K Number
K100490

Validate with FDA (Live)

Date Cleared
2010-03-16

(25 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Balloon dilatation of the femoral, popliteal and infra-popliteal arteries. These catheters are not designed to be used in the coronary arteries.

Device Description

The LitePAC PTA Catheter is a standard rapid exchange PTA catheter. The proximal end provides a mono luer hub connector which allows the attachment of an inflation device used to inflate and deflate the balloon through the outer lumen. The rapid exchange port is positioned at the transition area between the stiffer hypotube and the more flexible distal portion and facilitates single operator use for insertion into a 0.014" guidewire.

AI/ML Overview

This response is based on the provided text, which details a 510(k) submission for a medical device. This type of submission focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing, rather than clinical studies with human subjects or AI algorithms. Therefore, many of the requested categories related to clinical study design (sample size for test/training sets, expert ground truth, adjudication, MRMC studies, standalone performance) are not applicable to this document.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Performance Criterion)Reported Device Performance (Demonstrated by)
Visual and functional testingIn vitro testing (results demonstrated comparability to predicate devices)
Catheter body diameterIn vitro testing (results demonstrated comparability to predicate devices)
Measurement of balloon working surfaceIn vitro testing (results demonstrated comparability to predicate devices)
Inflation/deflation testingIn vitro testing (results demonstrated comparability to predicate devices)
Pull back testingIn vitro testing (results demonstrated comparability to predicate devices)
Balloon ComplianceIn vitro testing (results demonstrated comparability to predicate devices)
Average burst pressureIn vitro testing (results demonstrated comparability to predicate devices)
Tensile testing of proximal bondIn vitro testing (results demonstrated comparability to predicate devices)
Compliance with ISO 10555-1 and ISO 10555-4Testing performed in compliance with these standards (results demonstrated comparability to predicate devices)
Substantial Equivalence to Predicate DeviceComparison of technological characteristics and performance criterion, demonstrated through in vitro testing and numerous device similarities.

Study Proving Acceptance Criteria Met:

The study that proves the device meets the acceptance criteria is a non-clinical design verification and design validation testing and analysis. This involved a series of in vitro tests performed in accordance with FDA guidance documents on non-clinical testing of medical devices, as well as ISO 10555-1 and ISO 10555-4 standards.

The conclusion drawn from this testing is that "the technological characteristics and performance criteria of the LitePAC PTA catheter are comparable to the predicate devices and that it can perform in a manner equivalent to devices currently on the market for the same intended use." This demonstrates substantial equivalence, which is the primary criterion for 510(k) clearance.


Detailed Information as Requested (where applicable):

Since this is a 510(k) submission for a physical medical device (catheter) and not an AI/software device, many of the requested items related to study design for AI are not applicable.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: The document does not specify the exact number of units or samples tested for each in vitro test. It refers to "in vitro testing" without providing specific sample numbers.
  • Data Provenance: The testing was conducted by ClearStream Technologies Ltd. The data is retrospective in the sense that it's generated for submission after device design, but it's not retrospective clinical data. It is laboratory-generated test data. The country of origin for the testing itself is not explicitly stated, but the applicant's address is Moyne Upper, Enniscorthy, Co.Wexford, Ireland.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This is not applicable as the "ground truth" for a physical device's performance in in vitro testing is determined by engineering specifications, validated test methods, and measurement accuracy, not by expert human interpretation in the way it would be for medical image analysis or similar AI applications.

4. Adjudication method for the test set:

  • Not applicable. Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among human experts when establishing ground truth from subjective data. In vitro physical testing relies on objective measurements and established pass/fail criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a physical medical instrument, not an AI software intended to assist human readers. Therefore, an MRMC study comparing human reader performance with and without AI assistance was not performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is a catheter, not an algorithm.

7. The type of ground truth used:

  • The "ground truth" in this context is defined by engineering specifications, industry standards (ISO 10555-1 and ISO 10555-4), and FDA guidance documents for non-clinical testing of medical devices. The device's performance is measured against these established criteria.

8. The sample size for the training set:

  • Not applicable. There is no "training set" in the context of a physical device's non-clinical safety and performance testing. Training sets are relevant for machine learning algorithms.

9. How the ground truth for the training set was established:

  • Not applicable. As there is no training set, there is no ground truth to be established for it.

