(420 days)
The Hydron ActiFresh 400 (lidofilcon A) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and notaphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 diopters.
Eyecare practitioners may prescribe the lens for frequent replacement wear, with cleaning, disinfection and replacement. The lens may be disinfected using either a heat, chemical or hydrogen peroxide disinfection system.
The ActiFresh 400 soft contact lens is a hemispherical shell manufactured of a high water content (73%), polymerised material of n-vinyl pyrrolidone (nVP) and methy methacrylate (MMA) and other components which yield the appearance of a lens which is designed to fit over the corneal surface of the eye. A UV filter has been incorporated into the lens material. The lens is visibility tinted with a minute amount of poly (2-hydroxyethyl methacrylate and Reactive Blue Dye #4 copolymer. The lens is designed with varying base curves which conform to the shape of the radius of the cornea and centre over the apex of the cornea to provide corrective refraction for functional conditions of the eye including myopia (nearsightedness) and hyperopia (farsightedness). The lens provides corrective power which is to correspond to the refractive power of the eye to which it is being treated. The lens is designed with a spherical lenticulated front surface and spherical bicurve back surface. The lens is manufactured by cast molding.
The provided document describes the ActiFresh™ 400 Soft (hydrophilic) Contact Lens and its application for 510(k) clearance, asserting substantial equivalence to the predicate device, Hydron Omniflex SofBlue. The document does not describe acceptance criteria in the typical sense of a table with specific thresholds for performance metrics. Instead, it presents a comparative study against a predicate device and notes the absence of adverse events and comparability of clinical findings.
Here's an analysis based on the information provided:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state acceptance criteria in a table format with numerical thresholds for clinical outcomes. Instead, it relies on demonstrating substantial equivalence to a predicate device (Hydron Omniflex SofBlue) through a combination of non-clinical (physical, chemical, mechanical properties, toxicology, microbiology, stability) and clinical studies. The "acceptance criteria" are implicitly met if the ActiFresh 400 performs comparably to the predicate device and does not raise new questions of safety or effectiveness.
Here's a summary of the comparative performance as presented, alongside the predicate device, which can be interpreted as demonstrating "meeting acceptance criteria" by being comparable:
| Property | Acceptance Criteria (Implied: Comparable to Predicate) | Reported Device Performance (ActiFresh 400) | Predicate Device Performance (Omniflex SofBlue) |
|---|---|---|---|
| Non-Clinical | |||
| USAN | lidofilcon A | lidofilcon A | lidofilcon A |
| Monomers | nVP, MMA | nVP, MMA | nVP, MMA |
| Manufacturing Process | - (Different permitted if equivalent performance) | Cast Molding | Lathe-cut |
| Diameter (mm) | 14.3 | 14.3 | 14.3 |
| Center Thickness (mm) | 0.12 | 0.12 | 0.12 |
| Base Curves (mm) | - (Within specified range) | 8.4, 8.8 | 8.10 to 9.30 |
| Refractive Index | Comparable | $1.37 \pm 0.0003$ | $1.39 \pm 0.0007$ |
| Water Content (%) | Comparable | $73 \pm 0.3$ | $69 \pm 0.5$ |
| Oxygen Permeability (Dk) | Comparable | $28 \times 10^{-11}$ | $24 \pm 2.7 \times 10^{-11}$ |
| Light Transmittance | Comparable | $94.11\pm14.68%$ | $95.24\pm3.42%$ |
| Modulus (N/mm²) | Comparable | $0.31 \pm 0.04$ | $0.52 \pm 0.065$ |
| Tensile Strength (N/mm²) | Comparable | $0.48 \pm 0.14$ | $1.01\pm0.338$ |
| Elongation at Break (%) | Comparable | $126 \pm 27$ | $213\pm63$ |
| Toxicology | Non-toxic (according to ISO standards) | All results provide reasonable assurance of non-toxicity | Not explicitly stated but assumed for predicate |
| Microbiology | Sterilizable to SAL of 10⁻⁶ | Efficacy of sterilization checked, bioburden < 10 cfu/ml | Not explicitly stated but assumed for predicate |
| Lens Stability | 8-year shelf life | Demonstrated up to 8 years | Not explicitly stated but assumed for predicate |
| Clinical | |||
| Adverse Reactions | Absence of | Absent | Not explicitly stated but implied low incident in predicate |
| Clinically Significant Slit-lamp Findings | Low incidence, comparable to predicate | Low incidence, one subject discontinued with positive findings | Comparable to predicate |
| Changes in Keratometry, Refraction, BCVA | Very low incidence | Very low incidence | Not explicitly stated but assumed low incident in predicate |
