K Number
K093254

Validate with FDA (Live)

Date Cleared
2010-09-15

(334 days)

Product Code
Regulation Number
886.5928
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AMO brand Multi-Purpose Disinfecting Solution is indicated for the care of soft (hydrophilic) contact lenses, including silicone hydrogel lenses. Use this product as recommended by your eye care practitioner to:

  • Chemically (NOT HEAT) Disinfect ●
  • Clean ●
  • Rinse ●
  • Store
  • Remove Protein ●
  • Condition
Device Description

AMO Brand Multi-Purpose Disinfecting Solution is a sterile and buffered aqueous solution containing disinfectants / preservatives, buffers, a surfactant, a chelating agent, other ancillary agents, and purified water. The solution is packaged in tamper-resistant multi-dose plastic bottles with controlled dropper tips.

AI/ML Overview

Here's an analysis of the provided text, outlining the acceptance criteria and the studies that prove the device meets them:

Acceptance Criteria and Device Performance for AMO Brand Multi-Purpose Disinfecting Solution

1. Table of Acceptance Criteria and Reported Device Performance

The document describes several non-clinical and clinical studies conducted to demonstrate the safety and effectiveness of the AMO Brand Multi-Purpose Disinfecting Solution. The acceptance criteria are largely implied by the successful outcomes ("meets requirements," "compatible," "equivalent," "well tolerated," "not cytotoxic," "no ocular irritation").

Acceptance Criteria CategorySpecific Acceptance Criteria (Implied/Stated)Reported Device Performance
Solution CompatibilityCompatibility with various soft contact lenses, including silicone hydrogel lenses.Demonstrated product is compatible with soft contact lenses, including silicone hydrogel lenses (Groups 1, 4, and four representative silicone hydrogel lenses). Tests primarily followed FDA Guidelines, Premarket Notification (510(k)) Guidance Documents for Contact Lens Care Products, May 1, 1997.
Protein RemovalAbility to passively and effectively remove lysozyme protein from contact lens surfaces, comparable to the predicate device.Demonstrated ability to passively and effectively remove protein from soft (hydrophilic) contact lenses, including silicone hydrogels, comparable to the predicate device.
WettabilityEquivalent wettability, conditioning, and moisturizing of soft contact lenses (including silicone hydrogels) compared to the predicate device.Demonstrated equivalence to Opti-Free® RepleniSH® in soft (hydrophilic) contact lens, including silicone hydrogels, wettability, conditioning, and moisturizing (based on contact angle measurements).
Microbiological EfficacyMeets current FDA requirements for disinfection against bacteria, yeast, and mold at low shelf specifications. Meets USP <71> sterility test requirements.Product meets current FDA requirements for disinfection of contact lenses against bacteria, yeast, and mold at low shelf specifications for the disinfectant/preservative system. Formulated with dual-disinfection technology. Meets USP <71> sterility test requirements. All tests used solution at low shelf specifications.
StabilityRemains stable for the labeled shelf-life.Accelerated testing indicates the solution will remain stable for the labeled shelf-life.
Biocompatibility- Not cytotoxic to mouse fibroblasts. - No dermal sensitization reactions. - No 1-day acute ocular irritation in rabbits. - Well-tolerated in 22-day ocular safety studies with contact lenses in rabbits. - No acute oral toxicity in rats. - Less cytotoxic with pre-soaked contact lenses compared to predicate.- Neither cytotoxic to mouse fibroblasts cells after 24 hours (for test formulation and predicate). - No dermal reactions from either test or control in sensitization study. - 1-day acute ocular irritation study in rabbits demonstrated no ocular irritation. - Treatment was well tolerated in rabbits over 22 days (with PureVision™ or Acuvue® 2 lenses). - Neither caused an adverse effect when administered to rats in a single oral dose. - Overall results showed MPDS 9608X was less cytotoxic than the predicate device (with pre-soaked contact lenses). Tests followed FDA Guidance (1997) and Pre-IDE 1080487 (2008).
Clinical Safety & PerformanceSafe and effective for soft contact lens wearers, including silicone hydrogel. Comparable to predicate in patient acceptability, lens wearing comfort, and corneal staining.Demonstrated to be safe and effective for soft (hydrophilic) contact lenses, including currently marketed silicone hydrogel lenses. Comparable to the predicate device in terms of patient acceptability, lens wearing comfort, and corneal staining (PureVision™ lenses).

