K Number
K091984

Validate with FDA (Live)

Date Cleared
2009-10-28

(118 days)

Product Code
Regulation Number
866.5750
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in vitro diagnostic use with the IMMULITE® 2000 Analyzer - for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders.

Device Description

IMMULITE® 2000 3gAllergy™ Specific IgE is a solid-phase, two-step, chemiluminescent immunoassay that exploits liquid phase kinetics in a bead format. It represents a significant advance over conventional methods relying on allergens attached to a solid-phase support, such as a paper disk. The allergens are covalently bound to a soluble polymer matrix, which in turn is labeled with a ligand. The use of an amino acid co-polymer amplifies the amount of allergen that the matrix can support. Incubation Cycles: 2 × 30 minutes.

AI/ML Overview

The Siemens Healthcare Diagnostics Inc. IMMULITE® 2000 3gAllergy™ Specific IgE Assay is an in vitro diagnostic device intended for the quantitative measurement of allergen-specific IgE in human serum, to aid in the clinical diagnosis of IgE-mediated allergic disorders. The provided document focuses on the clearance of 27 additional specific allergens for use with this existing assay.

Here's an analysis of the acceptance criteria and study data provided:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for precision, linearity, or specificity (inhibition). Instead, it presents the study results and concludes that they demonstrate "acceptable analytical performance." For clinical performance, it discusses general agreement with clinical data.

Therefore, the table below will summarize the reported performance metrics, implying that these results were deemed acceptable by the manufacturer and the FDA for substantial equivalence.

Performance MetricAcceptance Criteria (Implied by Study Results & Approval)Reported Device Performance
PrecisionWithin-run CV and Total CV within acceptable ranges for immunoassays.Within-Run CV: Ranges from 2.68% (Pear, Positive #2) to 11.08% (Black Pepper, Positive #2). Most values are between 3-5%. Total CV: Ranges from 3.94% (Sole, Positive #3) to 12.30% (Black Pepper, Positive #2). Most values are below 9%.
LinearityRegression Slope close to 1 and Intercept close to 0, with narrow 95% Confidence Intervals.Slope: All slopes are very close to 1 (e.g., 0.9871 to 1.0233). Intercept: All intercepts are close to 0 (e.g., -0.2153 to 0.4035). 95% CI: The 95% CIs for both slope and intercept generally encompass 1 and 0 respectively or are reasonably tight.
Specificity (Inhibition)Demonstrated concentration-dependent inhibition by relevant extracts, with at least 50% inhibition at the highest inhibitor concentration.All tested allergens (Black Pepper, Broccoli, Cabbage, Cheese-Mold Type, Cherry, Chestnut, Chocolate, Coffee, Cucumber, Grape, Grapefruit, Haddock, Lamb, Lemon, Lima Bean, Peach, Pear, Pineapple, Plum, Pumpkin, Red Snapper, Sole, Spinach, Sweet Potato, Trout, Turkey Meat, and Watermelon) showed: - Concentration-dependent inhibition. - Target of 50% inhibition met at the highest inhibitor concentration tested (5 mg/mL, or the starting concentration for Haddock, Red Snapper, and Sole).
Clinical PerformanceGood agreement with clinical documentation of allergen sensitivity.Sensitivity: 66.2% (Lower Conf 64%, Upper Conf 69%) Specificity: 98.9% (Lower Conf 99%, Upper Conf 99%) Overall Agreement: 89.1% (Lower Conf 88%, Upper Conf 90%)

2. Sample Size Used for the Test Set and Data Provenance

  • Precision: Three positive samples and one negative sample were tested for each allergen. Each sample was assayed in two aliquots, two runs per day, on 20 different days, for three allergen lots. This setup aims to assess within-run and total precision across different batches and extended testing periods.
  • Linearity: For each allergen, two samples were diluted in 2-fold serial dilutions to 5 levels. Each diluted and undiluted sample was tested.
  • Specificity (Inhibition): A "single serum sample or pool of sera" was used for competitive inhibition testing for each allergen. A negative sample was used as a background control. Inhibition experiments involved 70uL of undiluted and minimally 4-5 levels of serially diluted inhibitor extract (5, 1, 0.2, 0.04, 0.008 mg/mL, and sometimes more dilute levels). Each sample mixture with inhibitor extract and controls was assayed with 1 lot of each allergen.
  • Clinical Performance: 4,118 samples were tested: 1,238 from clinically positive individuals and 2,880 from clinically normal (non-atopic) individuals.
  • Data Provenance: The document does not explicitly state the country of origin or whether the data is retrospective or prospective for any of these studies. It can be inferred that these are prospective studies conducted by the manufacturer, Siemens Healthcare Diagnostics Inc., for regulatory submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Precision, Linearity, Specificity: For these analytical performance studies, the "ground truth" is based on the inherent properties of the measurements and controls, rather than expert interpretation of a diagnostic outcome. The accuracy of the assay itself is being evaluated.
  • Clinical Performance: For the clinical performance study, the "ground truth" (clinical data for IgE-mediated allergic disorders) was established based on "clinically diagnosed atopic and non-atopic individuals." The document does not specify the number or qualifications of the experts (e.g., allergists, physicians) who performed these clinical diagnoses.

4. Adjudication Method for the Test Set

  • Precision, Linearity, Specificity: Not applicable. These are analytical studies where results are quantitative measurements against known standards or controls.
  • Clinical Performance: The document states that the ground truth for clinical performance was based on "clinical data" to define individuals as "clinically diagnosed atopic and non-atopic individuals." There is no mention of an adjudication method (like 2+1, 3+1 consensus) explicitly detailed for the clinical diagnoses themselves. It implies that these were established clinical classifications prior to testing with the IMMULITE® 2000.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in vitro diagnostic assay, not an imaging device or AI algorithm intended to assist human readers in interpretation. Therefore, the concept of "human readers improve with AI vs. without AI assistance" does not apply here.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, the studies evaluating the IMMULITE® 2000 3gAllergy™ Specific IgE Assay are fundamentally standalone. The device performs quantitative measurements of IgE levels in serum, and its performance (precision, linearity, specificity, and comparison to clinical diagnoses) is assessed based on these measurements. There is no "human-in-the-loop" component in the analytical operation of the assay or the interpretation of its quantitative output for the purpose of these performance studies. Clinicians then use these quantitative results, alongside other clinical information, for diagnosis.

7. Type of Ground Truth Used

  • Precision, Linearity: Ground truth is based on known concentrations or expected values derived from serial dilutions.
  • Specificity (Inhibition): Ground truth is based on the known presence and concentration of the inhibitor extract and the expected inhibition of the specific allergen reaction.
  • Clinical Performance: The ground truth was "clinical data," referring to the diagnosis of "clinically diagnosed atopic and non-atopic individuals." This would typically involve a combination of patient history, physical examination, and potentially other diagnostic tests (e.g., skin prick tests, other lab results, patient symptoms, and outcomes). It is not solely pathology or "outcomes data" in a broader sense but relies on a clinical determination of allergy status.

8. Sample Size for the Training Set

The document describes performance studies for the device itself, not the development of an algorithm that would require a distinct "training set" in the context of machine learning or AI. Therefore, the concept of a training set for an AI algorithm is not applicable to this submission. The studies detailed are for validation of the assay performance.

9. How the Ground Truth for the Training Set Was Established

As explained in point 8, the concept of a "training set" for an AI algorithm is not applicable to these studies for this in vitro diagnostic device.

{0}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS

Assigned 510(k) Number

OCT 2 8 2009

The assigned 510(k) number is K09 1984

Sponsor Name and Address

Siemens Healthcare Diagnostics Inc. 5210 Pacific Concourse Drive Los Angeles, CA 90045-6900 (310) 645-8200

Contact

Clare Santulli Sr.Regulatory Affairs Specialist (914) 524-2701 (914) 524-3579 fax clare,santulli@siemens.com

Device Name

Trade name: Classification: Classification Names: Regulation Number: Product Code: Catalog Numbers:

IMMULITE® 2000 3gAllergy™ Specific IgE Assay Class II Radioallergosorbent (RAST) Immunological Test System 866.5750 DHB L2KUN6 (600 tests)

Description of Device

IMMULITE® 2000 3gAllergy™ Specific IgE is a solid-phase, two-step, chemiluminescent immunoassay that exploits liquid phase kinetics in a bead format. 12 (U.S. Patent No. 4,778,751) It represents a significant advance over conventional methods relying on allergens attached to a solid-phase support, such as a paper disk.

