K Number
K063640

Validate with FDA (Live)

Device Name
KOWA SL-15
Manufacturer
Date Cleared
2006-12-18

(11 days)

Product Code
Regulation Number
886.1850
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

KOWA SL-15 is an ophthalmic device indicated for non-invasive illumination, magnification and observation of the human eye. It consists of a hand-held, battery powered slit-lamp biomicroscope with viewing and illumination optical systems and an AC-powered stand.

Device Description

KOWA SL-15 is an ophthalmic device indicated for non-invasive illumination, magnification and observation of the human eye. It consists of a hand-held, battery powered slit-lamp biomicroscope with viewing and illumination optical systems and an AC-powered stand.

AI/ML Overview

The provided document is a 510(k) premarket notification for the Kowa SL-15 Slit Lamp. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than presenting a clinical study with acceptance criteria, sample sizes, and ground truth establishment in the way typically seen for novel AI/medical device performance claims.

Therefore, many of the requested elements for describing "the acceptance criteria and the study that proves the device meets the acceptance criteria" are not found in this document. This submission primarily relies on engineering and design comparisons to an existing, legally marketed device (the Kowa Portable Slit Lamp SL-14).

Here's an analysis based on the provided text, addressing the points where information is available or noting where it is not applicable for this type of submission:


Summary of Acceptance Criteria and Device Performance (Based on Substantial Equivalence)

The acceptance criteria for the KOWA SL-15 are implicitly set by its substantial equivalence to the predicate device, the KOWA Portable Slit Lamp SL-14. This means the device must perform similarly and be as safe and effective as the predicate. The "study" is a comparative analysis of specifications and design changes.

1. A table of acceptance criteria and the reported device performance

Since this is a substantial equivalence submission for a traditional medical device (not an AI algorithm), the "acceptance criteria" are the specifications and performance characteristics of the predicate device. The "reported device performance" is the KOWA SL-15's specifications, which are deemed "substantially equivalent" for the majority of features.

Feature / Acceptance Criteria (from Predicate SL-14)Reported Device Performance (KOWA SL-15)Met/Not Applicable (NA)Notes
Indications For Use: Non-invasive illumination, magnification, and observation of the human eye.SameMetExplicitly stated as "Same".
Microscope Angle of convergence: 13 degreeSameMetExplicitly stated as "Same".
Magnification: 10X/16XSameMetExplicitly stated as "Same".
Reticles: Built-in both ocularsSameMetExplicitly stated as "Same".
Magnification change: 16X and 10X changed through movement of objectives.SameMetExplicitly stated as "Same".
Range of interpupillary distance adjustment: 50mm to 70mmSameMetExplicitly stated as "Same".
Range of examiner's dioptric protection adjustment: -8D to +4DSameMetExplicitly stated as "Same".
Slit projection unit Slit selection: TurretSameMetExplicitly stated as "Same".
Slit length: 12mm FixedSameMetExplicitly stated as "Same".
Slit width: 0.1, 0.2, 0.8mm and φ12 spotSameMetExplicitly stated as "Same".
Light Intensity selection: Three selection: Full, 1/2 and 1/4Three selection: Full, 1/4 and 1/16Partially MetModification: Light intensity options differ, but still offers three selections. Likely considered equivalent in function.
Slit projection angle: 60 degree for HorizontalSameMetExplicitly stated as "Same".
Filter: Built-in blue filterSameMetExplicitly stated as "Same".
Light source: 7.2V, 15W Halogen lamp7.5V, 15W Halogen lampPartially MetModification: Small voltage change for the halogen lamp.
Duration of illumination: Approximately 20 minutesApproximately 40 minutesMet (Improved)Modification: Improved battery life. This is a positive change.
Electric rating: DC 9V, 15WDC 8.4V, 15W (Lithium-ion rechargeable battery)Partially MetModification: Change due to new battery type.
Power unit Type and electric rating: Charger / Slit lamp standSameMetExplicitly stated as "Same".
Input voltage: AC 100/117/220/240V, 50/60HzAC 100-240V, 50/60HzPartially MetModification: Wider input voltage range, indicating improved compatibility.
Power consumption: 70VA90VAPartially MetModification: Increased power consumption. This likely relates to the new charging circuitry for the Li-ion battery or the new power unit.
EMC compliance with safety standards: -IEC60601-1-2: 2001NANew: The predicate did not list this, indicating an update to safety standards for the new device.
Safety compliance with safety standards: -IEC60601-1: 1988NANew: The predicate did not list this, indicating an update to safety standards for the new device.
Microscope Size: 95mm(W)x214mm(D)x213mm(H)221mm(W)x106mm(D)x214mm(H)Partially MetModification: Significant change in dimensions. While different, it is for a hand-held device and likely does not affect fundamental safety/effectiveness for its intended use.
Microscope Weight: 900g790gMet (Improved)Modification: Lighter weight. This is a positive change.
Power unit Size: Type A: 190mm(W)x250mm(D)x125mm(H), Type B: 150mm(W)x250mm(D)x97mm(H)124mm(W)x253mm(D)x68mm(H)Partially MetModification: Different dimensions for the power unit.
Power unit Weight: Type A: 3.5kg, Type B: 2.5kg1.25kgMet (Improved)Modification: Lighter power unit. This is a positive change.