| Efficacy (Visual Acuity) | High incidence of BCVA within one line of initial | High incidence | Not explicitly stated but assumed for predicate |
| Acceptability (Completion Rate, Discomfort) | High completion rate, low incidence of discomfort | High completion rate, low incidence of symptoms of discomfort | Comparable to predicate |
| Average Daily Wear Time | - (Relevant for daily wear claim) | 12.1-12.7 hours per day | Not explicitly stated but assumed for predicate |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Clinical Study):
- Enrolled: 39 subjects (78 eyes)
- Dispensed lenses: 38 subjects (76 eyes)
- Completed full 3 months: 33 subjects (66 eyes)
- Data Provenance: The study was conducted by Hydron Ltd., a company based in Farnborough, UK ("Hawley Lane, Famborough, Hants, GU14 8EQ, U.K."). The report itself is a submission to the FDA for market clearance in the US, but the studies were likely conducted in the UK or a similar location accessible to the company. The document does not explicitly state the country of origin for the clinical data in detail, beyond the company's location.
- Retrospective or Prospective: The clinical studies described (e.g., "Clinical Evaluation of Cast-Moulded ActiFresh 400 During a 3-month Period of Daily Wear," "High Power ActiFresh 400 Study," "Comparative Clinical Evaluation") appear to be prospective clinical trials. The description "3-month, 39 subject, open-label, uncontrolled, daily wear, dispensing study" confirms this.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify human experts establishing a "ground truth" for the test set in the sense of independent review of images or data. The "ground truth" is derived directly from clinical observations and measurements by the study investigators (presumably ophthalmologists, optometrists, or trained clinical staff) during the prospective clinical trials. The qualifications of these investigators are not provided.
4. Adjudication Method for the Test Set
No explicit adjudication method (e.g., 2+1, 3+1) is mentioned. The clinical study results are reported as direct observations and measurements from the study.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study, as typically understood for AI-assisted diagnostic devices, was not done.
- This device is a contact lens, not a diagnostic imaging AI. The "study" is a clinical trial evaluating the safety and efficacy of the contact lens itself, comparing its performance to a predicate device (Omniflex SofBlue) and demonstrating substantial equivalence, rather than comparing human readers with and without AI assistance. Therefore, no effect size of human readers improving with AI is reported.
6. Standalone Performance
- Yes, a standalone performance evaluation was conducted for the device. The clinical studies (LIDF-219, LIDF-220, DISP-203) evaluate the safety, efficacy, and acceptability of the ActiFresh 400 lens itself, independent of human interpretation nuances that would be present in an AI diagnostic device. However, "standalone" in this context refers to the device's inherent physical, mechanical, and clinical performance when worn by subjects, not an algorithm's performance.
7. Type of Ground Truth Used
The ground truth for the clinical studies was established through:
- Expert Clinical Observation and Measurement: Slit-lamp findings (e.g., corneal staining, palpebral follicles), keratometry, refraction, best-corrected visual acuity (BCVA), and patient-reported symptoms (discomfort) were directly assessed by clinical investigators.
- Patient Outcomes Data: Completion rates, discontinuation reasons, and daily wearing times.
For non-clinical properties, the ground truth was via laboratory measurements using standard industry methods (e.g., Abbe refractometer for refractive index, gravimetric determination for water content, Fatt method for oxygen permeability, UV/visible spectrophotometer for light transmittance, tensile measurements for mechanical properties).
8. Sample Size for the Training Set
- Not applicable in the AI sense. This submission is for a physical medical device (contact lens), not an AI algorithm. Therefore, there is no "training set" for machine learning. The "training" in the context of device development would refer to manufacturing process optimization and material formulation.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As there is no AI algorithm, there is no ground truth established for an AI training set. The clinical and non-clinical data collected serve to validate the performance of the manufactured device.