The allergens are covalently bound to a soluble polymer matrix, which in turn is labeled with a ligand. The use of an amino acid co-polymer amplifies the amount of allergen that the matrix can support.

Incubation Cycles: 2 × 30 minutes.

1 El Shami AS, Alaba O. Liquid-phase in vitro allergen-specific IgE assay with in situ immobilization. Adv Biosci 1989;74:191-201.

2 Alaba O, El Shami AS. Evaluation of non-specific IgE binding: comparison of two in vitro allergen assays. Adv Biosci 1989;74:203-14.

{1}------------------------------------------------

Indications for Use

For in vitro diagnostic use with the IMMULITE® 2000 Analyzer - for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders.

Establishment Information

IMMULITE® 2000 3gAllergy Specific IgE assay is manufactured by Siemens Healthcare Diagnostics Inc. at the following locations:

Siemens Healthcare Diagnostics Inc. 5210 Pacific Concourse Drive Los Angeles, CA 90045-6900 FDA Establishment #: 3005250747

Predicate

The purpose of this 510(k) submission is for clearance of twenty seven additional specific allergens, named in the table below, to be used with the IMMULITE® 2000 3gAllergy™ Specific IgE on the IMMULITE® 2000 analyzer.

1F280-Black Pepper14F208 - Lemon
2F260 - Broccoli15F182 - Lima Bean
3F216 - Cabbage16F95 - Peach
4F82 - Cheese-Mold Type17F94 - Pear
5F242 - Cherry18F210 - Pineapple
6F299 - Chestnut19F255 - Plum
7F105 - Chocolate20F225 - Pumpkin
8F221 - Coffee21F381 - Red Snapper
9F244 - Cucumber22F337 - Sole
10F259 - Grape23F214 - Spinach
11F209 - Grapefruit24F54 - Sweet Potato
12F42 - Haddock25F204 - Trout
13F88 - Lamb26F284 - Turkey Meat
27F329 - Watermelon

FDA clearance was previously obtained for the assay kit and 196 specific allergens and allergen panels (K013134, K021206, K013135 and K021208).

{2}------------------------------------------------

Please note that the FDA clearances indicated above were in the name of Diagnostic Products Corporation which was acquired by Siemens Medical Solutions Diagnostics in July 2006. Siemens Medical Solutions Diagnostics was renamed Siemens Healthcare Diagnostics Inc. on January 1, 2008.

Precision

Precision studies were performed in accordance with Clinical Laboratory Standard Institute (CLSI) guidance: Evaluation of Precision Performance of Quantitative Methods; Approved Guideline-Second Edition. CLSI document EPS-A2 (ISBN 1-56238-542-9). CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2004, assaying two aliquots of each test sample in two runs per day on 20 different days. Analysis of variance was used to estimate the within-run and total precision.

Three allergen lots were tested using three positive samples and one negative sample. Intraassay and inter-assay precision for the positive samples were evaluated by calculating the kU/L dose percent coefficients of variation (%CV) for each positive sample. Non-specific binding (NSB) was monitored by testing the negative control sample.

Representative precision claims for each allergen tested are presented below:

{3}------------------------------------------------

Allergen Precision Claims*
Within-RunTotal
SampleMeanSDCVSDCV
kU/LkU/L%kU/L%
Allergen = Black Pepper, Lot 115
Positive #11.410.0594.180.0735.18
Positive #23.430.38011.080.42212.30
Positive #36.810.2503.670.5718.38
Allergen = Broccoli, Lot 114
Positive #112.020.5464.540.7596.31
Positive #25.680.2514.420.3385.95
Positive #32.370.0913.840.1275.36
Allergen = Cabbage, Lot 117
Positive #10.780.0232.950.0536.79
Positive #214.400.5914.100.9716.74
Positive #322.920.8013.491.4226.20
Allergen = Cheese, Mold Type, Lot 116
Positive #11.200.0373.080.0594.92
Positive #21.660.0523.130.0674.04
Positive #36.040.2413.990.2874.75
Allergen = Cherry, Lot 118
Positive #13.310.1303.930.2216.68
Positive #21.100.0363.270.0706.36
Positive #311.190.3903.490.9818.77
Allergen = Chestnut, Lot 114
Positive #17.930.2383.000.3964.99
Positive #22.860.1083.780.1354.72
Positive #35.810.1813.120.2614.49
Allergen = Chocolate, Lot 118
Positive #10.430.0317.210.0388.84
Positive #21.130.0857.520.0958.41
Positive #30.410.0297.070.0389.27
Allergen = Coffee, Lot 116
Positive #10.580.0203.450.0417.07
Positive #21.840.0754.080.1286.96
Positive #30.610.0243.930.0487.87
Allergen = Cucumber, Lot 116
Positive #12.850.0973.400.1675.86
Positive #22.100.0823.900.1688.00
Positive #38.380.2563.050.6217.41
Allergen = Grape, Lot 117
Positive #15.210.1893.630.2925.60
Positive #21.290.0473.640.0644.96
Positive #314.200.4643.270.7074.98
Allergen = Grapefruit, Lot 116
Positive #12.680.0923.430.1605.97
Positive #21.070.0312.900.0888.22
Positive #313.790.5243.800.9887.16
Allergen =Haddock, Lot 110
Positive #10.840.0313.690.0384.52
Positive #25.210.1763.380.2685.14
Positive #38.660.3874.470.4134.77
Allergen = Lamb, Lot 115
Positive #12.960.1053.550.1806.08
Positive #29.500.3593.780.5916.22
Positive #37.870.3194.050.5096.47
Allergen = Lemon, Lot 116
Positive #112.130.4073.360.6215.12
Positive #21.160.0484.140.0655.60
Positive #33.880.1473.790.2666.86
Allergen = Lima Bean, Lot 111
Positive #13.890.1764.520.2576.61
Positive #25.850.2293.910.3205.47
Positive #314.770.5393.650.6684.52
Allergen = Peach, Lot 119
Positive #13.800.1594.180.1975.18
Positive #23.870.1764.550.2125.48
Positive #311.150.4423.960.6665.97
Allergen = Pear, Lot 116
Positive #11.630.0583.560.1056.44
Positive #24.360.1172.680.2686.15
Positive #36.610.2123.210.4286.48
Allergen = Pineapple, Lot 115
Positive #12.210.0984.430.1687.60
Positive #21.500.0493.270.1077.13
Positive #30.830.0242.890.0819.76
Allergen = Plum, Lot 113
Positive #11.380.1178.480.1369.86
Positive #23.930.1273.230.2115.37
Positive #32.810.0873.100.1475.23
Allergen = Pumpkin, Lot 115
Positive #10.980.0636.430.0767.76
Positive #23.330.1364.080.1875.62
Positive #38.920.3063.430.4134.63
Allergen = Red Snapper, Lot 110
Positive #115.920.6123.840.7234.54
Positive #23.740.1293.450.1824.87
Positive #315.140.4272.820.7004.62
Allergen = Sole, Lot 110
Positive #11.500.0513.400.0775.13
Positive #24.750.2194.610.2765.81
Positive #318.620.5773.100.7343.94
Allergen = Spinach, Lot 115
Positive #10.910.0272.970.0394.29
Positive #21.730.0543.120.0854.91
Positive #314.800.4052.740.6854.63
Allergen = Sweet Potato Lot 113
Positive #16.710.2384.220.3525.25
Positive #24.090.1784.350.2736.67
Positive #31.970.0814.110.1025.18
Allergen =Trout Lot 115
Positive #130.710.9453.081.8255.94
Positive #21.720.0583.370.0975.64
Positive #317.280.6463.740.9265.36
Allergen = Turkey Meat, Lot 116
Positive #11.360.0453.310.1178.60
Positive #24.800.153.250.2114.40
Positive #36.990.2333.330.6429.18
Allergen = Watermelon, Lot 112
Positive #11.620.0643.950.1227.53
Positive #25.030.1723.420.3456.86
Positive #319.250.5612.911.2506.49

:

{4}------------------------------------------------

and the comments of the comments of the comments of the comments of

:

and the comments of the comments of the country of

{5}------------------------------------------------

1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 -

and the control control control control control controllers and

,

:

:

.