{0}------------------------------------------------

K100490 page 1 of 3

Special 510(k) Submission - ClearStream Technologies Ltd

510(k) Summary

Applicant Information

MAR 1 6 2010

Date Prepared: February 10th 2010 Submitter: ClearStream Technologies Ltd

Address: Moyne Upper, Enniscorthy, Co.Wexford, Ireland.

Establishment Registration No: 9616666

Contact Person: Fiona Ni Mhullain, RA Manager

Telephone Number: +353 (0) 53 9237111 Fax Number: + 353 (0) 53 9237100

Device Information

Trade Name: LitePAC PTA Catheter Common Name: RX PTA Catheter Classification Name: Percutaneous Catheter Classification: Class II, 21 CFR 870.1250 Product Code: DQY

Predicate Device:

ClearStream Technologies Ltd, proposes its Sleek PTA Catheter cleared through the following 510(k) number submission: K072947, as the predicate device for this submission.

Device Description:

The LitePAC PTA Catheter is a standard rapid exchange PTA catheter. The proximal end provides a mono luer hub connector which allows the

{1}------------------------------------------------

K100490 page 2 of 3

Special 510(k) Submission - ClearStream Technologies Ltd

attachment of an inflation device used to inflate and deflate the balloon through the outer lumen.

The rapid exchange port is positioned at the transition area between the stiffer hypotube and the more flexible distal portion and facilitates single operator use for insertion into a 0.014" guidewire.

Intended Use:

Balloon dilatation of the femoral, popliteal and infra-popliteal arteries. These catheters are not designed to be used in the coronary arteries.

Comparison to Predicate Device:

The ClearStream Technologies LitePAC catheter (K100490) is substantially equivalent to the predicate device The Sleek PTA Catheter (K072947).

Test Data:

The safety and effectiveness of the ClearStream LitePAC Catheter has been demonstrated through data collected from non-clinical design verification and design validation tests and analyses.

{2}------------------------------------------------

clearStream
Technologies Ltd

Special 510(k)

LitePAC

Brief Comparison Summary

To demonstrate substantial equivalence of the applicant LitePAC PTA catheter to the predicate devices, the technological characteristics and performance criterion were evaluated using in vitro testing performed as outlined below:

In Vitro Testing

Using FDA guidance documents on non-clinical testing of medical devices the following in vitro tests were performed:

  • Visual and functional testing
  • Catheter body diameter .
  • Measurement of balloon working surface
  • Inflation/deflation testing .
  • Pull back testing .
  • Balloon Compliance .
  • Average burst pressure
  • Tensile testing of proximal bond

Testing was also performed in compliance with ISO 10555-1 and ISO 10555-4.

The results from the tests demonstrate that the technological characteristics and performance criteria of the LitePAC PTA catheter are comparable to the predicate devices and that it can perform in a manner equivalent to devices currently on the market for the same intended use.

Conclusion

ClearStream Technologies Ltd believes that the data and information presented in this application, including in vitro testing and numerous device similarities support a determination of substantial equivalence, and therefore market clearance of the LitePAC PTA catheter through this 510(k) premarket notification.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

MAR 1 6 2010

ClearStream Technologies, Ltd. c/o Fiona Ni Mhullain, Regulatory Affairs Manager Moyne Upper, Enniscorthy County Wexford Ireland

Re: K100490

Trade/Device Name: LitePAC PTA Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: LIT, DOY Dated: January 16, 2010 Received: February 19, 2010

Dear Ms. Mhullain:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Duma R. Lahner

Image /page/4/Picture/8 description: The image shows a handwritten symbol that resembles a cursive letter 'f' with an additional loop at the top. The symbol is drawn in black ink on a white background. The stroke begins with a curved line extending upwards, then loops back down before continuing into the main body of the 'f'. The overall impression is that of a quick, stylized signature or notation.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

page 1 of 1

Special 510(k) Submission - ClearStream Technologies Ltd

Indications for Use

510(k) Number (if known): K100490

Device Name: LitePAC RX PTA Catheters

Indications for Use:

:

Balloon dilatation of the femoral, popliteal and infra-popliteal arteries. These catheters are not designed to be used in the coronary arteries.

Prescription Use XX Over-The-Counter Use (Part 21 CFR 801 Subpart. AND/OR (21 CFR 801 Subpart C) D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

... . . . . . . . . . . . . . . . . . . ..

Duna R. Holmes
Division Sign Off

Division Sign-Off) Division of Cardiovascular Devices

510(k) Number_k100490

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).