K Number
K093795

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-03-12

(92 days)

Product Code
Regulation Number
880.2910
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Actherm Infrared Ear Thermometers ACT 8000R Series are used to measure body temperature from auditory canal. This device is intended for household and hospital use on pcople of all age, and is used with or without probe cover medACCU2010.

Device Description

Actherm Infrared Ear Thermometers (Model ACT 8000R Series) are electronic thermometers using a thermopile as the temperature sensor. The sensor's electric signal is then calculated and displayed by a Micro-Controller. These thermometers display the temperature decimal. To compare with ACT 8000 Series, Model ACT 8000R Series has the same indication for use. ACT 8000R Series can be made with various functions, such as Dual scale, Backlight, Scan LED, W/O cover, following with the model number indicates scale switchable, with Backlight, with Scan completed indicator, without probe cover, will be added following the model number. The infrared ear thermometer comprises: a thermopile for temperature sensing, a buzzer for sounding effect, a Micro Controller and a LCD display for calculating and displaying the target temperature digitally. Actherm Infrared Ear Thermometers ACT 8000R Series can be combined with accessories, base, wall mount and power adapter. The base is a tray for holding probe cover box, power input, power/charge indicator, portable/ fixed probe-cover dispenser and the area for placing digital thermometer and its probe cover. For the wall mount, it can be fixed on the wall and the base can be put on this wall mount. The system uses two 1.5V DC alkaline rechargeable batteries for the power supply and the battery power is automatically the Micro-Controller and displayed in LCD if the battery is exhausted. The batteries can be recharged only when the ear thermometer ACT8000R put on the base. ACT 8000R cannot be used when recharging.

AI/ML Overview

This document is a 510(k) summary for the Actherm Infrared Ear Thermometer ACT 8000R Series. It describes how the device meets acceptance criteria through conformance to various standards and provides limited information regarding specific study details.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Standard)Reported Device Performance
ASTM E 1965-98: 2003 (Standard Specification for Infrared Thermometer for Intermittent Determination of Patient Temperature)Fulfilled testing requirements; clinical test results conform to requirements.
EN 60601-1 (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance)Fulfilled testing requirements.
EN 60601-1-2 (Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic compatibility - Requirements and tests)Fulfilled testing requirements.
EN 12470-5 (Clinical thermometers - Part 5: Performance of infrared ear thermometers (with maximum device)Fulfilled testing requirements; clinical test results conform to requirements.
ISO 10993-1 (Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process)Requirements met.
ISO 10993-5 (Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity)Requirements met.
ISO 10993-10 (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization)Requirements met.
Electrical Safety (Performed test)Passed electrical safety performed test.

2. Sample size used for the test set and the data provenance

The document states "The clinical test results of Actherm Infrared Ear Thermometer ACT 8000 and Probe Cover medACCU 2010 conform to the requirements of ASTM E 1965-98 and EN 12470-5 that are the same performance for ACT 8000R Series." This implies that the testing was carried out on the predicate device (ACT 8000) and the results were deemed applicable to the ACT 8000R Series due to substantial equivalence.

  • Sample size for the test set: Not explicitly stated. The text refers to "clinical test results" but does not provide details on the number of subjects or measurements.
  • Data provenance: Not explicitly stated, but given that the submitter is Actherm Inc. in Taiwan, R.O.C., it is likely that the data originated from studies conducted in that region or through certified testing facilities. The study type (retrospective or prospective) is not mentioned.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The standards mentioned define performance requirements for device accuracy, but the method for establishing "ground truth" (e.g., using a reference thermometer and standard clinical protocols) and the involvement of specific experts are not detailed.

4. Adjudication method for the test set

This information is not provided in the document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This device is an infrared ear thermometer, a standalone measurement device, not an AI-assisted diagnostic tool for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance evaluation was done. The entire summary describes the device's technical specifications and its performance against established medical device standards (ASTM, EN, ISO). The device operates independently to measure temperature, meaning its performance is evaluated in a standalone manner without human subjective input affecting the measurement itself. The clinical tests mentioned (conforming to ASTM E 1965-98 and EN 12470-5) are standalone performance evaluations of the thermometer.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For a thermometer, the "ground truth" typically involves reference temperature measurements from highly accurate and calibrated devices. The ASTM E 1965-98 standard, which the device conforms to, outlines methods for clinical accuracy studies, usually involving comparing the device's readings against a known, highly accurate reference thermometer in a controlled clinical setting. The document does not explicitly state the specific type of ground truth used beyond conformity to these standards.

8. The sample size for the training set

This device is a traditional electronic thermometer, not an AI/machine learning device that requires a "training set." Therefore, the concept of a training set sample size does not apply.

9. How the ground truth for the training set was established

As mentioned above, this device does not utilize a training set in the context of machine learning.