. .

. . . . ·

. .

:

:

:

and the state of the state of the states and

{6}------------------------------------------------

  • data are representative of one lot on one instrument

Linearity

For each allergen, two samples were diluted in 2-fold serial dilutions to 5 levels. The undiluted (neat) and diluted samples were tested with the specific allergen to demonstrate linearity at concentrations within the assay limits. Regression statistics for each allergen comparing observed to expected data are presented below.

AllergenRegression EquationNSlope95% CIIntercept95% CI
Black PepperY=0.9975X +0.0404120.99750.9761-1.01880.0404-0.0165-0.0973
BroccoliY=1.0035X +0.2560121.00350.9666-1.04050.2560-0.0089-0.5209
CabbageY=0.9976X+0.0231120.99760.9883-1.00680.0231-0.0077-0.0540
Cheese-Mold TypeY=0.9929X +0.0656120.99290.9594-1.02640.06560.0110-0.1202
CherryY=1.0165X +0.2024121.01650.9719-1.06120.2024-0.1789-0.5838
ChestnutY=1.0074X +0.2158121.00740.9748-1.03990.2158-0.0857-0.5173
ChocolateY=1.0047X -0.016391.00470.9833-1.0261-0.0163-0.0349-0.0022
CoffeeY=0.9876X+0.0824110.98760.9168-1.05840.0824-0.0008-0.1655
CucumberY=1.0003X+0.1336121.00030.9756-1.02490.1336-0.1292-0.3963
GrapeY=1.0117X+0.1622121.01170.9811-1.04230.1622-0.1933-0.5177
GrapefruitY=0.9967X +0.0264120.99670.9816-1.01190.0264-0.0277-0.0804
HaddockY=1.0010X +0.0653121.00100.9852-1.01690.0653-0.1583-0.2888
LambY=1.0233X -0.2153121.02330.9966-1.0499-0.2153-0.4527-0.0221
LemonY=0.9995X +0.1424120.99950.9836-1.01530.14240.0300-0.2549
Lima BeanY=0.9973X +0.0791120.99730.9778-1.01670.07910.0067-0.1514
PeachY=0.9916X +0.1351120.99160.9647-1.01850.1351-0.2444-0.5145

Linearity

{7}------------------------------------------------

AllergenRegression EquationNSlope95% CIIntercept95% CI
PearY=1.0073X +0.4035121.00730.9667-1.04800.4035-0.1411-0.9481
PineappleY=0.9871X +0.0561120.98710.9782-0.99600.05610.0373-0.0749
PlumY=1.0175X +0.0408121.01750.9819-1.05310.0408-0.2704-0.3521
PumpkinY= 1.0033X +0.0589121.00330.9805-1.02620.0589-0.0278 -0.1455
Red SnapperY=0.9914X +0.0753120.99140.9822-1.00060.07530.0421-0.1085
SoleY=0.9966X +0.0610120.99660.9825-1.01060.0610-0.0042-0.1262
SpinachY= 1.0046X + 0.1871121.00460.9568-1.05230.1871-0.0821-0.4563
Sweet PotatoY=0.9999X +0.1379110.99990.9529-1.04690.13790.0074-0.2683
TroutY= 1.0038X -0.0478121.00380.9896-1.0181-0.0478-0.2083-0.1127
Turkey MeatY= 1.0129X +0.1070121.01290.9823-1.04360.1070-0.0919-0.3058
WatermelonY= 1.0138X +0.0405121.01380.9856-1.04200.0405-0.2677-0.3486

Specificity (Inhibition) Studies

..

Specificity of each allergen was verified through competitive inhibition testing using a single serum sample or pool of sera. A negative sample was used to measure the background response.

To initiate the inhibition experiment. 70uL of undiluted and minimally 4 levels of 5-fold serially diluted inhibitor extract (at 5. 1. 0.2, 0.04 and 0.008 mg/mL) were mixed with 250uL of sample or pool. In some instances additional levels of inhibitor extract (at 0.0016, 0.00032, and 0.000064 mg/mL) were also used. This mixture was incubated at room temperature (15-28 °C) for 1 hour allowing the immunological reaction to occur. Each sample mixture containing the inhibitor extract and the appropriate controls was assayed with 1 lot of each allergen. The percent (%) inhibition was calculated according to the following formula:

(Response of pos. control (pos. sample - ness sample response with inhibitor extract) (Response of pos. control (pos. sample - neg. sample) )

X 100

The inhibition plots demonstrate that the allergens tested are inhibited by the relevant inhibitor extract in a concentration dependent fashion. Also, the target % inhibition of 50% for the highest inhibitor concentration tested was met. These results indicate specificity of Black Pepper, Broccoli, Cabbage, Cheese- Mold Type, Chestnut, Chocolate, Coffee, Cucumber, Grape, Grapefruit, Haddock, Lamb, Lemon, Lima Bean, Peach, Pear, Pineapple, Plum, Pumpkin, Red Snapper, Sole, Spinach, Sweet Potato, Trout, Turkey Meat, and Watermelon allergens. Summary inhibition table is presented below.

{8}------------------------------------------------

InhibitorConcentration(mg/mL)%InhibitionInhibitorConcentration(mg/mL)%InhibitionInhibitorConcentration(mg/mL)%Inhibition
Black PepperBroccoliCabbage
593.01595.21596.75
177.62191.51195.56
0.256.990.285.560.291.34
0.0422.380.0474.170.0470.45
0.0081.400.00824.750.00816.30
Cheese-Mold TypeCherryChestnut
589.34598.94595.98
188.83192.40190.12
0.280.710.276.560.249.64
0.0474.110.0457.760.040.00
0.00816.750.00846.570.0082.70
ChocolateCoffeeCucumber
579.53574.53597.89
182.46166.98196.22
0.250.290.239.620.291.22
0.0427.490.0413.210.0484.76
0.0085.260.00810.850.00835.31
0.00169.35
GrapeGrapefruitHaddock(inhibitor conc. starting at 0.2 mg/mL)
585.56597.780.297.33
152.22193.130.0490.80
0.223.330.282.320.00883.47
0.0426.670.0415.760.001680.20
0.00814.440.0080.810.003270.73
0.00006418.47
InhibitorConcentration(mg/mL)%InhibitionInhibitorConcentration(mg/mL)%InhibitionInhibitorConcentration(mg/mL)%Inhibition
LambLemonLima Bean
595.86591.50596.94
192.88186.82191.68
0.284.860.283.360.282.49
0.0462.100.0459.910.0454.49
0.00834.020.0087.570.00821.66
PeachPearPineapple
597.28585.31588.36
192.41171.19183.60
0.283.460.257.060.272.49
0.0474.250.0449.720.0442.86
0.00845.010.00827.680.00820.11
PlumPumpkinRed Snapper(inhibitor conc. starting at 0.2 mg/mL)
597.86593.180.283.58
182.91183.360.0474.68
0.267.670.257.980.00876.29
0.0446.500.0433.970.001676.50
0.00816.700.00814.190.003246.46
0.0000645.79
Sole (inhibitor conc. starting at 1)SpinachSweet Potato
197.04580.10560.57
0.290.31181.20142.86
0.0477.550.262.020.20.00
0.00863.520.0436.580.040.00
0.001647.500.00815.750.0083.43
0.003239.34
InhibitorConcentration(mg/mL)%InhibitionInhibitorConcentration(mg/mL)%InhibitionInhibitorConcentration(mg/mL)%Inhibition
TroutTurkey MeatWatermelom
599.03596.525100.00
197.71193.99193.53
0.296.100.290.510.287.30
0.0491.360.0468.200.0483.37
0.00883.990.00820.090.00862.59
0.001678.250.00160.001612.70
0.003260.210.00320.00325.08
0.00006412.99

.