2. Sample Sizes Used for the Test Set and Data Provenance

The document does not explicitly state numerical sample sizes for most non-clinical tests.

  • Solution Compatibility: Tested with "conventional Group 1 and Group 4 soft (hydrophilic) contact lenses" and "four (4) representative marketed silicone hydrogel lenses."
  • Protein Removal: Tested using AMO Brand MPDS and the predicate device.
  • Wettability: Tested using "four (4) representative silicone hydrogel contact lenses and FDA Group IV contact lenses."
  • Microbiological Studies:
    • Disinfection: Evaluation against "bacteria, yeast, and mold." The specific number of organisms or replicates is not stated.
    • Sterility: No specific sample size given for USP <71> test.
  • Biocompatibility:
    • Cytotoxicity: Mouse fibroblasts (number not specified).
    • Sensitization: Dermal reactions (number of subjects/animals not specified).
    • Ocular Irritation: Rabbits (number not specified, but typically 3-6 per group for acute irritation).
    • 22-day Ocular Safety: Rabbits (number not specified, but uses "control" and "test formulation" with two lens types: PureVision™ or Acuvue® 2).
    • Acute Oral Toxicity: Rats (number not specified).
    • Cytotoxicity with pre-soaked contact lenses: Mouse fibroblasts (number not specified).
  • Clinical Trials:
    • Multi-center clinical trial: "Soft contact lenses, including silicone hydrogel lens wearers." No specific number of participants.
    • Single-center clinical trial: Used "PureVision™ lenses at specified time points over a period of four days." No specific number of participants.

Data Provenance: The studies appear to be prospective as they were conducted to test the specific device for regulatory submission. The country of origin is not explicitly stated for the test data, but the submitter (Abbott Medical Optics) is based in Santa Ana, CA, USA, and the submission is to the US FDA, implying the studies were conducted to US regulatory standards, likely within the US or by internationally recognized laboratories.

3. Number of Experts and Qualifications for Ground Truth

This information is not provided in the document. For laboratory-based non-clinical studies, "ground truth" is typically established by laboratory standards and protocols, not expert consensus in the same way it would be for clinical image interpretation. Clinical trials would involve ophthalmologists or optometrists for evaluations (e.g., corneal staining, adverse events), but their specific qualifications and number are not detailed.

4. Adjudication Method

This information is not applicable to most of the non-clinical tests described (e.g., microbiological, chemical analysis, animal studies) where results are quantitative or based on defined criteria. For the clinical trials, the document does not specify any adjudication method (e.g., 2+1, 3+1) for participant observations or outcomes.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a MRMC comparative effectiveness study was not done. This type of study (comparing human readers with and without AI assistance) is relevant for diagnostic imaging devices. The AMO Brand Multi-Purpose Disinfecting Solution is a contact lens care product, not a diagnostic imaging device.

6. Standalone (Algorithm Only) Performance Study

No, a standalone (algorithm only) performance study was not done. This concept is also primarily relevant for AI/ML-based diagnostic devices. The device described is a chemical solution; its "performance" is assessed through its physical, chemical, and biological interactions, not by an algorithm's output.

7. Type of Ground Truth Used

The "ground truth" for the various tests was established through:

  • Laboratory Standards and Protocols:
    • Microbiological: Quantitative measurements of microbial kill (against specific bacteria, yeast, mold) as per FDA guidelines and USP <71> for sterility.
    • Chemical/Physical: Measurement of protein removal, contact angles for wettability, stability metrics.
    • Biocompatibility: Observation of cytotoxicity, dermal reactions, ocular irritation, oral toxicity in animal models against defined endpoints (e.g., cell viability, irritation scale).
  • Clinical Observations: For the clinical trials, the "ground truth" for safety and performance (e.g., patient acceptability, comfort, corneal staining, adverse events) would be based on direct patient feedback and objective examinations by trained healthcare professionals (e.g., ophthalmologists, optometrists).