{0}------------------------------------------------
DEC I 0 1999
510 (k) SUMMARY FOR ACTIFRESH™ 400 SOFT (hydrophillc) CONTACT LENS !
K NUMBER: K 983637
APPLICANT INFORMATION:
| Date Prepared: | 13th October 1998 |
|---|---|
| Company Name: | Hydron Ltd |
| Company Address: | Hawley LaneFamboroughHantsGU14 8EQU.K. |
| Contact Person: | Julian HollowayHead of Quality Assurance and Regulatory Affairs |
| Phone No: | 011 44 1252 554500 |
| Fax No: | 011 44 1252 544040 |
DEVICE INFORMATION:
| Regulatory Classification: | Class II - Ophthalmic devices |
|---|---|
| Trade Name: | ActiFresh™ 400 (lidofilcon A) |
| Classification Name: | Soft (Hydrophilic) Contact Lens for Daily Wear |
EQUIVALENT DEVICE:
Hydron ActiFresh 400 (lidofilcon A) Soft (Hydrophilic) Contact Lens for Daily Wear is equivalent to Hydron Ornniflex SofBlue (lidofficon A) Hydrophilic Contact Lens for Daily Wear approved by the FDA under PMA application P830047 Suppl 6, April 1991.
ActiFresh 400 (lidofilcon A) Soft (Hydrophilic) Contact Lens for Daily Wear is substantially equivalent to the indication for use of the Hydron Omnifiex SofBlue (lidofilcon A) Hydrophilic Contact Lens for Daily Wear. This lens is in Group 2 non-ionic, high water content polymers as established by the FDA and located in the Guidance Document for Daily Wear Contact Lenses, Revised Edition May 1994. The physical, and chemical properties of the ActiFresh 400 (lidofilcon A) Soft (Hydrophilic) Contact Lens for Daily Wear are similar to those of the Hydron Omniflex SofBlue (lidofilcon A) Hydrophilic Contact Lens for Daily Wear. The subject device utilises a material of the same USAN and suffix as the predicate device but incorporates a UV filter, different handling tint and different method of manufacture.
A side by side comparison of the physical, optical and mechanical properties establishes the equivalency of these two contact lens products (see Table 2.1).
{1}------------------------------------------------
| Property | ActiFresh 400 | Omniflex SofBlue |
|---|---|---|
| Material: | ||
| USAN | lidofilcon A | lidofilcon A |
| Group | 2 (non-ionic) | 2 (non-ionic) |
| Monomers | nVP, MMA | nVP, MMA |
| Manufacturing: | ||
| Process | Cast Molding | Lathe-cut |
| Diameter (mm) | 14.3 | 14.3 |
| Center Thickness (mm) | 0.12 | 0.12 |
| Base curves (mm) | 8.4, 8.8 | 8.10 to 9.30 |
| Physical Properties: | ||
| Refractive Index | $1.37 \pm 0.0003$ | $1.39 \pm 0.0007$ |
| Water content (%) | $73 \pm 0.3$ | $69 \pm 0.5$ |
| Oxygen Permeability (Dk)*at 35° | $28 \times 10^{-11}$ | $24 \pm 2.7 \times 10^{-11}$ |
| Light Transmittance(from 385 to 780nm) | $94.11\pm14.68%$ | $95.24\pm3.42%$ |
| Mechanical Properties: | ||
| Modulus (N/mm²) | $0.31 \pm 0.04$ | $0.52 \pm 0.065$ |
| Tensile strength (N/mm²) | $0.48 \pm 0.14$ | $1.01\pm0.338$ |
| Elongation at break (%) | $126 \pm 27$ | $213\pm63$ |
Table 2.1. Summary of Properties of ActiFresh 400 and the Predicate Device, Omnifies SofBlue Hydrophilic Contact Lens.