,

. .


and the control control control control controllers and

and the comments of the comments of

{9}------------------------------------------------

and the comments of the comments of the comments of

and the comments of the comments of the comments of the comments of the contraction of the contribution of the contribution of the contribution of the contribution of the fir

"


{10}------------------------------------------------

Clinical Performance Studies

Clinical performance was demonstrated by testing serum samples against specific allergens from clinically diagnosed atopic and non-atopic individuals. Allergen-specific testing was obtained using the IMMULITE® 2000 3gAllergy™ assay.

Data summary agreement of the IMMULITE® 2000 3gAllergy results to clinical data is presented in the table below.

IMMULITE® 2000Clinical Data
ClinicalNormalTotal
Positive82032852
Negative4182,8483,266
Total1,2382,8804,118
66.2%98.9%89.1%
SensitivitySpecificityAgreement
Lower Conf64%99%88%
Upper Conf69%99%90%
Allergens included: Black Pepper, Broccoli, Cabbage, Cheese Mold-Type, Cherry, Chestnut, Chocolate, Coffee,Cucumber, Grape, Grapefruit, Haddock, Lamb, Lima Bean, Peach, Pear, Pineapple, Plum, Pumpkin,

Red Snapper, Sole, Spinach, Sweet Potato, Trout, Turkey Meat, and Watermelon

{11}------------------------------------------------

IMMULITE® 2000 3gAllergy assay results for all allergens compare well with clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity.

Conclusions for all Studies

Allergens including Black Pepper, Broccoli, Cabbage, Cheese Mold-Type, Cherry, Chestnut, Chocolate, Coffee, Cucumber, Grape, Grapefruit, Haddock, Lamb, Lemon, Lima Bean, Peach, Pear, Pineapple, Plum, Pumpkin, Red Snapper, Sole, Spinach, Sweet Potato, Trout, Turkey Meat, and Watermelon for use with the IMMULITE® 2000 3gAllergy Specific IgE assay demonstrate acceptable analytical performance including precision, linearity and specificity. IMMULITE® 2000 assay results compare well with clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity. Substantial equivalence was demonstrated to clinical data, supporting the following intended use of the IMMULITE® 2000 3gAllergys Specific IgE assay and the twenty seven previously listed allergens:

For in vitro diagnostic use with the IMMULITE® 2000 Analyzer - for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle with three wavy lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle symbol.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

OCT 2 8 2009

Siemens Healthcare Diagnostics, Inc c/o Ms. Clare Santulli Senior RA/ QS Specialist 5210 Pacific Concourse Dr. Los Angeles, CA 90045-6900

Re: K091984

Trade/Device Name; IMMULITE® 2000 3gAllergy™ Specific IgE assay kit Regulation Number: 21 CFR §866.5750

Regulation Name: Radioallergosorbent (RAST) immunological test system Regulatory Class: Class II Product Code: DHB Dated: October 19, 2009 Received: October 23, 2009

Dear Ms. Clare Santulli:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801);

{13}------------------------------------------------

medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.html for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mana M. Chan

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{14}------------------------------------------------

Indication for Use

510(k) Number (if known):

Device Name: IMMULITE 3gAllergy™ Specific IgE Assay

Indication For Use:

For in vitro diagnostic use with the IMMULITE 2000 Analyzer -- for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders.

Prescription Use 2 (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Geena Philip

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) 091984

§ 866.5750 Radioallergosorbent (RAST) immunological test system.