8. Sample Size for the Training Set

Not applicable. The device is a chemical solution, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. There is no training set for this type of device.

{0}------------------------------------------------

510(K) SUMMARY

AMO Brand Multi-Purpose Disinfecting Solution

This summary used the format provided in 21 CFR 807.92:

(a)(1) Submitter:Kesley E. GallagherSr. RA SpecialistGlobal Regulatory AffairsAbbott Medical Optics1700 E. St. Andrew PlaceSanta Ana, CA 92705
Phone: 714-247-8930Fax: 714-247-8784E-mail: kesley.gallagher@amo.abbott.com
(a)(2) Device Trade Name:AMO Brand Multi-Purpose DisinfectingSolution
Device Common Name:Soft (Hydrophilic) Contact Lens Solution
Device Classification/Panel:Class II (Special Controls)/OphthalmicDevice
Device Classification Names:Accessories to Contact Lens Solution(86LPN)

(a)(3) Identification of Predicate Device:

AMO Brand Multi-Purpose Disinfecting Solution (MPDS 9608X) is substantially equivalent to ALCON® Opti-free® RepleniSH® Multi-purpose disinfecting solution.

(a)(4) Device Description:

AMO Brand Multi-Purpose Disinfecting Solution is a sterile and buffered aqueous solution containing disinfectants / preservatives, buffers, a surfactant, a chelating agent, other ancillary agents, and purified water. The solution is packaged in tamper-resistant multi-dose plastic bottles with controlled dropper tips.

Intended Use (Indications for Use) (a)(5)

AMO Brand Multi-Purpose Disinfecting Solution is indicated for the care of soft (hydrophilic) contact lenses, including silicone hydrogel lenses. Use this product as recommended by your eye care practitioner to:

  • Chemically (NOT HEAT) Disinfect ●
  • Clean ●

{1}------------------------------------------------

  • Rinse ●
  • Store
  • Remove Protein ●
  • Condition

(a)(6) Comparison of Technological Characteristics

The technological characteristics of the new formulation remain the same as the predicate device.

(b)(1) Discussion of Non-Clinical Testing

Solution Compatibility: AMO Brand Multi-Purpose Disinfecting Solution was tested for compatibility with conventional Group 1 and Group 4 soft (hydrophilic) contact lenses, and with four (4) representative marketed silicone hydrogel lenses. The results demonstrate the product is compatible with soft contact lenses, including silicone hydrogel lenses. The tests were primarily carried out in accordance with the FDA Guidelines, Premarket Notification (510(k)) Guidance Documents for Contact Lens Care Products, May 1, 1997.

Protein Removal: AMO Brand Multi-Purpose Disinfecting Solution and the predicate device were tested for their abilities to remove lysozvme protein on contact lens surfaces. AMO® Brand Multi-Purpose Disinfecting Solution demonstrated its ability to passively and effectively remove protein from soft (hydrophilic) contact lenses, including silicone hydrogels, which is comparable to the predicate device.

Wettability of Contact Lenses: The contact angles of water drops on contact lenses were measured using the sessile drop method. Contact lenses, including four (4) representative silicone hydrogel contact lenses and FDA Group IV contact lenses, were presoaked in MPDS 9608X, or the predicate device, Opti-Free® RepleniSH®. The results demonstrate AMO® Brand Multi-Purpose Disinfecting Solution and Opti-Free® RepleniSH® are equivalent in soft (hydrophilic) contact lens, including silicone hydrogels, wettability, conditioning, and moisturizing.

Microbiological Studies: The product was evaluated for microbiological efficacy using studies outlined in FDA's Premarket Notification (510(k)) Guidance Documents for Contact Lens Care Products, May 1, 1997, modified to include organic soil in the stand alone test and test solution formulated to the low specification of the disinfectants / preservatives.