{0}------------------------------------------------

6F, No.18, Jhanye 2ndRd., Hsinchu Science Park, Hsinchu 30078, TAIWAN, R.O.C. e - mail: actherm@actherm.com.tw

율 886 36 66 95 96 @ 886 36 66 96 97

Image /page/0/Picture/3 description: The image shows a logo with the word "Aachen" written in a stylized font. Below the logo is a graphic element that resembles a stylized wing or swoosh. To the right of the wing, there is a certification mark with the letters "TUV" and "SOD" inside an octagon, and the letters "CE" followed by the number "0123" are present to the right of the certification mark.

1093795

MAR 1 2 2010

510(k) Summary

Submitter's Identification

Actherm Inc. 6F, No.18, Jhanye 2nd Rd., Hsinchu Science Park, Hsinchu 30078, Taiwan, R.O.C. +886 3 666 9596 Tel : Fax: +886 3 666 9697

Contact Person : Richard Hsieh Date the summary was prepared : December 04, 2009

Name of the device

Trade Name – Actherm Infrared Ear Thermometer ACT 8000R Series and Its Probe Cover medACCU 2010 Common Name - Infrared Ear Thermometer Classification name - Clinical Electronic Thermometer (21CFR 880.2910, Product FLL)

Predicate Device Information

Actherm Infrared Ear Thermometer ACT 8000 Series and Its Probe Cover medACCU 2010 510 (k) number : K090031.


ન્દ્ર

{1}------------------------------------------------

6F, No.18, Jhanye 2nd Rd., Hsinchu Science Park, Hsinchu 30078, TAIWAN, R.O.C. e - mail: actherm@actherm.com.tw

Aetherium

Device Description

Actherm Infrared Ear Thermometers (Model ACT 8000R Series) are electronic thermometers using a thermopile as the temperature sensor. The sensor's electric signal is then calculated and displayed by a Micro-Controller. These thermometers display the temperature decimal. To compare with ACT 8000 Series, Model ACT 8000R Series has the same indication for use. ACT 8000R Series can be made with various functions, such as Dual scale, Backlight, Scan LED, W/O cover, following with the model number indicates scale switchable, with Backlight, with Scan completed indicator, without probe cover, will be added following the model number.

8 886 36 66 95 96

886 36 66 96 97

The infrared ear thermometer comprises: a thermopile for temperature sensing, a buzzer for sounding effect, a Micro Controller and a LCD display for calculating and displaying the target temperature digitally.

Actherm Infrared Ear Thermometers ACT 8000R Series can be combined with accessories, base, wall mount and power adapter. The base is a tray for holding probe cover box, power input, power/charge indicator, portable/ fixed probe-cover dispenser and the area for placing digital thermometer and its probe cover. For the wall mount, it can be fixed on the wall and the base can be put on this wall mount.

The system uses two 1.5V DC alkaline rechargeable batteries for the power supply and the battery power is automatically the Micro-Controller and displayed in LCD if the battery is exhausted. The batteries can be recharged only when the ear thermometer ACT8000R put on the base. ACT 8000R cannot be used when recharging.

(Labeling please see Attachment 1)

Intended Use

Actherm Infrared Ear Thermometers ACT 8000R Series have the same intended use as the predicate device. They are used to measure body temperature from auditory canal. This device is intended for household and hospital use on people of all age, and is used with or without probe cover.

{2}------------------------------------------------

6F, No.18, Jhanye 2nd Rd., Hsinchu Science Park, Hsinchu 30078, TAIWAN. R.O.C. e - mail: actherm@actherm.com.tw

Image /page/2/Picture/2 description: The image shows two phone numbers: 886 36 66 95 96 and 886 36 66 96 97. There is a phone icon next to the first number and a document icon next to the second number. The image also contains a logo and a certification mark.

Technological Characteristics

Actherm Infrared Ear Thermometers ACT 8000R Series have the same mode of operation, design principle, and biological specifications as the predicate device. ACT 8000R Series can be made with various functions, such as Dual scale, Backlight, Scan LED, W/O cover, following with the model number indicates scale switchable, with Backlight, with Scan completed indicator, without probe cover, will be added following the model number.

Comparison to 510(k) Predicate Devices

The Actherm Infrared Ear Thermometers, the ACT 8000R Series are substantially equivalent to the following digital thermometers: Infrared Ear Thermometer Model ACT 8000 Series. Its 510(k) number is K090031.

To compare with ACT 8000 Series, the Actherm Infrared Ear Thermometers ACT 8000R Series are similar in design and intended use to the predicates differing only in supplied batteries.

The ACT8000 Series use two 1.5V DC alkaline batteries and the ACT 8000R Series use two 1.5V alkaline rechargeable batteries that can be recharged only when the infrared ear thermometer ACT 8000R is put on the base. ACT 8000R cannot be used when recharging.

All products use the same temperature sensing element - a thermopile, an LCD display, Micro-Controller and a buzzer.

Substantial Equivalence

Infrared Ear Thermometers ACT 8000R Series, and the claimed device are the same in intended use, essential component; sensor, power requirement, PCB circuit, software program, biological specification, performances, design principle, dimension of probe cover, and are also close in electrical safety, and materials, etc. However, in these characters, the slight difference of measuring range, battery type, labeling, won't affect product safety and effectiveness.