(a)
Identification. A radioallergosorbent immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the allergen antibodies (antibodies which cause an allergic reaction) specific for a given allergen. Measurement of specific allergen antibodies may aid in the diagnosis of asthma, allergies, and other pulmonary disorders.(b)
Classification. Class II (special controls). The device, when intended to detect any of the allergens included in Table 1 in this paragraph, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9. Table 1—Class II Exempt Allergens Under § 866.5750—Radioallergosorbent (RAST) Immunological Test Systems
Allergen code Allergen product Source (taxonomical name) Grass Pollens g1 Sweet vernal grass Anthoxanthum odoratum. g3 Cocksfoot grass, Orchard grass Dactylis glomerata. g4 Meadow fescue Festuca elatior. g5 Rye-grass (perennial rye grass) Lolium perenne. g7 Common reed (common reed grass) Phragmites communis. g8 Meadow grass, Kentucky blue (June grass) Poa pratensis. g9 Redtop, Bentgrass Agrostis stolonifera ,Agrostis gigantea (Agrostis alba ). g11 Brome grass Bromus inermis. g12 Cultivated rye (cultivated rye grass) Secale cereale. g13 Velvet grass Holcus lanatus. g14 Cultivated oat (cultivated oat grass) Avena sativa. g15 Cultivated wheat (cultivated wheat grass) Triticum aestivum (Triticum spp.). g16 Meadow foxtail (meadow foxtail grass) Alopecurus pratensis. g17 Bahia grass Paspalum notatum. g24 Wheat grass, Western Agropyron smithii (Elymus smithii ). g30 Bluegrass, annual Poa annua. g70 Wild rye grass Elymus triticoides Elymus condensatus. g71 Canary grass Phalaris arundinacea. g201 Barley, cultivated Hordeum vulgare. g202 Maize, corn (cultivated corn) Zea mays. g203 Salt grass Distichlis spicata. g204 False oat-grass Arrhenatherum elatius. g216 Cyn d 1 Cynodon dactylon. g701 Phl p 1.0102, Phl p 5.0101 Phleum pratense. g702 Phl p 7.0101 Phleum pratense. g703 Phl p 12.0101 Phleum pratense. Weed Pollens w2 Western ragweed Ambrosia psilostachya. w4 False ragweed Ambrosia acanthicarpa (Franseria acanthicarpa ). w5 Wormwood Artemisia absinthium Artemisia annua. w6 Mugwort Artemisia vulgaris. w7 Marguerite, ox-eye daisy Chrysanthemum leucanthemum. w8 Dandelion Taraxacum vulgare ,Taraxacum officinale. w9 Plantain (English), Ribwort Plantago lanceolata. w10 Goosefoot, lamb's quarters Chenopodium album. w11 Saltwort (prickly), Russian thistle Salsola kali (Salsola pestifer ). w12 Goldenrod Solidago virgaurea (Solidago spp.). w13 Cocklebur, common Xanthium commune. w14 Common pigweed (rough pigweed) Amaranthus retroflexus. w15 Scale, Lenscale Atriplex lentiformis. w16 Rough marsh elder Iva ciliate ,Iva annua. w17 Firebush (Kochia) Kochia scoparia. w18 Sheep sorrel Rumex acetosella. w19 Wall pellitory Parietaria officinalis. w20 Nettle (Common stinging nettle) Urtica dioica. w21 Wall pellitory Parietaria judaica. w22 Japanese hop (careless weed) Humulus japonicas (Humulus scandens ). w23 Yellow dock, Yellow dockweed Rumex crispus. w24 Spiny pigweed Amaranthus spinosus. w27 Carnation Dianthus spp. w28 Rose Rosa rugosa. w33 Clover Trifolium pratense. w35 Mexican tea Chenopodium ambrosioides. w36 Rabbit bush Ambrosia deltoidea (Franseria deltoides ). w37 Salt bush, annual Atriplex wrightii. w39 Water hemp, Western Amaranthus rudis (Acnida tamariscina ). w41 Burrobrush Hymenoclea salsola. w42 Poverty weed Baccharis neglecta. w43 Common sagebrush Artemisia tridentata. w45 Alfalfa Medicago sativa. w46 Dog fennel Eupatorium capillifolium. w53 Geranium Geranium spp. w67 Groundsel bush Baccharis halimifolia. w69 Iodine bush Allenrolfea occidentalis. w70 Ragweed, slender Ambrosia confertiflora. w75 Wing scale (wingscale) Atriplex canescens. w82 Careless weed Amaranthus palmeri ,Amaranthus hybridus. w90 Japanese hop Humulus japonicas (Humulus scandens ). w203 Rape (rape pollen) Brassica napus. w204 Sunflower Helianthus annuus. w206 Camomile Matricaria chamomilla. w207 Lupin Lupinus spp. w210 Sugar-beet Beta vulgaris. w211 Par j 2.0101 Parietaria judaica. w231 Art v 1 Artemisia vulgaris (Mugwort). w232 Sal k 1 Salsola kali. w233 Art v 3 Artemisa vulgaris (LTP, Mugwort). w234 Pla l 1 Plantago lanceolata. w235 Che a 1.0101 Chenopodium album. w236 Mer a 1.0101 Mercurialis annua. a753 Art v 1 Artemisia vulgaris (Mugwort weed). Tree Pollens t1 Box-elder (Maple) Acer negundo ,Acer saccharum. t2 Gray alder, speckled alder (alder) Alnus incana. t4 Hazel, hazelnut Corylus avellana ,Corylus americana. t5 American beech (beech) Fagus grandifolia (Fagus americana ). t6 Mountain juniper, Mountain cedar Juniperus ashei (Juniperus sabinoides ). t8 Elm Ulmus americana. t9 Olive Olea europaea. t10 Walnut Juglans californica ,Juglans nigra. t11 Maple leaf sycamore, London plane, Plane tree Platanus acerifolia. t61 Sycamore Platanus occidentalis. t12 Willow Salix caprea ,Salix nigra. t14 Cottonwood (Eastern Cottonwood/Black Cottonwood) Populus deltoides. t15 White ash Fraxinus americana. t16 White pine Pinus strobus. t18 Eucalyptus, gum-tree Eucalyptus globulus (Eucalyptus spp.). t19/t26 Acacia Acacia longifolia (Acacia spp.). t20 Mesquite Prosopis glandulosa/ Prosopis juliflora. t21 Melaleuca, cajeput tree Melaleuca quinquenervia (Melaleuca leucadendron ). t22 Pecan, hickory Carya illinoinensis (Carya pecan ). t23 Italian/Mediterranean/funeral cypress Cupressus sempervirens. t24 Japanese cypress Chamaecyparis obtusa (Chamaecyparis spp.). t25 Ash Fraxinus excelsior. t27 Maple, red Acer rubrum. t29 Acacia Acacia spp. t30 Birch, white Betula populifolia. t32 Willow, black Salix nigra. t33 Ash, Arizona Fraxinus velutina. t35 Cedar, salt Tamarix gallica. t37 Bald cypress (white bald cypress) Taxodium distichum. t38 Elm, Chinese/Siberian Ulmus pumila. t40 Hazelnut tree Corylus americana. t41 White hickory Carya alba (Carya tomentosa ). t42 Oak, red Quercus rubra. t43 Loblolly pine Pinus taeda. t44 Hackberry Celtis occidentalis. t45 Cedar elm Ulmus crassifolia. t47 Juniper, one seed Juniperus monosperma. t48 Pine, lodgepole Pinus contorta. t49 Pine, ponderosa Pinus ponderosa. t50 Beech, European Fagus sylvatica. t51 Tree of Heaven Ailanthus altissima. t52 Western white pine Pinus monticola. t54 Russian olive Elaeagnus angustifolia. t55 Scotch broom Cytisus scoparius. t56 Bayberry Myrica cerifera. t57 Red cedar Juniperus virginiana. t60 Western juniper Juniperus occidentalis. t61 Sycamore Platanus occidentalis. t70 Mulberry (white mulberry) Morus alba. t71 Red mulberry Morus rubra. t72 Queen palm Arecastrum romanzoffiamon. t73 Australian pine Casuarina equisetifolia. t77 Oak mix (red, white, black) Quercus spp. t80 Japanese cypress Chamaecyparis obtusa. t81 Japanese alder Alnus japonica. t83 Mango tree Mangifera indica. t90 Walnut, black Juglans nigra. t96 Poplar, white (poplar) Populus alba. t103/t218 Virginia live oak (live oak) Quercus virginiana. t105 Pepper tree Schinus molle. t110 Orange tree Citrus sinensis. t201 Spruce, Norway spruce Picea abies (Picea excelsa ). t202 Alder, smooth Alnus incana spp.Rugosa (Alnus rugosa ). t203 Horse chestnut Aesculus hippocastanum. t205 Elder Sambucus nigra. t206 Chestnut Castanea sativa. t207 Douglas fir Pseudotsuga menziesii (Pseudotsuga taxifolia ). t208 Linden Tilia cordata. t209 Horn beam Carpinus betulus. t210 Privet Ligustrum vulgare. t211 Sweet gum Liquidambar styraciflua. t212 Cedar Libocedrus decurrens. t213 Pine Pinus radiata. t214 Date palm