms. The "comparative study with its predicative devices" mentioned seems to refer to a technical and functional comparison rather than a human reader study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable as the Keeler PSL Classic Slit Lamp is a physical optical instrument, not an AI algorithm. Its performance is always "standalone" in the sense that it is the direct observation tool, with the human eye being the "reader."

7. The Type of Ground Truth Used

For the safety and technical performance tests (Optical Radiation Safety, Electrical Safety, EMC), the "ground truth" would be established by compliance with recognized international and national standards (e.g., IEC standards for medical electrical equipment).

For the "comparative study with its predicative devices" and "field trial in clinical environment," the "ground truth" would implicitly be the established clinical utility and performance of existing slit lamps, particularly the predicate KOWA SL-15, as interpreted by healthcare professionals using the device for its intended purpose. The document states its performance was "judged to be as safe, as effective and performs as well to predicate devices," implying informal clinical observation and feedback.

8. The Sample Size for the Training Set

This is not applicable. The Keeler PSL Classic Slit Lamp is an optical-mechanical device, not a machine learning or AI system that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

This is not applicable, as there is no "training set" for this type of device.

{0}------------------------------------------------

K100500

MAR - 9 2010

510(k) Summary

ﻲ ﺃ

Submitter's information

The submitter of this pre-market notification is:

Mr. Eugene VanArsdale (Marketing Manager). Keeler Instruments Inc 456, Parkway, Broomall, PA 19008 100/1 antituty, Broomail, FT-10
Company Phone No:
Company Fax No: (610) 353 4350 (610) 353 7814

Date summary prepared: January 04, 2010

Device Identification

Device Trade Name:Keeler PSL Classic Slit Lamp
Common Name:Portable Slit Lamp
Class:II
Classification Panel:86
Product Code:HJO
Regulation Number:886.1850

Device description

The Keeler Portable Slit Lamp comprises a rechargeable hand held portable illuminated biomicroscope system and a desk mounted base charger unit that is powered from a low voltage (12V) power supply.

The hand held unit incorporates a built-in rechargeable battery powering the illumination system. The illumination system and fixation targets are activated using a single/double click trigger located on the front of the grip/handle. To increase or reduce the light intensity there is a rheostat located below the eyepieces on the rear of the grip /handle.

The 10 x and 16 x magnifications optical system is controlled using the lever located under the adjustable eyepieces.

Basic Principle of operation

The illumination system and fixation targets are activated using a single/double click trigger located on the front of the grip/handle. The construction of this slit lamp is such that light from a 15W filament lamp is relayed to the corneal plane via a lens and prism, the corneal plane being situated at 50mm from the output face of the prism. Filters or slits can be placed in the beam to vary the output in terms of colour and illuminated shape. Pupillary distance between the eye pieces, dioptric power of each eye piece and magnification can be adjusted as per requirement.

This instrument is used for non-invasive illumination, magnification and observation of the human eye.

Intended use

The slit lamp is an instrument consisting of a light source that can be focused to shine a thin sheet (slit) of light into the eye. It is used in conjunction with a biomicroscope. The lamp facilitates an examination of the anterior segment, or frontal structures and posterior segment, of the human eye, which includes the eyelid, sclera, conjunctiva, iris, natural crystalline lens, and cornea. The binocular slit-lamp examination provides stereoscopic magnified view of the eve structures in

{1}------------------------------------------------

detail, enabling anatomical diagnoses to be made for a variety of eye conditions. The Slit Lamp should be used only by suitably trained and authorised healthcare personnel.

The device and accessories are indicated as noninvasive aid in the examination and diagnosis of eye conditions and indications are the same as those claimed for predicate device

Substantial equivalence

ﻟﺔ ﺍﻟ

Predicate Device: KOWA SL-15 510(k) No: K063640 Manufacturer: KOWA Company Ltd.

The Kowa Portable Slit Lamp SL-15 is chosen as a substantially equivalent device, with its similar intended use and its classification as a slit lamp. The predicate device is a hand-held slit lamp and is equipped with rechargeable battery. Similar to predicate device, Keeler Portable Slit Lamp is equipped with a rechargeable battery.

The Keeler Portable slit lamp has following similarities to those of the predicate device:

-has the same indication for use,

-uses the same operating principle,

-incorporates the same basic optical design.

Technological characteristics

The technological characteristics are the same or similar to those found with the predicate device where eye is examined by projecting light onto it.

Declaration of device design

The Keeler Portable Slit Lamp has been designed using latest methods of evaluation and validation to ensure the risk of physical injury whilst performing the intended functions has been reduced as much is practicably possible.

Keeler Quality Approvals

:

Keeler is an ISO 9001:2000 and BS EN 13485:2003 compliant company and is obliged to ensure that our operating and design practices fully comply. As part of the design process, all necessary risk analysis, risk management, design verification and validation are conducted to ensure that the safety and effectiveness of product designs either meet or surpass the product specification and comply with the relevant standards.

Summary of Performance Testing

The performance of Keeler PSL Classic has been evaluated in Optical Radiation Safety Test, Electrical Safety Test, Electromagnetic Compatibility Test, comparative study with its predicative devices and field trial in clinical environment. The units under test met the acceptance criteria of the operating range and usability.

Based on the results of performance testing, the performance of Keeler PSL Classic is judged to be as safe, as effective and performs as well to predicate devices.

{2}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

MAR ~ 9 2010

Keeler Instruments, Inc. c/o Mr. Jeff Rongero Senior Project Engineer Underwriters Laboratories, Inc. 12 Laboratory Dr. Research Triangle, NC 27709

Re: K100500

Trade Name: Keeler PSL Classic Slit Lamp Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slitlamp biomicroscope Regulatory Class: Class II Product Code: HJO Dated: February 19, 2010 Received: February 22, 2010

Dear Mr. Rongero:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 - Mr. Jeff Rongero

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Kesia Alexander for

Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K100500

Device Name: Keeler PSL Classic Slit Lamp

Indications For Use:

Keeler PSL Classic Slit Lamp is an AC-powered Slit lamp biomicroscope. It is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior eye segment.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Dege

3/5/2010

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K100500

Page 1 of 1

§ 886.1850 AC-powered slitlamp biomicroscope.

(a)
Identification. An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.(b)
Classification. Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.