Key modifications identified:

  • Change to Lithium-ion rechargeable battery.
  • Modification of charging circuit for Lithium-ion battery.
  • Modification of power unit to switching regulator.
  • Change of lamp bulb from 7.2V/15W to 7.5V/15W Halogen lamp.
  • Improved illumination duration (40 min vs 20 min).
  • Lighter microscope and power unit.
  • Updated compliance with EMC and Safety standards.

2. Sample sized used for the test set and the data provenance

  • Sample Size for Test Set: Not applicable. This submission relies on engineering comparison and compliance with standards rather than a clinical test set of patient data with a specific sample size.
  • Data Provenance: Not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. There was no clinical "test set" and thus no ground truth established by experts in this context. The substantial equivalence argument rests on the predicate device's established safety and effectiveness.

4. Adjudication method for the test set

  • Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not an AI device and no MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable. The "ground truth" here is the long-established safety and effectiveness of the predicate device (KOWA Portable Slit Lamp SL-14) based on its market history and regulatory clearance. The KOWA SL-15 is demonstrated to be substantially equivalent through design and specification comparison.

8. The sample size for the training set

  • Not applicable. This is not an AI device that requires a training set.

9. How the ground truth for the training set was established

  • Not applicable.

{0}------------------------------------------------

510(k) Notification

063640

9. Certification 9.1 Summary for public disclosure

Submitter information:

Applicant:Kowa Company, Ltd.4-14, Nihonbashi-honcho 3-ChomeChuo-ku, Tokyo, 103-8433 JapanPhone: +81-3-3279-7329Fax: +81-3-3245-1109DEC 1 8 2006
Contact:E-mail address:Satohiko Takanashi, PEs-takana@kowa.co.jp
Date summary prepared:December 6, 2006

Device identification:

Device trade name:KOWA SL-15
Classification name:Biomicroscope, Slit lamp, AC-powered
Product code:HJO

Intended use:

KOWA SL-15 is an ophthalmic device indicated for non-invasive illumination, magnification and observation of the human eye. It consists of a hand-held, battery powered slit-lamp biomicroscope with viewing and illumination optical systems and an AC-powered stand.

Comparison:

The KOWA Portable Slit Lamp SL-14 was chosen as a substantially equivalent device. The predicate device is a hand-held slit lamp and is equipped with rechargeable battery. Similar to the predicate device the KOWA SL-15 is equipped with a rechargeable battery.

The KOWA SL-15 has the following similarities to those of the predicate device:

  • has the same indicated use,
  • uses the same operating principle,
  • incorporates the same basic optical design,
  • incorporates the similar materials,
  • incorporates the similar light source.

The only modifications that were made are:

  • Change the battery to Lithium-ion rechargeable battery from.
    Kowa Company, Ltd., Modification of KOWA Portable Slit Lamp SL-14

{1}------------------------------------------------

  • Modify the charging circuit for Lithium-ion battery,
  • Modify the power unit to switching regulator.
  • Modify lamp bulb to 7.5V/15W Halogen lamp from 7.2V/15W Halogen lamp.

The comparison tables of KOWA SL-15 and KOWA Portable Slit Lamp SL-14 are shown in Table 4-1 and 4-2.

In summary, the KOWA SL-15 described in this submission is, in our opinion, substantially equivalent to the predicate device.