Phoenix canariensis. t215 Lilac Syringa vulgaris. t217 Pepper tree Schinus molle. t217 Red alder Alnus rubra. t218 Virginia live oak Quercus virginiana. t218 Bayberry (bayberry/sweet gale) Myrica gale. t219 Palo verde Cercidium floridum. t219 Red cedar Juniperus virginiana. t220 Bet v 4 Betula verrucosa (Birch). t221 Bet v 2.0101, Bet v 4 Betula verrucosa (Birch). t222 Cypress (Arizona cypress) Cupressus arizonica. t223 Oil palm Elaeis guineensis. t224 Ole e 1 Olea europaea. t225 Bet v 6 Betula verrucosa (Birch). t226 Cup a 1 Cupressus arizonica. t227 Ole e 7 Olea Europaea. t228 Aspen, quaking Populus tremuloides. t229 Eastern hemlock Tsuga canadensis. t230 Redwood (sequoia) Sequoia sempervirens. t232 Pussy willow Salix discolor. t240 Ole e 9.0101 Olea Europaea. t241 Pla a 1.0101 Platanus acerifolia. t242 Pla a 2 Platanus acerifolia. t243 Pla a 3.0101 Platanus acerifolia. t244 Cor a 1.0103 Corylus avellana. t245 Aln g 1.0101 Alnus glutinosa. t246 Cry j 1 Cryptomeria japonica. t280 Locust tree Robinia pseudoacacia. t401 Brazilian peppertree Schinus terebinthifolius. t402 Mastic tree Pistacia lentiscus. t404 Tree of heaven Ailanthus altissima. t406 Date palm Phoenix dactylifera. a482 Ole e 1 Olea europaea (Olive Oil). Mites d207 Blo t 5.0101 Blomia tropicalis. d208 Lep d 2.0101 Lepidoglyphus destructor. Microorganisms, Molds, Yeast m1 Penicillium chrysogenum (Penicillium notatum ) Penicillium chrysogenum (Penicillium notatum ). m2 Cladosporium herbarum (Hormodendrum ) Cladosporium herbarum (Hormodendrum ). m3 Aspergillus fumigatus Aspergillus fumigatus. m4 Mucor racemosus Mucor racemosus. m5 Candida albicans Candida albicans. m7 Botrytis cinerea Botrytis cinerea. m8 Drechslera halodes (Setomelanomma rostrata, Helminthosporium halodes, Helminthosporium interseminatum ) Drechslera halodes (Setomelanomma rostrata, Helminthosporium halodes. m9 Fusarium moniliforme (Fusarium proliferatum ) Fusarium moniliforme (Fusarium proliferatum ). m10 Stemphylium botryosum Stemphylium herbarum (Stemphylium botryosum ). m11 Rhizopus nigricans Rhizopus nigricans. m12 Aureobasidium pullulans Aureobasidium pullulans. m13 Phoma betae Phoma betae. m14 Epicoccum purpurascens Epicoccum purpurascens (Epicoccum nigrum ). m15 Trichoderma viride Trichoderma viride. m16 Curvularia lunata Curvularia lunata ,Curvularia specifera ([K923044](https://510k.innolitics.com/device/[K923044](https://510k.innolitics.com/device/K923044)) ). m17 Cladosporium fulvum Cladosporium fulvum. m18 Fusarium culmorum Fusarium culmorum. m19 Aspergillus versicolor Aspergillus versicolor. m20 Mucor mucedo Mucor mucedo. m21 Aspergillus clavatus Aspergillus clavatus. m22 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m23 Thermoactinomyces vulgaris Thermoactinomyces vulgaris. m24 Stachybotrys atra Stachybotrys chartarum (Stachybotrys atra ). m24 Paecilomyces spp Paecilomyces spp. m25 Aspergillus versicolor Aspergillus versicolor. m25 Penicillium brevicompactum Penicillium brevicompactum. m26 Cladosporium cladosporioides Cladosporium cladosporioides. m26 Penicillium citrinum Penicillium citrinum. m27 Penicillium spp Penicillium spp. m29 Aspergillus repens Aspergillus repens. m30 Penicillium roqueforti Penicillium roqueforti. m32 Cladosporium cladosporioides Cladosporium cladosporioides. m34 Serpula lacrymans Serpula lacrymans. m36 Aspergillus terreus Aspergillus terreus. m37 Trichophyton mentagrophytes Trichophyton mentagrophytes. m40 Aspergillus amstelodami Aspergillus amstelodami. m43 Saccharomyces Carlsberg Saccharomyces carlsbergensis. m44 Saccharomyces cerevisiae Saccharomyces cerevisiae. m45 Hormodendrum hordei Hormodendrum hordei. m46 Bipolaris spicifera Bipolaris spicifera. m47 Aspergillus nidulans Aspergillus nidulans. m48 Aspergillus oryzae Aspergillus oryzae. m49 Fusarium oxysporum Fusarium oxysporum. m50 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m51 Thermoactinomyces vulgaris Thermoactinomyces vulgaris. m53 Microspora canis Microsporum canis (Microspora canis ). m54 Aspergillus flavus Aspergillus flavus. m63 Helminthosporium intersemin Helminthosporium intersemin. m66 Mucor plumbeus Mucor plumbeus. m67 Mycogone Mycogone perniciosa. m68 Nigrospora oryzae Nigrospora oryzae. m69 Rhodotorula Rhodotorula rubra (Rhodotorula mucilaginosa ). m70 Malassezia furfur (Pityrosporum orbiculare ) Malassezia furfur (Pityrosporum orbiculare ). m71 Spondylocladium Spondylocladium spp. m72 Epidermophyton Epidermophyton floccosum. m73 Epicoccum nigrum Epicoccum purpurascens (Epicoccum nigrum ). m80 Staphylococcal enterotoxin A (Sta a SEA ) Staphylococcus aureus. m80 Helminthosporium spp Helminthosporium spp. m81 Staphylococcal enterotoxin B (Sta a SEB ) Staphylococcus aureus. m88 Stemphylium solani Stemphylium solani. m93 Gliocladium fimbriatum Gliocladium fimbriatum. m94 Phycomyces blakesleeanus Phycomyces blakesleeanus. m201 Tilletia tritici (Ustilago nuda, Ustilago tritici ) (Barley smut) Tilletia tritici (Ustilago nuda, Ustilago tritici ). m202 Acremonium kiliense (Cephalosporium acremonium ) Acremonium kiliense (Cephalosporium acremonium ). m203 Trichosporon pullulans Trichosporon pullulans. m204 Ulocladium chartarum Ulocladium chartarum. m205 Trichophyton rubrum Trichophyton rubrum. m207 Aspergillus niger Aspergillus niger. m208 Chaetomium globosum Chaetomium globosum. m209 Penicillium frequentans Penicillium glabrum (Penicillium frequentans ). m209 Stachybotrys chartarum Stachybotrys chartarum (Stachybotrys atra ). m210 Trichophyton mentagrophytes var. goetzii Trichophyton mentagrophytes var. goetzii. m211 Trichophyton mentagrophytes var. interdigitale Trichophyton mentagrophytes var. interdigitale. m211 Oat smut Ustilago avenae. m212 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m212 Geotrichum candidum Geotrichum candidum. m213 Bermuda grass smut Ustilago cynodontis. m214 Johnson grass smut Sphacelotheca cruenta. m215 Corn smut Ustilago maydis. m218 Asp f 1.0101 Aspergillus fumigatus. a3050 Asp r 1 Aspergillus restrictus. m219 Asp f 2 Aspergillus fumigatus. m220 Asp f 3.0101 Aspergillus fumigatus. m221 Asp f 4 Aspergillus fumigatus. m222 Asp f 6.0101 Aspergillus fumigatus. m223 Staphylococcal enterotoxin C (Sta a SEC ) Staphylococcus aureus. m224 Staphylococcal enterotoxin D (Sta a SED ) Staphylococcus aureus. m226 Staphylococcal enterotoxin TSST (Sta a TSST ) Staphylococcus aureus. m227 Malassezia spp. (Pityrosporum spp.) Malassezia spp. (Pityrosporum spp.). m228 Aspergillus flavus. m229 Alt a 1.0101 Alternaria alternata (Alternaria tenuis ). m230 Alt a 6.0101 Alternaria alternata (Alternaria tenuis ). m231 Cla h 8.0101 Cladosporium herbarum (Hormodendrum ). m300 Eurotium spp Eurotium spp. m304 Aspergillus oryzae Aspergillus oryzae. m305 Penicillium brevicompactum Penicillium brevicompactum. m309 Aspergillus terreus Aspergillus terreus. m310 Aspergillus nidulans Aspergillus nidulans. m311 Aspergillus flavus Aspergillus flavus. m312 Aspergillus clavatus Aspergillus clavatus. Epidermal & Animal e6 Guinea pig epithelium Cavia porcellus. e7 Pigeon droppings Columba palumbus ,Columba livia. e25 Chicken serum Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e26 Parrot serum Psittacoidea spp. e62 Camel Camelus dromedaries. e70 Goose feathers Anser anser. e71 Mouse epithelium Mus musculus (Mus spp.). e73 Rat epithelium Rattus norvegicus. e74 Rat urine proteins Rattus norvegicus ,Rattus rattus. e75 Rat serum