  • The product meets current FDA requirements for disinfection of contact ● lenses against bacteria, yeast, and mold at low shelf specifications for the disinfectant / preservative system.
  • AMO brand Multi-Purpose Disinfecting Solution is formulated with dualdisinfection technology designed specifically for soft contact lenses, including silicone hydrogel lenses. Dual disinfectants work to destroy microbes on the surface of the lens and provide safe contact lens storage up to 30-days. All

{2}------------------------------------------------

tests used AMO® Brand Multi-Purpose Disinfecting Solution at the low shelf specifications for the disinfectant / preservative system.

  • . The product meets USP <71> sterility test requirements.
    Stability: Accelerated testing indicates AMO Brand Multi-Purpose Disinfecting Solution will remain stable for the labeled shelf-life.

Biocompatibility: The safety of the product was evaluated with a worst case formulation at the upper specification for the disinfectants / preservatives using studies outlined in the FDA's Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products, May 1, 1997 and Pre-IDE 1080487 Investigational Multi-Purpose Solution for Contact Lens Care, July 31, 2008.

  • . Cytotoxicity: Neither the test formulation, nor the predicate device, were cytotoxic to mouse fibroblasts cells after 24 hours.
  • Sensitization: There were no dermal reactions from either test or control.
  • Ocular irritation: 1-day acute ocular irritation study in rabbits demonstrated ● no ocular irritation.
  • 22-day Ocular Safety Studies with Contact Lenses: Treatment with either the . test formulation or control with either PureVision™ or Acuvue® 2 contact lenses was well tolerated in rabbits over the indicated period of time.
  • . Acute Oral Toxicity: Neither the test formulation or the control caused an adverse effect when administered to rats in a single oral dose.
  • . Cytotoxicity with pre-soaked contact lenses - Direct Contact. Overall results showed MPDS 9608X was less cytotoxic than the predicate device.

(b)(2) Discussion of Clinical Data:

A multi-center clinical trial was conducted using AMO Brand Multi-Purpose Disinfecting Solution. The safety and performance of the product were evaluated in soft contact lenses, including silicone hydrogel lens wearers. lt was demonstrated the product is safe and effective when using soft (hydrophilic) contact lenses, including currently marketed silicone hydrogel lenses.

An additional single-center clinical trial was conducted using AMO Brand Multi-Purpose Disinfecting Solution. The safety and performance of the product were evaluated in PureVision™ lenses at specified time points over a period of four days. It was demonstrated that the product is comparable to the predicate device in terms of patient acceptability, lens wearing comfort, and corneal staining.

(b)(3) Conclusions Drawn from Data Supporting Equivalence Determination

It is concluded that the safety, efficacy, and performance of AMO Brand Multi-Purpose Disinfecting Solution is substantially equivalent to the predicate device, ALCON® Opti-Free® RepleniSH®.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract emblem resembling an eagle or bird in flight, composed of three curved lines. The logo is rendered in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Abbott Medical Optics, Inc. c/o Ms. Kesley E. Gallagher Sr. RA Specialist Global Regulatory Affairs 1700 E. St. Andrews Place Santa Ana, CA 92705

Re: K093254

SEP 1 5 2010

Trade/Device Name: AMO brand Multi-Purpose Disinfecting Solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) contact lens care products Regulatory Class: Class II Product Code: LPN Dated: July 29, 2010 Received: July 30, 2010

Dear Ms. Gallagher:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2 - Ms. Kesley Gallagher

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

17

Sincerely yours,

Kesia Alexander

Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological, and Ear. Nose and Throat Devices Office of Device Eyaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) NUMBER: (IF KNOWN)

DEVICE NAME:

K093254

AMO brand Multipurpose Disinfecting
Solution

INDICATIONS FOR USE:

AMO brand Multi-Purpose Disinfecting Solution is indicated for the care of soft (hydrophilic) contact lenses, including silicone hydrogel lenses. Use this product as recommended by your eye care practitioner to:

  • Chemically (NOT HEAT) Disinfect ●
  • Clean ●
  • Rinse .
  • Store
  • Remove Protein
  • Condition

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR K. Warbnb Over-the-Counter-Use ✓
(Optional Format 1-2-96)

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K093254

SEP 1 5 2010

§ 886.5928 Soft (hydrophilic) contact lens care products.

(a)
Identification. A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.(b)
Classification. Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”