Conclusion:

KOWA SL-15 is equipped with the same fundamental technology features equivalent to the predicate devices, and also delivers the equivalent level safety and effectiveness. Thus it is concluded that there is no significant difference in the basic functions, safety and effectiveness between KOWA SL-15 and the predicate device.

Predicate DeviceManufacturer510(k) No.Date Cleared
KOWA Portable Slit LampSL-14Kowa Company, Ltd.K954782Nov.24, 1995

Table A: Predicate device

Kowa Company, Ltd., Modification of KOWA Portable Slit Lamp SL-14

{2}------------------------------------------------

:

KOWA SL-15KOWA Portable Slit Lamp SL-14
IndicationsFor UseSameKOWA Portable Slit Lamp SL-14is an ophthalmic device indicatedfor non-invasive illumination,magnification and observation ofthe human eye. It consists of ahand-held, battery powered slit-lamp biomicroscope with viewingand illumination optical systemsand an AC-powered stand.
Microscope
Angle ofconvergenceSame13 degree
MagnificationSame10X/16X
RetilesSameBuilt-in both oculars
MagnificationchangeSame16X and 10X Magnification ischanged through movement ofobjectives to two positions.
Range ofinterpupillarydistance adjustmentSame50mm to 70mm
Range of examiner'sdioptric protectionadjustment.Same-8D to +4D
KOWA SL-15KOWA Portable Slit Lamp SL-14
Slit projection unit
Slit selectionSameTurret
Slit lengthSame12mm Fixed
Slit widthSame0.1, 0.2, 0.8mm and φ12 spot
Light IntensityselectionThree selection:Full, 1/4 and 1/16Three selection:Full, 1/2 and 1/4
Slit projection angleSame60 degree for Horizontal
FilterSameBuilt-in blue filter
Light source7.5V, 15W Halogen lamp7.2V, 15W Halogen lamp
Duration ofilluminationApproximately 40 minutesApproximately 20 minutes
Electric ratingDC 8.4V, 15W(Lithium-ion rechargeable battery)DC 9V, 15W
Dimensions
Power unit
Type and electricratingSameCharger / Slit lamp stand
Input voltageAC 100-240V, 50/60HzAC 100/117/220/240VSelected 50/60Hz
Power consumption90VA70VA
Compliance with safety standards
EMCIEC60601-1-2: 2001-
SafetyIEC60601-1: 1988-
Dimension
Microscope
Size221mm(W)x106mm(D)x214mm(H)95mm(W)x214mm(D)x213mm(H)
Weight790g900g
Power unit
Size124mm(W)x253mm(D)x68mm(H)Type A:190mm(W)x250mm(D)x125mm(H)Type B:150mm(W)x250mm(D)x97mm(H)
Weight1.25kgType A: 3.5kgType B: 2.5kg

Table B Predicate device comparison

.

:

{3}------------------------------------------------

:

Table C Predicate Device Comparison

: :

24

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three tail feathers, representing the department's commitment to health, services, and people. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

KOWA Company, Ltd. c/o Satohiko Takanashi, PE, Chief Analyst 4-14. Nihonbashi-Honcho 3-chome Chuo-ku, Tokyo, 103-8433 Japan

Re: K063640

DEC 1 8 2006

Trade/Device Name: Kowa SL-15 Slit Lamp Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slit lamp biomicroscope Regulatory Class: Class II Product Code: HJO Dated: December 6, 2006 Received: December 7, 2006

Dear Mr. Takanashi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

M.B. Eychlemu SimMD

Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Notification

Indications for Use

510(k) Number (if know):

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications for Use:

KOWA SL-15 is an ophthalmic device indicated for non-invasive illumination, magnification and observation of the human eye. It consists of a hand-held, battery maginneation and 00001 croscope with viewing and illumination optical systems and an AC-powered stand.

Over-The-Counter Use Prescription Use (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) AND/OR

(PLEASE DO WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device

Office of Device

Dexyhuhr 12/15/06

vision Sian-Off vision of Ophthalmic Ear, Throat Devis

510(k) Number _

Kowa Company, Ltd., Modification of KOWA Portable Slit Lamp SL-14

§ 886.1850 AC-powered slitlamp biomicroscope.

(a)
Identification. An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.(b)
Classification. Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.