proteins Rattus norvegicus ,Rattus rattus. e76 Mouse serum proteins Mus musculus (Mus spp.). e77 Budgerigar droppings Melopsittacus undulatus. e78 Budgerigar feathers Melopsittacus undulatus. e79 Budgerigar serum proteins Melopsittacus undulatus. e80 Goat epithelium Capra hircus. e81 Sheep epithelium Ovis aries (Ovis spp.). e82 Rabbit epithelium Oryctolagus cuniculus (Oryctolagus spp.). e83 Swine epithelium Sus scrofa (Sus scrofa domesticus; Sus spp.). e84 Hamster epithelium Cricetus cricetus, Mesocricetus auratus, and Phodopus sungorus. e85 Chicken feathers Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e86 Duck feathers Anas platyrhynchos. e87 Rat epithelium, serum proteins, and urine proteins Rattus norvegicus Rattus rattus. e88 Mouse epithelium, serum proteins, and urine proteins (mouse) Mus musculus (Mus spp.). e89 Turkey feathers Meleagris gallopavo. e90 Budgerigar serum proteins, feathers, and droppings Melopsittacus undulatus. e91 Pigeon serum proteins, feathers, and droppings Streptopelia roseogrisea ,Psittacidae spp. e92 Parrot serum proteins, feathers, and droppings Ara spp. e93 Pigeon serum proteins Streptopelia roseogrisea. e94 Fel d 1.0101 Felis domesticus. a345 Fel d 1 Felis domesticus. e98 Parrot droppings Psittacoidea spp. e101 Can f 1.0101 Canis familiaris (Canis domesticus ). a174 Can f 1 Canis familiaris (Canis domesticus ). e102 Can f 2.0101 Canis familiaris (Canis domesticus ). e196 Parakeet feathers Nymphicus hollandicus. e197 Parakeet droppings Nymphicus hollandicus. e198 Parakeet serum Nymphicus hollandicus. e199 Canary bird serum Serinus canarius. e200 Canary bird droppings Serinus canarius. e201 Canary bird feathers (Canary feathers) Serinus canarius. e202 Reindeer epithelium Rangifer tarandus. e203 Mink epithelium Mustela spp. e204 Bos d 6 Bos domesticus (Bos taurus; Bos spp.). e205 Horse, serum proteins Equus caballus (Equus spp.). e206 Rabbit, serum proteins Oryctolagus cuniculus (Oryctolagus spp.). e208 Chinchilla epithelium Chinchilla laniger. e209 Gerbil epithelium Meriones unguiculatus. e210 Fox epithelium Vulpes vulpes. e211 Rabbit, urine proteins Oryctolagus cuniculus (Oryctolagus spp.). e212 Swine, urine proteins Sus scrofa (Sus scrofa domesticus; Sus spp.). e213 Parrot feathers Ara spp. e214 Finch feathers Lonchura domestica. e215 Pigeon feathers Streptopelia roseogrisea (Streptopelia spp.),Columbia spp. e216 Deer epithelium Dama dama. e217 Ferret epithelium Mustela putorius. e218 Chicken droppings Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e219 Chicken, serum proteins Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e220 Fel d 2, Cat serum albumin Felis domesticus. e221 Can f 3 Canis familiaris (Canis domesticus ) (Dog serum albumin ). e222 Swine serum albumin (Sus s PSA) Sus scrofa (Sus scrofa domesticus; Sus spp.). e225 Lovebird feathers Psittacoidea agapomis. e226 Can f 5.0101 Canis familiaris. e227 Equ c 1.0101 Equus caballus. e228 Fel d 4.0101 Felis domesticus. e230 Equ c 3 Equus caballus. e231 Mus m 1 Mus musculus. Food f9 Rice Oryza sativa. f12 Pea (green pea) Pisum sativum. f15 White bean Phaseolus vulgaris. f19 Cayenne pepper Capsicum frutescens (Capsicum annum ). f21 Sugar cane Saccharum officinarum. f22 Raspberry Rubus idaeus. f26 Pork Sus scrofa (Sus scrofa domesticus; Sus spp.). f29 Watermelon Citrullus lanatus (Citrullus vulgaris ). f31 Carrot Daucus carota. f32 Oyster mushroom Pleurotus ostreatus. f33 Orange Citrus sinensis. f35 Potato Solanum tuberosum. f43 Mother's milk Homo sapiens. f44 Strawberry Fragaria vesca (Fragaria spp.). f45 Yeast, baker's Saccharomyces cerevisiae. f46 Pepper, Red Capsicum annuum. f47 Garlic Allium sativum. f48 Onion Allium cepa. f49 Apple Malus x domestica (Malus spp.). f51 Bamboo shoot Phyllostachys pubescens. f52 Cacao/chocolate Theobroma cacao. f54 Sweet potato Ipomoea batatas. f55 Common millet Panicum miliaceum. f56 Foxtail millet Setaria italica. f57 Japanese millet Echinochloa crus-galli. f58 Pacific squid Todarodes pacificus. f59 Octopus Octopus vulgaris (Octopus spp.). f63 Kefir NA. f67 Parmesan cheese NA. f81 Cheese, cheddar type NA. f82 Cheese, mold type NA. f83 Chicken Gallus domesticus (Gallus gallus domesticus; Gallus spp.). f86 Parsley Petroselinum crispum. f87 Melon Cucumis melo Cucumis melo + Citrullus lanatus. f88 Mutton (lamb) Ovis aries (Ovis spp.). f90 Malt Hordeum vulgare. f92 Banana Musa spp. f93 Cacao Theobroma cacao. f94 Pear Pyrus communis (Pyrus spp.). f97 Yam Dioscorea spp.Dioscorea opposita. f97 Chamomile tea Matricaria chamomilla. f98 Gliadin Triticum aestivum (Triticum spp.). f102 Cantaloupe Cucumis melo var. cantalupensis. f105 Chocolate Theobroma cacao. f109 Cottonseed Gossypium hirsutum. f110 Giant radish Raphanus sativus. f118 Zucchini Cucurbita pepo. f119 Radish Raphanus sativus. f120 Venison Capreolus capeolus. f121 Pinto bean Phaseolus vulgaris. f122 Cheese, American NA. f127 Black-eyed pea Vigna unguiculata. f131 Black Olive Olea europaea. f136 Red beet Beta vulgaris var. conditiva. f139 Goat's Cheese Capra aegagrus. f140 Bran NA. f141 Corn (vegetables) Zea mays. f152 Green bell pepper Capsicum annuum. f155 Brewer's yeast Saccharomyces carlsbergensis. f157 Duck Anas domesticus. f158 Goose Anser anser. f160 Camembert cheese NA. f162 Nectarine Prunus persica var. nucipersica. f163 Kohlrabi Brassica oleracea var. gongylodes. f65 Perch f166 Leek Allium porrum. f170 Cheese (Switzerland) (Swiss cheese) NA. f174 Fig Ficus carica. f177 Cranberry Vaccinium macrocarpon. f179 Raisin Vitis spp. f182 Lima bean Phaseolus lunatus. f198 Flaxseed (bruised grain) Linum usitatissimum. f199 Untreated native milk Bos domesticus (Bos taurus; Bos spp.). f208 Lemon Citrus limon. f209 Grapefruit Citrus paradisi. f210 Pineapple Ananas comosus. f211 Blackberry Rubus fruticosus. f212 Mushroom (champignon) Agaricus hortensis (Agaricus spp.). f213 Rabbit Oryctolagus cuniculus (Oryctolagus spp.). f214 Spinach Spinacia oleracea. f215 Lettuce Lactuca sativa. f216 Cabbage Brassica oleracea var. capitata. f217 Brussels sprouts Brassica oleracea var. gem. f218 Paprika, sweet pepper Capsicum annuum. f219 Fennel seed Foeniculum vulgare. f219 Sage Salvia officinalis. f220 Cinnamon Cinnamomum spp. f221 Coffee Coffea spp. f222 Tea Camellia sinensis. f223 Green olive Olea europaea. f225 Summer squash, pumpkin Cucurbita pepo. f225 Pumpkin Cucurbita maxima. f226 Pumpkin seed Cucurbita pepo. f227 Sugar-beet seed Beta vulgaris. f229 Safflower Seed Carthamus tinctorius. f231 Milk, boiled Bos domesticus (Bos taurus; Bos spp.). f234 Vanilla Vanilla planifolia. f237 Apricot Prunus armeniaca. f241 Gouda cheese NA. f242 Cherry Prunus avium. f244 Cucumber Cucumis sativus. f246 Guar, guar gum Cyamopsis tetragonoloba. f247 Honey NA. f248 Rosemary Rosmarinus officinalis. f254 Plaice Pleuronectes platessa. f255 Plum Prunus domestica ,Prunus americana. f258 Squid Loligo spp. f259 Grape Vitis vinifera (Vitis spp.). f260 Broccoli Brassica oleracea var. italica (Brassica oleracea var. cultivar ). f261 Asparagus Asparagus officinalis. f262 Aubergine, eggplant Solanum melongena. f263 Green pepper Piper nigrum ,Capsicum annuum. f264 Eel Anguilla anguilla. f265 Caraway Carum carvi. f265 Cumin Cuminum cyminum. f266 Mace Myristica fragrans. f267 Cardamon Elettaria cardamomum. f268 Clove Syzygium aromaticum. f269 Basil Ocimum basilicum. f270 Ginger Zingiber officinale. f271 Anise Pimpinella anisum. f272 