- measured on single lenses by the original Fatt method. Units: (cm² x ml x mmHg)
DESCRIPTION OF THE DEVICE:
The ActiFresh 400 soft contact lens is a hemispherical shell manufactured of a high water content (73%), polymerised material of n-vinyl pyrrolidone (nVP) and methy methacrylate (MMA) and other components which yield the appearance of a lens which is designed to fit over the corneal surface of the eye. A UV filter has been incorporated into the lens material. The lens is visibility tinted with a minute amount of poly (2-hydroxyethyl methacrylate and Reactive Blue Dye #4 copolymer. The lens is designed with varying base curves which conform to the shape of the radius of the cornea and centre over the apex of the cornea to provide corrective refraction for functional conditions of the eye including myopia (nearsightedness) and hyperopia (farsightedness). The lens provides corrective power which is to correspond to the refractive power of the eye to which it is being treated. The lens is designed with a spherical lenticulated front surface and spherical bicurve back surface. The lens is manufactured by cast molding.
INDICATIONS FOR USE:
ﺎ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤ
Device Name: Hydron ActiFresh 400 (lidofilcon A) Soft (Hydrophilic) Contact Lens for Frequent Replacement Daily Wear.
{2}------------------------------------------------
The Hydron ActiFresh 400 (lidofilcon A) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and notaphakic persons with non-diseased eyes that may exhibit refractive and/or corneal astigmatism up to 2.00 diopters.
Eyecare practitioners may prescribe the lens for frequent replacement wear, with cleaning, disinfection and replacement. The lens may be disinfected using either a heat, chemical or hvdrogen peroxide disinfection system.
The lenses are available in the spherical power range of -15.00 to +10.00 diopters.
TECHNICAL SUMMARY:
A range of tests have been carried out to support this submission. They are briefly summarised below.
-
- NON CLINICAL
Physical and Mechanical Properties
The refractive indices of the ActiFresh 400 lenses and the predicate device were measured using a calibrated Abbe refractometer. Water content was determined gravimetrically while oxygen permeability was measured using the Fatt method that employs the Clark electrode. Light transmittance was determined using a UV/visible spectrophotometer averaging the results over the range 385 to 780 nm. The mechanical properties were determined from tensile measurements made on samples cut from the lenses. The tests being made with the samples immersed in saline solution. For each test, 30 lenses were measured to obtain the quoted results shown in table 2.1.
Toxicology
Agar overlay toxicity, acute systemic toxicity and acute ocular irritation studies were carried out according to the appropriate ISO standards. All of the results provide reasonable assurance that the ActiFresh 400 lenses are non-toxic.
Microbiology
Lenses are sterilised by steam at 121°C for 15 minutes. The efficacy of the sterilisation process is checked during the validation of the autoclave using biological indicators.
Bioburden testing of pre-sterile product is performed on a weekly basis with levels below 10 cfu/ml to ensure a final SAL of 10°.
Compatibility
Compatibility testing is not required as recommended lens care products have been approved for the same lens material (lidofilcon A).
Lens Stability
Accelerated shelf life and sterility testing has demonstrated a shelf life of up to 8 years for ActiFresh 400 lenses.
{3}------------------------------------------------
2. CLINICAL
The following summaries give details of the dinical studies undertaken to date with the ActiFresh 400 lens.
:
- Clinical Evaluation of Cast-Moulded ActiFresh 400 During a 3-month Period of Daily Wear . (LIDF-219)
- High Power ActiFresh 400 Study (LIDF-220) .
- Comparative Clinical Evaluation of the Hydron ActiFresh 400 Soft Contact Lens (DISP-. 203).
STUDY OBJECTIVE
The study was designed to evaluate the safety, efficacy, and acceptability of the visibility tinted, cast molded, ActiFresh 400 lens incorporating a UV filter.
STUDY DESIGN
This was a 3-month, 39 subject, open-label, uncontrolled, daily wear, dispensing study, Subjects were equally divided between two groups, each group using a different care system: Oxysept 1-Step or Hydron Multi chemical multi-purpose system.
STUDY LENSES AND LENS CARE REGIMENS
Study lenses were provided in 8.80mm and 8.40mm base curves. 14.3mm diameter, and in powers ranging from +8.00 to -8.00D. Subjects wore the study lenses on a daily wear basis and were assigned to either a chemical or oxidative lens care disinfection system. The following regimens were used:
OXIDATIVE Allergan Oxysept 1 Step and LC65 Allergan Oxysept Saline
CHEMICAL Hydron Multi, a multi-purpose system
STUDY POPULATION
A total of 39 (26 females) subjects (78 eves) were enrolled and 38 subjects (76 eyes) were dispensed lenses. Subjects included both new and existing contact lens wearers. Thirty-three subjects (66 eyes) completed the full three months of study lens wear (Table 1). Of the 38 subjects (76 eyes) who had study lenses dispensed, 20 (40 eyes) were assigned a chemical disinfection regimen, and 18 (36 eyes) an oxidative disinfection regimen. Regimen changes during the study were allowed, if necessary: but no subjects required such a change,
RESULTS
Five subjects (10 eves) were discontinued from the study before completing three-months of lens wear (Table 8). Of the five subjects discontinued, two (4 eves) were discontinued for reasons unrelated to the study lenses, and three (6 eyes) for reasons related to the study lenses.
