K Number
K092361

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-10-29

(86 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The REEF HP 0.035" OTW PTA Balloon Dilatation Catheters are intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Device Description

The Reef HP 0.035"OTW PTA Balloon Catheter is an over-the-wire percutaneous transluminal angioplasty (PTA) catheter consisting of a proximal hub, dual lumen shaft, and a distal non-compliant dilatation balloon. The REEF HP 0.035" OTW PTA Balloon Dilatation Catheter is compatible with guidewires with a maximum diameter of 0.035" and with 5, 6 or 7 FR introducer sheaths, depending on the diameter/length of the dilatation balloon. The catheter is provided in useable catheter lengths of 50 cm, 80cm and 120 cm.

AI/ML Overview

The provided documentation is a 510(k) summary for a medical device (REEF HP 0.035" OTW PTA Balloon Dilatation Catheter), not a study evaluating an AI/ML powered device. Therefore, it does not contain the specific information requested in the prompt regarding acceptance criteria, study design parameters (sample size, data provenance, ground truth establishment, expert qualifications, adjudication methods), or comparative effectiveness studies typically associated with AI/ML device evaluations.

The document states:

  • Performance Data: "Bench testing of the REEF HP 0.035" OTW PTA Balloon Dilatation Catheter demonstrated that the device met the bench test acceptance criteria."
  • Biocompatibility: "Biocompatibility testing was performed in accordance with ISO 10993-Part 1."

However, it does not provide:

  1. A table of acceptance criteria or reported device performance metrics and their values.
  2. Details about sample sizes for test sets, data provenance, or anything related to "test sets" in the context of an AI/ML algorithm.
  3. Information on experts, ground truth establishment (beyond general bench testing), or adjudication methods, as these are not relevant to the traditional medical device being described.
  4. Any mention of a multi-reader multi-case (MRMC) comparative effectiveness study.
  5. Any mention of a standalone algorithm performance study.
  6. The type of ground truth used (e.g., pathology, expert consensus) for algorithm training or testing.
  7. Sample size for a training set.
  8. How ground truth for a training set was established.

This document is a regulatory submission for a conventional medical device (a balloon catheter), focusing on its substantial equivalence to predicate devices based on design, materials, technological characteristics, and bench performance, rather than an AI/ML diagnostic or assistive device.

{0}------------------------------------------------

K092361

OCT 2 9 2009

510(k) Summary

This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807, Section 807.92(c).

Invatec S.p.A. Via Martiri della Libertá. 7 25030 Roncadelle (Brescia) Italy Peter Muster PhD, VP QA & RA Invatec S.p.A. Tel: +39 030 258 93 11 Fax: +39 030 258 93 12 Email: Peter.Muster@invatec.com

Contact Person:

Applicant:

John Clay, Director, QA&RA Invatec Inc., US Emrick Blvd. 3101, Suite 113 Bethlehem, PA 18020 Tel: (610) 625-1429 Fax: (610) 625-1436 Email: john.clay@invatec-us.com

Date Prepared: July 30, 2009

Device Information

Trade Name:REEF HP 0.035" OTW PTA Balloon Dilatation Catheter
Common Name:Percutaneous Transluminal Angioplasty Catheter
Regulation Name:Percutaneous Catheter

Predicate Devices

  • Invatec Admiral Extreme 0.035" OTW PTA Balloon Dilatation Catheter ◆ (K062809)
  • Cordis Powerflex Extreme PTA Balloon Catheter (K032737) .

Device Description

The Reef HP 0.035"OTW PTA Balloon Catheter is an over-the-wire percutaneous transluminal angioplasty (PTA) catheter consisting of a proximal hub, dual lumen shaft, and a distal non-compliant dilatation balloon. The REEF HP 0.035" OTW PTA Balloon Dilatation Catheter is compatible with guidewires with a maximum diameter of 0.035" and with 5, 6 or 7 FR introducer sheaths, depending on the diameter/length of the

{1}------------------------------------------------

dilatation balloon. The catheter is provided in useable catheter lengths of 50 cm, 80cm and 120 cm.

Indication for Use

The REEF HP 0.035" OTW PTA Balloon Dilatation Catheters are intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Technological Characteristics

The REEF HP 0.035" OTW PTA Balloon Dilatation Catheter have the same or similar design, materials and fundamental technology as the previously cleared Admiral Extreme 0.035" OTW PTA Balloon Dilatation Catheters. In addition, the REEF HP 0.035" OTW PTA Balloon Dilatation Catheter has Rated Burst Pressures (RBP) similar to the Cordis Powerflex Extreme PTA Balloon Catheter, and other PTA catheters.

Performance Data.

Bench testing of the REEF HP 0.035" OTW PTA Balloon Dilatation Catheter demonstrated that the device met the bench test acceptance criteria. Biocompatibility testing was performed in accordance with ISO 10993-Part 1.

Conclusion

Based on the same intended use, technological characteristics, and performance characteristics, Invatec believes the REEF HP 0.035" OTW PTA Balloon Dilatation Catheter is substantially equivalent to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

Invatec, Inc., US c/o Mr. John Clay Director of Regulatory Affairs and Quality 3101 Emrick Boulevard, Suite 113 Bethlehem, PA 18020

OCT 2 9 2009

Re: K092361

REEF HP 0.035" OTW PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: July 30, 2009 Received: August 4, 2009

Dear Mr. Clay:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 - Mr. John Clay

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

W. A. Wilhelm

Bram D. Zuckerman, M.D. Director

Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): Kog 236)

Device Name: REEF HP 0.035" OTW PTA Balloon Dilatation Catheter

Indications for Use:

The REEF HP 0.035" OTW PTA Balloon Dilatation Catheters are intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Prescription Use _ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE iF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M.A. Yillehem

ision Sian-C Division of Cardiovascular Devices

510(k) Number

Page 1 of

12

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).