K Number
K093023

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-10-29

(30 days)

Product Code
Regulation Number
870.1340
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PTFE Peelable Introducer is intended for use in the percutaneous insertion of pacing leads or catheters in the venous system.

Device Description

The Greatbatch Medical PTFE Peelable Introducer kit consists of a disposable needle, a disposable syringe, a guidewire, and a peelable introducer set consisting of a dilator and sheath with integrated proximal handles. The Introducer is available in various lengths in 3.5F through 16.5F. The dilator is designed to be delivered over a guidewire. The dilator and hub is constructed of high density polyethylene. The outer sheath and handle is constructed of PTFE and TPX. The inner dilator and outer sheath are radiopaque so that they are visible under fluoroscopy. The hub of the dilator has a spin lock design that connects to the luer of the sheath when rotated. The PTFE Peelable Introducer sheath has a "peel-away" feature which allows the user to remove the sheath without removing the inserted catheter or pacing lead. The PTFE Peelable Introducer is provided sterile and is intended for single use. The device is not intended for sterilization by the user.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Greatbatch Medical PTFE Peelable Introducer, as derived from the provided text:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Device functions in a safe and effective mannerTest results verify that the device performs per specification requirements and is equivalent to the predicate device without creating additional risk to the patient or user.
Meets device specificationsTest results verify that the device performs per specification requirements.
Functional and dimensional testing of the PTFE Peelable IntroducerTest results verify that the device performs per specification requirements.
Equivalent to the predicate device (Greatbatch Medical PTFE Peelable Introducer K081394)Test results verify that the device is equivalent to the predicate device.
Will perform as intendedThe testing performed confirms that the PTFE Peelable Introducer will perform as intended.

Study Information

  1. Sample size used for the test set and the data provenance:

    • The document does not specify a separate "test set" in the context of clinical or image-based studies.
    • The "Performance Testing" section states, "The performance testing for this device included testing to verify that the device continues to functions in a safe and effective manner." This refers to non-clinical, in-vitro testing rather than human or animal studies.
    • Data Provenance: Not applicable as no human or image data was used. The testing described is non-clinical/in-vitro.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The "ground truth" for this device's performance was established through engineering specifications and non-clinical functional testing rather than expert review of clinical cases.
  3. Adjudication method for the test set:

    • Not applicable. There was no clinical test set requiring adjudication.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a medical device (introducer), not an AI-powered diagnostic or image analysis system. Therefore, an MRMC study related to AI assistance is not relevant.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical medical device, not an algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Engineering specifications and functional performance metrics. The ground truth for proving the device's efficacy and safety relies on meeting pre-defined physical and functional requirements established through non-clinical testing.
  7. The sample size for the training set:

    • Not applicable. This is a physical medical device, not a machine learning model, so there is no "training set."
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set.

{0}------------------------------------------------

K093023

Greatbatch Medical Special Premarket Notification 510(k)

Image /page/0/Picture/3 description: The image shows the logo for Greatbatch Medical. The logo consists of a hexagon shape made up of six smaller hexagons, with one of the hexagons filled in black. To the right of the hexagon shape is the text "Greatbatch" in a bold, sans-serif font, with the word "Medical" below it in a smaller font. The logo is simple and modern, and the use of the hexagon shape suggests a connection to science or technology.

510(K) Summary

4.1 Applicant Information

OCT 2 9 2009

Submitter's Name:Greatbatch Medical
Address:2300 Berkshire Lane North
Minneapolis, MN 55441

Establishment Registration No: 2183787

Contact Person: Kristi Fox Regulatory Affairs Associate kfox@greatbatchmedical.com (763) 951-8181 (phone) (763) 559-0148 (fax)

4.2 Device Information

Trade Name:PTFE Peelable Introducer
Classification Name:Introducer, Catheter
Product Code:DYB
Regulation:Class II, 21 CFR 870.1340
Panel:Cardiovascular

4.3 Performance Standards

No performance standards applicable to this product have been developed under Section 514 of the Act.

4.4 Predicate Device

· Greatbatch Medical PTFE Peelable Introducer (K081394)

4.5 Device Description

The Greatbatch Medical PTFE Peelable Introducer kit consists of a disposable needle, a disposable syringe, a guidewire, and a peelable introducer set consisting of a dilator and sheath with integrated proximal handles. The Introducer is available in various lengths in 3.5F through 16.5F. The dilator is designed to be delivered over a guidewire.

{1}------------------------------------------------

The dilator and hub is constructed of high density polyethylene. The outer sheath and handle is constructed of PTFE and TPX. The inner dilator and outer sheath are radiopaque so that they are visible under fluoroscopy. The hub of the dilator has a spin lock design that connects to the luer of the sheath when rotated.

The PTFE Peelable Introducer sheath has a "peel-away" feature which allows the user to remove the sheath without removing the inserted catheter or pacing lead. The PTFE Peelable Introducer is provided sterile and is intended for single use. The device is not intended for sterilization by the user.

4.6 Indications for Use

The PTFE Introducer is intended for use in the percutaneous insertion of pacing leads or catheters in the venous system.

4.7 Summary of Non-Clinical Testing

Performance Testing: The performance testing for this device included testing to verify that the device continues to functions in a safe and effective manner. The performance testing included the device specifications, functional and dimensional testing of the PTFE Peelable Introducer and other testing as applicable to the device. Test results verify that the device performs per specification requirements and is equivalent to the predicate device without creating additional risk to the patient or user.

4.8 Summary of Clinical Testing

No clinical evaluations of this product have been performed.

4.9 Statement of Equivalence

Through the information presented, Greatbatch Medical considers the PTFE Peelable Introducer to remain substantially equivalent to the currently marketed Greatbatch Medical PTFE Peelable Introducers based on a comparison of the indications for use and the technological characteristics. The testing performed confirms that the PTFE Peelable Introducer will perform as intended.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES * USA" arranged in a circular pattern around the eagle. The eagle is black, and the text is in a smaller font size.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

OCT 2 9 2009

Greatbatch Medical % Ms. Kristi Fox 2300 Berkshire Lane North Minneapolis, MN 55441

Re: K093023

Trade/Device Name: PTFE Peelable Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Introducer, Catheter Regulatory Class: Class II Product Code: DYB Dated: September 25, 2009 Received: September 29, 2009

Dear Ms. Fox:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing

{3}------------------------------------------------

Page 2-Ms. Kristi Fox

practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Le From Zuckerman, M.

Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

Kog3023 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: PTFE Peelable Introducer

Indications for Use:

The PTFE Peelable Introducer is intended for use in the percutaneous insertion of pacing leads or catheters in the venous system.

Prescription Use____ X

AND/OR

Over-The-Counter Use_

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLÉASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices

Division of Cardiovascular Devices
510(k) Number K0943023

Confidential ·

4-7

§ 870.1340 Catheter introducer.

(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).