{3}------------------------------------------------

r

6F, No.18, Jhanye 2nd Rd., Hsinchu Science Park, Hsinchu 30078, TAIWAN. R.O.C. e - mail: actherm@actherm.com.tw

会 886 36 66 95 96 图 886 36 66 96 97

The change is in power supply. The ACT 8000 Series use two 1.5V DC alkaline batteries and the ACT 8000R Series use two 1.5V alkaline rechargeable batteries that can be recharged only when the infrared ear thermometer ACT 8000R is put on the base is a tray for holding probe cover box, power input, power/charge indicator, portable/ fixed probe-cover dispenser and the area for placing digital thermometer and its probe cover. ACT 8000R cannot be used when recharging. It has passed electrical safety performed test, therefore, it is safe for human body.

According to the description above, the two devices are substantial equivalent.

Performance Data

Actherm Infrared Ear Thermometers ACT 8000R Series meets the ASTM Standard Specification for Infrared Thermometer for Intermittent Determination of Patient Temperature (ASTM E 1965-98: 2003), as well as EN 60601-1 (IEC 60601-1), EN 60601-1-2 (IEC 60601-1-2), EN 12470-5 and ISO 10993-1, ISO 10993-5, ISO 10993-10 requirements. For all body contacting materials, the analysis has been made and the identical materials have been used in other legally marketed devices under the same use conditions.

Safety and Effectiveness

Actherm Infrared Ear Thermometer ACT 8000 and ACT 8000R and Probe Cover medACCU 2010 fulfill the testing requirements of ASTM E 1965-98, EN 60601-1), EN 60601-1-2 (IEC 60601-1-2) and EN 12470-5. ACT 8000R just add the function of recharge, when recharging ACT 8000R, it must be put on the base and cannot take temperature. In addition, ACT 8000R Series also meets electrical Safety performed test such as EN 60601-1 (IEC 60601-1-2 (IEC 60601-1-2), so it is safe for human body. Moreover, Infrared Ear Thermometers ACT 8000R Series, and the claimed device are the same in intended use, essential component, sensor, power requirement, PCB circuit, Software program, biological specification, performances, design principle, dimension of probe and probe cover, and also close in electrical safety, and materials, etc. The clinical test results of Actherm Infrared Ear Thermometer ACT 8000 and Probe Cover medACCU 2010 conform to the requirements of ASTM E 1965-98 and EN 12470-5 that are the same performance for ACT 8000R Series. The product quality is standard-guaranteed. The compliance tests demonstrate precise measuring function and the clinical accuracy of the thermometer. Therefore, these thermometers are safety and effectiveness.

4

{4}------------------------------------------------

ACTIFICI

6F, No.18, Jhanye 2ndRd., Hsinchu Science Park, Hsinchu 30078, TAIWAN, R.O.C. e - mail: actherm@actherm.com.tw

Image /page/4/Picture/3 description: The image shows a logo with the word "Acaborn" written in a stylized font. Below the word is a horizontal line. To the right of the line is a certification mark with the letters "TUV" inside of an octagon. Next to the certification mark is the text "CE0123".

Conclusion

ﺮ ﮐ

Actherm Infrared Ear Thermometers ACT 8000R Series have intended use, principles of operation, and similar technological characteristics as predicate devices. Moreover, bench testing contained in this submission and clinical testing supplied demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the Actherm Infrared Ear Thermometers ACT 8000R Series are substantially equivalent to the predicate devices.

5

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Richard Hsieh General Manager Actherm, Incorporated 6F, No. 18 Jhanye 2nd Road Hsinchu Science Park Hsinchu China Taiwan 30078

MAR ] 2 2010

Re: K093795

Trade/Device Name: Actherm Infrared Ear Thermometer ACT 8000R Series and Its Probe Cover medACCU2010 Regulation Number: 21CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: February 4, 2010 Received: February 16, 2010

Dear Mr. Hsieh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2- Mr. Hsieh

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm 1.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

ACTHERM Inc.

6F, No.18, Jhanye 2nd Rd., Hsinchu Science Park, Hsinchu 30078, TAIWAN, R.O.C. e - mail: actherm@actherm.com.tw

☎ 886 36 66 95 96 B 886 36 66 96 97

Image /page/7/Picture/3 description: The image shows a logo with the word "Aetherium" in a stylized font. Below the word is a graphic element that appears to be a stylized representation of a bird or wing. To the right of the graphic element are two certification marks: a TÜV SÜD octagon logo and a CE mark followed by the number 0123.

Indication for Use

510(k) Number (if known) : K093795

Device Name : Actherm Infrared Ear Thermometer ACT 8000R Series and Its Probe Cover medACCU2010

Indication For Use :

Actherm Infrared Ear Thermometers ACT 8000R Series are used to measure body temperature from auditory canal. This device is intended for household and hospital use on pcople of all age, and is used with or without probe cover medACCU2010.

Prescription Use Over-The-Counrter Use ____x AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Th

(Division Sign-Q (Offision of Anesthesiology. General Hospital Infection Control, Dental Devices

Page 1 of 1

16 093 510(k) Number:

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.