{4}------------------------------------------------
Of the 39 qualified subjects (78 eyes), 26 (52 eyes) had previous contact lens wearing experience and 13 (26 eyes) did not. Twenty-five (50 eyes) previous wearing experience was with hydrogel contact lenses and one (2 eyes) with rigid gas permeable (RGP) lenses.
There were significant differences between the two care system user groups with two slit-lamp variables: comeal staining and inferior palpebral follicles. More staining was recorded with the chemical users than the oxidative users. At the baseline visit 32% and 38% of the Hydron Multi and Oxysept 1-step subjects respectively showed some corneal staining. At the 3-month visit 69% of the Hydron Multi eyes and 43% of the Oxysept 1-step eyes showed some corneal staining.
There were no other notable differences in the findings arnong the two lens care regimens; consequently pooled tabulations are presented for all variables,
DISCUSSION
The safety of the study lens was demonstrated in these studies by the absence of adverse reactions, and the low incidence of clinically significant slit-lamp findings. Only one subject was discontinued with positive slit-lamp findings. The safety of the study lens is also supported by the very low incidence of changes in keratometry, refraction, and best-corrected visual acuity observed after 3-months wear.
The efficacy of the study lens was demonstrated by the high incidence of subjects whose study lens corrected visual acuity after three months of lens wear was within one line of their initial best corrected visual acuity.
The acceptability of the study lenses was demonstrated by the high completion rate and the low incidence of symptoms of discomfort. Furthermore, daily lens wearing times for those who completed the study averaged between 12.1 after adaptation and 12.7 hours per day at the final visit, even with 33% (13/39) new lens wearers in the population.
The incidence of clinically significant slit-lamp findings and symptoms of discomfort in this study was comparable to that observed in the predicate device, OmniFiex SofBlue (PMA P830047), which is lathe-cut, incorporating a different handling tint, and without a UV filter. This suggests that the basic material and design are not adversely affected by the addition of UV filter and different handling tint, or by the different manufacturing process.
CONCLUSION
The results of this study indicate that the cast molded, visibility tinted, ActiFresh 400 lens incorporating a UV filter worn on a daily wear basis is safe, effective, and acceptable when used with chemical or oxidative disinfection regimens.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized lines forming its body and wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 0 1999
Hydron, Limited C/O Julian B. Holloway Head of Quality and Regulatory Affairs Hawley Lane, Farneborough Hampshire GU14 8EQ United Kingdom
Re: K983637
Trade Name: ActiFresh™ 400 (lidofilcon A) Soft (Hydrophilic) Contact Lens for Daily Wear (Cast molded, Visibility tinted, with UV absorber) Regulatory Class: II Product Code: LPL Dated: November 3, 1999 Received: November 11, 1999
Dear Mr. Holloway:
We have reviewed your Section 5101k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2 - Mr. Julian B. Holloway
This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in a classification for your devices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its tollfree number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Afvey L. Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known): K 983637
Device Name: Hydron ActiFresh 400 (Lidofilcon A) soft (hydrophilic) Contact Lens for Daily Wear
Indications For Use:
The Hydron ActiFresh 400 (lidofilcon A) Soft (Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and notaphakic persons with non-diseased eyes that may exhibit refractive and/or comeal astigmatism up to 2.00 diopters.
Eyecare practitioners may prescribe the lens for frequent replacement wear, with cleaning, disinfection and replacement. The lens may be disinfected using either a heat, chemical or hydrogen peroxide disinfection system.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Karen Warburton
510(k) Number
(Optional Format 3-10-98)
Prescription Use
(Per 21 CFR 801.109)
(Optional Format 3-10-98)
N/A