Tarragon Artemisia dracunculus. f273 Thyme Thymus vulgaris. f274 Marjoram Origanum majorana. f275 Lovage Levisticum officinale. f276 Fennel, fresh Foeniculum vulgare. f277 Dill Anethum graveolens. f278 Bay leaf Laurus nobilis. f279 Chili pepper Capsicum frutescens. f280 Black pepper Piper nigrum. f281 Curry (Santa Maria) NA. f282 Nutmeg Myristica fragrans. f283 Oregano Origanum vulgare. f284 Turkey meat Meleagris gallopavo. f285 Elk/moose meat Alces spp. f286 Mare's milk Equus caballus (Equus spp.). f287 Red kidney bean Phaseolus vulgaris. f288 Blueberry Vaccinium myrtillus (Vaccinium spp.). f289 Date Phoenix dactylifera. f291 Cauliflower Brassica oleracea var. botrytis. f292 Guava Psidium guajava. f293 Papaya Carica papaya. f294 Passion fruit, Maracuja Passiflora edulis (Passiflora spp.). f295 Carambola Averrhoa carambola. f296 Carob Ceratonia siliqua. f297 Gum Arabic Acacia senegal (Acacia spp.). f298 Tragacanth Astragalus spp. f299 Sweet chestnut (chestnut) Castanea sativa. f300 Pinto bean Phaseolus spp. f301 Persimmon (kaki fruit, sharon) Diospyros kaki. f302 Mandarin (tangerine, clementine, satsumas) Citrus reticulata. f305 Fenugreek Trigonella foenum-graecum. f306 Lime Citrus aurantifolia. f307 Hake Merluccius merluccius. f308 Sardine (pilchard) Sardina pilchardus. f310 Blue vetch Lathyrus sativus. f311 Megrim Lepidorhombus whiffiagonis. f315 Green bean Phaseolus vulgaris. f316 Rape seed Brassica napus. f317 Coriander Coriandrum sativum. f318 Jack fruit Artocarpus heterophyllus. f319 Beetroot Beta vulgaris. f320 Crayfish Astacus astacus. f321 Horse meat Equus caballus (Equus spp.). f322 Red currant Ribes sylvestre. f324 Hop (fruit cone) Humulus lupulus. f325 Saffron Colchicum autumnale. f328 Fig Ficus carica. f329 Watermelon Citrullus lanatus. f330 Rose hip Rosa spp. f331 Saffron Crocus sativus. f332 Mint Mentha piperita. f333 Linseed Linum usitatissimum. f336 Jujube Ziziphus jujuba. f336 Wine vinegar Vitis vinifera (Vitis spp.). f337 Sole Solea solea. f337 English sole Parophrys vetulus. f338 Wine, white Vitis vinifera (Vitis spp.). f339 Allspice Pimenta dioica. f339 Wine, red Vitis vinifera (Vitis spp.). f341 Cranberry Vaccinium oxycoccus ,Vaccinium macrocarpon. f342 Olive (black, fresh) Olea europaea. f343 Raspberry Rubus idaeus. f344 Sage Salvia officinalis. f346 Chives Allium schoenoprasum. f347 Quinoa Chenopodium quinoa. f348 Litchi Litchi chinensis. f349 Chum salmon roe Oncorhynchus keta. f358 Artichoke Cynara scolymus. f360 Yogurt NA. f368 Black bass Micropterus dolomieu (Micropterus dolomieui ). f374 Karaya gum Sterculia urens. f375 Horseradish Armoracia rusticana. f377 Maple syrup NA. f379 Okra Abelmoschus esculentus. f382 Beet, sugar Beta vulgaris var. altissima. f401 Loquat Eriobotrya japonica. f402 Fig Ficus carica. f403 Brewer's yeast Saccharomyces cerevisiae. f405 Mint Mentha spp. f406 Arugula Eruca vesicaria. House Dust h1 Greer Labs., Inc NA. h2 Hollister-Stier Labs NA. h6 Japan NA. Venoms & Insects i7 Midge Chironomus yoshimatsui. i8 Moth Bombyx mori ,Heterocera spp. i47 Water flea Daphnia spp. i49 Deer fly Chrysops spp. i51 Black ant Camponotus pennsylvanicus. i54 Flea mix (dog/cat), common flea Ctenocephalides spp. i71 Mosquito Aedes communis ,Aedes spp.and Culex spp. i72 Green nimitti Cladotanytarsus lewisi. i73 Blood worm Chironomus thummi ,Chironomusri parius ,Chironomus spp. i75 European hornet Vespa crabro. i76 Berlin beetle Trogoderma angustum. i77 European paper wasp Polistes dominulus. i78 Fly Musca domestica. i80 Bumblebee Bombus pennsylvanicus. i201 Horse bot fly Gasterophilus intestinalis. i202 Grain weevil Sitophilus granarius. i203 Mediterranean flour moth Ephestia kuehniella (Anagasta kuehniella ). i204 Horse fly Tabanus spp. i205 Bumblebee Bombus terrestris. i208 Api m 1.0101 Apis mellifera. a45 Api m 1 Apis mellifera. i209 Ves v 5.0101 Vespula vulgaris. a670 Ves v 5 Vespula vulgaris. i210 Pol d 5.0101 Polistes dominulus. i211 Ves v 1.0101 Vespula vulgaris. i213 Api m 4 Apis mellifera. i214 Api m 2 Apis mellifera. i215 Api m 3 Apis mellifera. i216 Api m 5 Apis mellifera. i217 Api m 10 Apis mellifera. i220 Bla g 1.0101 Blattella germanica. i221 Bla g 2.0101 Blattella germanica. i222 Bla g 5.0101 Blattella germanica. i223 Bla g 7 Blattella germanica. a46 Api m 2 Apis mellifera. Miscellaneous o1 Cotton, crude fibers Gossypium spp. o3 Cotton (treated) Gossypium spp. o70 Seminal fluid Homo sapiens. o71 Staphylococcus aureus Staphylococcus aureus. o72 Pichia pastoris crude extract customer specific Pichia pastoris. o72 Sperm-sediment Homo sapiens. o73 Pichia pastoris crude extr. vector customer specific Pichia pastoris. o74 Pichia pastoris with vector customer specific Pichia pastoris. o201 Tobacco leaf, tobacco dust Nicotiana tabacum. o202 Artemia salina, fish feed Artemia salina. o203 Tetramin, fish feed NA. o207 Daphnia, fish feed Daphnia spp. o211 Mealworm Tenebrio molitor. o212 Streptavidin Streptomyces avidini. o213 MBP (maltose binding protein) Escherichia coli. o214 CCD; MUXF3 from bromelain Ananas comosus. o72 Enterotoxin A (Sta a SEA) Staphylococcus aureus. o73 Enterotoxin B (Sta a SEB) Staphylococcus aureus. Parasites p1 Ascaris Ascaris suum. p2 Echinococcus Echinococcus granulosus. p3 Schistosoma Schistosoma mansoni. p4 Anisakis (Herring Worm) Anisakis simplex (Anisakis spp.). p5 Toxocara canis Toxocara canis. p10 Ani s 3.0101 Anisakis simplex (Anisakis spp.). p11 Ani s 1 Anisakis simplex (Anisakis spp.). Occupational k4 Threshing dust NA. k5 Flax NA. k7 Hay Dust NA. k8 Hop (hops) Humulus lupulus. k12 Grain mill dust NA. k14 Kapok NA. k20 Sheep's wool (treated) (wool) Ovis aries (Ovis spp.). k21 Sheep's wool (Untreated) Ovis aries (Ovis spp.). k23 Straw Dust NA. k33 Oak NA. k70 Green coffee bean Coffea spp. k71 Castor bean Ricinus communis. k72 Ispaghula Plantago psyllium/Plantago ovata. k73 Silk waste NA. k74 Silk Bombyx mori. k75 Isocyanate TDI (Toluene diisocyanate) NA. k76 Isocyanate MDI (Diphenylmethane diisocyanate) NA. k77 Isocyanate HDI (Hexamethylen diisocyanate) NA. k78 Ethylene oxide NA. k79 Phthalic anhydride NA. k80 Formaldehyde/Formalin NA. k81 Ficus Ficus benjamina (Ficus spp.). k83 Cotton seed Gossypium hirsutum. k84 Sunflower seed Helianthus annuus. k85 Chloramin T NA. k86 Trimellitic anhydride, TMA NA. k87 Asp o 21, alpha-amylase Aspergillus oryzae. k89 Orris root Iris florentina. k99 HSA (Human Serum Albumin) (Hom s HSA) Homo sapiens. k201 Car p 1, Papain Carica papaya. k202 Ana c 2, Bromelain Ananas comosus. k204 Maxatase Bacillus licheniformis. k205 Alcalase Bacillus spp. k206 Savinase, Protease 1 (Bac l Subtilisin) Bacillus spp. k208 Gal d 4, Lysozyme Gallus domesticus (Gallus gallus domesticus; Gallus spp.). k209 Hexahydrophtalic anhydrid NA. k210 Maleic anhydride NA. k211 Methyltetrahydrophtalic anhydrid NA. k212 Abachi wood dust Triplochiton scleroxylon. k213 Pepsin (Sus s Pepsin) Sus scrofa (Sus scrofa domesticus; Sus spp.). k213 TCPA NA. k214 Bougainvillea Bougainvillea spp. k225 Horse radish peroxidase (Arm r HRP) Armoracia rusticana. k226 Ascorbate oxidase (Cuc p ascorbate oxidase) Cucurbita pepo. k301 Flour dust Triticum spp. k501 Savinase customer specific Proprietary knowledge of customer. k502 Lipolase customer specific Proprietary knowledge of customer. k503 Termamyl customer specific Proprietary knowledge of customer. k504 Clazinase customer specific Proprietary knowledge of customer.