K Number
K093822

Validate with FDA (Live)

Date Cleared
2010-01-28

(45 days)

Product Code
Regulation Number
870.1330
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AngioAssist is a set of manual instruments intended to facilitate alignment and introduction of interventional devices (i.e., guide wires and catheters) within the coronary and peripheral vasculature.

Device Description

The AngioAssist consists of the Docking Station , Teirstein Edge and AngioAssist Combination Device. The devices are provided sterile for single use.

AI/ML Overview

The provided document is a 510(k) summary for the AngioAssist device. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than conducting a full clinical study with specific acceptance criteria and detailed device performance metrics typically found in clinical trials.

Therefore, many of the requested sections below cannot be fully extracted or are not applicable based on the content of this 510(k) summary.

Here's an analysis based on the available information:

1. A table of acceptance criteria and the reported device performance

This document does not specify quantitative acceptance criteria or detailed performance metrics from a study in the way a clinical trial report would. The "performance" assessment relies on non-clinical testing to ensure the device functions safely and effectively and demonstrating substantial equivalence to predicate devices.

Acceptance Criteria (Inferred)Reported Device Performance
Device functions safelyNon-clinical testing results "provide reasonable assurance that the devices have been designed and tested to assure conformance to the requirements for the indications for use."
Device functions effectivelyNon-clinical testing results "provide reasonable assurance... conformance to the requirements for the indications for use."
BiocompatibilityBiocompatibility testing performed to verify safe function.
Substantial Equivalence to Predicate Devices"The AngioAssist instruments are substantially equivalent to the predicate devices listed above based on a comparison of the indications for use and the technological characteristics. The testing performed confirms that the AngioAssist will perform as intended."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "non-clinical testing" and "biocompatibility testing." These generally refer to bench testing, mechanical testing, and material testing, not human patient data. Therefore, there is no "sample size" in terms of patient cases, nor "country of origin" or "retrospective/prospective" data provenance in the context of a clinical study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. "Ground truth" established by expert consensus is relevant for diagnostic devices or AI algorithms analyzing complex data (e.g., medical images). This document describes a manual instrument (catheter guide wire accessory) and its evaluation through non-clinical testing.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are used in clinical studies with multiple readers or experts to resolve discrepancies in diagnoses or assessments. This is not mentioned in the context of the non-clinical testing described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a manual instrument, not an AI or imaging diagnostic device. The concept of "human readers improve with AI" does not apply.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a manual instrument for human use, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For a device like AngioAssist, "ground truth" would relate to its functional performance (e.g., does it successfully guide a wire or catheter, does it withstand certain forces, is it biocompatible). This "ground truth" is established through engineering specifications, material standards, and predefined functional tests, rather than expert consensus on patient data, pathology, or outcomes data.

8. The sample size for the training set

Not applicable. This document describes a traditional medical device (manual instrument) and its non-clinical evaluation, not an AI algorithm that requires a "training set."

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device submission.

{0}------------------------------------------------

5. 510(K) SUMMARY AS REQUIRED BY 21 CFR 807.92

ﻨﮧ

JAN 2 8 2010

510(k) Summary

510(k) Number: _ KU93822

Date PreparedAugust 26, 2009
Submitter Information
Submitter's Name/Address:Shepherd Scientific, Inc.1515 Coast WalkLa Jolla, CA 92037
Establishment Registration:TBA
Contact Person:Paul Teirstein, M.D.President(858) 554-9905 telephone(858) 459-7285 faxpteirstein@scrippsclinic.com
Device Information
Trade Name:AngioAssist
Common Name:Catheter Guide Wire Accessory
Classification Name:Wire, Guide, Catheter
Product Code:DQX
Regulation:Class II, 21 CFR 870.1330
Panel:Cardiovascular

Performance Standards

No performance standards applicable to this product have been developed under Section 514 of the Act.

Predicate DeviceManufacturer510(k) Status
Cinch QR Steerable GuidewireExtension AccessoriesCordis CorporationK963171
Ostial Pro Stent Positioning SystemOstial Solutions, LLCK062192
Torque DeviceMerit Medical Systems, Inc.K072552
WireClip TorquerBoston ScientificK003398
KEEP Accessory OrganizerMerit Medical Systems, Inc.Exempt

5.1. Device Description

The AngioAssist consists of the Docking Station , Teirstein Edge and AngioAssist Combination Device. The devices are provided sterile for single use.

{1}------------------------------------------------

5.2. Intended Use/Indications for Use

The AngioAssist is a set of manual instruments intended to facilitate alignment and introduction of interventional devices (i.e., guide wires and catheters) within the coronary and peripheral vasculature.

-1,

, ,

:

ﻧﺪ.

5.3. Summary of Non-Clinical Testing

The AngioAssist instruments underwent performance and biocompatibility testing to verify that the device functions in a safe and effective manner. The results of the tests provide reasonable assurance that the devices have been designed and tested to assure conformance to the requirements for the indications for use.

5.4. Statement of Equivalence

The AngioAssist instruments are substantially equivalent to the predicate devices listed above based on a comparison of the indications for use and the technological characteristics. The testing performed confirms that the AngioAssist will perform as intended.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a circular design with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, which is a common symbol used in government seals to represent freedom and strength.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

JAN 2 8 2010

Shepherd Scientific, Inc. c/o Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313

Re: K093822

Trade/Device Name: AngioAssist Family Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Accessory Regulatory Class: Class II (two) Product Code: DQX Dated: December 31, 2009 Received: January 6, 2010

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 -- Mr. Mark Job

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing

practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

suma R. lahner

Image /page/3/Picture/8 description: The image shows a black and white drawing of a cursive letter. The letter appears to be an 'L' or a stylized 'E'. The letter has a loop at the top and a curved line extending to the right. The drawing is simple and lacks any shading or texture.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and · Radiological Health

Enclosure

{4}------------------------------------------------

4. INDICATIONS FOR USE STATEMENT

The Indications for Use Statement is provided on the following page. Indications for Use Statement

510(k) Number (if known): _ K093822_

Device Name: AngioAssist

Indications for Use:

The AngioAssist is a set of manual instruments intended to facilitate alignment and introduction of interventional devices (i.e., guide wires and catheters) within the caronary and perinteral vasculature.

Prescription Use V (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

ourner 2. In time

(Division Sign-Off) Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Cardiovascular Devices

510(k) Number_ko93822

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.